K Number
K033418

Validate with FDA (Live)

Date Cleared
2003-11-10

(14 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TSX-101A CT Scanners are indicated for head and whole body computed tomography applications.
X-ray imaging of whole body - Computerized Tomography Including: Axial Volumetric (Helical) CT Fluoroscopy

Device Description

The Aquilion TSX101A/A and TSX-101A/C CT Scanner Systems are composed of a gantry, patient table, image acquisition hardware and software, an operator console and associated accessories. Materials and construction are similar to the Aquilion CT Scanner Multi-slice 16 (CGS-31A) [K022050] and the TSX-101A/7 Aquilion Super 4 CT Scanner [K031469], and are compliant with IEC60601-1 and associated collateral and particular standards, and applicable sections of 21 CFR Subchapter J.

AI/ML Overview

The provided text is a 510(k) summary for the Toshiba Aquilion Multislice CT Scanners (TSX-101A/A and TSX-101A/C). This document describes the device, its intended use, and argues for its substantial equivalence to previously marketed predicate devices.

Crucially, this document does not contain information about specific acceptance criteria or a study designed to prove the device meets those criteria in the way a clinical performance study would for an AI/CAD device. This is a Computed Tomography (CT) scanner, which is a medical imaging device, not an AI or CAD software that produces diagnostic interpretations or measurements.

The "Summary of Studies" section in the K033418 510(k) document states:
"These devices have been evaluated for electrical, mechanical, and radiation safety. Furthermore they conform to applicable medical device safety and performance standards."

This indicates that the studies conducted were primarily focused on safety and engineering standards compliance rather than clinical performance metrics like sensitivity, specificity, or reader improvement with AI.

Therefore, for many of your requested points, the answer based on this document will be that the information is not applicable or not provided.

Here's a breakdown based on the information available:


1. Table of Acceptance Criteria and Reported Device Performance

Not Applicable/Not Provided. This document does not describe specific clinical acceptance criteria (e.g., sensitivity, specificity, accuracy) for a diagnostic task or algorithm performance. The device is a CT scanner, and the evaluation focuses on its safety and conformance to general engineering and radiation safety standards.


2. Sample Size Used for the Test Set and Data Provenance

Not Applicable/Not Provided. The document does not describe a "test set" in the context of clinical performance evaluation for a diagnostic algorithm. The evaluations were for electrical, mechanical, and radiation safety of the hardware.


3. Number of Experts Used to Establish Ground Truth and Qualifications

Not Applicable/Not Provided. There is no mention of experts establishing ground truth for a clinical test set, as this was not a clinical performance study of a diagnostic algorithm.


4. Adjudication Method for the Test Set

Not Applicable/Not Provided. Since there was no clinical test set for diagnostic performance, no adjudication method is described.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document does not mention an MRMC comparative effectiveness study. This type of study is typically done for AI/CAD devices to evaluate reader performance with and without AI assistance, which is outside the scope of a CT scanner's hardware safety and performance evaluation.


6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

No. The document does not describe a standalone performance study. This is a hardware device (CT scanner) and does not have a standalone algorithm for diagnostic interpretation in the sense of AI/CAD software.


7. Type of Ground Truth Used

Not Applicable/Not Provided. There is no mention of ground truth in the context of diagnostic accuracy, as the studies focused on safety and functional compliance of the CT scanner hardware.


8. Sample Size for the Training Set

Not Applicable/Not Provided. This device is a CT scanner, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI development.


9. How the Ground Truth for the Training Set Was Established

Not Applicable/Not Provided. As there is no training set mentioned, this point is not applicable.

{0}------------------------------------------------

NOV 1 0 2003

510(k) Summary

This 510(k) summary of safety and effectiveness is submitted in accordance with the requirements set forth in 21 CFR 807.87(h).

Date:October 17, 2003
Submitter:Toshiba America Medical Systems, Inc.P.O. Box 2068, 2441 Michelle Drive,Tustin, CA 92781-2068
Submitter's Contact:Paul BigginsSenior Manager, Regulatory AffairsTelephone: (714) 730-5000; Fax: (714) 730-1310
Device Proprietary Name:TSX-101A/A and TSX-101A/C Aquilion Multislice CT Scanners
Common Name:Scanner, Computed Tomography, X-Ray[Fed. Reg. No. 892.1750, Pro. Code: 90JAK]
Predicate Device(s):TSX-101A/8 Aquilion 16 CTTSX-101A/7 Aquilion Super 4 CT

Description of this Device:

The Aquilion TSX101A/A and TSX-101A/C CT Scanner Systems are composed of a gantry, patient table, image acquisition hardware and software, an operator console and associated accessories. Materials and construction are similar to the Aquilion CT Scanner Multi-slice 16 (CGS-31A) [K022050] and the TSX-101A/7 Aquilion Super 4 CT Scanner [K031469], and are compliant with IEC60601-1 and associated collateral and particular standards, and applicable sections of 21 CFR Subchapter J.

Indications for Use:

The TSX-101A CT Scanners are indicated for head and whole body computed tomography applications.

Comparison with Predicate Devices

The TSX-101A/A and TSX-101A/C Computed Tomography Systems are modifications of and are of comparable type and substantially equivalent to the currently marketed Toshiba TSX-101A/7 and TSXC-101A/8 CT Systems. They employ the same technological characteristics, are comparable in regards to primary concerns of safety and effectiveness, employ the same basic design, construction, materials and have the same intended use as the predicate devices.

Summary of Studies:

These devices have been evaluated for electrical, mechanical, and radiation safety. Furthermore they conform to applicable medical device safety and performance standards.

Conclusion

Intended use and technologies are the predicate Toshiba devices listed above. The design and development process of the manufacturer conforms with 21 CFR 820 and ISO 9001/13485 quality systems. Additionally these devices conform to applicable medical device safety and performance standards. It is the opinion of Toshiba America Medical Systems that these devices are substantially equivalent to the legally marketed predicate devices.

,

{1}------------------------------------------------

Image /page/1/Picture/11 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure with outstretched arms, with three lines representing the figure's body and arms. The figure is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" surrounding the top half of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 10 2003

Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th NW BUFFALO MN 55313

Re: K033418 Trade/Device Name: TSX-101A/A and TSX-101A/C Aquilion CT Scanners Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 23, 2003 Received: October 27, 2003

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

l of Page 1

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

TSX-101A/A and TSX-101A/C Aquilion CT Scanners Device Name:

Indications for Use:

X-ray imaging of whole body - Computerized Tomography Including: Axial Volumetric (Helical)


CT Fluoroscopy

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)Division of Reproductive, Abdominal,and Radiological Devices510(k) NumberK033418
Prescription Use(Per 21 CFR 801.109) OROver-The-Counter Use ______