K Number
K043366

Validate with FDA (Live)

Date Cleared
2004-12-17

(9 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Permacol® Surgical Implant is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists or ruptured soft tissue membranes. It is specifically indicated for the repair of abdominal wall defects and hernias, including but not limited to parastomal hernias.

Device Description

Permacol® Surgical Implant Crosslinked Porcine Dermal Collagen Mesh

AI/ML Overview

This 510(k) summary (K043366) is for Tissue Science Laboratories, plc's Permacol® Surgical Implant Crosslinked Porcine Dermal Collagen Mesh. The submission primarily focuses on establishing substantial equivalence to predicate devices based on intended use, technological characteristics, and performance data from biocompatibility and bench testing.

Here's an analysis of the provided information against your requested criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
BiocompatibilityNon-toxic to biological tissues"The biocompatibility test results show that the materials used in the design and manufacture of the device are non-toxic..."
Non-sensitizing to biological tissues"...non-sensitizing to biological tissues."
Material PerformanceMaterials consistent with intended use"...new techniques and results demonstrate that the materials are consistent with the intended use."
Physical PerformanceMeet performance criteria during intended use"Laboratory tests demonstrate that the Permacol® Surgical Implant will meet chosen and the design utilized in manufacturing criteria of performance during their intended use."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify sample sizes for biocompatibility or bench testing. It only generally refers to "test results" and "laboratory tests."

The data provenance is not explicitly stated as retrospective or prospective, nor does it mention a country of origin for the data. However, the manufacturer is in the UK, suggesting the tests were conducted under their guidance, possibly in the UK or a contract lab.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the given document. Biocompatibility and bench testing usually rely on standardized protocols and laboratory measurements rather than expert human assessment for ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable and therefore not provided. Adjudication methods are typically used in clinical studies involving human interpretation or subjective assessment, which is not the case for the described biocompatibility and bench tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

No MRMC comparative effectiveness study was mentioned or indicated. The submission focuses on substantial equivalence based on material and mechanical properties, not clinical efficacy comparisons with human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

Not applicable. This device is a surgical mesh, not a medical imaging or AI diagnostic algorithm. Therefore, "standalone algorithm performance" is not relevant.

7. The Type of Ground Truth Used:

For biocompatibility, the ground truth is established by standardized biological assays and observations for cellular response, toxicity, and sensitization, compared against established biological safety limits.

For physical performance and material characterization (bench testing), the ground truth is established by objective measurements of properties like tensile strength, elasticity, pore size, degradation rate, etc., compared against pre-defined engineering specifications or comparisons to predicate devices' known properties.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical medical implant, not an AI algorithm. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As there is no training set for a physical device, there is no ground truth to establish for it.


Summary of Device Performance Study:

The provided 510(k) summary indicates that the substantial equivalence of the Permacol® Surgical Implant is supported by:

  • Biocompatibility testing: This demonstrated that the materials used are non-toxic and non-sensitizing. The "ground truth" here is the biological response observed in standardized tests, confirming the material's safety for biological contact.
  • Bench testing (laboratory tests): These tests confirmed that the device's materials are "consistent with the intended use" and that the product "will meet chosen and the design utilized in manufacturing criteria of performance during their intended use." The "ground truth" here is objective physical and mechanical measurements compared against predetermined specifications and predicate device characteristics.

The submission does not include human clinical trial data, MRMC studies, or AI algorithm performance data, as these are not relevant to the type of device and the pathway chosen for regulatory clearance (510(k) substantial equivalence). The focus is on the device's physical and biological properties.

{0}------------------------------------------------

K043366$_{Pi/2}$

510(k) SUMMARY

DEC 17 2004

Tissue Science Laboratories, plc's Permacol® Surgical Implant Crosslinked Porcine Dermal Collagen Mesh

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:

Victoria Taylor Tissue Science Laboratories, plc 1141 Clark Street, Suite D Covington, GA 30014 Phone: (678) 342-7808 (678) 342-7844 Fax: Email: vtaylor@tissuescience.com

Contact Person: Victoria Taylor

December 06, 2004 Date Prepared:

Name of Device and Name/Address of Sponsor

Tissue Science Laboratories, plc 7th Floor, Victoria House Victoria Road Aldershot Hampshire GU111 1EJ United Kingdom

Common or Usual Name

Surgical Mesh

Classification Name

Surgical Mesh

Predicate Devices

Permacol® Crosslinked Porcine Dermal Coilagen Surgical Mesh (K992556), BioSugar Sungers (19), Cross 110), Box 1002162), Cook Biotech 1 V ascular Inc S Supple Felf-Guard® (11705163).
SurgiSIS™ (K980431) and Bard's® CK Parastomal Hernia Patch (K042026).

Intended Use

Permacol® Surgical Implant is intended for use as a soft tissue patch to reinforce Permiacol® Surgical implant is intensed or ruptured soft tissue

{1}------------------------------------------------

K043366

P2/2

membranes. It is specifically indicated for the repair of abdominal wall defects and hernias, including but not limited to parastomal hernias.

Technological Characteristics and Substantial Equivalence

Permacol® is substantially equivalent to its predicates because it has the same intended use and very similary equivalent with to tax to the soft tiseve intended use and very simmal teenge of surgical procedures for soft tissues and are intended for use in a broad fange of saige of saige of Sarge of Service, includes include repairremforcement. Permacol® and Burd 50 Corrent wall defects and hernias, including parastomal hernias.

The technological characteristics of Permacol® are very similar to its predicate The technological characteristics of I climits on I cimensions, technological characteristics, and thickness.

Performance Data

Biocompatibility and bench testing have been completed and support the safety and effectiveness of the Permacol® Surgical Implant for its intended use.

The biocompatibility test results show that the materials used in the design and manufacture of the device are non-toxic and non-sensitizing to biological tissues manufacture of the device are non toxic and new teen tresults demonstrate that the materials materials consistent with then michded use. Laboratory termacol® Surgical Implant will meet chosen and the design unfized in manufacturing i critation of tormance during their intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized lines above a wavy line.

Public Health Service

DEC 17 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Victoria Taylor Associate Director of Regulatory Affairs Tissue Science Laboratories plc North American Division 1141 Clark Street, Suite D Covington, Georgia 30014

Re: K043366

Trade/Device Name: Permacol® Surgical Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Polymeric surgical mesh Regulatory Class: II Product Code: FTL Dated: December 6, 2004 Received: December 8, 2004

Dear Ms. Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device w & nave rowled above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrease, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and cosmetie fore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Victoria Taylor

This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter with and my of substantial equivalence of your device of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

( ol Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K043366

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Permacol® Surgical Implant Device Name:

Indications for Use:

Permacol® is intended of use as a soft tissue patch to reinforce soft tissue where weakness exists Pennacol® is intended of ass a box or ruptured soft tissue membranes. It is specifically alld for the surgical repair of damages wall defects and hernias, including but not limited to parastomal hernias.

Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

iriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page _of _
--------------

510(k) Number K643366
59

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.