K Number
K032129

Validate with FDA (Live)

Date Cleared
2003-07-17

(7 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Radius .018 Cougar Wire is intended to facilitate the placement of balloon dilitation catheters during PTCA and/or PTA. The Radius .018 Cougar Wires are compatible with all currently approved and marketed PTCA/PTA balloon catheters which are labeled for use with an .018 guidewire.

Device Description

The Radius .018 Cougar Wire is constructed from a composite stainless steel and Nitinol core to which a coil is attached to the tapered distal section. The proximal section of the wire is coated with PTFE, and the distal coil portion of the wire is hydrophilic coated. The device is packaged in a protective hoop sealed into a Tyvek/mylar pouch, and is sterilized using ETO gas.

AI/ML Overview

Here's the analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

Device: Radius .018 Cougar Wire (PTCA Guidewire)

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (Performance Tests Performed)Reported Device Performance
Tensile StrengthPerformed and met criteria
Torque StrengthPerformed and met criteria
TorqueabilityPerformed and met criteria
Tip FlexibilityPerformed and met criteria
Coating Adherence/IntegrityPerformed and met criteria

Note: The document states, "the proposed Radius .018 Cougar Wire meets the minimum requirements that are considered adequate for its intended use." This indicates that the device's performance in these in vitro tests met the pre-defined acceptance criteria, even though the specific numerical results or thresholds are not provided in this summary.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The document only mentions that "the following in vitro performance tests were performed." It does not provide the number of devices or iterations tested for each performance criterion.
  • Data Provenance: The tests are described as "in vitro performance tests," meaning they were conducted in a laboratory setting. There is no mention of country of origin for the data or whether it was retrospective or prospective, as these terms are typically relevant for clinical studies, not in vitro device performance testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as this document describes in vitro performance testing of a medical device, not a study involving human interpretation or subjective assessment where expert-established ground truth would be necessary. The ground truth for these tests would be determined by objective physical measurements against established engineering specifications.

4. Adjudication method for the test set:

This information is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used in studies where there is subjective interpretation and a need to resolve discrepancies between human readers. The tests performed here (Tensile Strength, Torque Strength, etc.) are objective physical measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The document describes the premarket notification for a PTCA guidewire, which is a physical medical device. It does not involve AI or human readers, so an MRMC comparative effectiveness study for AI assistance would not be relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. As mentioned in point 5, the device is a physical guidewire and does not involve an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for the performance tests (Tensile Strength, Torque Strength, Torqueability, Tip Flexibility, Coating Adherence/Integrity) would be based on engineering specifications and established industry standards for guidewire performance. These are objective, measurable characteristics, not subjective interpretations requiring expert consensus, pathology, or outcomes data.

8. The sample size for the training set:

This information is not applicable. The document describes the performance testing of a physical medical device, not a machine learning model that would require a training set.

9. How the ground truth for the training set was established:

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

JUL 1 7 2003

Attachment 6 510(k) Summary of Safety and Effectiveness July 9, 2003

GENERAL INFORMATION A.

Submitter's Name:Radius Medical Technologies, Inc.
Address:15 Craig RoadActon, MA 01720
Contact Person:Maureen A. Finlayson
Device Generic Name:PTCA Guidewire
Device Trade Name:Radius .018 Cougar Wire
Classification Name:Wire, Guide, Cardiovascular (74DQX)

INDICATIONS B.

The Radius .018 Cougar Wire is intended to facilitate the placement of balloon dilitation catheters during PTCA and/or PTA. The Radius .018 Cougar Wires are compatible with all currently approved and marketed PTCA/PTA balloon catheters which are labeled for use with an .018 guidewire.

DESCRIPTIVE CHARACTERISTICS C.

The Radius .018 Cougar Wire is constructed from a composite stainless steel and Nitinol core to which a coil is attached to the tapered distal section. The proximal section of the wire is coated with PTFE, and the distal coil portion of the wire is hydrophilic coated. The device is packaged in a protective hoop sealed into a Tyvek/mylar pouch, and is sterilized using ETO gas.

COMPARATIVE INFORMATION D.

The Radius .018 Cougar Wire is substantially equivalent to the currently marketed Radius Cougar Guidewire (K011287).

E. PERFORMANCE TESTING

The following in vitro performance tests were performed on the Radius .018 Cougar Wire:

    1. Tensile Strength
    1. Torque Strength
    1. Torqueability
    1. Tip Flexibility
    1. Coating Adherence/Integrity

CONCLUSION:

Based on the indications for use, technological characteristics, and safety and performance testing, the proposed Radius .018 Cougar Wire meets the minimum requirements that are considered adequate for its intended use.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The bird is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 17 2003

Radius Medical Technologies, Inc. c/o Debbie Iampiedro 7 Tiffany Trail Hopkinton, MA 01748

Re: K032129

Radius 018 Cougar Wire Regulation Number: 870.1330 Regulation Name: Cardiovascular Guidewire Regulatory Class: Class II Product Code: DQX Dated: July 9, 2003 Received: July 10, 2003

Dear Ms. Iampiedro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{2}------------------------------------------------

Page 2 - Debbie Iampietro

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

Bram D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Radius .018 Cougar Wire Device Name:

Indications For Use:

Radius .018 Cougar Wire is intended for placement of balloon dilatation catheters during PTCA and/or PTA.

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Prescription Use (Per 21 CFR 801.109) 510(k

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.