K Number
K033866

Validate with FDA (Live)

Date Cleared
2004-04-07

(117 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay is a homogeneous enzyme immunoassay with 300 ng/mL cutoff for cocaine metabolite, 1000 ng/mL cutoff for amphetamines, 300 ng/mL cutoff for opiates, and 25 ng/mL cutoff for phencyclidine. The assay will produce a positive result if any of the four analyte are present at a concentration at or above their respective cutoffs but will not identify which drug is present. The assay is solely intended for the qualitative screening of human urine for these analytes. Measurements obtained by this device are used in the diagnosis and treatment of individuals who have used cocaine, amphetamines, opiates, or phencyclidine. The assay is designed for professional use with a number of automated clinical chemistry analyzers,

The Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method for the individual drugs must be used to obtain a confirmed analytiral result, Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Description

LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibodies that can detect cocaine metabolite, amphetamines, opiate r ic assay uses specific analo rine with minimal cross-reactivity to various, common prescription drugs and abused drugs.

The assay is based on competition between drug labeled with glucose-6-phosphate I he ussay is based on componities and free drug from the urine sample for a fixed amount of ucific antibody. In the absence of free drug from the urine sample the specific antibody specific antiboury. In the dobtDH enzyme causing a decrease in enzyme activity. It is therefore the drug concentration is proportional to the enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to covert nicotinamide adenine dinucleotide (NAD) to NADH.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay (CAMP):

The provided document describes a 510(k) premarket notification for a new device, comparing it to existing predicate devices. It focuses on demonstrating substantial equivalence rather than setting new, specific acceptance criteria for each performance characteristic of the new device in isolation. Instead, the "acceptance criteria" are implied to be comparable performance to the predicate devices. The study aims to show that the new multi-analyte assay performs as well as, or equivalently to, the individual previously cleared assays.

Let's break down the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

As mentioned, explicit, quantitative acceptance criteria for all performance metrics are not clearly defined as distinct thresholds for the new device. Instead, the document consistently states "100% agreement" or "Comparable to the predicate device" as the desired outcome for performance characteristics (accuracy, analytical recovery, specificity) when compared to either the individual LZI predicate EIAs or commercial EIAs (which are implicitly the standard). For precision, specific CV% values are provided for both the new and predicate devices, which can be interpreted as demonstrating "comparable" precision.

Here's the table, inferring acceptance criteria based on the comparison to predicate devices:

FeatureAcceptance Criteria (Inferred from comparison to predicates)LZI's CAMP EIA Performance (Reported)
Accuracy (vs. LZI Predicate EIA)100% agreement expected100% agreement (Positive Samples)
100% agreement (Negative Samples)
Accuracy (vs. Commercial EIA)Comparable agreement (implied often 100% or very high, sometimes with GC/MS/HPLC confirmation for predicates). The CAMP EIA is only compared to LZI predicates for accuracy, not directly to commercial EIAs.Not directly reported for CAMP EIA vs. commercial EIA. The predicate devices achieved 100% agreement for positive samples vs. commercial EIA and GC/MS/HPLC, and 93.8% to 100% for negative samples. CAMP device implicitly aims for similar performance as the individual LZI-predicates.
Analytical Recovery100% agreement on positive vs. negative tests expected100% agreement on positive vs. negative tests
SpecificityComparable to the predicate deviceSee attached Assay package insert (implying comparability)
Within Run Precision (Cocaine Metabolite)% CV values comparable to LZI Cocaine Metabolite EIA (e.g., typically <1% at various concentrations)Negative: 0.75%, 225 ng/mL: 0.65%, 300 ng/mL: 0.96%, 375 ng/mL: 0.86%, 3000 ng/mL: 0.61%
Run-To-Run Precision (Cocaine Metabolite)% CV values comparable to LZI Cocaine Metabolite EIA (e.g., typically <1% at various concentrations)Negative: 0.48%, 225 ng/mL: 0.65%, 300 ng/mL: 0.55%, 375 ng/mL: 0.85%, 3000 ng/mL: 0.62%
Within Run Precision (Amphetamines)% CV values comparable to LZI Amphetamines EIA (e.g., typically <1% at various concentrations)Negative: 0.85%, 750 ng/mL: 0.81%, 1000 ng/mL: 0.60%, 1250 ng/mL: 0.73%, 2000 ng/mL: 0.69%
Run-To-Run Precision (Amphetamines)% CV values comparable to LZI Amphetamines EIA (e.g., typically <1% at various concentrations)Negative: 0.48%, 750 ng/mL: 0.54%, 1000 ng/mL: 0.47%, 1250 ng/mL: 0.76%, 2000 ng/mL: 0.40%
Within Run Precision (Opiate)% CV values comparable to LZI Opiate EIA (e.g., typically <1% at various concentrations)Negative: 0.90%, 225 ng/mL: 0.92%, 300 ng/mL: 0.92%, 375 ng/mL: 0.78%, 1000 ng/mL: 0.67%
Run-To-Run Precision (Opiate)% CV values comparable to LZI Opiate EIA (e.g., typically <1% at various concentrations)Negative: 0.489%, 225 ng/mL: 0.81%, 300 ng/mL: 0.01%, 375 ng/mL: 0.37%, 1000 ng/mL: 0.69%
Within Run Precision (Phencyclidine)% CV values comparable to LZI Phencyclidine EIA (e.g., typically <1% at various concentrations)Negative: 0.84%, 18 ng/mL: 0.78%, 25 ng/mL: 0.66%, 32 ng/mL: 0.87%, 100 ng/mL: 0.76%
Run-To-Run Precision (Phencyclidine)% CV values comparable to LZI Phencyclidine EIA (e.g., typically <1% at various concentrations)Negative: 0.48%, 18 ng/mL: 0.63%, 25 ng/mL: 0.63%, 32 ng/mL: 0.57%, 100 ng/mL: 0.56%
Sensitivity (Cocaine Metabolite)50 ng/mL (same as predicate)50 ng/mL
Sensitivity (Amphetamines)100 ng/mL (same as predicate)100 ng/mL
Sensitivity (Opiate)50 ng/mL (same as predicate)50 ng/mL
Sensitivity (Phencyclidine)3 ng/mL (same as predicate)3 ng/mL

Note: The reported performance for the CAMP EIA consistently shows similar (and often identical) precision values to the individual LZI predicate EIAs, demonstrating that the multi-analyte assay maintains the precision characteristics of its single-analyte counterparts. Accuracy, Analytical Recovery, and Specificity are all reported as 100% agreement or "Comparable to the predicate device," indicating the new device meets the implied acceptance criterion of equivalence.

2. Sample Size Used for the Test Set and Data Provenance

The document focuses on the performance characteristics of the new assay compared to its predicate devices, primarily based on laboratory testing rather than large-scale patient test sets.

  • Precision Studies:

    • Within Run Precision: n=21 replicates for each concentration level tested (for each of the four analytes).
    • Run-To-Run Precision: n=12 replicates (presumably per concentration over multiple runs) for each analyte.
    • The document does not specify the provenance (country of origin, retrospective/prospective) of the samples used for these precision studies, but they are typically controlled laboratory samples spiked with known concentrations of analytes.
  • Accuracy, Analytical Recovery, Specificity:

    • The document states "100% agreement" for accuracy and analytical recovery when comparing the LZI CAMP EIA to the individual LZI predicate EIAs. It doesn't break down the specific number of samples for these direct comparisons.
    • For the predicate devices, accuracy compared to a commercial EIA and GC/MS/HPLC showed varying numbers of samples (e.g., for Opiate EIA, 97.1% agreement vs. commercial EIA and 100% vs. GC/MS/HPLC, implies a specific number of samples used in that predicate study).
    • The document does not provide a specific sample size for a "test set" of clinical samples for the LZI CAMP EIA's overall accuracy, analytical recovery, or specificity against "ground truth" using a confirmatory method like GC/MS/HPLC. The primary evidence presented is the comparison to the already-cleared predicate devices.
    • Data Provenance: Not specified, but generally, these types of IVD studies are conducted in a laboratory setting, sometimes with clinical samples, but the source details are omitted here. These appear to be retrospective comparisons against established methods.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • No human "experts" are explicitly mentioned as establishing ground truth in this context. For in vitro diagnostic (IVD) devices like this, the ground truth for drug testing is typically established by:
    • Analytical Chemistry Methods: The gold standard is Gas Chromatography/Mass Spectrometry (GC/MS) or High-Performance Liquid Chromatography (HPLC), which are objective analytical methods, not dependent on human expert interpretation.
    • Comparison to a Predicate Device: The performance is benchmarked against a legally marketed device that has already established its accuracy, sensitivity, and specificity.
  • Therefore, the concept of "number of experts" and "qualifications of experts" does not apply directly to how ground truth was established for this type of chemical assay.

4. Adjudication Method for the Test Set

  • Since the ground truth is established by objective analytical methods (GC/MS/HPLC for predicate device claims, and the predicate devices themselves for the new device), there's no human adjudication method (like 2+1, 3+1 consensus) involved as there might be for image-based diagnostic AI, for example. The results are compared quantitatively and qualitatively (positive/negative agreement).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. This type of study (comparing human reader performance with and without AI assistance) is relevant for diagnostic imaging AI devices where human interpretation is the primary method. This device is an in-vitro diagnostic assay for detecting drugs in urine; it does not involve human "readers" interpreting results in the same way as an imaging device, and there is no AI component.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, this is a standalone device in the context of an IVD. The "LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay" is the algorithm/device itself. Its performance is measured directly by analyzing urine samples and producing a detectable signal, which is then interpreted as positive or negative based on a defined cutoff. There is no "human-in-the-loop" to enhance or modify the device's analytical output. The "algorithm" here refers to the chemical assay itself and its spectrophotometric measurement rather than a machine learning algorithm.

7. The Type of Ground Truth Used

  • The ground truth used for demonstrating the performance of the predicate devices was primarily:
    • Commercial EIAs: Other commercially available enzyme immunoassays.
    • Confirmatory Analytical Methods: Gas Chromatography/Mass Spectrometry (GC/MS) and High-Performance Liquid Chromatography (HPLC). These are objective, highly precise chemical analytical techniques considered the gold standard for drug confirmation.
  • For the LZI's CAMP EIA (the new device), the primary "ground truth" against which its performance was demonstrated for the purpose of the 510(k) submission was the individual LZI predicate enzyme immunoassays themselves. The new device showed "100% agreement" with these already-cleared devices for accuracy and analytical recovery.

8. The Sample Size for the Training Set

  • This device is an immunoassay (chemical assay), not an AI/machine learning device. Therefore, the concept of a "training set" in the context of machine learning (where an algorithm learns from data) does not apply.
  • The "training" for such a chemical assay typically involves optimizing the reagent concentrations, reaction conditions, and assay parameters during product development and manufacturing. This is an engineering and chemistry process, not a computational "training set" like that used for AI.

9. How the Ground Truth for the Training Set Was Established

  • As explained above, there is no "training set" in the AI sense for this chemical immunoassay. The development and optimization of the assay rely on established biochemical principles, analytical chemistry validation, and quality control procedures, rather than on a labeled training dataset.

{0}------------------------------------------------

K033866

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter name, Address, and Contact

Lin-Zhi International, Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085 Phone: (408) 732-3856 (408) 732-3849 Fax:

Contact:Cheng-I Lin, Ph.D.
President, R&D Director

Device Name and Classification

1. Classification Name:Cocaine Metabolite, Amphetamines, Opiate, andPhencyclidine test system,
The Cocaine and Cocaine Metabolite test system has beenplaced in Class II by the Bureau of Medical Devices.Classification Number: DIO (21 CFR 862.3250)Panel: 91 Toxicology
The Amphetamine test system has been placed in Class II bythe Bureau of Medical Devices.Classification Number: DKZ (21 CFR 862.3100)Panel: 91 Toxicology
The Opiate test system has been placed in Class II by theBureau of Medical Devices.Classification Number: DJG (21 CFR 862.3650)Panel: 91 Toxicology
The Phencyclidine test systemClassification Number: LCMPanel: 91 Toxicology

{1}------------------------------------------------

Common Name:Homogeneous enzyme immunoassay for the detection ofcocaine metabolite, amphetamines, opiate, and phencyclidinein human urine.
Proprietary Name:CAMP Enzyme Immunoassay

Legally Marketed Predicate Device(s)

Lin-Zhi International, Inc.'s Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Life-Zill Intentational, Inc. 3 Dimananovas Occaratially equivalent to Cocaine Metabolite Multiple Anaryto Enzynio Iminetamines Enzyme Immunoassay, Opiate Enzyme Enzyme Inimunoassay, Aniphotamilizzyme Immunoassay by Lin-Zhi International, Inc., minunoasay, and I nonoyonano izazjo2763 (Cocaine Metabolite Enzyme Immunoassay), cleared under premanter nouncalitation Prouvers (Opiate Enzyme Immunoassay), and K020254 (Phencyclidine Enzyme Immunoassay).

LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay is similar to their predicates in terms of intended use, method principle, device components, and clinical performance.

Device Description

LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibodies that can detect cocaine metabolite, amphetamines, opiate r ic assay uses specific analo rine with minimal cross-reactivity to various, common prescription drugs and abused drugs.

The assay is based on competition between drug labeled with glucose-6-phosphate I he ussay is based on componities and free drug from the urine sample for a fixed amount of ucific antibody. In the absence of free drug from the urine sample the specific antibody specific antiboury. In the dobtDH enzyme causing a decrease in enzyme activity. It is therefore the drug concentration is proportional to the enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to covert nicotinamide adenine dinucleotide (NAD) to NADH.

Intended Use

The Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay (CAMP) is a homogeneous enzyme immunoassay with 300 ng/mL Enzyme Innnunousber (Critics) 100 ng/mL cutoff for methamphetamine, 300 ng/mL cutoff for opiate, and 25 ng/mL cutoff for phencyclidine. The assay is solely intended for use in the qualitative screening for negative human urine samples for cocaine metabolite, amphetamines, opiate, and phencyclidine drugs.

{2}------------------------------------------------

Comparison to Predicate Device

LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay (CAMP) is substantially equivalent to the individual single analyte assay products in commercially distribution intended for similar use. Most notably it is substantially equivalent to the Cocaine Metabolite Enzyme Immunoassay, Amphetamines Enzyme Immunoassay, Opiate Enzyme Immunoassay, and Phencyclidine Enzyme Immunoassay By Lin-Zhi International, Inc., cleared under premarket notification K020763 (Cocaine Metabolite Enzyme Immunoassay), K020369 (Amphetamines Enzyme Immunoassay), K020368 (Opiate Enzyme Immunoassay), and K020254 (Phencyclidine Enzyme Immunoassay).

The following table compares LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay (CAMP) with the predicate devices, Cocaine Metabolite Enzyme Immunoassay, Amphetamines Enzyme Immunoassay, Opiate Enzyme Immunoassay, and Phencyclidine Enzyme Immunoassay by Lin-Zhi International, Inc.

Similarities:

  • Both assays are used for qualitative detection of drug in human urine. .
  • Both have same cutoff design (300 ng/mL for cocaine metabolite, 1000 ng/mL . for methamphetamine, 300 ng/mL for opiate, and 25 ng/mL for phencyclidine).
  • Both assays use same control concentration (+/- 25% of cut-off value). .
  • Both assays use the same method principle, and device components. ●

Difference:

  • Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple . Analyte Enzyme Immunoassay is designed for qualitative screening purpose only.
  • Multiple analyte calibrators/controls cannot be use in this assay. .
  • The CAMP assay should be calibrated with Opiate calibrators only. ●

{3}------------------------------------------------

FeatureLZI s CAMP EIALZI's Cocaine Metabolite EIAFeatureLZI's CAMP EIALZI's Amphetamines EIA
Within Run Precision:(n=21)(mA/min.)MeanSD% CV(mA/min.)MeanSD% CVWithin Run Precision:(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
Negative736.35.50.75Negative243.70.90.37(n=21)Negative734.76.20.85Negative273.01.00.35
225 ng/mL786.25.10.65225 ng/mL356.81.80.50750 ng/mL793.06.40.81750 ng/mL390.01.40.37
300 ng/mL803.67.70.96300 ng/mL380.21.80.481000 ng/mL804.04.90.601000 ng/mL415.71.40.34
375 ng/mL811.97.00.86375 ng/mL397.32.20.561250 ng/mL810.25.90.731250 ng/mL439.01.60.37
3000 ng/mL876.85.30.613000 ng/mL488.41.80.362000 ng/mL825.55.70.692000 ng/mL480.51.40.30
Run-To-Run Precision:(n=12)(mA/min.)MeanSD% CV(mA/min.)MeanSD% CVRun-To-Run Precision:(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
Negative742.43.50.48Negative243.10.90.36(n=12)Negative742.43.50.48Negative272.92.30.8
225 ng/mL793.45.10.65225 ng/mL354.82.40.67750 ng/mL792.84.30.54750 ng/mL390.52.90.7
300 ng/mL806.94.40.55300 ng/mL377.03.50.931000 ng/mL799.03.80.471000 ng/mL415.93.00.5
375 ng/mL822.17.00.85375 ng/mL394.62.10.531250 ng/mL808.96.10.761250 ng/mL439.13.60.4
3000 ng/mL882.35.50.623000 ng/mL486.02.40.492000 ng/mL828.43.30.402000 ng/mL479.63.50.4
Sensitivity:50 ng/mL4 ng/mLSensitivity:100 ng/mL30 ng/mL
Accuracy:Vs. LZI Cocaine metabolite EIAVs. a commercial EIAAccuracy:Vs. LZI Amphetamines EIAVs. a commercial EIA
Positive Samples:100 % agreement100 % agreementPositive Samples:100 % agreement100 % agreement (100% vs. GC/MS /HPLC)
Negative Samples:100 % agreement100 % agreementNegative Samples:100 % agreement100 % agreement
Analytical Recovery:100 % agreement on positive vs. negative tests100 % agreement on positive vs. negative testsAnalytical Recovery:100 % agreement on positive vs. negative tests100 % agreement on positive vs. negative tests
Specificity:See attached Assay package insertComparable to the predicate device.Specificity:See attached Assay package insertComparable to the predicate device.

Performance Characteristics

{4}------------------------------------------------

FeatureLZI's CAMP EIALZI's Opiate EIA
Within Run Precision:(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
(n=21)Negative736.06.60.90Negative291.62.190.75
225 ng/mL782.47.20.92225 ng/mL374.43.010.80
300 ng/mL801.87.40.92300 ng/mL399.83.420.86
375 ng/mL817.56.40.78375 ng/mL421.83.200.76
1000 ng/mL878.75.90.671000 ng/mL530.85.170.97
Run-To-Run Precision:(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
(n=12)Negative742.43.50.489Negative292.81.810.62
225 ng/mL793.06.50.81225 ng/mL375.83.610.96
300 ng/mL809.14.60.01300 ng/mL400.83.340.83
375 ng/mL819.83.10.37375 ng/mL421.12.870.68
1000 ng/mL886.06.20.691000 ng/mL528.64.840.92
Sensitivity:50 ng/mL15 ng/mL
Accuracy:Vs. LZI Opiate EIAVs. a commercial EIA
Positive Samples:100 % agreement97.1 % agreement (100% vs. GC/MS /HPLC)
Negative Samples:100 % agreement93.8 % agreement
Analytical Recovery:100 % agreement on positive vs. negative tests100 % agreement on positive vs. negative tests
Specificity:See attached Assay package insertComparable to the predicate device

·

FeatureLZI's CAMP EIALZI's Phencyclidine EIA
Within Run Precision:(n=21)(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
Negative735.16.20.84Negative168.00.680.41
18 ng/mL789.66.20.7818 ng/mL238.61.160.49
25 ng/mL799.25.20.6625 ng/mL264.61.190.45
32 ng/mL809.37.00.8732 ng/mL282.81.420.50
100 ng/mL833.56.30.76100 ng/mL341.20.730.21
Run-To-Run Precision:(n=12)(mA/min.)MeanSD% CV(mA/min.)MeanSD% CV
Negative742.43.40.48Negative168.00.900.54
18 ng/mL794.15.00.6318 ng/mL238.80.550.23
25 ng/mL804.95.10.6325 ng/mL264.60.790.30
32 ng/mL814.04.60.5732 ng/mL284.10.870.31
100 ng/mL829.14.70.56100 ng/mL340.11.070.32
Sensitivity:3 ng/mL1 ng/mL
Accuracy:Vs. LZI Phencyclidine EIAVs. a commercial EIA
Positive Samples:100 % agreement100 % agreement(100% vs. GC/MS /HPLC)
Negative Samples:100 % agreement100 % agreement
Analytical Recovery:100 % agreement on positive vs. negative tests100 % agreement on positive vs. negative tests
Specificity:See attached Assay package insertComparable to the predicate device

{5}------------------------------------------------

Conclusion

The LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay was evaluated for several performance characteristics including precision, sensitivity, accuracy, analytical recovery, and specificity. All the studies showed acceptable results when compared to the individual predicate device.

We trust the information provided in this Premarket Notification [510(k)] submission will support a determination of substantial equivalence of the LZI's Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay to other individual test systems for screening purpose currently marketed in the United States.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures. The symbol is composed of thick, black lines and curves, giving it a modern and minimalist appearance.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

APR - 7 2004

Cheng-I Lin, Ph.D. President, R&D Director Lin-Zhi International. Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085

Re: K033866

Trade/Device Name: Simultaneous Cocaine- Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO, DJG, DZK, LCM Dated: March 19, 2004 Received: March 23, 2004

Dear Dr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indictions for use stated in the enclosure) to legally marketed predicate devices marketed in intenstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendonstity, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application(PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, I isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition FDA may publish further announcements concerning your device in the Federal Do- ' ur .

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{7}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Radiological Health

Jean M. Cooper, MS, DVM.

Yean M. Cooper, MS. D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and

Enclosure

{8}------------------------------------------------

Premarket Notification

Indications for Use

510(k) Number (if known): K033866

Device Name: Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay

Indications for Use:

The Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay is a homogeneous enzyme immunoassay with 300 ng/mL cutoff for cocaine metabolite, 1000 ng/mL cutoff for amphetamines, 300 ng/mL cutoff for opiates, and 25 ng/mL cutoff for phencyclidine. The assay will produce a positive result if any of the four analyte are present at a concentration at or above their respective cutoffs but will not identify which drug is present. The assay is solely intended for the qualitative screening of human urine for these analytes. Measurements obtained by this device are used in the diagnosis and treatment of individuals who have used cocaine, amphetamines, opiates, or phencyclidine. The assay is designed for professional use with a number of automated clinical chemistry analyzers,

The Simultaneous Cocaine-Amphetamines-Morphine-Phencyclidine Multiple Analyte Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method for the individual drugs must be used to obtain a confirmed analytiral result, Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033866

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).