K Number
K042475

Validate with FDA (Live)

Date Cleared
2004-11-10

(58 days)

Product Code
Regulation Number
866.5510
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only. VITROS Chemistry Products IgG Reagent is used to quantitatively measure immunoglobulin G (IgG) concentration in human serum and plasma.

For in vitro diagnostic use only. VITROS Chemistry Products IgA Reagent is used to quantitatively measure immunoglobulin A (IgA) concentration in human serum and plasma.

For in vitro diagnostic use only. VITROS Chemistry Products IgM Reagent is used to quantitatively measure immunoglobulin M (IgM) concentration in human serum and plasma.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of transferrin, C3, C4, IgG, IgA and IgM.

For in vitro diagnostic use only. VITROS Chemistry Products Protein Performance Verifiers I, II, and III are assayed controls used to monitor the performance of TRFRN, C3, C4, IgG, IgA and IgM Reagents on VITROS 5,1 FS Chemistry Systems.

Device Description

The VITROS 5,1 FS Chemistry System is a fully automated clinical chemistry analyzer intended for use in the in vitro determination of various analytes in human specimens (serum, plasma, urine, and cerebrospinal fluid). The VITROS 5,1 FS is designed for use with VITROS Chemistry Products MicroTip and Thin Film assays.

The system is comprised of four main elements:

  1. The VITROS 5,1 FS Chemistry System instrumentation, which provides automated use of the chemistry reagents.
  2. The VITROS Chemistry Products MicroTip range of liquid reagent products (in this case VITROS Chemistry Products IgG, IgA, and IgM Reagents and VITROS Chemistry Products Calibrator Kit 20, and VITROS Chemistry Products Protein Performance Verifiers I, II, and III), which are combined by the VITROS 5,1 FS Chemistry System to perform the VITROS IgG, IgA, and IgM assays.
  3. The VITROS Chemistry Products Thin Film range of dry products, which are dry, multilayered, analytical elements, coated on polyester supports.
  4. Common reagents used by multiple assays on the VITROS System (in this case, VITROS Chemistry Products FS Diluent Pack 2 (BSA/ Saline).

The VITROS System and reagents are designed specifically for use with the VITROS Chemistry Products range of products.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the VITROS Chemistry Products IgG, IgA, and IgM Reagents, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for substantial equivalence are primarily demonstrated through method comparison studies, exhibiting strong correlation and linear relationships with a legally marketed predicate device. The performance metrics focus on accuracy and analytical characteristics rather than diagnostic outcome metrics (like sensitivity/specificity for a particular disease).

MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (VITROS IgG, IgA, IgM assays)
Method Comparison (Correlation)High correlation to predicate device (implied r ≥ 0.97, given the presented values)IgG: r = 0.995
IgA: r = 0.999
IgM: r = 0.998
Method Comparison (Slope)Slope near 1 (implied, indicating proportional agreement)IgG: 0.98
IgA: 1.07
IgM: 1.02
Method Comparison (Intercept)Intercept near 0 (implied, indicating constant agreement)IgG: 91.2 mg/dL
IgA: -3.03 mg/dL
IgM: 1.1 mg/dL
Intended UseSame as predicate (quantitative measurement of immunoglobulins in human serum/plasma).Same as predicate.
MethodSame as predicate (Immunoturbidimetric).Same as predicate.
Reportable Range (IgG)Similar to predicate (e.g., within a clinically acceptable range).270 - 2700 mg/dL (Predicate: 140 - 4600 mg/dL)
Reportable Range (IgA)Similar to predicate.40 - 800 mg/dL (Predicate: 25 - 800 mg/dL)
Reportable Range (IgM)Similar to predicate.25 - 400 mg/dL (Predicate: 20 - 640 mg/dL)
Sample TypeSame as predicate (Human Serum).Same as predicate.
Reactive IngredientsSimilar (e.g., Goat vs. Rabbit antisera to human immunoglobulins).Goat antisera to human IgG, IgA, IgM (Predicate: Rabbit antisera)
InstrumentationCompatible with the overall VITROS 5,1 FS Chemistry System.VITROS 5,1 FS Chemistry System (Predicate: DADE BN ProSpec)

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states that "Equivalence to predicate(s) was demonstrated using commercially available reagents along with patient samples." However, it does not specify the sample size of these patient samples used in the method comparison studies.

The document also does not specify the country of origin of the data or whether the study was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The studies performed are analytical performance studies comparing the new device to a predicate device, not diagnostic performance studies requiring ground truth established by experts. The "ground truth" in this context is the measurement obtained from the predicate device.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided in the document. The studies are method comparison assays against a predicate device, not studies that would require expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for imaging or diagnostic tools where human interpretation plays a significant role in the outcome. The devices described are automated in vitro diagnostic assays for quantitative measurement.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The studies described are inherently standalone (algorithm only) in nature. The VITROS 5,1 FS Chemistry System is a fully automated clinical chemistry analyzer, and the reagents are designed for automated use. The performance characteristics of the assay itself are being evaluated, without human interpretation of raw data.

7. The Type of Ground Truth Used

The "ground truth" for the performance studies (method comparison) was the measurements obtained from the legally marketed predicate devices:

  • Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) assay
  • Dade Behring N Protein Standard SL (human)

For other analytical studies like precision, specificity, linearity, antigen excess, lower limit of detection, and expected values, the ground truth would be established through appropriate reference materials, calibrators, and experimental design according to recognized laboratory standards, though specific details are not provided in this summary.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided in the document. These are in vitro diagnostic reagents and an automated chemistry system that perform quantitative measurements based on well-established immunoturbidimetric principles. There is no "training set" in the context of machine learning algorithms for image analysis or other similarly complex data interpretation tasks. The development of such systems involves assay optimization and validation rather than algorithm training on a distinct dataset.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Chapter 1 - Summary Information

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

64 2475 The assigned 510(k) number is: _

  • Ortho-Clinical Diagnostics, Inc. 1. Submitter 100 Indigo Creek Drive name, Rochester, New York 14626-5101 address. (585) 453-4041 contact Contact Person: Marlene A. Hanna Email: mhanna1 @ocdus.jnj.com
  • Date 510(k) prepared: September 10, 2004 2. Preparation Date
  • Trade or Proprietary Name(s): Device 3. VITROS Chemistry Products IgG Reagent name(s) VITROS Chemistry Products IgA Reagent VITROS Chemistry Products IgM Reagent VITROS Chemistry Products Calibrator Kit 20 VITROS Chemistry Products Protein Performance Verifiers I, II, and III

Common Name(s): Immunoglobulin G (IgG) assay Immunoglobulin A (IgA) assay Immunoglobulin M (IgM) assay

Classification Name(s): Immunoglobulins A, G, M, D, and E immunological test system (866.5510): Class: II (performance standards).

Calibrator (862.1150): Calibrator Kit 20: Class II.

Quality Control material (assayed and unassayed) (862.1660): Class I (general controls). Since these devices (Protein Performance Verifiers I, II, and III) are assayed controls, they meet the reserved criteria under Section 510(1) of the Food, Drug, and Cosmetic Act.

{1}------------------------------------------------

510(k) Summary, Continued

  • a. The VITROS Chemistry Products IgG, IgA, and IgM Reagents and 4. Predicate Calibrator Kit 20 are substantially equivalent to the Dade Behring N Devices Antisera to Human Immunoglobulins (IgG, IgA, and IgM) and Dade Behring N Protein Standard SL (human).
    • b. The VITROS Chemistry Products Protein Performance Verifiers I, II, and III are substantially equivalent to the Dade Behring N/T Protein Controls S/L.
  • The VITROS 5,1 FS Chemistry System is a fully automated clinical chemistry 5. Device analyzer intended for use in the in vitro determination of various analytes in human description specimens (serum, plasma, urine, and cerebrospinal fluid). The VITROS 5,1 FS is designed for use with VITROS Chemistry Products MicroTip and Thin Film assays.

The system is comprised of four main elements:

    1. The VITROS 5,1 FS Chemistry System instrumentation, which provides automated use of the chemistry reagents. The VITROS 5,1 FS Chemistry System was cleared for market by a separate 510(k) premarket notification (K031924).
    1. The VITROS Chemistry Products MicroTip range of liquid reagent products (in this case VITROS Chemistry Products IgG, IgA, and IgM Reagents and VITROS Chemistry Products Calibrator Kit 20, and VITROS Chemistry Products Protein Performance Verifiers I, II, and III), which are combined by the VITROS 5,1 FS Chemistry System to perform the VITROS IgG, IgA, and IgM assays.
    1. The VITROS Chemistry Products Thin Film range of dry products, which are dry, multilayered, analytical elements, coated on polyester supports. The thin film products each have their own 510(k) clearance numbers and were cleared for market for use on the VITROS 5.1 FS Chemistry System through submission of information required by the ODE Guidance Document: "Data For Commercialization Of Original Equipment Manufacturer, Secondary and Generic Reagents For Automated Analyzers". The required information was provided in the VITROS 5,1 FS Chemistry System premarket notification (K031924).
    1. Common reagents used by multiple assays on the VITROS System (in this case, VITROS Chemistry Products FS Diluent Pack 2 (BSA/ Saline).

The VITROS System and reagents are designed specifically for use with the VITROS Chemistry Products range of products.

{2}------------------------------------------------

510(k) Summary, Continued

VITROS Chemistry Products IgG Reagent: For in vitro diagnostic use only. 6. Device(s) VITROS Chemistry Products IgG Reagent is used to quantitatively measure intended immunoglobulin G (IgG) concentration in human serum and plasma. use

VITROS Chemistry Products IgA Reagent: For in vitro diagnostic use only. VITROS Chemistry Products IgA Reagent is used to quantitatively measure immunoglobulin A (IgA) concentration in human serum and plasma.

VITROS Chemistry Products IgM Reagent: For in vitro diagnostic use only. VITROS Chemistry Products IgM Reagent is used to quantitatively measure immunoglobulin M (IgM) concentration in human serum and plasma.

VITROS Chemistry Products Calibrator Kit 20: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of transferrin, C3, C4, IgG, IgA and IgM.

VITROS Chemistry Products Protein Performance Verifiers I, II, and III: For in vitro diagnostic use only. VITROS Chemistry Products Protein Performance Verifiers I, II, and III are assayed controls used to monitor the performance of TRFRN, C3, C4, IgG, IgA and IgM Reagents on VITROS 5,1 FS Chemistry Systems.

7. Comparison

to predicate

device(s):

IgG, IgA, and IgM Reagents and Calibrator Kit 20

The VITROS Chemistry Products IgG, IgA, and IgM Reagents and VITROS Chemistry Products Calibrator Kit 20 are substantially equivalent to K860894, the Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM), and K012470, Dade Behring N Protein Standard SL (predicate devices) which were cleared by the FDA for IVD use.

{3}------------------------------------------------

IgG Reagent and Calibrators

The relationship between the VITROS IgG and the predicate device, determined by least squares linear regression, is:

$$\text{VITROS IgG} = 0.98\text{x} + 91.2 \text{ (mg/dL)}, 1$$

with a correlation coefficient of 0.995, where X is the Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) assay.

IgA Reagent and Calibrators

The relationship between the VITROS IgA and the predicate device, determined by least squares linear regression, is:

$$\text{VIIIROS IgA} = 1.07\text{x} - 3.03 \text{ (mg/dL)},$$

with a correlation coefficient of 0.999, where X is the Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) assay.

IgM Reagent and Calibrators

The relationship between the VITROS IgM and the predicate device, determined by least squares linear regression, is:

VITROS IgM = $1.02x + 1.1$ (mg/dL),

with a correlation coefficient of 0.998, where X is the Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) assay.

In addition to the above mentioned correlation study, studies were performed to determine the precision, specificity, linearity, antigen excess, lower limit of detection, and expected values of the VITROS IgG, IgA and IgM assays, (refer to the VITROS IgG, IgA, and IgM Reagent's Instructions for Use for summaries of the results of these studies).

{4}------------------------------------------------

Table 1 lists the characteristics of the assays performed using the VITROS IgG, IgA, and IgM assays and the Dade Behring N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) assay.

  • Table 1 lists the characteristics of the VITROS IgG, IgA, and IgM Reagent (new Table 1 devices) and the DADE IgG, IgA, and IgM assays (predicate device).
DeviceCharacteristicVITROS IgG, IgA and IgM(New device)DADE IgG, IgA, and IgM(Predicate device)
Intended UseFor in vitro diagnostic use only.VITROS Chemistry ProductsIgG , IgA, and IgM Reagents areused to quantitatively measureimmunoglobulin G (IgG),immunoglobulin A (IgA), andimmunoglobulin M(IgM)concentration in human serum.In vitro diagnostic reagents forthe quantitative determinationof immunoglobulins (IgG,IgA, and IgM) in humanserum as well as of IgG inhuman CSF.
MethodImmunoturbidimetricImmunoturbidimetric
ReportableRange:
IgG:270 - 2700 mg/dL140 - 4600 mg/dL
IgA:40 - 800 mg/dL25 - 800 mg/dL
IgM:25 - 400 mg/dL20 - 640 mg/dL
Sample TypeHuman SerumHuman Serum
ReactiveIngredients
IgG:Goat antisera to human IgGRabbit antisera to human IgG
IgA:Goat antisera to human IgARabbit antisera to human IgA
IgM:Goat antisera to human IgMRabbit antisera to human IgM
InstrumentationVITROS 5,1 FS ChemistrySystemDADE BN ProSpec

{5}------------------------------------------------

510(k) Summary, Continued

Performance Verifiers

The VITROS Chemistry Products Protein Performance Verifiers I, II, and III are substantially equivalent to the Dade Behring N/T Protein Controls SL (predicate device) which was cleared by the FDA (K012468) for IVD use.

Table 2 lists the similarities and differences of the device characteristics between the VITROS Protein Performance Verifiers I, II, and III with the predicate device, the Dade Behring N/T Protein Controls SL.

DeviceCharacteristicVITROS Protein PerformanceVerifiers(New Device)DADE Protein Controls(Predicate Device)
Intended UseFor in vitro diagnostic use only.VITROS Chemistry ProductsProtein Performance Verifiers I, II,and III are assayed controls used tomonitor the performance ofTRFRN, C3, C4, IgG, IgA andIgM Reagents on VITROS 5,1 FSChemistry Systems.For in vitro diagnostic use only.N/T Protein Controls S/L is for useas accuracy and precision assayedcontrols in the determination of thefollowing human serum proteinsby immunonephelometry withBN™ Systems: IgG1-4, IgA, IgM,C3c, C4, Transferrin, Albumin, α1-antitrypsin, α2-macroglobulin,Haptoglobin, α1-acid glycoprotein,Prealbumin, Hemopexin,Ceruloplasmin, RbP, Ig/L-chainlambda & Kappa, β2-microglobulin, soluableTransferrin Receptor (sTfR),Ferritin, IgE, Total protein.
MatrixProcessed human serum to whichinorganic salts, buffers, andpreservatives have been added.Human serum with stabilizers andpreservatives.
FormLiquidLiquid
Volume1 mL per vial1 mL per vial

Table 2

The data presented in the premarket notification provide a reasonable assurance Conclusions that the VITROS IgG, IgA, and IgM reagents, Calibrator Kit 20, and the VITROS Chemistry Products Protein Performance Verifiers are safe and effective for the stated intended uses and are substantially equivalent to the cleared predicate devices.

Equivalence to predicate(s) was demonstrated using commercially available reagents along with patient samples.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the left side of the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Public Health Service

NOV 1 0 2004

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Marlene A. Hanna Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY 14626

K042475 Re:

Trade/Device Name: VITROS Chemistry Products IgG Reagent VITROS Chemistry Products IgM Reagent VITROS Chemistry Products IgA Reagent VITROS Chemistry Products Calibrator Kit 20 VITROS Chemistry Products Protein Performance Verifiers I, II, and III Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, E immunological test system Regulatory Class: Class II Product Code: CFN, JIX, JJY Dated: September 10, 2004 Received: September 13, 2004

Dear Ms. Hanna:

We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to such additional controls. Existing major regulations affecting vour device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your devices in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your devices as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed

{7}------------------------------------------------

Page 2 -

predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific information about the application of labeling requirements to your devices, or questions on the promotion and advertising of your devices, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Totallary of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Robert H. Becker f

Robert L. Becker, Jr., MD, Ph.D Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use for the IgG assay
510(k) Number (if known):K042475
Device Name(s):VITROS Chemistry Products IgG ReagentVITROS Chemistry Products Calibrator Kit 20VITROS Chemistry Products Protein Performance Verifiers I, II, and III
Indications for Use:For in vitro diagnostic use only. VITROS Chemistry Products IgG Reagent is used to quantitatively measure immunoglobulin G (IgG) concentration in human serum and plasma.For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of transferrin, C3, C4, IgG, IgA and IgM.For in vitro diagnostic use only. VITROS Chemistry Products Protein Performance Verifiers I, II, and III are assayed controls used to monitor the performance of TRFRN, C3, C4, IgG, IgA and IgM Reagents on VITROS 5,1 FS Chemistry Systems.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D)

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Maria M Chan


Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K042475

{9}------------------------------------------------

Indications for Use for the IgA assay
---------------------------------------
510(k) Number (if known):K042475
Device Name(s):VITROS Chemistry Products IgA ReagentVITROS Chemistry Products Calibrator Kit 20VITROS Chemistry Products Protein Performance Verifiers I, II, and III
Indications for Use:For in vitro diagnostic use only. VITROS Chemistry Products IgA Reagent is used to quantitatively measure immunoglobulin A (IgA) concentration in human serum and plasma.For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of transferrin, C3, C4, IgG, IgA and IgM.For in vitro diagnostic use only. VITROS Chemistry Products Protein Performance Verifiers I, II, and III are assayed controls used to monitor the performance of TRFRN, C3, C4, IgG, IgA and IgM Reagents on VITROS 5,1 FS Chemistry Systems.

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (Per 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)


Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

ia m chan

Division Sign-Off

ാffice of In Vitro Diagnostic Device Evaluation and Safety

310(k) K042475

{10}------------------------------------------------

Indications for Use for the IgM assay

510(k) Number (if known):K042475
Device Name(s):VITROS Chemistry Products IgM ReagentVITROS Chemistry Products Calibrator Kit 20VITROS Chemistry Products Protein Performance Verifiers I, II, and III
Indications for Use:For in vitro diagnostic use only. VITROS Chemistry Products IgM Reagent is used to quantitatively measure immunoglobulin M (IgM) concentration in human serum and plasma.For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of transferrin, C3, C4, IgG, IgA and IgM.For in vitro diagnostic use only. VITROS Chemistry Products Protein Performance Verifiers I, II, and III are assayed controls used to monitor the performance of TRFRN, C3, C4, IgG, IgA and IgM Reagents on VITROS 5,1 FS Chemistry Systems.

ﺃ Prescription Use _ (Per 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

m cher
Division Sign-Off

Division Sign-Off

Office of In Vitro Dlagnostic Davice Evaluation and Safety

s10(k) K04 2475

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).