K Number
K032903

Validate with FDA (Live)

Date Cleared
2004-02-17

(153 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are rapid chromatographic immunoassays for the qualitative detection of Benzoylecgonine in urine at a cutoff concentration of 150 ng/mL. They are intended for healthcare professionals including professionals at point-of-care sites.

Device Description

The ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Benzoylecgonine in a urine sample. The test is based on the principle of antigenantibody immunochemistry. It utilizes the antibody to selectively detect elevated levels of cocaine and its metabolite (Benzoylecgonine) in urine at a cutoff concentration of 150 ng/mL. These tests can be performed without the use of an instrument.

A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Benzoylecgonine at the concentration below the cutoff level will generate a colored-line in the test region. To serve as a procedural control, a colored-line should always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

AI/ML Overview

Here's an analysis of the provided text regarding the ACON COC-150 One Step Cocaine Test Strip and Test Device, structured according to your request:

Acceptance Criteria and Study for ACON COC-150 One Step Cocaine Test

The document describes a clinical evaluation to demonstrate the substantial equivalence of the ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device. The primary acceptance criteria appear to be high agreement rates (ideally 100% or very close with tight confidence intervals) when compared to both a predicate FDA-cleared cocaine test and Gas Chromatography/Mass Spectrometry (GC/MS) analysis at specified concentration ranges.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance (vs. FDA-cleared Cocaine Test)Reported Device Performance (vs. GC/MS at Cutoff) - ACON Test StripReported Device Performance (vs. GC/MS at Cutoff) - ACON Test Device
High Positive Agreement100% (97.4% - 100% CI)98.6% (95.0% - 99.8%) CI for Positive samples > Cutoff98.6% (95.0% - 99.8%) CI for Positive samples > Cutoff
High Negative Agreement100% (97.7% - 100% CI)98.7% (95.5% - 99.6%) CI for Negative samples < Cutoff98.7% (95.5% - 99.6%) CI for Negative samples < Cutoff
High Overall Agreement100% (98.8% - 100% CI)Not directly stated as 'overall' but implied high in the tablesNot directly stated as 'overall' but implied high in the tables
Low false positives near cutoff(Not explicitly stated or broken out for predicate)0 false positives for negative and -25% to Cutoff0 false positives for negative and -25% to Cutoff
Low false negatives near cutoff(Not explicitly stated or broken out for predicate)0 false negatives for Cutoff to +25% Cutoff0 false negatives for Cutoff to +25% Cutoff

Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria (e.g., "must achieve >95% agreement"). Instead, it presents the results, implying that the achieved performance, particularly the 100% agreement with the predicate device and very high agreement with GC/MS, met the necessary standards for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 300 clinical urine specimens.
    • Approximately 10% of these specimens contained Benzoylecgonine concentrations that fell within -25% cutoff to +25% cutoff range, indicating a focus on the critical decision point around the cutoff.
  • Data Provenance: The document states "clinical urine specimens," implying human-derived samples. However, it does not specify the country of origin of the data. It also does not explicitly state if the data was retrospective or prospective. Given it's a "clinical evaluation," it's likely prospective collection of samples or retrospective use of stored clinical samples, but this is not definitively stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • The document does not specify the number of experts used and their qualifications for establishing ground truth.
  • The ground truth for the comparison was established by an "FDA-cleared cocaine test" and "customary Gas Chromatography/Mass Spectrometry analysis." These are analytical methods, not human experts making subjective diagnoses.

4. Adjudication Method for the Test Set

  • None in the context of human expert adjudication. The ground truth was established by laboratory methods (predicate device and GC/MS), which are objective and do not require adjudication in the way subjective human assessments would.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • No, a MRMC comparative effectiveness study was NOT done. This study focuses on the performance of the device itself against established methods, not on how human readers perform with or without AI assistance. The device is a "rapid chromatographic immunoassay" that provides a visual, qualitative end result, and is not an AI-assisted diagnostic tool for interpretation by human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, a standalone study was done. The entire evaluation described is a standalone performance assessment of the ACON COC-150 test strip and device. It measures the device's ability to detect Benzoylecgonine in urine specimens independently, without human interpretation influencing the result beyond the visual reading of a line (which is inherent to the device's design). The comparison against GC/MS is a direct measure of its standalone analytical performance.

7. The Type of Ground Truth Used

  • The primary ground truth used was Gas Chromatography/Mass Spectrometry (GC/MS) analysis. This is considered the "gold standard" confirmatory method for drug testing and provides highly accurate quantitative data, from which qualitative "positive" or "negative" results are derived based on the 150 ng/mL cutoff.
  • Additionally, an FDA-cleared predicate device was used for comparative purposes, serving as a secondary ground truth reference.

8. The Sample Size for the Training Set

  • The document does not mention a training set sample size. This is expected because the ACON COC-150 devices are immunoassays, not machine learning or AI algorithms that require a separate training phase. The manufacturing process and quality control for these devices ensure their performance.

9. How the Ground Truth for the Training Set Was Established

  • As no training set is described or implied for this type of immunoassay device, the question of how its ground truth was established is not applicable.

{0}------------------------------------------------

page 1 of 4

FEB 1 7 2004

510(k) SUMMARY 8.

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

-032903 The Assigned 510(k) number is

Submitter:

ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121

Tel .: 858-535-2030 Fax: 858-535-2038

Date:

September 12, 2003

Contact Person:

Edward Tung, Ph.D.

Product Names:

ACON® COC-150 One Step Cocaine Test Strip ACON® COC-150 One Step Cocaine Test Device

Common Name:

Immunochromatographic test for the qualitative detection of Benzoylecgonine in urine.

Classification Name:

Cocaine test system.

Device Classification:

{1}------------------------------------------------

page 2 of 4

The Cocaine test systems have been classified as Class II devices with moderate complexity. The ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are similar to other FDA-cleared devices for the qualitative detection of Benzoylecgonine, a major cocaine metabolite, in urine specimens. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis.

Intended Use:

The ACON® COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are rapid chromatographic immunoassays for the qualitative detection of Benzoylecgonine in urine at a cutoff concentration of 150 ng/mL. They are intended for healthcare professionals including professionals at point-of-care sites.

Description:

The ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Benzoylecgonine in a urine sample. The test is based on the principle of antigenantibody immunochemistry. It utilizes the antibody to selectively detect elevated levels of cocaine and its metabolite (Benzoylecgonine) in urine at a cutoff concentration of 150 ng/mL. These tests can be performed without the use of an instrument.

A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Benzoylecgonine at the concentration below the cutoff level will generate a colored-line in the test region. To serve as a procedural control, a colored-line should always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

Comparison to a Predicate Device:

A comparison of the features of the ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device versus a FDA-cleared cocaine test with 150 ng/mL benzoylecgonine cutoff is shown below:

  • . Both tests are assays intended for the qualitative detection of Benzovlecgonine in urine samples.
  • . Both tests are intended as a screening method that provides a preliminary analytical test result.
  • Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Benzoylecgonine with a visual, qualitative end result.
  • . Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody interactions to indicate a positive or negative result.
  • Both tests have a cut-off Benzoylecgonine concentration of 150 ng/mL. .

{2}------------------------------------------------

page 3 of 4

Safety and Effectiveness Data:

Accuracy

A clinical evaluation was conducted using 300 clinical urine specimens including approximately 10% of the specimens containing Benzoylecgonine concentration fell between -25% cutoff to +25% cutoff range. This evaluation compared the test results between ACON® COC-150 One Step Cocaine Test Strip and ACON® COC-150 One Step Cocaine Test Device with a FDA-cleared cocaine test; as well as compared against data obtained from the customary Gas Chromatography/Mass Spectrometry analysis. These comparisons of data yielded the following results:

ACON COC-150 One Step Cocaine Test Strip versus FDA-cleared Cocaine Test:

Positive Agreement: 141 / 141 = 100% (97.4% - 100%) Negative Agreement: 159 / 159 = 100% (97.7% - 100%) Overall Agreement: 300/300 = 100% (98.8% - 100%*)

  • Since the proportion (%) cannot go above 100%, this is really a 97.5% Confidence interval.

ACON COC-150 One Step Cocaine Test Device versus FDA-cleared Cocaine Test:

Positive Agreement: 141 / 141 = 100% (97.4% - 100%) Negative Agreement: 159 / 159 = 100% (97.7% - 100%) Overall Agreement: 300/300 = 100% (98.8% - 100%*)

  • Since the proportion (%) cannot go above 100%, this is really a 97.5% Confidence interval.

ACON COC-150 One Step Cocaine Test Strip versus data obtained with GC/MS at the cutoff concentration of 150 ng/mL:

Benzoylecgonine Conc.vs. CutoffNegative-25%Cutoff toCutoffCutoff to+25% Cutoff> +25%Cutoff% Agreement withGC/MS
ACONCOC-150Test StripPositive00813398.6%(95.0% - 99.8%)**
Negative15070298.7%(95.5% -99.6%)**
  • Denotes 95% Confidence Interval.

{3}------------------------------------------------

page 4 of 4

ACON COC-150 One Step Cocaine Test Strip versus data obtained with GC/MS at the cutoff concentration of 150 ng/mL

Benzoylecgonine Conc. vs. CutoffNegative-25% Cutoff to CutoffCutoff to +25% Cutoff> +25% Cutoff% Agreement with GC/MS
ACON COC-150Test DevicePositive00813398.6%(95.0% - 99.8%)**
ACON COC-150Test DeviceNegative15070298.7%(95.5% - 99.6%)**
  • Denotes 95% Confidence Interval

Performance Characteristics and Other information:

The performance characteristics of ACON COC-150 One Step Cocaine Test Strip, ACON COC-150 One Step Cocaine Test Device were verified by analytical sensitivity study, specificity and cross reactivity study, interference studies, precision study, read time flex study, temperature flex study, specimen storage and stability study. Study results indicate that these test devices are robust and can perform satisfactorily when used according to the "Indication for Use" statement specified in their package inserts.

Conclusion:

These clinical studies demonstrated substantial equivalency on performance between the ACON COC-150 One Step Cocaine Test Strip, ACON COC-150 One Step Cocaine Test Device and a FDA-cleared cocaine test with the same cocaine cutoff concentration. It is also demonstrated that these tests are safe and effective in qualitatively detecting Benzoylecgonine at a concentration of 150 ng/mL. The POL study demonstrated that these tests are suitable for healthcare professionals including professionals at point-of-care sites.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an abstract human figure with three arms or lines extending upwards, representing health and well-being. The graphic is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)", which is arranged in a circular fashion around the graphic.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MO 20850

FEB 1 7 2004

Edward Tung, Ph.D. Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121

K032903 Re:

Trade/Device Name: ACON® COC-150 One Step Cocaine Test Strip ACON® COC-150 One Step Cocaine Test Device Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: November 25, 2003 Received: January 6, 2004

Dear Dr. Tung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Jean M. Cooper, MS, DVM.

Jean M. Cooper, MS. D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

11. INDICATIONS FOR USE

510(k) Number:

ACON® COC-150 One Step Cocaine Test Strip Device Name: ACON® COC-150 One Step Cocaine Test Device

Indications for Use:

The ACON COC-150 One Step Cocaine Test Strip and ACON COC-150 One Step Cocaine Test Device are rapid chromatographic immunoassays for the qualitative detection of Benzoylecgonine levels in urine at a designated cutoff concentration of 150 ng/mL. They are intended for healthcare professionals including professionals at point-of-care sites.

(Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_

(Per 21 CFR 801.109)

Or

Over-The-Counter Use

Haleh Zandi
Revision Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K032923

વર્સ

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).