Search Results
Found 6 results
510(k) Data Aggregation
(521 days)
IXK
Ask a specific question about this device
(74 days)
IXK
The PortaVision PVMed DDR 2520 Digital Imaging System is intended for use in generating radiographic images of human anatomy. (The system is not intended to be use for mammography). It is intended to replace film/screen or computed radiography in extremity and general-purpose procedures appropriate to the input field of view. This imager is intended for incorporation into a complete x-ray system by a qualified equipment manufacturer. This device is intended for use by qualified medical personnel and is contraindicated when, in the judgment of the physician, procedures would be contrary to the best interest of the patient.
The PortaVision PVMed DDR 2520 Digital Imaging System consists of two components, an amorphous silicon digital x-ray imager and software for viewing the captured images on a Windows-based computer. The digital imager is composed of an amorphous silicon flat panel imager which uses a large-area amorphous silicon sensor array with a gadolinium oxysulfide scintillator. The PVMed DDR 2520 Digital Imaging System will display high quality images in less that 5 seconds over a wide range of X-Ray dose settings. The System is not intended to be use for mammography.
"The provided text does not contain detailed information about specific acceptance criteria or a dedicated study proving the device meets those criteria. Instead, it focuses on the device's substantial equivalence to a predicate device. Therefore, I cannot extract the requested information directly.
However, I can infer from the text that the primary "acceptance criterion" for this 510(k) submission is substantial equivalence to existing legally marketed devices. The study proving this "acceptance criterion" is the 510(k) submission itself, which presents a comparison of the technological characteristics and intended use of the PortaVision PVMed DDR 2520 Digital Imaging System with those of predicate devices.
Here's an attempt to answer your questions based on the available text, highlighting what's explicitly stated and what cannot be determined:
Acceptance Criteria and Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Substantial Equivalence to Predicate Devices | The PortaVision PVMed DDR 2520 Digital Imaging System has the same intended use and the same technological characteristics (amorphous silicon digital x-ray imager and software for viewing images on a Windows-based computer) as the predicate device (Paxscan 4030 Medical Digital Imaging Systems, Varian Medial Systems, Inc.). The only difference noted is the size, with the proposed device being smaller than the predicate. |
Safety and Effectiveness | The submission concludes that "the PortaVision PVMed DDR 2520 Digital Imaging System does not raise any questions regarding safety and effectiveness" and "is as safe and effective as the predicate device." |
Image Display Quality and Speed | The device "will display high quality images in less than 5 seconds over a wide range of X-Ray dose settings." (This is a performance characteristic but not explicitly framed as an acceptance criterion in the provided text.) |
Intended Use | Intended for use in generating radiographic images of human anatomy (excluding mammography), replacing film/screen or computed radiography in extremity and general-purpose procedures appropriate to the input field of view. This matches the intended use of the predicate device implicitly. |
Study Details (Based on available information)
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- The provided text does not mention a specific test set, its sample size, or data provenance related to a performance study. The submission focuses on a comparison of technical specifications and intended use against a predicate device for substantial equivalence.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- The provided text does not mention a test set, ground truth establishment, or experts for such a purpose.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- The provided text does not mention a test set or any adjudication method.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- The provided text does not indicate that an MRMC comparative effectiveness study was done. The device described is a digital imaging system, not an AI-assisted diagnostic tool, so such a study would not be expected in this context.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The provided text does not describe a standalone algorithm performance study. The device is an imaging system designed for use by qualified medical personnel.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The provided text does not mention a specific ground truth type as no performance study with a test set is detailed. The "ground truth" for this 510(k) submission is essentially the legally marketed status and established safety/effectiveness of the predicate device.
-
The sample size for the training set
- The provided text does not mention a training set sample size. This type of information is typically relevant for machine learning or AI-based devices, which this imaging system is not described as.
-
How the ground truth for the training set was established
- The provided text does not mention a training set or how a ground truth for it would be established.
In summary: The provided 510(k) summary is for an imaging system seeking substantial equivalence to a predicate device. It primarily relies on demonstrating that the new device shares the same intended use and technological characteristics as the already-cleared predicate, rather than presenting a detailed clinical performance study with a test set, ground truth, or expert evaluations in the manner typically seen for AI/ML-driven diagnostic aids."
Ask a specific question about this device
(15 days)
IXK
The Radiotherapy Solution Based on CR25.0 is indicated for producing simulation and quality control images for use in radiation therapy planning and quality control.
The Radiotherapy Solution Based on CR allows the application of Portal Imaging in a very wide dose range (1 MU - 400 MU's and higher) by using two different Portal Imaging Cassette types, which are optimised for image quality at their intended dose range. The Radiotherapy Solution Based on CR supports both low- and high-dose applications (sometimes called localisation and verification portal imaging). Not only does the system enable the acquisition of the images under the typical Radiotherapy conditions, the specific requirements for these images are also met which allows their use by the typical "next-in-line" radiotherapy applications. Typical "next-in-line" applications for simulation imaging are, for instance, image comparison and bloc compensator/MLC calculations. For portal imaging, a typical "next-in-line" application is image comparison with a reference image (this can be a simulation image or DRR: comparisons are made between hardcopy prints or on a digital workstation).
The provided text describes a 510(k) premarket notification for a medical device called "Radiotherapy Solution Based on CR" (later referred to as "Radiotherapy Solution Based on CR25.0"). This submission focuses on demonstrating substantial equivalence to previously marketed predicate devices, rather than establishing specific performance criteria through a detailed clinical study with acceptance criteria.
The submission is for an accessory to a CR system and emphasizes its technical characteristics and intended use being similar to existing cleared devices. Therefore, the information typically found in a clinical study demonstrating performance against acceptance criteria in terms of accuracy, sensitivity, specificity, or other quantitative metrics is largely absent.
Here's an analysis based on the provided text, highlighting what is available and what is not:
1. A table of acceptance criteria and the reported device performance:
This information is not explicitly provided in the submission. The submission states that the device "has been tested for proper performance to specifications through various in-house reliability and imaging performance demonstration tests." However, the specific acceptance criteria for these "specifications" and the quantitative results are not detailed.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
The text mentions "Clinical performance has been tested in the typical environment of a clinical radiotherapy department, and sample clinical images have been provided in this 510(k)." The sample size for this "clinical performance" test is not specified, nor is the exact data provenance (country of origin, retrospective/prospective). The emphasis is on demonstrating "sample clinical images" as confirmatory evidence of equivalence.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This information is not provided. Given the nature of a 510(k) submission focused on substantial equivalence rather than a de novo clinical performance study, the establishment of ground truth by multiple experts is not detailed. The "clinical performance" testing seems to be more about confirming the device functions as expected in a real-world setting rather than a rigorous diagnostic accuracy study.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
An MRMC comparative effectiveness study was not conducted or reported. The device is an image acquisition and processing system, not an AI-assisted diagnostic tool for human readers. Its primary function is to enable "Portal Imaging" and produce images for "simulation and quality control" in radiotherapy.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
A standalone performance evaluation of the "algorithm" in the sense of a diagnostic accuracy study is not explicitly detailed. The device is an integrated system (cassettes, image acquisition, and processing) for producing images. The "performance to specifications" would likely involve objective image quality metrics (e.g., spatial resolution, contrast, noise, dose linearity) which are inherent to the algorithm and hardware working together. However, specific acceptance criteria and detailed results for these are not provided in the summary.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Given the device's function in radiotherapy imaging (localization and verification, simulation, quality control), the "ground truth" for its performance would likely relate to the accuracy of anatomical representation, consistency of image quality, and ability to meet the requirements for "next-in-line" radiotherapy applications. However, the specific methodology for establishing this "ground truth" (e.g., comparison to known anatomical landmarks, phantoms, established dosimetry) is not detailed in the provided summary. There is no mention of expert consensus, pathology, or outcomes data in the context of establishing ground truth for the clinical performance.
8. The sample size for the training set:
The concept of a "training set" in the context of machine learning is not applicable to this device. This device is an imaging system, not a machine learning model.
9. How the ground truth for the training set was established:
As the concept of a "training set" is not applicable, this information is not relevant.
In summary:
This 510(k) summary is typical for a device demonstrating substantial equivalence, where the focus is on comparing the new device's intended use and technological characteristics to already cleared predicate devices. It does not provide the detailed performance metrics, acceptance criteria, multi-reader studies, or comprehensive ground truth establishment methodologies that would be expected for a de novo device or an AI-based diagnostic tool requiring rigorous clinical validation. The "testing" mentioned is primarily for "proper performance to specifications" and "clinical performance" to confirm its functionality in a radiotherapy environment, without specific quantitative results against predefined acceptance criteria being disclosed in this summary.
Ask a specific question about this device
(42 days)
IXK
The PoRT Cassette is an accessory to the PcCR Digital Imaging Device and is indicated for portal imaging (Radiation Therapy Quality Control).
The PoRT Cassette is an optional cassette to be used with the PcCR Digital Imaging Devices for portal imaging. The PcCR Digital Imaging Devices are filmless systems intended for digital radiography using a phosphor storage screen which were cleared under K003256. The PcCR Digital Imaging device enables the clinician to scan or import images for display, review, or storage in a database. The PcCR device consists of reusable phosphor storage screens for recording radiographic images, an image reader/digitizer, and software for displaying, enhancing, and storing radiographs using a user-provided personal computer. The PoRT Cassette is optional cassette to be used with the PcCR System and is also a reusable phosphor screen. It is offered in 14" x 17" size. Other than the addition of the PoRT Cassette and a software upgrade, the PcCR Digital Imaging Device is identical to that described in K003256.
The provided 510(k) summary for the PoRT Cassette (K032654) contains very limited information regarding detailed acceptance criteria and the specifics of the study conducted. The summary focuses on establishing substantial equivalence to a predicate device and does not include an in-depth clinical or performance study breakdown typically found in a full submission.
Based on the information provided, here's what can be extracted and what is missing:
Acceptance Criteria and Reported Device Performance
The document states: "Bench testing and clinical validation demonstrate that the PoRT Cassette performs according to specifications." However, it does not explicitly define what those specifications or acceptance criteria are, nor does it provide specific numerical results for device performance against those criteria. It only makes a general statement of compliance.
Therefore, a table of acceptance criteria and reported device performance cannot be generated from the given text.
Study Details
The provided text mentions "Bench testing and clinical validation," but offers almost no specifics about these studies.
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size (Test Set): Not specified.
- Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
-
Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
- Not specified. The document does not mention the use of experts for ground truth establishment.
-
Adjudication Method for the Test Set:
- Not specified.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No mention of an MRMC study being performed.
- No data on the effect size of human readers improving with AI vs. without AI assistance. This device is an accessory to a digital imaging system, not explicitly an "AI" device as typically understood today, so such a study would likely not be relevant in this context.
-
Standalone Performance Study (Algorithm Only Without Human-in-the-Loop):
- The document implies performance testing of the device itself ("bench testing"). However, it does not explicitly detail a "standalone" study in the context of an algorithm's performance without human interaction. The PoRT Cassette is a hardware component (reusable phosphor screen) and an associated software upgrade, not a standalone diagnostic algorithm.
-
Type of Ground Truth Used:
- Not specified. Given the nature of a portal imaging accessory for radiation therapy quality control, ground truth would likely involve comparison to known radiation fields, phantom measurements, or possibly expert assessment of image quality for therapy verification. However, this is speculative as the document is silent on this point.
-
Sample Size for the Training Set:
- Not specified. The document does not mention a "training set" in the context of machine learning, as this device is not presented as an AI/ML algorithm that would typically require such a set. Its "software upgrade" likely refers to functionality rather than an AI model.
-
How the Ground Truth for the Training Set Was Established:
- Not applicable/Not specified, as no training set (in the ML sense) is mentioned or implied.
Summary of Missing Information:
The provided 510(k) summary is extremely high-level and does not contain the detailed study results, acceptance criteria, or ground truth establishment methods that would typically be expected for a comprehensive clinical study report. It primarily serves to establish substantial equivalence based on intended use, technological characteristics, and operational characteristics, stating that "Bench testing and clinical validation demonstrate that the PoRT Cassette performs according to specifications" without providing those specifications or results.
Ask a specific question about this device
(398 days)
IXK
Ask a specific question about this device
(28 days)
IXK
Ask a specific question about this device
Page 1 of 1