K Number
K032768

Validate with FDA (Live)

Date Cleared
2003-10-09

(31 days)

Product Code
Regulation Number
876.5010
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic, Inc. RACER Biliary Stent System is indicated for use in the palliation of malignant neoplasms in the biliary tree.

Device Description

The RACER Biliary Stent System consists of a MP35N (Cobalt, Nickel, Chromium, Molybdenum Alloy) stent mounted on an over-the-wire delivery system. The device is equipped with an AV100 balloon mounted on the distal portion of the catheter. Diameters of the device include 4.0, 5.0, 6.0, and 7.0mm and the lengths are either 12mm or 18mm, all premounted on 80cm or 130cm catheters.

AI/ML Overview

The Medtronic, Inc. RACER Biliary Stent System received 510(k) clearance based on substantial equivalence to a predicate device, the Bridge Constant Biliary Stent System (K030633). This type of clearance typically relies on a comparison of device characteristics and performance to a legally marketed device rather than a comprehensive study with pre-defined acceptance criteria for a new clinical endpoint.

However, based on the provided text, here's a breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission based on substantial equivalence, there are no explicit "acceptance criteria" for a new clinical outcome study as might be found for a PMA (Pre-Market Approval). Instead, the "acceptance criteria" are implied by demonstrating that the new device is as safe and effective as the predicate device. The "performance" is therefore measured against the predicate.

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (as described in the 510(k) summary)
Material Identity: Materials are identical to predicate.Constructed from identical biocompatible materials (MP35N alloy).
Technological Identity: Technology is identical to predicate.Balloon expandable and premounted on a sheathless delivery system, identical to predicate technology.
Intended Use Equivalence: Same indications for use as predicate.Indicated for use in the palliation of malignant neoplasm in the biliary tree, same as predicate.
Clinical Needs Equivalence: Meets clinical needs as well as predicate.Meets the clinical needs of physicians (assertion of substantial equivalence).
Safety and Effectiveness Equivalence: Safe and effective performance comparable to predicate.Preclinical testing conducted to confirm safe and effective performance and biocompatibility.
Lower Crossing Profile: The subject device should offer a lower crossing profile.The subject device offers a lower crossing profile (stated as a minor difference not relevant to the ability to palliate malignant neoplasms).
Sterilization: Provided sterile and not intended for reuse or resterilization.Provided sterile. Not intended for reuse or resterilization.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a clinical "test set" in the context of human subjects or patient data. The basis for clearance is preclinical testing and comparison to an existing device.

  • Sample Size for Test Set: Not applicable in the context of a human clinical test set for this 510(k) submission. Preclinical testing was mentioned but no sample sizes for those tests are provided.
  • Data Provenance: Not applicable in the context of a human clinical test set for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This 510(k) submission relies on preclinical testing and comparison to a predicate device, not on expert-adjudicated ground truth from a clinical test set.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring adjudication in the context described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No. This document describes a 510(k) submission for a medical device based on substantial equivalence, not an MRMC study comparing human reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

No. This device is a physical medical device (stent system), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for this 510(k) submission is established by:

  • The performance and safety profile of the predicate device (Bridge Constant Biliary Stent System, K030633) which has already been deemed safe and effective.
  • Preclinical testing results demonstrating the device's safe and effective mechanical and biological performance (e.g., biocompatibility, mechanical integrity).

8. The Sample Size for the Training Set

No "training set" in the context of AI/ML or a clinical study is mentioned. The device's design and manufacturing are based on established engineering principles and materials.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the described context. The performance and safety of the device are benchmarked against an existing, cleared predicate device and supported by standard preclinical engineering and biocompatibility tests.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the Medtronic logo along with the date October 9, 2003. The logo consists of a circular emblem with three figures in motion, followed by the company name "Medtronic" in a bold, sans-serif font. The date is positioned above the logo, indicating a specific time reference associated with the image.

: 上一篇:

K032768

510(k) Summary for the Medtronic, Inc. RACER Biliary Stent System

510(k) SummaryThis summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R. § 807.92.
SubmitterMedtronic, Inc.Peripheral Technologies3576 Unocal PlaceSanta Rosa, California 95407
Contact PersonSarah SheppardRegulatory Affairs ManagerPeripheral TechnologiesPhone: (707) 591-7465FAX: (707) 591-7406e-mail: sarah.sheppard@medtronic.com
Date PreparedSeptember 5, 2003
Trade NameMedtronic, Inc. RACER Biliary Stent System ("RACER")
Common NameBiliary Stent and Delivery System
Classification NameBiliary Catheter and Accessories
Device ClassificationClassification: Class IIClassification Panel: 78FGERegulation Number: 21 C.F.R. §876.5010
Predicate DeviceBridge Constant Biliary Stent System (K030633)(Previously named Bridge Symbiant)
Performance StandardsPerformance standards have not been established by the FDA under section 514 of the Federal, Food, Drug and Cosmetic Act

{1}------------------------------------------------

510(k) Summary for the Medtronic, Inc. RACER Biliary Stent System

Device DescriptionThe RACER Biliary Stent System consists of a MP35N (Cobalt, Nickel, Chromium, Molybdenum Alloy) stent mounted on an over-the-wire delivery system. The device is equipped with an AV100 balloon mounted on the distal portion of the catheter. Diameters of the device include 4.0, 5.0, 6.0, and 7.0mm and the lengths are either 12mm or 18mm, all premounted on 80cm or 130cm catheters.
Indications for UseThe RACER Biliary Stent System is indicated for use in the palliation of malignant neoplasm in the biliary tree.
TechnologicalCharacteristicsThe RACER Biliary Stent System is substantially equivalent to the currently cleared Bridge Constant Biliary Stent System (K030633). The subject and predicate stents are identical both in material and technology and are intended for palliation of malignant neoplasms in the biliary tree. The subject and predicate stents are constructed from identical biocompatible materials. The subject and predicate stents are both balloon expandable and premounted on a sheathless delivery system. The subject device offers a lower crossing profile. The subject and predicate stents are both intended to meet clinical needs. The difference between the subject and predicate devices are minor and are not relevant to the ability of the subject device to palliate malignant neoplasms in the biliary tree.
NonclinicalPerformancePreclinical testing was conducted to confirm the safe and effective performance of this device as well as the biocompatibility of the device.
SterilizationThe RACER Biliary Stent System is provided sterile. The device is not intended for reuse or resterilization.
ConclusionThe RACER Biliary Stent System is substantially equivalent to the currently cleared Bridge Constant device (K030633) and meets the clinical needs of the physicians.

:

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with three curved lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 9 2003

Ms. Sarah Sheppard Regulatory Affairs Manager Peripheral Technologies Medtronic AVE 3576 Unocal Place SANTA ROSA CA 95403

Re: K032768

Trade/Device Name: Medtronic, Inc. RACER Biliary Stent System Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: September 5, 2003 Received: September 9, 2003

Dear Ms. Sheppard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{3}------------------------------------------------

Page 2 - Ms. Sarah Sheppard

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Daniel G. Schultz, M.D.

Director Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Page _ 1 of 1

510(k) Number (if known): K032768

Device Name: Medtronic, Inc. RACER Biliary Stent System

FDA's Statement of the Indications For Use for device:

The Medtronic, Inc. RACER Biliary Stent System is indicated for use in the palliation of malignant neoplasms in the biliary tree.

Prescription Use OR
(Per 21 CFR 801.109)

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

Nancy C Brogdon

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.