K Number
K042336

Validate with FDA (Live)

Date Cleared
2004-10-14

(45 days)

Product Code
Regulation Number
864.8625
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Meter Trax is an assayed whole blood control used in monitoring the precision of laboratory testing procedures for the analytes listed in this package insert.

Device Description

Meter Trax Control is a suspension of stabilized human red blood cell components with added constituents of animal origin, chemicals, stabilizers, and preservatives. This product is provided in liquid form for convenience.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Meter Trax Control." This document focuses on establishing substantial equivalence to a predicate device, rather than presenting a study design with acceptance criteria for a new, innovative device that would typically involve clinical performance evaluation.

Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set details, type of ground truth) are not applicable in this context as the submission is not for a performance study evaluating a diagnostic algorithm's accuracy against human interpretation or a specific clinical endpoint. The document is about a quality control material whose performance is assessed through stability studies.

Here's an attempt to answer the applicable questions based on the provided text:

Acceptance Criteria and Device Performance

1. A table of acceptance criteria and the reported device performance

The document focuses on stability claims as the primary performance metrics for the Meter Trax Control. The predicate device's performance already established the acceptable ranges for these analytes.

Performance MetricAcceptance Criteria (Claim)Reported Device Performance
Open Vial Stability
HematocritStable for 31 days at 2°C to 8°CClaims stability for 31 days at 2°C to 8°C
GlucoseStable for 31 days at 2°C to 8°C or 18°C to 30°CClaims stability for 31 days at 2°C to 8°C or 18°C to 30°C
HemoglobinStable for 31 days at 2°C to 8°C or 18°C to 30°CClaims stability for 31 days at 2°C to 8°C or 18°C to 30°C
Shelf Life Stability
All AnalytesStable for 180 days at 2°C to 8°C (until expiration date unopened)Claims stability for 180 days at 2°C to 8°C

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated. The stability studies would involve a series of measurements over time, but the number of unique "samples" (e.g., batches, vials) tested is not provided.
  • Data Provenance: Not explicitly stated, but assumed to be internal laboratory data generated by Bio-Rad Laboratories. The studies are prospective in nature for stability testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This device is a quality control material, not a diagnostic device requiring expert interpretation for ground truth. The stability of the analytes is measured directly using laboratory equipment.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a quality control material and does not involve AI or human readers for diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a quality control material, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Analytical Measurement: The "ground truth" for stability studies is defined by the analytical measurements of glucose, hemoglobin, and hematocrit over time, compared against their initial values and pre-defined acceptable ranges for precision. The stability is demonstrated if the analytes remain within these acceptable ranges for the stated duration.

8. The sample size for the training set

  • Not Applicable. This device uses stability studies for performance assessment, not a training set for an algorithm.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

K042336

Summary of Safety and Effectiveness : Meter Trax Control

1.0 Submitter

Bio-Rad Laboratories " 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 Fax: (949) 598-1557

Contact Person

Maria Zeballos Regulatory Affairs Specialist Telephone: (949) 598-1367

Date of Summary Proparation

August 25, 2004

2.0 Device Identification

Product Trade Name:Meter Trax Control
Common Name:Hematology and Pathology Devices(Hematology quality control mixture)
Classifications:Class II
Product Code:JPK
Regulation Number:21 CFR 864.8625

Device to Which Substantial Equivalence is Claimed 3.0

Meter Trax is an assayed whole blood control used in monitoring the precision of faboratory procedures for glucose, hemoglobin and hematocrit testing. This control is substantially equivalent to the following quality control materials that are currently in the market:

Meter Trax Bio-Rad Laboratories (formerly Hematronix, Inc.) Plano, Texas 75074 | 510 (k) Number: K904461

4.0 Description of Device

Meter Trax Control is a suspension of stabilized human red blood cell components with added constituents of animal origin, chemicals, stabilizers, and preservatives. This product is provided in liquid form for convenience.

Intended Use

Meter Trax is an assayed whole blood control used in monitoring the precision of laboratory procedures for glucose, hemoglobin and hematocrit testing.

{1}------------------------------------------------

Comparlson of the new device with the Predicate Device 5.0

Meter Trax Control is substantial equivalent to Meter Trax (K904461) currently in commercial distribution The similarities and differences are listed below.

CharacteristicsBio-RadMeter Trax Control(New Device)Bio-Rad Laboratories(formerly Hematronix, Inc.)Meter Trax(Predicate Device K904481)
MatrixHuman Whole Blood basedHuman Whole Blood based
StabilizersContains stabilizersDoes not contain stabilizers
FormLiquidLiquid
Storage(Unopened)2°C to 8°CUntil expiration date2°C to 8°CUntil expiration date
Open vial StabilityHematocrit: 31days at 2 to 8°CGlucose and Hemoglobin: 31days at 2 to 8°Cor 18 to 30°CHematocrit: 31 days at 2 to 8°CGlucose and Hemoglobin: 31days at 2 to 8°Cor 18 to 30°C
AnalytesContains the following parameters:GlucoseHemoglobinHematocritContains the following parameters:GlucoseHemoglobinHematocrit
Intended UseMeter Trax is an assayed whole blood control used inmonitoring the precision of laboratory procedures forglucose, hemoglobin and hematocrit testing.Meter Trax™ Control is intended for use as a wholeblood reference control for glucose, hemoglobin, andhematocot

Table 1. Similarities and Differences between new and predicate device.

1.0 STATEMENT OF SUPPORTING DATA

Stability studies have been performed to determine the open vial stability and shelf life for the Meter Trax Control. Product claims are as follows:

  • 1.1 Open vial Stability:
    Claims: Hernatocrit will be stable for 31 days at 2 to 8°C. Glucose and Hemoglobin will be stable for 31 days at 2 to 8°C or 18 to 30°C. 1

1.2 Shelf Life Stability

Claims: 180 days at 2 to 8℃

All supporting data is retained on file at Bio-Rad Laboratories.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

OCT 1 4 2004

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Maria Zeballos Regulatory Affairs Specialist Bio-Rad Laboratories 9500 Jeronimo Road Irvine, CA 92618-2017

K042336 Trade/Device Name: Meter Trax Control Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology Quality Control Mixture Regulatory Class: Class II Product Code: JPK, JJY Dated: August 25, 2004 Received: August 30, 2004

Dear Ms. Zeballos:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Robert L. Becker Jr.

Robert L. Becker, Jr., M.D., Ph. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

K042336

Device Name:

Meter Trax Control

Indications For Use:

Meter Trax is an assayed whole blood control used in monitoring the precision of laboratory testing procedures for the analytes listed in this package insert.

The following parameters are listed in the package insert:

  • · Glucose
  • Hemoglobin
  • · Hematocrit

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Xitro Diagnostic Devices (OIVD)

Josephine Bautista

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510 K042336

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.