K Number
K040121

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-03-17

(57 days)

Product Code
Regulation Number
890.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine.

The indications of the DermaWave 5000 Combi Max System also include the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate postsurgical stimulation of calf rnuscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissucs, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound.

The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.

Device Description

The DermaWave 5000 Combi Max System is comprised of the following main components:

  • A system console including software and control electronics. .
  • A control and display panel. .
  • · Delivery device accessories including patient cables and electrodes, ultrasound cables, soundheads, and infrared heating accessories.
AI/ML Overview

The DermaWave 5000 Combi Max System is a medical device incorporating multiple modalities, including ultrasound, electrical stimulation, and infrared heating. The provided document describes its 510(k) submission, asserting its substantial equivalence to previously cleared predicate devices rather than detailing a specific clinical study with acceptance criteria in the traditional sense for a novel device.

The "study" referenced in the document is a comparative analysis of technical characteristics and features, and, for the infrared component, a heating characteristics test.

Here's a breakdown of the requested information based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria (Implied)Reported Device Performance
General Indications for Use: Similar to predicate devices in physical medicine, general and plastic surgery, and neurology applications.The DermaWave 5000 Combi Max System shares the same general indications for use in physical medicine, general and plastic surgery, and neurology applications.
Basic Characteristics & Features: Similar to predicate devices.The DermaWave 5000 Combi Max System shares the same or similar basic characteristics and features as the predicate devices (Chatanooga Group Inc., Dynatronics Corporation, Excel Tech). It utilizes identical accessories as MedX Health products (MedX LCT 100, MedX Duolight 500, MedX 600).
Infrared Heating Characteristics: Elevate tissue temperature above 40°C and below 45°C for a minimum 10-minute period.The DermaWave 100 Accessory has almost identical heating characteristics as the MedX line of infrared lamp accessories and falls within the recommended guidelines for such devices in raising the temperature of tissue to no more than 45°C in continuous operation over a set time period (implying the 10-minute duration is also met). Comparative temperature charts were included in Section 10 of the submission.
Safety and Effectiveness: Demonstrate safety and effectiveness for indicated applications.Validation documentation and a comparison of technical characteristics and features were provided to demonstrate that the DermaWave 5000 Combi Max System is safe and effective when indicated in specific applications.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a traditional clinical test set with human subjects. For the infrared component, the "test set" would be the DermaWave 100 Accessory undergoing heating characteristic measurements. The sample size for this specific test is not detailed (e.g., number of devices tested, number of measurement points).
    • Data Provenance: The document does not specify the country of origin for any specific test data. It refers to "widely accepted clinical data" for infrared heating guidelines. The submission itself is from a US company (DermaWave LLC, Loxahatchee, FL). The "Comparative temperature charts" were included in Section 10 of the submission, implying generated data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k) submission. The "ground truth" for the infrared heating characteristic was established by "widely accepted clinical data" (i.e., established medical guidelines for therapeutic heating devices) rather than a panel of experts specifically adjudicating measurements for DermaWave's accessory. For substantial equivalence, the "ground truth" essentially is the "predicate device's performance and indications."
  3. Adjudication method for the test set:

    • Not applicable as no expert adjudication process is described for a clinical test set. The evaluation primarily centered on comparing the device's technical specifications and heating characteristics to established guidelines and predicate devices.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted diagnostic device. The DermaWave 5000 Combi Max System is a therapeutic physical medicine device. Therefore, no MRMC study, or assessment of human reader improvement with AI assistance, was conducted or applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, to a limited extent for specific components. The assessment of the DermaWave 100 Accessory's heating characteristics described in the document is a standalone performance test. It evaluates the device's ability to achieve specific temperature ranges independently of human interaction during the measurement process, adhering to pre-defined clinical guidelines. However, this is not a "standalone algorithm" in the context of an AI device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the infrared heating component, the ground truth was established medical/clinical guidelines for therapeutic heating devices: "above 40°C and below 45°C for a minimum 10 minute period."
    • For the overall device, the "ground truth" for substantial equivalence is the performance and indications for use of the legally marketed predicate devices.
  7. The sample size for the training set:

    • Not applicable. This device does not use machine learning or AI that would require a "training set." The closest analogy might be the data used to develop the predicate devices, but this is not discussed.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Attachment 14

K040121
1 OF 3

510(k) Summary Statement for the DermaWave 5000 Combi Max System

1. General Information

Submitter:DermaWave LLC15693 83rd Lane NorthLoxahatchee, FL 33470
-----------------------------------------------------------------------------

Contact Person:

Alan Bunting 15693 83td Lane North Loxahatchee, FL 33470

Summary Preparation Date: December 4, 2003

2. Names

Device Names: DermaWave 5000 Combi Max System (K040121)

Primary Classification Names:

21 CFR 890.5860, IMG Class 2 Stimulator, Ultrasound

  • 21 CFR 882.5300, IMI Class 2 Diathermy, Ultrasonic
    21 CFR 882.5890 84GZJ Class 2 Transcutaneous Electrical Nerve Stimulator (TENS)

21 CFR 882.1320 GXY Class 2 Electrode, Cutaneous

21CFR 882. 5890 LIH Interferential Current Therapy

21 CFR 890.5500 ILY Class 2 Infrared Lamp

3. Predicate Devices

  • Chatanooga Group Inc., Vectra 2C & Vectra 4C (K982317) .
  • Chaatanooga Group Inc. Vectra Pro 2 & Vectra Pro 4 (K982324) .
  • Dynatronics Corporation, Dynatron 150 Plus Ultrasound (K935728) .
  • Excel Tech, XL Tek Ultra VM Ultrasound (K001166) .
  • MedX 1000 Console (K020017) .
  • MedX LCT 100 Laser Accessory (K032231) .
  • MedX/Duolight 500 (K032231) .
  • MedX 600 LED Accessory (K020017) .

{1}------------------------------------------------

K040121
20F3

4. Product Description

The DermaWave 5000 Combi Max System is comprised of the following main components:

  • A system console including software and control electronics. .
  • A control and display panel. .
  • · Delivery device accessories including patient cables and electrodes, ultrasound cables, soundheads, and infrared heating accessories.

5. Indications for Use

The DermaWave 5000 Combi Max System and the delivery accessories that are used with them are indicated for use in the following:

Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine.

The indications of the DermaWave 5000 Combi Max System also includes the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with posttraumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound.

The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.

6. Rationale for Substantial Equivalence

The DermaWave 5000 Combi Max System shares the same general indications for use in physical medicine, general and plastic surgery and neurology applications and share the same or similar basic characteristics and features and, therefore, are substantially equivalent to the:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a combination of alphanumeric characters. The text includes 'K040121' and '30F3', with the latter positioned below and slightly to the right of the former. The handwriting is somewhat stylized, and the text is oriented diagonally in the frame.

  • Chatanooga Group Inc., Vectra 2C & Vectra 4C (K982317) ♥
  • Chaatanooga Group Inc. Vectra Pro 2 & Vectra Pro 4 (K982324) .
  • Dynatronics Corporation, Dynatron 150 Plus Ultrasound (K935728) .
  • Excel Tech, XL Tek Ultra VM Ultrasound (K001166) .

The DermaWave 5000 Combi Max System utilizes identical accessories as those produced by MedX Health and share identical indications for use in physical medicine. Further, the MedX accessories comprising the MedX LCT 100 (K032231), MedX Duolight 500 (K032231) and the MedX 600 (K020017) are already cleared. The DermaWave 100 Accessory has almost identical heating characteristics as the MedX line of infrared lamp accessories and falls within the recommended guidelines for such devices in raising the temperature of tissue to no more then 45C in continuous operation over a set time period. The parameters used to achieve the comparative data are based on widely accepted clinical data that require heat lamp devices to elevate the temperature to above 40C and below 45C for a minimum 10 minute period. Comparative temperature charts are included in Section 10 of this submission.

The MedX and Duolight cleared accessories are connected to a control console that provides treatment time information to the user. The control console supplied by MedX and named the MedX 1000/1100 Console, is limited in its control features. The DermaWave 5000 Combi Max System employs sophisticated controls and control systems within system architecture to enhance the safety of the device in applications for which it is used.

7. Safety and Effectiveness Information

Validation documentation and a comparison of the technical characteristics and features were provided to demonstrate that the DermaWave 5000 Combi Max System is safe and effective, when indicated in specific applications in the medical specialties of physical medicine, general and plastic surgery and neurology.

8. Conclusion

The DermaWave 5000 Combi Max System was found to be substantially equivalent to the predicate devices and share similar indications for use and characteristics and functional features and are substantially equivalent to the currently marketed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is black and white.

MAR 1 7 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Dermawave, LLC C/o Mr. Derwyn Reuber Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, Massachusetts (01719 Re: K040121 Trade/Device Name: DermaWave 5000 Combi Max System Regulation Numbers: 21 CFR §890.5860, 21 CFR §890.5300, 21 CFR §882.5890, 21 CFR §882.1320, 21 CFR §890.5500 Regulation Names: Ultrasound and Muscle Stimulator, Ultrasonic diathermy, Transcutaneous electrical nerve stimulator for pain relief, Cutaneous electrode, Infrared lamp Regulatory Class: II Product Codes: IMG, IMI, GZJ, GXY, LIH, ILY Dated: March 3, 2004 Received: March 5, 2004

Dear Mr. Reuber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Dcrwyn Reuber

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark N. Millican

Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K040121 Device Name: DermaWave :500 Combi Max Indications for Use:

Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine.

The indications of the DermaWave 5000 Combi Max System also include the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate postsurgical stimulation of calf rnuscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissucs, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound.

The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.

Mark A Millhum

and Neurologi

510(k) Number K090121

Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

N/A