K Number
K030234

Validate with FDA (Live)

Date Cleared
2003-04-28

(95 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and analytical laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/ml. in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample.

This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.

Device Description

The Cozart EIA Cocaine Oral Fluid Kit is a laboratory based test for the detection of cocaine and metabolites in human oral fluid using a cutoff equivalent to 30ng/mL. The Cozart EIA Cocaine Oral Fluid Kit is a competitive ELISA for the detection of cocaine and its metabolites in human oral fluid. The kit supplies the following reagents - a microtitre plate coated with antibody, enzyme conjugate reagent, wash buffer, substrate solution, stop solution and four calibrators (0, 5, 10 and 50ng/ml benzoylecgonine in oral fluid matrix).

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Cozart EIA Cocaine Oral Fluid Kit, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria CategoryReported Device PerformanceComments / Context
Method Comparison96% overall agreementCompared to GC/MS (Gas Chromatography/Mass Spectrometry), the gold standard.
PrecisionCVs less than 9% (kit)Acceptable for qualitative manual ELISA assay (CVs less than 10% acceptable).
Total precision (kit) CVs less than 11%
Sensitivity1.8 ng/mLDetection limit for cocaine in oral fluid.
SpecificityNo cross-reactivity with 29 potential interferentsThree out of four related compounds showed some cross-reactivity.
Cutoff Concentration ValidationAbsorbances for 5 ng/mL higher than 10 ng/mL cutoff calibrator; Absorbances for 15 ng/mL lower than 10 ng/mL cutoff calibratorConfirmed appropriate response around the 10 ng/mL cutoff.
Interference StudiesNo interference observedTested against alcohol, sample adequacy indicator dye, hemoglobin, smoking, coffee, tea, water, food, orange juice, hard candy, chewing gum, and mouthwash.
Stopped Assay StabilityMust be read within 15 minutes at 450 nmAbsorbance readings at various time points (0, 5, 10, 15, 30, 45, 60 minutes) showed stability up to 15 minutes.
Assay DriftLittle change across the plateSample addition must occur within 25 minutes.

Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 177 samples.
    • Data Provenance: Not explicitly stated, but the context of clinical and forensic samples suggests human samples. The country of origin is not specified, but the manufacturer is based in the UK. The study appears to be retrospective, as existing samples were tested.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable for this type of in-vitro diagnostic (IVD) device. The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS), which is an analytical method, not human expert consensus, for drug detection.
  3. Adjudication Method for the Test Set:

    • Not applicable. The reference method (GC/MS) served as the definitive ground truth, and the device's results were compared directly to it. There was no mention of human adjudication between different readers or methods beyond the GC/MS.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC study was not done. This device is an in-vitro diagnostic kit, not an imaging or diagnostic algorithm requiring human interpretation of output.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, this was a standalone performance study. The Cozart EIA Cocaine Oral Fluid Kit is a laboratory-based test that provides a qualitative result based on its biochemical reactions and spectrophotometric reading. Its performance was evaluated directly against GC/MS without human intervention in the interpretation of the kit's results.
  6. The Type of Ground Truth Used:

    • analytical reference method, specifically Gas Chromatography/Mass Spectrometry (GC/MS). The text explicitly states: "The device detailed above was compared to Gas Chromatography/Mass Spectrometry (GC/MS)." and "96% overall agreement as compared with GC/MS". GC/MS is widely considered the "gold standard" for confirmatory drug testing.
  7. Sample Size for the Training Set:

    • Not explicitly stated. The document describes a "method comparison" study as the primary performance evaluation. For a competitive ELISA kit, internal calibration and validation (which might involve a "training" equivalent) would have been performed by the manufacturer during its development, but details regarding a distinct "training set" sample size are not provided in this 510(k) summary.
  8. How the Ground Truth for the Training Set Was Established:

    • Not explicitly stated. Given that performance for the test set was established against GC/MS, it is highly probable that any internal development, calibration, or "training" (if applicable for this type of device) would also have referenced GC/MS or highly purified standards of cocaine and its metabolites at known concentrations.

{0}------------------------------------------------

K030234

PREMARKET NOTIFICATION [510(k)] SUMMARY

APR 2 8 2003

SubmitterCozart Bioscience Ltd45 Milton ParkAbingdonOxfordshire OX14 4RUUKTel No: 01235 861483Fax No: 01235 835607
Contact PersonDr Roberto LiddiQuality Assurance & Regulatory Affairs Manager
Date20th January 2003
Device NameCozart EIA Cocaine Oral Fluid Kit

Trade Name Cozart EIA Cocaine Oral Fluid Kit

Classification Name Cocaine Test System

Classification

Class II

Code of Federal Regulations Title 21 Food and Drugs Part 862 Clinical Chemistry and Clinical Toxicology Devices Subpart D Clinical Toxicology Test Systems 862.3250 Cocaine and cocaine metabolite test system

Establishment Registration No

3002336046

Performance Standards

BS EN ISO 9001:1994; EN 46001:1996

Substantial Equivalence

STC Opiates Micro-Plate EIA (Orasure® Application), 510(k) no. K981341

{1}------------------------------------------------

ParameterCozart EIA Cocaine Oral Fluid KitCocaine Metabolite Micro-Plate EIA(Orasure® Application), 510(k) no.K982061
Intended UseQualitative test for Cocaine in human oralfluid with a 30ng/ml cutoff. Recommendconfirmation of positive results by GC/MS.Qualitative test for Cocaine in human oralfluid with a 10ng/ml cutoff. Recommendconfirmation of positive results by GC/MS.
Target PopulationClinical and forensic samples.Clinical samples.
DesignCompetitive ELISACompetitive ELISA
EnzymeHorse Radish PeroxidaseHorse Radish Peroxidase
ResultsRead spectrophotometrically at 450nm.Read spectrophotometrically at 450nm.
Calibrators0, 5, 10, 50ng/mL0, 5, 10, 50ng/mL
MatrixHuman Oral FluidHuman Oral Fluid
ControlsNone supplied but Cozart recommendsusing external controls.Unknown
Method Comparison177 samples were tested, 100 screenedpositive for cocaine and metabolites, ofwhich 100 were confirmed positive byGC/MS. 77 samples screened negative forcocaine and metabolites and 71 wereconfirmed negative by GC/MS.92% Agreement as compared to GC/MS
PrecisionCV (%) of 1.9 – 10.7%CV (%) of 5 – 11.2%
Sensitivity1.8ng/mL<5ng/mL
Specificity29 potential interferents tested – nonecross-reactedUnknown

Introduction

The Cozart EIA Cocaine Oral Fluid Kit is a laboratory based test for the detection of cocaine and metabolites in human oral fluid using a cutoff equivalent to 30ng/mL. The device detailed above was compared to Gas Chromatography/Mass Spectrometry (GC/MS).

Cozart Bioscience Ltd is the manufacturer of the Opiates Oral Fluid Kit. We have not purchased this device from another manufacturer and the device is not marketed under another product name.

Intended Use

The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and forensic laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/mL in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample.

The remainder of this document will refer to the 10ng/mL cutoff concentration.

This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive

{2}------------------------------------------------

result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.

Target Population

The target population for the Cozart EIA Cocaine Oral Fluid Kit is clinical and forensic samples.

Where Used

The Cozart EIA Cocaine Oral Fluid Kit is designed for use in clinical and forensic laboratories. For professional use only.

Design

As can be seen from the Principle of the Test section in the pack insert, the Cozart EIA Cocaine Oral Fluid Kit is a competitive ELISA for the detection of cocaine and its metabolites in human oral fluid.

Materials

The Cozart EIA Cocaine Oral Fluid kit supplies the following reagents - a microtitre plate coated with antibody, enzyme conjugate reagent, wash buffer, substrate solution, stop solution and four calibrators (0, 5, 10 and 50ng/ml benzoylecgonine in oral fluid matrix).

Performance

Method Comparison

Cocaine Oral Fluid The Cozart EIA Kit was compared to Gas Chromatography/Mass Spectrometry (GC/MS). All the samples were tested with the Cozart EIA Cocaine Oral Fluid Kit according to the pack insert enclosed.

177 samples were tested through the Cozart EIA Cocaine Oral Fluid Kit, 100 screened positive and 100 were confirmed positive by GC/MS. 77 samples screened negative and 71 were confirmed negative by GC/MS. Of the 177 samples tested 15 were between -50% cutoff and +50% cutoff.

New DeviceGC/MSNegsNear CutoffGC/MSNegs *Near CutoffGC/MSPos **GC/MSPos***Percent AgreementwithGC/MS
Pos100004100100
Neg777165692
  • Between -50% Cutoff and the Cutoff.

** Between +50% Cutoff and the Cutoff.

*** Total number of positives (includes near cutoff samples).

96% overall agreement as compared with GC/MS

Cozart EIA Cocaine Oral Fluid Kit

{3}------------------------------------------------

Precision

The precision obtained for the Cozart EIA Cocaine Oral Fluid Kit produced CVs less than 9%. The total precision for the kit produced CVs less than 11%. The Cozart EIA Cocaine Oral Fluid Kit is a qualitative manual ELISA assay and CVs of less than 10% are acceptable for this assay type.

Sensitivity

The sensitivity of the Cozart EIA Cocaine Oral Fluid Kit is 1.8ng/ml.

Specificity

Twenty-nine potentially interfering unrelated substances were tested for cross reactivity in the Cozart Cocaine Oral Fluid Kit and none were found to cross react. Four related compounds were tested and three showed a level of cross reactivity.

Cutoff Concentration

Testing samples at the cutoff concentration, 50% above and 50% below were carried out to validate the cutoff concentration. The absorbances obtained for the 5ng/ml sample were all higher than the 10ng/ml cutoff calibrator. Similarly the absorbances obtained for the 15ng/ml sample were all lower than the 10ng/ml cutoff calibrator.

Interference Studies

A range of parameters including alcohol, sample adequacy indicator dye, haemoglobin, smoking, coffee, tea, water, food, orange juice, hard candy, chewing gum and mouthwash were tested for interference in the Cozart Cocaine Oral Fluid Kit. No interference was observed with any of the parameters.

Stopped Assay Stability

The stability of the stopped assay was investigated by reading the absorbance at 450nm at times 0, 5, 10, 15, 30, 45 and 60 minutes. The Cozart EIA Cocaine Oral Fluid Kit must be read within 15 minutes at 450mm.

Assay Drift

Sample addition at time 0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20, 22.5 and 25 minutes was investigated. Little change was observed across the plate and therefore sample addition to a Cozart EIA Cocaine Oral Fluid Kit must take place within 25 minutes.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.

APR 2 8 2003

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Dr. Roberto Liddi Quality Assurance & Regulatory Affairs Manager Cozart Bioscience Ltd 45 Milton Park Abingdon Oxfordshire OX14 4RU UK

Re: K030234

Trade/Device Name: Cozart EIA Cocaine Oral Fluid Kit Regulation Number: 21 CFR 862.3250 Regulation Name: Specimen transport and storage container Regulatory Class: Class II Product Code: DIO Dated: April 15, 2003 Received: April 18 2003

Dear Dr. Liddi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of

510(k) Number (if known): K 030 Z3 4

Device Name:__________________________________________________________________________________________________________________________________________________________________ Cozart EIA Cocaine Oral Fluid Kit

Indications For Use:

The Cozart EIA Cocaine Oral Fluid Kit is intended for use in clinical and analytical laboratories when used in conjunction with the Cozart RapiScan Oral Fluid collection system. Using this collection system it provides qualitative screening results for cocaine and metabolites in human oral fluid at a cutoff concentration of 10ng/ml. This is equal to 30ng/ml. in undiluted oral fluid as the collection system involves a 1:3 dilution of the sample.

This assay is for professional use only and provides only a preliminary analytical test result. Clinical consideration and professional judgement must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a more confirmed analytical result a more specific alternative chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method.

Sean Cooper

Division Sign-Off)
Clinical Labora
510(k) Number K 030234

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ry

(Optional Format 3-10-98)

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).