K Number
K041799

Validate with FDA (Live)

Date Cleared
2004-08-25

(54 days)

Product Code
Regulation Number
866.5270
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Reagent is used to quantitatively measure C-reactive protein (CRP) in human serum and plasma. CRP is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with values of CRP that exceed 3 mg/L.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.

For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Performance Verifiers are assayed controls used to monitor performance of hsCRP Reagent on VITROS 5,1 FS Chemistry Systems.

Device Description

The VITROS 5.1 FS Chemistry System is a fully automated clinical chemistry analyzer intended for use in the in vitro determination of various analytes in human specimens (serum, plasma, urine, and cerebrospinal fluid). The VITROS 5,1 FS Chemistry System is designed for use with VITROS Chemistry Products MicroTip and Thin Film assays.

The system is comprised of four main elements:

  1. The VITROS 5.1 FS Chemistry System instrumentation, which provides automated use of the chemistry reagents.
  2. The VITROS Chemistry Products range of MicroTip assays, in this case the VITROS Chemistry Products hsCRP Reagent, VITROS Chemistry Products Calibrator Kit 17, VITROS Chemistry Products FS Calibrator 1, and the VITROS Chemistry Products hsCRP Performance Verifiers, which are combined by the VITROS 5,1 FS Chemistry System to perform the VITROS hsCRP assay.
  3. The VITROS Chemistry Products Thin Film range of products, which are dry, multilayered, analytical elements, coated on polyester supports.
  4. Common reagents used by multiple assays on the VITROS System (in this case, VITROS Chemistry Products FS Diluent Pack 2).

The VITROS System and reagents are designed specifically for use with the VITROS Chemistry Products range of products.

AI/ML Overview

Here's an analysis of the provided text regarding the VITROS Chemistry Products hsCRP Reagent and related components, focusing on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating acceptance criteria and then proving the device meets them with specific values in a separate study.

Instead, the summary reports results from a "correlation study" comparing the new device to the predicate. The stated goal for regulatory approval is "substantial equivalence."

Acceptance Criteria (Implied from Correlation Study)Reported Device Performance (vs. Predicate)
Correlation with Predicate Device: A strong linear relationship with the predicate device, evidenced by a high correlation coefficient and a slope close to 1 with a small y-intercept.Correlation Coefficient: 0.993 Linear Regression Equation: VITROS hsCRP assay = 0.954 X + 0.063 mg/L (where X is the predicate device)
Additional Performance Characteristics: Acceptable precision, expected values, linearity, and specificity.Studies were performed to determine precision, expected values, linearity, and specificity. (Details are referenced as being in the Instructions for Use, but not provided in this summary.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document mentions "patient samples" were used for the equivalence study, but does not specify the exact number of samples used for the correlation with the predicate device.
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable to this submission. The "ground truth" in this context is established by the measurements of the predicate device and the new device, not by expert consensus on visual assessment or diagnostic interpretation. This is a quantitative chemical assay, where "truth" is measured by the concentration of C-reactive protein.

4. Adjudication Method for the Test Set

This is not applicable as the study is a comparative quantitative measurement against a predicate device, not an interpretation task requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for image-based diagnostic aids where human readers interpret medical images. The device in question is an in-vitro diagnostic assay for C-reactive protein (CRP) measurement.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the primary study effectively functions as a standalone performance assessment. The comparison is between the new VITROS hsCRP assay (algorithm/device only) and the predicate N High Sensitivity CRP assay (device only). There is no "human-in-the-loop" component in the direct measurement comparison presented.

7. The Type of Ground Truth Used

The "ground truth" for the comparison study was effectively the measurements provided by the legally marketed predicate device (N High Sensitivity CRP assay). The new device's measurements were compared against these predicate measurements to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

This information is not provided in the summary. For an IVD assay, "training set" might refer to samples used during the development and optimization of the assay's reagents and calibration parameters. However, the document does not detail this.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the summary. Similar to point 8, if a "training set" was used during development, the ground truth would likely have been established through a highly accurate reference method or a carefully characterized set of control materials.

{0}------------------------------------------------

1.0 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: Ko41799

1.1 Submitter name, address, contact

Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, New York 14626-5101 Phone: (585) 453-3482 Fax: (585) 453-3368

Contact Person: Carey A. Mayo, M.S., RAC

1.2 Date of Preparation: July 1, 2004

1.3 Device Proprietary Name(s)

Trade Name(s) VITROS Chemistry Products hsCRP Reagent VITROS Chemistry Products Calibrator Kit 17 VITROS Chemistry Products FS Calibrator 1 VITROS Chemistry Products hsCRP Performance Verifier I, II, and III

Common Name C-Reactive Protein assay and controls

1.4 Classification Name(s)

C-Reactive Protein Immunological Test System: Class II (21 CFR 866.5270) Assayed Controls: Class I, reserved (21 CFR 862.1660)

1.5 Predicate device

The VITROS Chemistry Products hsCRP assay is substantially equivalent to the N High Sensitivity CRP assay (Dade Behring, Inc.).

The VITROS Chemistry Products hsCRP Performance Verifiers are substantially equivalent to the VITROS Chemistry Products CRP Performance Verifiers.

{1}------------------------------------------------

1.6 Device description

The VITROS 5.1 FS Chemistry System is a fully automated clinical chemistry analyzer intended for use in the in vitro determination of various analytes in human specimens (serum, plasma, urine, and cerebrospinal fluid). The VITROS 5,1 FS Chemistry System is designed for use with VITROS Chemistry Products MicroTip and Thin Film assays.

The system is comprised of four main elements:

    1. The VITROS 5.1 FS Chemistry System instrumentation, which provides automated use of the chemistry reagents. The VITROS 5,1 FS Chemistry System was cleared for market by a separate 510(k) premarket notification (K031924).
    1. The VITROS Chemistry Products range of MicroTip assays, in this case the VITROS Chemistry Products hsCRP Reagent, VITROS Chemistry Products Calibrator Kit 17, VITROS Chemistry Products FS Calibrator 1, and the VITROS Chemistry Products hsCRP Performance Verifiers, which are combined by the VITROS 5,1 FS Chemistry System to perform the VITROS hsCRP assay.
    1. The VITROS Chemistry Products Thin Film range of products, which are dry, multilayered, analytical elements, coated on polyester supports. The thin film products each have their own 510(k) clearance numbers and were cleared for market for use on the VITROS 5,1 FS Chemistry System through submission of information required by the ODE Guidance Document: "Data For Commercialization Of Original Equipment Manufacturer, Secondary and Generic Reagents For Automated Analyzers". The required information was provided in the VITROS 5,1 FS Chemistry System premarket notification (K031924).
    1. Common reagents used by multiple assays on the VITROS System (in this case, VITROS Chemistry Products FS Diluent Pack 2).

The VITROS System and reagents are designed specifically for use with the VITROS Chemistry Products range of products.

{2}------------------------------------------------

1.7 Device intended use

VITROS Chemistry Products hsCRP Reagent: For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Reagent is used to quantitatively measure Creactive protein (CRP) in human serum and plasma. CRP is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with values of CRP that exceed 3 mg/L.

VITROS Chemistry Products Calibrator Kit 17: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.

VITROS Chemistry Products hsCRP Performance Verifier I, II and III: For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Performance Verifiers are assayed controls used to monitor performance of hsCRP Reagent on VITROS 5,1 FS Chemistry Systems.

Comparison to predicate device: Reagent Kit and Calibrators 1.8

The VITROS Chemistry Products hsCRP Reagent, VITROS Chemistry Products Calibrator Kit 17 and VITROS Chemistry Products FS Calibrator 1 are substantially equivalent to the N High Sensitivity CRP Assay (Dade Behring, Inc.), which was cleared by FDA (K033908) for IVD use.

The relationship between the VITROS hsCRP assay and the predicate device, determined by least squares linear regression, is:

VITROS hsCRP assay = 0.954 X + 0.063 mg/L with a correlation coefficient of 0.993, where X is the predicate device.

In addition to the above mentioned correlation study, studies were performed to determine the precision, expected values, linearity, and specificity of the VITROS hsCRP assay, (refer to the VITROS Chemistry Products hsCRP Reagent Instructions for Use for summaries of the results of these studies).

{3}------------------------------------------------

The table below lists the characteristics of the VITROS hsCRP Assay and the N High Sensitivity CRP Assay (Dade Behring, Inc.).

DeviceCharacteristicVITROS hsCRP Assay(New device #1)N High Sensitivity CRP Assay(Predicate device #1)
Intended UseQuantitative measurementof C-reactive protein (CRP)Same
Basic principleLatex particle enhancedImmunoturbidimetryParticle enhancedimmunonephelometry
TraceabilityCRM 470CRM 470
ReagentsLiquid ready to useLiquid ready to use
InstrumentationVITROS 5,1 FS ChemistrySystemDade Behring BN ProSpecSystem
Sample typeSerum and plasma (Heparin)Serum and plasma (heparin andEDTA)

Comparison to predicate device: Performance Verifiers 1.9

VITROS Chemistry Products hsCRP Performance Verifiers are substantially equivalent to VITROS Chemistry Products CRP Performance Verifiers (predicate device) which were cleared by the FDA (K953197) for IVD use.

The table below lists the characteristics of the VITROS Chemistry Products hsCRP Performance Verifiers and the VITROS Chemistry Products CRP Performance Verifiers.

DeviceCharacteristicVITROS hsCRPPerformance Verifiers(New device #2)VITROS CRP PerformanceVerifiers(Predicate device #2)
Intended UseVITROS hsCRPPerformance Verifiers areassayed controls used tomonitor performance ofhsCRP Reagent on VITROS5,1 FS Chemistry Systems.VITROS CRP PerformanceVerifier is an assayed controlused to monitor performanceon VITROS ChemistrySystems.
MatrixA base matrix of humanplasma and plasma proteinsto which stabilizers andpreservative have beenadded.A base matrix of humanserum to which purifiedhuman C-reactive protein,inorganic salts andpreservatives have beenadded.
LevelsLow, Medium, and HighLow and High

{4}------------------------------------------------

1.10 Conclusions

The data presented in the premarket notification provide a reasonable assurance that the VITROS Chemistry Products hsCRP assay and the VITROS hsCRP Performance Verifiers are safe and effective for the stated intended uses and are substantially equivalent to the cleared predicate devices. Equivalence to the predicates was demonstrated using a commercially available assay along with patient samples.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird-like figure.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Carey A. Mayo, M.S., RAC Regulatory Affairs Associate Ortho-Clinical Diagnostics 100 Indigo Creek Drive Rochester, NY 14626-5101

AUG 2 5 2004

K041799 Re:

Trade/Device Name: VITROS Chemistry Products hsCRP Reagent VITROS Chemistry Products Calibrator Kit 17 VITROS Chemistry Products FS Calibrato VITROS Chemistry Products hsCRP Performance Verifiers I. II and III Regulation Number: 21 CFR 866.5270 Regulation Name: C-Reactive protein immunological test system Regulatory Class: Class II Product Code: NQD, JIT, JJX Dated: July 1, 2004 Received: July 2, 2004

Dear Ms. Mayo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

2.0 Indications for Use Statement

Page 1 of 1

510(k) Number (if known):K041799
Device Name:VITROS Chemistry Products hsCRP ReagentVITROS Chemistry Products Calibrator Kit 17VITROS Chemistry Products FS Calibrator 1VITROS Chemistry Products hsCRP Performance Verifiers I, II, and III
Indications for Use:For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Reagent is used to quantitatively measure C-reactive protein (CRP) in human serum and plasma. CRP is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with values of CRP that exceed 3 mg/L.& 3. For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.For in vitro diagnostic use only. VITROS Chemistry Products hsCRP Performance Verifiers are assayed controls used to monitor performance of hsCRP Reagent on VITROS 5,1 FS Chemistry Systems.
Prescription Use(Part 21 CFR 801 Subpart D)AND/OROver-The-Counter Use(21 CFR 807 Subpart C)
____________________

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Carol Benem
Division Sign Off

Division Sign-Off

Office of In Vitro Dlagnostic Device Evaluation and Safety

510(k) K041799

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).