K Number
K042779

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-10-21

(15 days)

Product Code
Regulation Number
892.1630
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Radiotherapy Solution Based on CR25.0 is indicated for producing simulation and quality control images for use in radiation therapy planning and quality control.

Device Description

The Radiotherapy Solution Based on CR allows the application of Portal Imaging in a very wide dose range (1 MU - 400 MU's and higher) by using two different Portal Imaging Cassette types, which are optimised for image quality at their intended dose range. The Radiotherapy Solution Based on CR supports both low- and high-dose applications (sometimes called localisation and verification portal imaging). Not only does the system enable the acquisition of the images under the typical Radiotherapy conditions, the specific requirements for these images are also met which allows their use by the typical "next-in-line" radiotherapy applications. Typical "next-in-line" applications for simulation imaging are, for instance, image comparison and bloc compensator/MLC calculations. For portal imaging, a typical "next-in-line" application is image comparison with a reference image (this can be a simulation image or DRR: comparisons are made between hardcopy prints or on a digital workstation).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Radiotherapy Solution Based on CR" (later referred to as "Radiotherapy Solution Based on CR25.0"). This submission focuses on demonstrating substantial equivalence to previously marketed predicate devices, rather than establishing specific performance criteria through a detailed clinical study with acceptance criteria.

The submission is for an accessory to a CR system and emphasizes its technical characteristics and intended use being similar to existing cleared devices. Therefore, the information typically found in a clinical study demonstrating performance against acceptance criteria in terms of accuracy, sensitivity, specificity, or other quantitative metrics is largely absent.

Here's an analysis based on the provided text, highlighting what is available and what is not:

1. A table of acceptance criteria and the reported device performance:

This information is not explicitly provided in the submission. The submission states that the device "has been tested for proper performance to specifications through various in-house reliability and imaging performance demonstration tests." However, the specific acceptance criteria for these "specifications" and the quantitative results are not detailed.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

The text mentions "Clinical performance has been tested in the typical environment of a clinical radiotherapy department, and sample clinical images have been provided in this 510(k)." The sample size for this "clinical performance" test is not specified, nor is the exact data provenance (country of origin, retrospective/prospective). The emphasis is on demonstrating "sample clinical images" as confirmatory evidence of equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not provided. Given the nature of a 510(k) submission focused on substantial equivalence rather than a de novo clinical performance study, the establishment of ground truth by multiple experts is not detailed. The "clinical performance" testing seems to be more about confirming the device functions as expected in a real-world setting rather than a rigorous diagnostic accuracy study.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

An MRMC comparative effectiveness study was not conducted or reported. The device is an image acquisition and processing system, not an AI-assisted diagnostic tool for human readers. Its primary function is to enable "Portal Imaging" and produce images for "simulation and quality control" in radiotherapy.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

A standalone performance evaluation of the "algorithm" in the sense of a diagnostic accuracy study is not explicitly detailed. The device is an integrated system (cassettes, image acquisition, and processing) for producing images. The "performance to specifications" would likely involve objective image quality metrics (e.g., spatial resolution, contrast, noise, dose linearity) which are inherent to the algorithm and hardware working together. However, specific acceptance criteria and detailed results for these are not provided in the summary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Given the device's function in radiotherapy imaging (localization and verification, simulation, quality control), the "ground truth" for its performance would likely relate to the accuracy of anatomical representation, consistency of image quality, and ability to meet the requirements for "next-in-line" radiotherapy applications. However, the specific methodology for establishing this "ground truth" (e.g., comparison to known anatomical landmarks, phantoms, established dosimetry) is not detailed in the provided summary. There is no mention of expert consensus, pathology, or outcomes data in the context of establishing ground truth for the clinical performance.

8. The sample size for the training set:

The concept of a "training set" in the context of machine learning is not applicable to this device. This device is an imaging system, not a machine learning model.

9. How the ground truth for the training set was established:

As the concept of a "training set" is not applicable, this information is not relevant.

In summary:

This 510(k) summary is typical for a device demonstrating substantial equivalence, where the focus is on comparing the new device's intended use and technological characteristics to already cleared predicate devices. It does not provide the detailed performance metrics, acceptance criteria, multi-reader studies, or comprehensive ground truth establishment methodologies that would be expected for a de novo device or an AI-based diagnostic tool requiring rigorous clinical validation. The "testing" mentioned is primarily for "proper performance to specifications" and "clinical performance" to confirm its functionality in a radiotherapy environment, without specific quantitative results against predefined acceptance criteria being disclosed in this summary.

{0}------------------------------------------------

K042779

OCT 2 1 2004

510(k) Summary

Radiotherapy Solution Based on CR

Computed Common/Classification Name: Radiotherapy Cassettes for Radiography, 21 CFR 892.1630

Agfa Corporation 10 South Academy Street Greenville, SC 29602-9048

Contact: Jeffery A. Jedlicka, Prepared: August 9, 2004

LEGALLY MARKETED (PREDICATE) DEVICES A.

The Radiotherapy Solution Based on CR is substantially equivalent to the PoRT Cassette (K032654) in regard to its indications for use and cassette characteristics, and to the CR25.0 in regard to the image acquisition and processing characteristics (K041701). The Radiotherapy Solution Based on CR should be considered as an accessory to the CR25.0, as was the case for the PoRT Cassette with respect to the PcCR device. K032654 was cleared by FDA as an accessory under regulation 21 CFR 892.1630 (product code IXK), so presumably this is the appropriate regulation and product code for the present product as well.

B. DEVICE DESCRIPTION

The Radiotherapy Solution Based on CR allows the application of Portal Imaging in a very wide dose range (1 MU - 400 MU's and higher) by using two different Portal Imaging Cassette types, which are optimised for image quality at their intended dose range. The Radiotherapy Solution Based on CR supports both low- and high-dose applications (sometimes called localisation and verification portal imaging).

Not only does the system enable the acquisition of the images under the typical Radiotherapy conditions, the specific requirements for these images are also met which allows their use by the typical "next-in-line" radiotherapy applications. Typical "next-in-line" applications for simulation imaging are, for instance, image comparison and bloc compensator/MLC calculations. For portal imaging, a typical "next-in-line" application is image comparison with a reference image (this can be a simulation image or DRR: comparisons are made between hardcopy prints or on a digital workstation).

{1}------------------------------------------------

C. INTENDED USE

The Radiotherapy Solution Based on CR is substantially equivalent to the PoRT Cassette by Orex (K032654) in regard to its indications for use and cassette characteristics, and to the Agfa CR25.0 in regard to the image acquisition and processing characteristics (K041701). The Radiotherapy Solution Based on CR should be considered as an accessory to the CR25.0, as was the case for the PoRT Cassette with respect to the PcCR device (K003256). K032654 was cleared by FDA as an accessory under regulation 21 CFR 892.1630 (product code IXK), so presumably this is the appropriate regulation and product code for the present product as well.

SUBSTANTIAL EQUIVALENCE SUMMARY D.

The Radiotherapy Solution Based on CR has a similar, but not identical, indications for use statement as the legally marketed predicate devices. However, the differences are minor and do not alter the fact that the devices clearly have the same intended use. The Radiotherapy Solution Based on CR has the same technological characteristics as the predicate devices. This premarket notification has described the characteristics of the device in sufficient detail to assure substantial equivalence. Sample clinical images are provided as confirmatory evidence that the device is equivalent to the legally marketed devices.

TECHNOLOGICAL CHARACTERISTICS E.

The technological characteristics are the same in the proposed and legally marketed predicate devices.

F. TESTING

The Radiotherapy Solution Based on CR has been tested for proper performance to specifications through various in-house reliability and imaging performance demonstration tests. Clinical performance has been tested in the typical environment of a clinical radiotherapy department, and sample clinical images have been provided in this 510(k).

G. CONCLUSIONS

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a circular logo for the Department of Health & Human Services - USA. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of three stylized lines that resemble a bird or a symbol representing health and human services.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 1 2004

Agfa Corporation % Mr. William Sammons Project Engineer ·Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709 Re: K042779 Trade/Device Name: Radiotherapy Solution Based on CR25.0 Regulation Number: 21 CFR 892.1630 Regulation Name: Electrostatic x-ray imaging system Regulatory Class: II

Product Code: 90 IXK Dated: October 1, 2004 Received: October 6, 2004

Dear Mr. Sammons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have it viewed your botten box(s) p
above and have determined the device is substantially equivalent (for the indications for use stated in above and have decemii... Parketed predicate devices marketed in interstate commerce prior to the enclosure) to regally maneted problem. Device Amendments, or to devices that have been May 20, 1770, the chacinent acto of the Federal Food, Drug, and Cosmetic Act (Act) that it classified in accordantes approval application (PMA). You may, therefore, market the do not require approval or a provisions of the Act. The general controls provisions of the Act device, subject to the general esting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket 11 your de rice is elassilied (200 2011 additional controls. Existing major regulations affecting your Apployally, It may or babyer to on Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA device ban of round mouncements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease oe advisod that i by vion that your device complies with other requirements of the Act or any FDA has masternination and regulations administered by other Federal agencies. You must comply with all the I cacraf states and regulation but not limited to registration and listing (21 CFR Part 807); labeling Act 3 requirements, morating practice requirements as set forth in the quality systems (QS) (21 CFR Part 8017), good not if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin maneting of substantial equivalence of your device to a legally premarked notheation. The PDF mining of Cassification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please II you desire specific acvice for your de roche following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, Tormation on your responsibilities under the Act from the 807.77). Tou may obtain careers, International and Consumer Assistance at its toll-free number (800) DVNSloll or (301) 443-6597 or at its Internet address http://www.fda.ggv/cdrb/dsmadsmamain.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. I Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K 04 2779

Radiotherapy Solution Based on CR25.0 Device Name:__________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The Radiotherapy Solution Based on CR25.0 is indicated for producing simulation and quality The Radiotherapy Solution Based on ORLO.6 In Indiation therapy planning and quality control.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David h. Dyson

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number __

Page 1 of 1

00037

§ 892.1630 Electrostatic x-ray imaging system.

(a)
Identification. An electrostatic x-ray imaging system is a device intended for medical purposes that uses an electrostatic field across a semiconductive plate, a gas-filled chamber, or other similar device to convert a pattern of x-radiation into an electrostatic image and, subsequently, into a visible image. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.