K Number
K030621

Validate with FDA (Live)

Device Name
ATAC CALIBRATOR
Manufacturer
Date Cleared
2003-03-19

(20 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ATAC Calibrator Kit is intended for use with the ATAC Clinical Systems to establish points of reference that are used in the determination of albumin, calcium, cholesterol, creatinine, glucose, magnesium, phosphorus, total bilirubin, total protein and urea in human specimens. This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.

Device Description

The ATAC Calibrator Kit is intended to calibrate the ATAC Clinical Systems for the quantitative determination of albumin, calcium, cholesterol, creatinine, glucose, magnesium, phosphorus, total protein and urea nitrogen.

AI/ML Overview

The provided document describes the ATAC Calibrator Kit and its performance in demonstrating substantial equivalence to a predicate device. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria for the regression statistics (e.g., specific ranges for slope and y-intercept). However, the implicit acceptance criterion is that the regression statistics demonstrate substantial equivalence to the comparative method, indicating accurate calibration.

AnalyteComparative MethodReported Device Performance (Regression Statistics)
AlbuminHitachi 704 / Roche Albumin Reagent, product 1970569Roche c.f.a.s Calibrator, product 759350y = 0.25 + 0.908x, n = 59
MagnesiumBeckman Synchron CX Magnesium Reagent, product 445360Beckman Synchron Multi Calibrator, product 442600y = 0.03 + 0.967x, n = 55
PhosphorusBeckman Synchron CX Phosphorus Reagent, product 465145Beckman Synchron Multi Calibrator, product 442600y = - 0.10 + 0.992x, n = 58
Total BilirubinBeckman Synchron CX Total Bilirubin Reagent, product 442745Beckman Synchron Bilirubin Calibrator, product 465915y = - 0.06 + 1.031x, n = 54
Total ProteinBeckman Synchron CX Total Protein Reagent, product 442740Beckman Synchron Multi Calibrator, product 442600y = 0.00 + 1.000x, n = 52

Additionally, for reconstituted stability:

  • Acceptance Criteria (Implicit): Statistically insignificant changes or changes less than round-off error/minimal clinical significance.
  • Reported Device Performance:
    • Changes in albumin, magnesium, bilirubin, and total protein over 3 days at 2°C to 8°C were statistically insignificant or less than the round-off error of the assay.
    • Observed change for phosphorus over three days was less than 0.2 mg/dL.

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size:
    • Albumin: 59 sera
    • Magnesium: 55 sera
    • Phosphorus: 58 sera
    • Total Bilirubin: 54 sera
    • Total Protein: 52 sera
  • Data Provenance: The data provenance is not explicitly stated (e.g., country of origin). The studies appear to be prospective in nature, as they involve assaying "at least 50 sera" for comparison with existing methods. It does not indicate the data is retrospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This type of study does not involve "experts" establishing ground truth in the way a medical imaging study might. Instead, the "ground truth" for the calibrator's performance is established by comparing its calibration results to those obtained using established, commercially available comparative methods with their respective calibrators and reagents. The document does not specify the qualifications of the individuals who performed these laboratory assays, but it would typically be trained laboratory personnel.

4. Adjudication Method for the Test Set

Not applicable. This is a quantitative laboratory calibration study comparing results to established methods, not a subjective interpretation task requiring adjudication by multiple readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

Not applicable. This is a calibration study for an in vitro diagnostic device, not a study involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this study essentially demonstrates the "standalone" performance of the ATAC Calibrator Kit in the context of the ATAC 8000 Random Access Chemistry System. It assesses the calibrator's accuracy by comparing results from the ATAC system (calibrated with the ATAC Calibrator) against established laboratory methods.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth used is the results obtained from established, commercially available comparative methods (e.g., Hitachi 704, Beckman Synchron CX reagents) which themselves are calibrated with their respective, validated calibrators. This serves as the reference standard for evaluating the accuracy of the ATAC Calibrator Kit.

8. The Sample Size for the Training Set

Not applicable. This is a calibration and method comparison study, not a machine learning study that typically involves distinct training and test sets in the same way. The calibrator itself (ATAC Calibrator Kit) is used to establish reference points, and its accuracy is then validated against other calibrated methods. There isn't an "algorithm" being trained in the conventional sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there isn't a "training set" in the context of a machine learning algorithm. The "ground truth" for the overall system's performance is established by the well-defined, validated methodologies of the comparative reference instruments and reagents.

{0}------------------------------------------------

MAR 1 9 2003

Image /page/0/Picture/1 description: The image shows the word "élan" in a stylized font. The "é" has a unique design, with the top part of the letter replaced by a textured, gradient-like effect that resembles a burst or spray. The rest of the letters are in a simple, sans-serif font, creating a contrast between the artistic "é" and the more straightforward "lan".

1075 W. Lambert Road Building D Brea, California 92821 T (714) 672 3553 F (714) 672 3554

SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS INFORMATION

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The ATAC Calibrator Kit is intended to calibrate the ATAC Clinical Systems for the quantitative determination of albumin, calcium, cholesterol, creatinine, glucose, magnesium, phosphorus, total protein and urea nitrogen. The ATAC Calibrator is substantially equivalent to the Beckman Synchron Multi-Calibrator, product no. 465915, which is currently marketed by Beckman Coulter, Inc. of Brea, CA. The effectiveness of ATAC Calibrator Kit on the ATAC 8000 Random Access Chemistry System is shown by the following studies.

The reconstituted stability claim is confirmed through the assay of albumin, total bilirubin and total protein in calibrators of increasing ages. The observed changes in the concentrations of albumin, magnesium, bilirubin and total protein over 3 days at 2°C to 8°C were statistically insignificant or less than the round-off error of the assay. The observed change for phosphorus over three days was less than 0.2 mg/dL.

The accuracy of the assigned set points was documented through method comparison studies. At least 50 sera were assayed for albumin, magnesium, phosphorus, total bilirubin and total protein over at least 4 analytical runs using a commercially available method and ATAC reagent applications calibrated with the ATAC Calibrator.

AnalyteComparative MethodRegression Statistics
AlbuminHitachi 704 / Roche Albumin Reagent, product 1970569Roche c.f.a.s Calibrator, product 759350y = 0.25 + 0.908x, n = 59
MagnesiumBeckman Synchron CX Magnesium Reagent, product 445360Beckman Synchron Multi Calibrator, product 442600y = 0.03 + 0.967x, n = 55
PhosphorusBeckman Synchron CX Phosphorus Reagent, product 465145Beckman Synchron Multi Calibrator, product 442600y = - 0.10 + 0.992x, n = 58
TotalBilirubinBeckman Synchron CX Total Bilirubin Reagent, product 442745Beckman Synchron Bilirubin Calibrator, product 465915y = - 0.06 + 1.031x, n = 54
Total ProteinBeckman Synchron CX Total Protein Reagent, product 442740Beckman Synchron Multi Calibrator, product 442600y = 0.00 + 1.000x, n = 52

Deming regression statistics are summarized below.

Wyman Stoching

n Stocking Manager of Regulatory Affairs Elan, Brea California

Image /page/0/Picture/12 description: The image shows a blurred, close-up view of a document or printed material. The text "www.jelar" is visible along the left edge, oriented vertically. The overall impression is that the image is capturing a detail from a larger document, possibly a website address or company name.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of flowing lines, possibly representing human profiles or interconnectedness.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 1 9 2003

Mr. Wynn Stocking Manager of Regulatory Affairs Elan Diagnostics 1075 W. Lambert Road Building D Brea, California 92821

Re: K030621

Trade/Device Name: ATAC Calibrator Kit Regulation Number: 21 CFR § 862.1150 Regulation Name: Calibrator, Multi - Analyte Mixture Regulatory Class: II Product Code: JIX Dated: February 26, 2003 Received: February 27, 2003

Dear Mr. Stocking:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{2}------------------------------------------------

Page 2 --

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

K030621

Device Name:

ATAC Calibrator

Indications for Use:

The ATAC Calibrator Kit is intended for use with the ATAC Clinical Systems to establish points of reference that are used in the determination of albumin, calcium, cholesterol, creatinine, glucose, magnesium, phosphorus, total bilirubin, total protein and urea in human specimens.

This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jean Cooper

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K030624

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.