K Number
K032981

Validate with FDA (Live)

Date Cleared
2004-04-22

(211 days)

Product Code
Regulation Number
868.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPpro™ ESP is an anesthesia conduction needle incorporating a needle tip shielding mechanism to aid in the prevention of needle-stick injuries after use to deliver an anesthetic and/or therapeutic agent to provide regional anesthesia. The needle can be used for a single dose administration or in conjunction with a catheter for continuous regional anesthesia.

Device Description

The SPpro™ ESP anesthesia conduction needle is a single use device for injection of local anesthetics into body cavities other than blood vessels. The device will be packaged as sterile single units or incorporated into kits. The device is designed with a tip shield and sheath mechanism integrated with the introducer hub to aid in the prevention of sharps injuries. After activation the shield covers the introducer tip. The device is disposed of according to institutional procedure in a sharps container.

AI/ML Overview

"The provided document is limited in detail regarding explicit acceptance criteria and a comprehensive study report. However, based on the information available, here's a structured response:

Acceptance Criteria and Device Performance

The document does not provide a table with specific quantitative acceptance criteria (e.g., minimum percentage of successful shield activations) or detailed reported device performance metrics in a tabular format. It broadly states that the device "met all functional requirements and specifications" and that "functional and simulated clinical testing was performed to support that there are no new issues or effectiveness raised by adding the sharps injury protection feature."

The primary performance claim is related to the sharps injury protection feature and its ability to "aid in the prevention of needle-stick injuries after use."

Implied Acceptance Criteria (Derived from text):

  • Functional requirements: The sharps injury protection feature must activate and shield the needle tip effectively after use.
  • Sharps Injury Prevention: The device design must aid in preventing needle-stick injuries.
  • Biocompatibility: Materials must demonstrate biocompatibility for transient external tissue contact.
  • Substantial Equivalence: The device must demonstrate substantial equivalence to predicate devices in its therapeutic function and the safety feature's technological equivalence.

Reported Device Performance (Derived from text):

  • "The SPpro™ ESP met all functional requirements and specifications." (Section 2: Device Performance/Product Testing)
  • "Functional and simulated clinical testing was performed to support that there are no new issues or effectiveness raised by adding the sharps injury protection feature." (Section 2: Device Performance/Product Testing)
  • The safety feature is described as "an active shielding system that incorporates a second hub section integrated with the luer portion of the needle that slides up from the needle hub and locks over the needle tip." This describes the mechanism intended to achieve the sharps injury prevention.

Study Information

The document describes functional and simulated clinical testing rather than a formal clinical study with a specified sample size, ground truth, or expert involvement as typically seen in efficacy studies for AI/diagnostic devices.

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document refers to "functional and simulated clinical testing" but does not quantify the number of devices tested or the number of simulated cases.
    • Data Provenance: Not specified. It can be inferred that the testing was conducted in a controlled environment as "simulated clinical testing." There is no mention of country of origin or whether it was retrospective or prospective data from actual patient use.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the testing described is functional and simulated clinical testing, not a study requiring expert-established ground truth for a diagnostic outcome. The "ground truth" here would be the successful activation and shielding of the needle, observable during testing.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable due to the nature of the functional and simulated clinical testing. Adjudication methods are typically used when interpreting subjective findings or reconciling discrepancies in expert opinions for diagnostic assessments.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This type of study is relevant for diagnostic or AI-powered devices where human readers (e.g., radiologists) interpret cases with and without AI assistance. The SPpro™ ESP is a medical device with a sharps injury prevention feature, not a diagnostic or AI device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The device is a physical medical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the performance of the sharps injury protection feature would be objective observation during functional and simulated clinical tests: whether the shield successfully activated and covered the needle tip as designed. The document refers to "functional requirements and specifications," implying that the device was tested against pre-defined performance standards for its mechanical function.
  7. The sample size for the training set

    • Not applicable. This device is not an AI/ML algorithm that requires a training set.
  8. How the ground truth for the training set was established

    • Not applicable. This device is not an AI/ML algorithm that requires a training set.

Additional Context from the Document:

The nature of the submission (K032981) is a 510(k) premarket notification, which establishes "substantial equivalence" to legally marketed predicate devices. The focus of the performance testing for this type of submission is often to demonstrate that the new or modified feature (the sharps injury protection in this case) does not raise new questions of safety or effectiveness and performs comparably to or better than previously cleared devices, especially considering relevant guidance documents. The document explicitly states: "Clinical simulation testing was in accordance with the Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features: Guidance for Industry and FDA issued on December 31, 2002." This indicates adherence to regulatory guidance for this specific type of safety feature.

{0}------------------------------------------------

K032981

APR 2 2 2004

Image /page/0/Picture/2 description: The image contains the text "InVaMed" in a large, bold font at the top left of the image. Below this text, there is another line of text that reads "These are wondrous". The background of the image is black, and the text is white, creating a high contrast. There is also a small, blurry image in the center of the image.

Tim Duvall President Office Number (925) 820-3128

InVaMed Technologies, Inc. 9955 Titan Park Circle / Littleton, CO 80125 Corp. Phone 303-346-5300 Fax 303-346-9120

510(k) Summary - Dated September 19, 2003

General Information

Classification:Class II
Trade Name:SPpro™ - ESP
Submitter:InVaMed Technologies, Inc.9955 Titan Park CircleLittleton, CO 80125
Contact:Tim DuvallPresident(925) 820-3128

{1}------------------------------------------------

Device Identification

  • A. Common Name
    Epidural Needle with Sharps Injury Protection Feature Spinal Needle with Sharps Injury Protection Feature

  • B. Classification Name
    Needle, Conduction, Anesthetic (W/WO Introducer)

C. Regulation Number

858.5150

D. Panel

Anesthesiology

E. Product Code

BSP

  • F. Class
    Class II

Predicate Devices

Manan Epidural Needle from Manan Medical Products, Inc. K980536 Manan Spinal Needles from Manan Medical Products, Inc. K983620 Perifix Safety Epidural Needle from B. Braun Medical, Inc. K013610

Device Description Information

Intended Use

The SPpro™ ESP is an anesthesia conduction needle incorporating a needle tip shielding mechanism to aid in the prevention of needle-stick injuries after use to deliver an anesthetic and/or therapeutic agent to provide regional anesthesia. The needle can be used for a single dose administration or in conjunction with a catheter for continuous regional anesthesia.

{2}------------------------------------------------

Device Description

The SPpro™ ESP anesthesia conduction needle is a single use device for injection of local anesthetics into body cavities other than blood vessels. The device will be packaged as sterile single units or incorporated into kits. The device is designed with a tip shield and sheath mechanism integrated with the introducer hub to aid in the prevention of sharps injuries. After activation the shield covers the introducer tip. The device is disposed of according to institutional procedure in a sharps container.

Biocompatibility

Although the SPpro™ ESP sharps injury protection device has no direct contact with internal body fluids the materials and assembly methods were selected to provide for transient external tissue contact biocompatibility.

Biocompatibility testing on the therapeutic portion of the device included all tests performed on the predicate devices. The therapeutic devices are already approved devices made for InVaMed Technologies by Manan Medical Products, Inc.

Device Performance/Product Testing

Functional and simulated clinical testing was performed to support that there are no new issues or effectiveness raised by adding the sharps injury protection feature. The SPpro™ ESP met all functional requirements and specifications. Clinical simulation testing was in accordance with the Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features: Guidance for Industry and FDA issued on December 31, 2002.

{3}------------------------------------------------

Technology Characteristics

The therapeutic device is equivalent technologically to the devices mentioned under predicate devices above. The therapeutic devices are already approved devices made for InVaMed Technologies by Manan Medical Products, Inc.

The safety feature, which has been integrated with the therapeutic portion of the device, is the reason for this submission. The safety feature is an active shielding system that that incorporates a second hub section integrated with the luer portion of the needle that slides up from the needle hub and locks over the needle tip. A second feature incorporates a plastic sheath attached to the sliding portion and the luer hub, which covers the previously exposed canula of the needle.

Substantial Equivalence

The InVaMed SPpro ESP needle incorporates an active needle tip cover mechanism to minimize needle stick injuries when used to access non-vascular areas of the body to administer local anesthetics. The basic design, methods of manufacturing, and materials used in the therapeutic portion of the device are substantially equivalent to the predicate devices, Manan Epidural Needle and Manan Spinal Needles from Manan Medical Products, Inc. While basic design, methods of manufacturing, and materials used in the safety portion of the device differ the technological equivalence exist between the predicate devices, Perifix® Safety Epidural Needle from B. Braun Medical, Inc. Our application for this device is substantially equivalent to the aforementioned devices already approved for use. The clinical indications for use remain unchanged. InVaMed believes the SPpro ESP anesthesia conduction needle is substantially equivalent to currently marketed anesthesia conduction needles and an anesthesia conduction needle with sharps injury protection features.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines representing the body and arms.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 2 2004

Mr. Tim Duvall President InVaMed Technologies; Incorporated 9955 Titan Park Circle Littleton, Colorado 80125

Re: K032981

Trade/Device Name: SPpro ESP, Modcls ESP 603-18-05 and ESP 603-18-09 Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: March 16, 2004 Received: March 19, 2004

Dear Mr. Duvall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Mr. Tim Duvall

You must comply with all the Act's requirements, including, but not limited to: registration r ou indist configry with a 807); labeling (21 CFR Part 801); good manufacturing practice and home (2) as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as bet form nroduct radiation control provisions (Sections 531-542 of the Act); appreadio, and other will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence in your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), n 900 donto the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the maj over it Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Ravis

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

K032981

Device Name: SPpro ESP

Indications For Use:

The SPpro™ ESP is an anesthesia conduction needle incorporating a needle tip shielding mechanism to aid in the prevention of needle-stick injuries after use to deliver an anesthetic and/or therapeutic agent to provide regional anesthesia. The needle can be used for a single dose administration or in conjunction with a catheter for continuous regional anesthesia.

Prescription Use_ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

A.W. Westerhem

esthesiology, General Hospital,

Page 1 of i

510(k) Number. K032981

Section 1 - 9

§ 868.5150 Anesthesia conduction needle.

(a)
Identification. An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.(b)
Classification. Class II (performance standards).