K Number
K030219

Validate with FDA (Live)

Date Cleared
2003-08-05

(195 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital temperature monitor, model KY-01 is the battery-operated electronic devices with intended use of measuring human body temperature precisely. This device is reusable and intended for abdominal temperature measurement of a body less then eight ages old, and the measurement of oral, armpit and rectal temperature for all ages.

Device Description

The Digital temperature monitor, model KY-01, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control. From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts. This system uses a 1.5V DC battery for operation of complete system whenever the battery is low. the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of KY-01, it was designed and verified according to the US standard ASTM E1112-98.

AI/ML Overview

Here's an analysis of the provided 510(k) summary, specifically focusing on acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document refers to compliance with ASTM E1112-98 as the primary standard for performance verification. This standard is for electronic thermometers. Therefore, the acceptance criteria are derived from this standard. The document doesn't explicitly state all performance metrics or their specific values from the standard, but it asserts compliance.

Acceptance Criteria (Based on ASTM E1112-98)Reported Device Performance (Model KY-01)
Accuracy (e.g., maximum error allowed at specific temperatures)"designed and verified according to the US standard ASTM E1112-98."
Response Time (e.g., time to stable reading)Implied to meet ASTM E1112-98 requirements. The document states it's for "measuring human body temperature precisely."
Repeatability/ReproducibilityImplied to meet ASTM E1112-98 requirements.
Stability (e.g., drift over time)Implied to meet ASTM E1112-98 requirements.
Operational Lifetime (e.g., battery life)The ASIC circuit detects low battery automatically.
Environmental Performance (e.g., temperature, humidity limits)Implied to meet ASTM E1112-98 requirements and EN standards (EN 60601-1, EN 60601-1-1, EN 60601-1-2).
Display Characteristics (e.g., readability)"LCD for displaying the measuring temperature digitally."
Material Biocompatibility (if applicable to patient contact)Not explicitly stated but implied by substantial equivalence and general safety requirements.
Electrical SafetyCompliance with EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirements.
Electromagnetic Compatibility (EMC)Compliance with EN 60601-1-2 requirements.

Study Proving Device Meets Acceptance Criteria:

The document states: "Regarding the performance of KY-01, it was designed and verified according to the US standard ASTM E1112-98." And also, "Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows: Compliance to applicable voluntary standards includes ASTM E1112: 1998, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement."

This indicates that the device underwent non-clinical testing to demonstrate compliance with these standards. The specific details of these tests (e.g., number of measurements, specific temperature points tested) are not provided in this summary.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set or any details about human subjects. The testing described is "non-clinical tests" and compliance with standards, which would typically involve laboratory measurements on the device itself, potentially using simulated body temperatures or calibrated constant temperature baths. Given the nature of a digital thermometer, a "test set" in the context of patient data (prospective or retrospective) is not the primary means of demonstrating compliance with performance standards like ASTM E1112.

The data provenance is not applicable in the human subject data sense, as the performance evaluation referenced appears to be engineering and laboratory-based. The device's manufacturer is Kang Ying Medical Appliances Inc. in Taiwan.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This question is not applicable in the traditional sense of clinical imaging or diagnostic device ground truth. The "ground truth" for a thermometer's accuracy would be established by reference to highly accurate, calibrated temperature standards in a laboratory setting, not by human experts. The involved "experts" would be metrology engineers or technicians. The document does not specify their number or qualifications.

4. Adjudication Method for the Test Set

This is not applicable as the evaluation described is laboratory-based compliance testing against a standard, not expert adjudication of a diagnostic finding.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a digital thermometer, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The device itself is a standalone electronic thermometer. Its performance, as measured by compliance with ASTM E1112-98, would be its "standalone" performance. The document doesn't describe separate "algorithm only" performance, as the device's functionality is directly tied to its physical components (thermistor, ASIC).

7. The Type of Ground Truth Used

The ground truth used for assessing the thermometer's performance would be:

  • Calibrated Reference Temperatures: Highly accurate and traceable temperature standards (e.g., from a national metrology institute) used in a laboratory setting to verify the accuracy of the thermometer's readings.
  • Compliance with ASTM E1112-98: This standard itself defines the acceptable deviations from true temperature for different temperature ranges, acting as the ground truth for performance.

8. The Sample Size for the Training Set

This is not applicable. The device is a digital thermometer that uses a thermistor and an ASIC for calculation. It is not an AI/machine learning device that requires a "training set" of data in the computational sense. Its "training" is in its design and calibration processes.

9. How the Ground Truth for the Training Set was Established

This is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Kang Ying Medical Appliances Inc. AUG - 5 2003

510(K) SUMMARY

ATTACHMENT >

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807,92.

The assigned 510(k) number is: __ K030219 Submitter's Identifications: Kang Ying Medical Appliances Inc. No. 44, Laen 1, Gangyan Rd., Gangshan Jen, Kaohsiung County, Taiwan, 820, R.O.C.

Contact: Mr. Ching-Chu, Chen / QA Manager Date of Summary Preparation: May 6, 2003

    1. Name of the Device: Digital temperature monitor, model KY-01.
    1. Information of the 510(k) Cleared Device (Predicate Device): Wireless thermometer monitor; model RT-910I (K003326).
    1. Device Description:

The Digital temperature monitor, model KY-01, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control.

From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

This system uses a 1.5V DC battery for operation of complete system whenever the battery is low. the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of KY-01, it was designed and verified according to the US standard ASTM E1112-98.

    1. Intended Use:
      The Digital temperature monitor, model KY-01 is the battery-operated electronic devices with intended use of measuring human body temperature precisely. This device is reusable and intended for abdominal temperature measurement of a body less then eight ages old, and the measurement of oral, armpit and rectal temperature for all ages.
    1. Comparison to the 510(k) Cleared Device (Predicate Device): The Digital temperature monitor, model KY-01 is substantially equivalent to the Funai model RT-9101 (K003326).
  • Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows: ଚ. Compliance to applicable voluntary standards includes ASTM E1112: 1998, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.

    1. Conclusions

The Kang Ying digital temperature monitor, model KY-01, has the same intended use and technological characteristics as the cleared device of Funai model RT-9101. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.

Page 1 of 1

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of three wavy lines, which represent the department's mission to protect the health of all Americans and provide essential human services.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 5 2003

Mr. Chen Ching-Chu Quality Assurance Manager Kang Ying Enterprise, Incorporated No. 25, Tai-I 1 Street Jen-Te Hsiang, Tainan County TAIWAN, R.O.C

Re: K030219

Trade/Device Name: Digital Temperature Monitor/Model: KY-01 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 24, 2003 Received: July 24, 2003

Dear Mr. Ching-Chu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Ching-Chu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runser

Susan Runner, DDDS, MA Interim Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

ATTACHMENT /

Indications for Use Statement

Page 1 of

510(k) Number (if known):_K030219

Device Name: Digital temperature monitor / Model: KY-01

Indications For Use :

The Digital temperature monitor, model KY-01 is the battery-operated electronic devices with intended use of measuring human body temperature precisely. This device is reusable and intended for abdominal temperature measurement of a body less then eight ages old, and the measurement of oral, armpit and rectal temperature for all ages.

Patricia Cuccente

ision Sian-Off ion of Anesthesiology, General Hospital. tion Control. Dental Devices

510(k) Number: K030219

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

1

(Optional Format 1-2-96)

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.