K Number
K040146

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-05-04

(103 days)

Product Code
Regulation Number
864.7415
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Interlab Acid Hemoglobin Electrophoresis is a qualitative test system intended for the electrophoretic separation of hemoglobins to confirm the identity of clinically relevant hemoglobins such as A, F, S and C. It is to be used in conjunction with the Interlab Alkaline Hemoglobin Electrophoresis test kit. The Acid Hemoglobin test kit employs cellulose acetate supported on Mylar® as the medium and is for in vitro diagnostic use. The test can be automated on the Microtech 672 PC and Microtech 648 ISO instruments.

Device Description

The InterLab Acid Hemoglobin test kits for the electrophoretic separation of hemoglobin in whole blood and are intended for In-Vitro diagnostic use only. The InterLab Acid Hemoglobin Electrophoresis Test Devices provide qualitative identification of abnormal hemoglobin bands visualized by staining of the fractions. The principle of hemoglobin electrophoresis is based upon the visualization of specific hemoglobin bands following separation by electrophoresis. Dilutions of a patient's specimen are placed on separate tracks (fingers) on a cellulose acetate slide six fingers shaped, and the major hemoglobin groups are separated by electrophoresis. The migration rate depends on the temperature, pH, ionic force of the solution and proportions of the reactants. After electrophoresis, the slide is processed to remove excess soluble materials through a washing step. Fractionated hemoglobins are stained. The excess of stain is removed by a destaining step. The slides visually read to identify the bands present.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the InterLab Acid Hemoglobin Electrophoresis device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Qualitative identification of abnormal hemoglobin fractions."The InterLab Acid Hemoglobin test demonstrated equivalent band patterns to the reference tests with no false negative or false positive bands observed by visual inspection for the seventy-one (71) normal and pathological samples evaluated. This study resulted in a 100% agreement to the reference methods for the observed bands at both sites."
Detection of specific hemoglobin bands at given concentrations (Analytical Sensitivity)."The InterLab Hemoglobin Test systems will detect hemoglobin bands at concentrations greater than 2.4 g/L for HbA, 1.9 g/L for Hb F, 1.6 g/L for HbS, and 1.54 g/L for HbC."
Precision (Within Slide) - qualitatively identical patterns with correct band identification and no false fractions."The patterns were visually inspected and found to be qualitatively identical. In each lane the bands were correctly identified. No false fractions or aberrations were observed." (Typical results provided for specific samples for SRE158K and SRE164K kits).
Precision (Slide to Slide) - qualitatively identical patterns with correct band identification and no false fractions."The patterns were visually inspected and found to be qualitatively identical. In each lane the bands were correctly identified. No false fractions or aberrations were observed." (Typical results provided for specific samples for SRE158K and SRE164K kits).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Comparative Study (Primary Equivalence Study): 71 samples (combined normal and pathological).
  • Data Provenance: Samples were obtained from "normal and suspected pathological patients submitted for routine testing to the clinical or hospital laboratory." The studies were run at two clinical facilities. The raw geographical origin of the samples (e.g., country) is not specified. The data is retrospective as it involved samples "submitted for routine testing."
  • Sample Size for Analytical Sensitivity: Not explicitly stated as a single number. Samples were "serially diluted" to determine detection limits.
  • Sample Size for Precision (Within Slide): 3 different serum samples, each run in replicate (12 or 8 replicates per sample, depending on the kit).
  • Sample Size for Precision (Slide to Slide): 3 different serum samples, each run in triplicate on multiple separate slides (30, 33, or 36 replicates per sample, depending on the kit).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not explicitly stated. The determination of ground truth relied on "the laboratory's routine commercially available test system following the laboratory's procedure." This implies trained laboratory personnel interpreted the results of these reference methods, but specific numbers or qualifications of individual experts are not provided.
  • Qualifications of Experts: Not specified beyond the implication that they are trained laboratory personnel operating "routine commercially available test system" in clinical or hospital laboratories.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly described. The text states "Comparison was accomplished exclusively by visual inspection of the electrophoretic patterns" against reference methods. For the primary equivalence study, there was a "100% agreement to the reference methods for the observed bands at both sites," suggesting a direct comparison or concordance rather than a separate adjudication process beyond the visual comparison.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done. This device is an in-vitro diagnostic test, not an AI-assisted diagnostic tool for human readers. The assessment focused on the device's performance against established laboratory methods.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • This device is not an algorithm, but a laboratory test system. The "InterLab Acid Hemoglobin Electrophoresis" system involves a physical process of electrophoresis, followed by visual inspection of stained bands. Therefore, a standalone algorithm performance study is not applicable. The device itself, when operated, performs the separation and produces the bands that are then visually interpreted. The performance reported (e.g., 100% agreement) is of the combined system's output.

7. The Type of Ground Truth Used

  • Ground Truth Type: For the primary equivalence study, the ground truth was established by comparison to existing "commercially available test system[s]" (reference methods) in the clinical laboratory setting.
    • Site one's reference: "HPLC analysis of blood samples displays values of hemoglobin fractions that are outside the reference ranges...electrophoresis is required as an additional confirmatory test."
    • Site two's reference: "commercially available Alkaline Hemoglobin Electrophoresis system to screen the patient samples. Fractions identified as potentially abnormal are run on the Acid Hemoglobin test for confirmation."

8. The Sample Size for the Training Set

  • The document implies that this is a 510(k) submission for a device, and the studies described are validation studies for device performance, not training for a machine learning algorithm. Therefore, there is no explicit "training set" mentioned in the context of the device's development or evaluation as described here.

9. How the Ground Truth for the Training Set Was Established

  • Since there is no explicit "training set" for an algorithm, this question is not applicable in the context of the provided document. The device operates based on the physical-chemical principles of electrophoresis, not on a machine learning model trained on data.

{0}------------------------------------------------

K040146

MAY - 4 2004

ATTACHMENT 8 1 8 1

SUMMARY OF SAFETY AND EFFECTIVENESS

{1}------------------------------------------------

STATEMENT OF STAFETY AND EFFECTIVENESS

The sponsor, InterLab srl. has developed, manufactured, and tested under Good Laboratory Practices guidelines, in vitro diagnostic (IVD) devices for qualitative identification of abnormal hemoglobin fractions for the electrophoretic separation of whole blood. The device trade name is the InterLab Acid Hemoglobin Electrophoresis having FDA assigned name: Electrophoretic Abnormal Hemoglobin Analysis systems, and a classification as a Class II device per 21 CFR Sec. 864.7415, with product code: GKA.

The InterLab Acid Hemoglobin test kits for the electrophoretic separation of hemoglobin in whole blood and are intended for In-Vitro diagnostic use only. The InterLab Acid Hemoglobin Electrophoresis Test Devices provide qualitative identification of abnormal hemoglobin bands visualized by staining of the fractions. The principle of hemoglobin electrophoresis is based upon the visualization of specific hemoglobin bands following separation by electrophoresis. Dilutions of a patient's specimen are placed on separate tracks (fingers) on a cellulose acetate slide six fingers shaped, and the major hemoglobin groups are separated by electrophoresis. The migration rate depends on the temperature, pH, ionic force of the solution and proportions of the reactants. After electrophoresis, the slide is processed to remove excess soluble materials through a washing step. Fractionated hemoglobins are stained. The excess of stain is removed by a destaining step. The slides visually read to identify the bands present.

Studies were performed using samples from both normal and suspected pathological patients submitted for routine testing to the clinical or hospital laboratory. The studies were run at 2 facilities and the samples were tested with the InterLab Acid Hemoglobin test system run on the Microtech instrument and compared to the laboratory's routine commercially available test system following the laboratory's procedure.

Standard procedure at site one for Acid Hb is based on the qualitative evaluation of the electrophoretic results, when HPLC analysis of blood samples displays values of hemoglobin fractions that are outside the reference ranges (HbA2, HbF ) or indicates the presence of abnormal fractions, electrophoresis is required as an additional confirmatory test. Abnormal results obtained by HPLC analysis, including either elevated concentration of normal hemoglobins (e.g. HbA2 or HbF) or appearance of variants, namely hemoglobins not normally present in red blood cells (e.g. HbS), require further investigation to assess the type of additional Hemoglobin(s) present in the sample.

Site two uses a commercially available Alkaline Hemoglobin Electrophoresis system to screen the patient samples. Fractions identified as potentially abnormal are run on the Acid Hemoglobin test for confirmation. Comparison was accomplished exclusively by visual inspection of the electrophoretic patterns. Complete agreement between the two systems was obtained.

The InterLab Acid Hemoglobin test demonstrated equivalent band patterns to the reference tests with no false negative or false positive bands observed by visual inspection for the seventy-one (71) normal and pathological samples evaluated. This study resulted in a 100% agreement to the reference methods for the observed bands at both sites. These results demonstrate that the InterLab Acid Hemoglobin test system is substantially equivalent to the commercially available reference tests.

Analytical Sensitivity

The InterLab Hemoglobin Test systems will detect hemoglobin bands at concentrations greater than 2.4 g/L for HbA, 1.9 g/L for Hb F, 1.6 g/L for HbS, and 1.54 g/L for HbC. Samples were serially diluted and run on the intert ab method. The patterns were visually inspected to see

{2}------------------------------------------------

when the bands were no longer visible. That sample's dilution was back calculated to determine its concentration.

Precision (Within Slide)

Three different serum samples were run in replicate on single slide for each instrument system. The patterns were visually inspected and found to be qualitatively identical. In each lane the bands were correctly identified. No false fractions or aberrations were observed. The following typical results were obtained for the kits.

SRE158K

SAMPLEN.Detected bands
B12HbF, HbA, HbC
C12HbA, HbC
A12HbF, HbA, HbS

SRE164K

SAMPLEN.Detected bands
AA8HbA-HbS
BB8HbF-HbA
CC8HbA-HbC

Precision (Slide to Slide)

Three different serum samples were run in triplicate on multiple separate slides for each instrument system. The patterns were visually inspected and found to be qualitatively identical. In each lane the bands were correctly identified. No false fractions or aberrations were observed. The following typical results were obtained for the kits.

SRE158K

SAMPLEN.Detected bands
A30HbF, HbA, HbS, HbC
S30HbA, HbS
F30HbF, HbA

SRE164K

SAMPLEN.Detected bands
XX33HbF, HbA1, HbS, HbC
YY33HbF, HbA
ZZ36HbA, HbS

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Public Health Service

ood and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY - 4 2004

InterLab S.R.L. Mr. Gary Lehnus Trouble Shooter Consulting, Inc. 150 Cherry Lane Road East Stroudsburg, PA 18301

Re: K040146

Trade/Device Name: Interlab Acid Hemoglobin Test Systems By Eletrophoresis Regulation Number: 21 CFR 864.7415 Regulation Name: Abnormal Hemoglobin assay Regulatory Class: Class II Product Code: GKA Dated: March 30, 2004 Received: April 9, 2004

Dear Mr. Lehnus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Joseph L. Arallett

Joseph L. Hackett, Ph.D. Acting Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): __ K040146

Interlab Acid Hemoglobin Test Systems Device Name: _________________________________________________________________________________________________________________________________________________________________

by Electrophoresis

Indications For Use:

The Interlab Acid Hemoglobin Electrophoresis is a qualitative test system intended for the electrophoretic separation of hemoglobins to confirm the identity of clinically relevant hemoglobins such as A, F, S and C. It is to be used in conjunction with the Interlab Alkaline Hemoglobin Electrophoresis test kit. The Acid Hemoglobin test kit employs cellulose acetate supported on Mylar® as the medium and is for in vitro diagnostic use. The test can be automated on the Microtech 672 PC and Microtech 648 ISO instruments.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Josephine Bautista

Division &

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K040/46

x Prescription Use: (Per 21 CFR 801.109)

OR Over-The-Counter Use: (Optional Format 1-2-96)

§ 864.7415 Abnormal hemoglobin assay.

(a)
Identification. An abnormal hemoglobin assay is a device consisting of the reagents, apparatus, instrumentation, and controls necessary to isolate and identify abnormal genetically determined hemoglobin types.(b)
Classification. Class II (special controls). A control intended for use with an abnormal hemoglobin assay is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.