(83 days)
Atramat ® Polyglycolic Acid Sutures are indicated for use as absorbable sutures in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures.
sorb-FAST ® Polyglycolic Acid Sutures are indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. These sutures are not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.
Atramat® Polyglycolic Acid and sorb-FAST® Polyglycolic Acid are synthetic absorbable sterile surgical sutures composed of a synthetic polyglycolic acid polymer.
Atramat® Polyglycolic acid and sorb-FAST® Polyglycolic Acid meet all requirements as described in the United States Pharmacopeia (USP) monograph for Absorbables Surgical Sutures. These products are offered as monofilament or multifilament and it is offered uncoated or coated with polycaprolactone and calcium sterarate, it could also be undyed or dyed with D&C Violet No. 2.
Sizes offered are U.S.P. 8-0 through 2 ( metric equivalent 0.4 through 5).
Materials: Polyglycolic acid. USP & Polycaprolactone and Calcium Stearate
The provided text describes a 510(k) summary for Atramat® Polyglycolic Acid and sorb-FAST® Polyglycolic Acid surgical sutures. This submission is for a medical device and therefore the acceptance criteria and study information provided are related to demonstrating substantial equivalence to a predicate device, rather than the performance of a diagnostic or AI-driven device. As such, the typical structure for reporting on acceptance criteria for diagnostic devices (like those involving AI) does not directly apply here.
However, I can extract the information relevant to device performance and the "study" conducted to demonstrate equivalence, framed within the context of a 510(k) submission for a non-AI surgical suture.
Acceptance Criteria and Reported Device Performance
For both Atramat® Polyglycolic Acid and sorb-FAST® Polyglycolic Acid, the core acceptance criteria revolve around meeting or exceeding the performance requirements of the United States Pharmacopeia (USP). This is the key "performance" standard for these devices.
| Acceptance Criteria Category | Criteria for Atramat® Polyglycolic Acid | Reported Device Performance for Atramat® Polyglycolic Acid | Criteria for sorb-FAST® Polyglycolic Acid | Reported Device Performance for sorb-FAST® Polyglycolic Acid |
|---|---|---|---|---|
| General Performance Standard | Meet or exceed the performance requirements for "Absorbable Surgical Suture" as defined in the official Monograph of the United States Pharmacopoeia (USP) 24. | Meets or exceeds the performance requirements for "Absorbable Surgical Suture" as defined in the official Monograph of the United States Pharmacopoeia (USP) 24. | Meet or exceed the performance requirements for "Absorbable Surgical Suture" as defined in USP 26 monographs. | Meets or exceeds the performance requirements for "Absorbable Surgical Suture" as defined in USP 26 monographs. |
| Tensile Strength Retention | Implicitly defined by USP 24, but specific percentages are given as characteristic for comparison to predicate. | Retains approximately 70% of its original tensile strength at two weeks post implantation, with approximately 50% remaining at three weeks. | Implicitly defined by USP 26, but specific percentages are given as characteristic for comparison to predicate. | Sorb-FAST® retains approximately 50% of its original tensile strength at one week post implantation. |
| Absorption Rate | Implicitly defined by USP 24, but specific timeframe is given as characteristic for comparison to predicate. | Absorption of the suture is essentially complete between 60 and 90 days. | Implicitly defined by USP 26, but specific timeframe is given as characteristic for comparison to predicate. | Absorption of the suture complete in 42 days. |
| Suture Diameter | Meet USP requirements. | Meets or exceeds the performance requirements (USP 24). | Meet USP requirements. | Meets or exceeds the performance requirements (USP 26). |
| Suture Length | Meet USP requirements. | Meets or exceeds the performance requirements (USP 24). | Meet USP requirements. | Meets or exceeds the performance requirements (USP 26). |
| Knot Pull Tensile Strength | Meet USP requirements. | Meets or exceeds the performance requirements (USP 24). | Meet USP requirements. | Meets or exceeds the performance requirements (USP 26). |
| Needle Attachment Strength | Meet USP requirements. | Meets or exceeds the performance requirements (USP 24). | Meet USP requirements. | Meets or exceeds the performance requirements (USP 26). |
Study Information (Non-Clinical Testing for Substantial Equivalence)
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document states "Non-clinical testing was conducted on the subject devices to prove conformance to the requirements of USP standards". It does not specify the exact sample sizes used for each test (e.g., how many sutures were tested for tensile strength or absorption rate).
- Data Provenance: Not specified, but generally, such non-clinical testing would be performed by the manufacturer (Internacional Farmaceutica S. A. de C.V.) in Mexico, or by a contract testing laboratory. It is retrospective in the sense that the testing was completed before the 510(k) submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable in the context of this 510(k) submission. "Ground truth" established by experts is relevant for diagnostic performance studies, often comparing human interpretation to a definitive answer. For medical devices like surgical sutures, "ground truth" is defined by established engineering and material science standards (e.g., USP monographs) rather than expert consensus on a diagnostic outcome. The "experts" involved would be those performing the standardized physical and chemical tests according to USP guidelines.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods are used to resolve discrepancies among multiple expert readers in diagnostic studies. Here, the "test results" are objective measurements against defined standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a 510(k) submission for a surgical suture, not a diagnostic device or an AI-powered system that assists human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device. The "performance" is the physical and chemical properties of the suture itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The "ground truth" is adherence to United States Pharmacopeia (USP) standards for absorbable surgical sutures. These are objective, quantitative specifications for material composition, physical dimensions, tensile strength, and absorption rates. For Atramat® Polyglycolic Acid, the standard is USP 24, and for sorb-FAST® Polyglycolic Acid, it is USP 26.
-
The sample size for the training set:
- Not applicable. There is no "training set" as this is not an AI/machine learning device. The product development and manufacturing processes would involve internal testing and quality control, but this is distinct from a "training set" in an algorithm development context.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set. The "ground truth" for the device's performance is established by the well-defined and published USP monographs.
{0}------------------------------------------------
APR 2 9 2004
Exhibit #2 K040282
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR§807.92.
The assigned 510(k) number is: K040282
1. Submitter's identification:
Internacional Farmaceutica S. A. de C.V. Carreteraco 44 Col. Parque San Andrés Coyoacan 04040 México D.F. Contact: Alejandro von Mohr, General Director Telephone: (52) 55 44 87 60 to 62 ext. 227, 219 Fax: (52) 55 49 42 34
Date Summary prepared: April 23, 2004
2. Name of the Device:
Trade Name: Atramat ® Polyglycolic Acid sorb-FAST ® Polyglycolic Acid
Common or Usual Name:
Polyglycolic Acid Surgical Suture
Classification Name:
Atramat ® Polyglycolic Acid and sorb-FAST ® Polyglycolic Acid based sutures are not specifically categorized or defined in 21CFR, parts 800-1299, the responsible device panel provided the following class, classification panel and product code for this product.
Device Class: Class II
Classification Panel: General & Plastic Surgery Devices Panel
Product Code: 79GAM
3. Predicate Device Information:
{1}------------------------------------------------
Atramat @ Polyglycolic Acid Absorbable suture is substantially equivalent to the following absorbable suture marketed by Davis & Geck:
DEXON II ® Polyglycolic Acid Synthetic Absorbable Surgical Sutures with Polycaprolactone Coating System. (K972566) and Ethicon Absorbable Poly (Llactide/glycolide) Surgical Sutures with Poliglactina 370 and Calcium Stearatc Coating System (K964345).
With respect to substantial equivalence, the comparation device represents a virtually identical device. Materials, packaging, sterilization method, sizes, multi and monofilament, dyed and undyed as well as functional characteristics (absorption rate, strength, diameter, etc) equivalence can also be drawn with respect to the design, material composition, performance and intended use, Atramat ® and Dexon II both meet or exceed the performance requirements of USP 24.
sorb-FAST® Polyglycolic Acid Absorption sutures are substantially equivalent to the following absorbable suture marketed by Ethicon (K962480 and K944110), Victyl Rapide® is prepared with a mixture of copolymer of glycolide and L-lactide with Poliglactin 370 and Calcium Stearate Coating System.
With respect to substantial equivalence, the predicate device represents a virtually identical device. Materials, packaging, sterilization method, sizes, multifilament, dyed and undyed as well as functional characteristics. Equivalence also demonstrated in material composition, performance, and intended use. Sorb-FAST® and Vicryl Rapide® both meet or exceed the performance requirements of USP 26.
4. Device Description:
Atramat® Polyglycolic Acid and sorb-FAST® Polyglycolic Acid are synthetic absorbable sterile surgical sutures composed of a synthetic polyglycolic acid polymer.
Atramat® Polyglycolic acid and sorb-FAST® Polyglycolic Acid meet all requirements as described in the United States Pharmacopeia (USP) monograph for Absorbables Surgical Sutures. These products are offered as monofilament or multifilament and it is offered uncoated or coated with polycaprolactone and calcium sterarate, it could also be undyed or dyed with D&C Violet No. 2.
Sizes offered are U.S.P. 8-0 through 2 ( metric equivalent 0.4 through 5), The following table illustrates the sizes offered.
| USP Size(MetricSize) |
|---|
| 8.0 (0.4) |
| 7.0 (0.55) |
| 6.0 (0.7) |
{2}------------------------------------------------
| 5.0 | (1) |
|---|---|
| 4.0 | (1.5) |
| 3.0 | (2) |
| 2.0 | (3) |
| 0 | (3.5) |
| 1 | (4) |
| 2 | (5) |
Materials: Polyglycolic acid. USP & Polycaprolactone and Calcium Stearate
-
- Intended Use:
Atramat ® Polyglycolic Acid Sutures are indicated for use as absorbable sutures in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures.
- Intended Use:
sorb-FAST ® Polyglycolic Acid Sutures are indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. These sutures are not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.
6. Comparison to Predicate Devices:
| Atramat® Polyglycolic acid | Predicate Device | |
|---|---|---|
| Intended Use | General soft tissue approximationand/or ligation, including use inophthalmic procedures, but not foruse in cardiovascular andneurological procedures. | Same Intended Use |
| Suture Material | Composed of glycolide polymer | Davis & Geck:composed of glycolidepolymer |
| SutureCharacteristics | Retains approximately 70% of itsoriginal tensile strength at twoweeks post implantation, withapproximately 50% remaining atthree weeks. Absorption of thesuture is essentially completebetween 60 and 90 days. | Same |
| SterilizationMethod | Ethylene Oxide gas | Same |
| Suture Diameter,Suture length, knotpull tensile,Strength and needleattachment strength | Meets or exceeds the performancerequirements for “AbsorbableSurgical Suture” as defined in theofficial Monograph of the UnitedStates Pharmacopoeia (USP) 24 | Meet USPrequirements |
| Packaging | Dry packaged in Aluminum Foil and | Same or equivalent |
{3}------------------------------------------------
| ్రాల్యేy彩 | . | |
|---|---|---|
| ------------------- | --- | -- |
| CARD O CONSULTION LA CORPORALand and the charge of the first and the first of the first the first the first the first the first the first the collection | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A |
|---|---|
| was such who " ".------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.A Property of Children------------------------------------------------------------------------------------------------------------------------------------------------------------------------------I | montorSection of Property of Children your Printers of Probeccapes------------------------------------------------------------------------------------------------------------------------------------------------------------------------------I |
| Sorb-FAST® Polyglycolic Acid | Predicate Device | |
|---|---|---|
| Intended Use | Indicated only for use as absorbablesutures in superficial general softtissue approximation of the skin andmucosa, where only short termwound support, (7-10 days) isrequired, but not for use in ligation,ophthalmic, cardiovascular andneurological procedures. | Same Intended Use |
| Suture Material | Composed of glycolide polymer | Vicryl Rapidecomposed ofcopolymer of glycolideand L-lactide |
| SutureCharacteristics | Sorb-FAST® retains approximately50% of its original tensile strength atone week post implantation.Absorption of the suture completein 42 days. | Same |
| SterilizationMethod | Ethylene Oxide gas | Same |
| Suture Diameter,Suture length, knotpull tensile,Strength and needleattachment strength | Meets or exceeds the performancerequirements for "AbsorbableSurgical Suture" as defined in USP26 monographs. Pharmacopeia(USP) 26 | Meet USPrequirements |
| Packaging | Dry packaged in Aluminum Foil andpolyester tear open | Same or equivalentmanner |
7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Non-clinical testing was conducted on the subject devices to prove conformance to the requirements of USP standards to demonstrate substantial equivalence to the predicate device. Physical properties and functionality testing assured the safety and effectiveness of the subject device within its intended uses.
Results of the non-climical testing demonstrate conformance with the USP standards and requirements for Absorbable surgical suture.
8. Discussion of Clinical Tests Performed:
No clinical trials were conducted.
{4}------------------------------------------------
9. Conclusions:
Based on the technological characteristics and physical properties of the polyglycolic acid sutures , the description, the intended use of the device and conformance with the following performance standards like:
USP 26
ISO 9002 AND EN 46002
FDA Guidance for Surgical Suture 510(k)
Internacional Farmaceutica believes that the subject devices demonstrate a substantial equivalence to the predicate devices and are safe and effective for their intended use.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or abstract shapes.
APR 2 9 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Internacional Farmaceutica S.A. de C.V. c/o Ms. Carolann Kotula MDI Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K040282
Trade/Device Name: Atramat and SuperSorb Polyglycolic Acid Sutures Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable polyglycolic surgical suture Regulatory Class: II Product Code: GAM Dated: February 5, 2004 Received: February 6, 2004
Dear Ms. Kotula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are varior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 Cl·R Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Ms. Carolann Kotula
This letter will allow you to begin marketing your device as described in your Section 510(k) rmarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Exhibit 1
510(k) Number (if known): K040282
Device Name: Atramat and SuperSorb Polyglycolic Acid Sutures
Indications For Use:
Atramat ® Polyglycolic Acid Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures.
sorb-FAST ® Polyglycolic Acid Sutures are indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular and neurological procedures.
Prescription Use × (Per 21 CFR 801.109)
Over-The Counter Use (Optional Format 1-2-96) OR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
510(k) Number K040282
§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.
(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.