K Number
K024048

Validate with FDA (Live)

Date Cleared
2003-01-06

(28 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Express™ Biliary LD Premounted Stent System is indicated for the palliation of malignant neoplasms in the biliary tree.

Device Description

The 27 mm Express Biliary LD Premounted Stent System will be identical to the currently marketed Express Biliary LD Premounted Stent Systems (K021630). The Express Biliary LD Stent is a balloon expandable metallic stent intended to maintain patency of biliary strictures produced by malignant neoplasms. The stent will be available in a variety of sizes to address clinician needs. The Express Biliary LD Stent is made from 316L surgical grade stainless steel tubing. The seamless tubing is initially extruded to a cylindrical shape, and is then drawn down in a series of steps to the final tubing dimension. The stent is formed by laser cutting the pattern from the tube, after which it is cleaned and electropolished to obtain smooth rounded struts. The geometry is a continuous pattern consisting of large and small sinusoidal bands connected by axial struts. The deployed stent provides radial strength while conforming to the natural curvature of the anatomy. The stent provides a ghost like image using conventional radiographic imaging equipment. The currently marketed Express Biliary LD Premounted Stent System is offered with stent diameters of 6 - 10 mm, in one mm increments. The 6 - 8 mm diameter stents has lengths of 17 mm, 37 mm, and 57 mm. The 27 mm length is proposed for these stent diameters and is the subject of this submission. These are considered to be the Small Lumen (SL) models. The 9 mm and 10 mm diameter stents have lengths of 25 mm. 37 mm, and 57 mm. These are considered to be the Large Lumen (LL) models. The balloon delivery catheter for the 27 mm Express Biliary LD Premounted Stent System will be identical to the one utilized on the currently marketed Express Biliary LD Premounted Stent Systems (K021630). The Balloon Delivery Catheter is an over-the-wire catheter offered in a two lumen catheter shaft design. One lumen is used to pass the catheter over a guide wire. The proposed device is designed to be placed over guide wires which have outer diameters of 0.035" or smaller. This lumen can also be used for infusion of contrast medium. The second lumen communicates with the balloon and is used to inflate and deflate the balloon during the stenting procedure. The guide wire lumen and the balloon lumen terminate at the proximal end of the catheter by means of a bifurcated hub with luer lock fittings.

AI/ML Overview

This document describes the regulatory submission (K024048) for the Boston Scientific Express Biliary LD Premounted Stent System (27 mm). The submission asserts substantial equivalence to a previously marketed device (K021630) and does not contain detailed primary study data or explicit acceptance criteria with numerical performance targets and reported device performance.

Therefore, many of the requested sections regarding specific study details, such as sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, cannot be extracted from the provided text.

Based on the provided text, here's what can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of numerical acceptance criteria with corresponding device performance metrics. Instead, it describes general functional testing and states that "Test results verified that the Express Biliary LD Premounted Stent System is adequate for its intended use."

Acceptance Criteria CategoryReported Device Performance
Stent Component Functional Testing:Test results verified adequacy for intended use.
- Tensile and elongation(Implicitly met standards)
- Bile corrosion resistance(Implicitly met standards)
- Dimensional verification (strut widths, stent lengths)(Implicitly met standards)
- Foreshortening(Implicitly met standards)
- Recoil(Implicitly met standards)
- Expansion uniformity(Implicitly met standards)
- Compression resistance(Implicitly met standards)
- Over expansion(Implicitly met standards)
- Deployment accuracy(Implicitly met standards)
Balloon Delivery Catheter Functional Testing:Test results verified adequacy for intended use.
- Multiple balloon inflation within a stent(Implicitly met standards)
- Balloon burst within a stent(Implicitly met standards)
- Balloon inflation and deflation time(Implicitly met standards)
- Balloon proximal bond tensile(Implicitly met standards)
- Stent / balloon crossing profile(Implicitly met standards)
- Sheath insertion / withdrawal force(Implicitly met standards)
- Stent securement force(Implicitly met standards)
- Tracking(Implicitly met standards)
- Hub to shaft tensile(Implicitly met standards)
BiocompatibilityTest results verified adequacy for intended use.
2-year product shelf lifeTest results verified adequacy for intended use.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not specified.
  • Data Provenance: Not specified. The document describes non-clinical (functional and biocompatibility) testing. It does not mention clinical data or human subject data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not applicable. The document describes non-clinical engineering and biological testing, not expert-derived ground truth from clinical cases.
  • Qualifications of Experts: Not applicable.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Adjudication Method: Not applicable. This concept applies to expert review of clinical cases. The testing described is objective, functional testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No, an MRMC comparative effectiveness study was not done.
  • Effect Size of AI Assistance: Not applicable, as this is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance: Not applicable, as this is a medical device (stent system), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Ground Truth: For the non-clinical tests described, the "ground truth" would be established engineering specifications, material standards, and validated test methods (e.g., tensile strength, corrosion resistance, dimensional tolerances). There is no reference to expert consensus, pathology, or outcomes data for establishing ground truth in this non-clinical submission.

8. The sample size for the training set:

  • Sample Size for Training Set: Not applicable. This is a medical device, not an AI model, so there is no concept of a "training set" in the context of machine learning.

9. How the ground truth for the training set was established:

  • Ground Truth for Training Set: Not applicable.

{0}------------------------------------------------

KO24048

Summary of Safety and Effectiveness Section 6

(Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)

6.1 General Provisions

Submitter's Name andAddressBoston Scientific CorporationOne Scimed PlaceMaple Grove, Minnesota 55311
Contact PersonTodd Kornmann(763) 494-2467
Classification NameBiliary Catheter and AccessoriesProduct Code – 78 FGERegulation Number 21 CFR Part 876.5010
Common or Usual NameBiliary Stent and Balloon DilatationCatheter
Proprietary NameBoston Scientific CorporationExpress Biliary LD Premounted Stent System
Name of Predicate DeviceBoston Scientific Express Biliary LDPremounted Stent System

6.3 Device Description

Stent Description

6.2

The 27 mm Express Biliary LD Premounted Stent System will be identical to the currently marketed Express Biliary LD Premounted Stent Systems (K021630).

The Express Biliary LD Stent is a balloon expandable metallic stent intended to maintain patency of biliary strictures produced by malignant neoplasms. The stent will be available in a variety of sizes to address clinician needs.

The Express Biliary LD Stent is made from 316L surgical grade stainless steel tubing. The seamless tubing is initially extruded to a cylindrical shape, and is then drawn down in a series of steps to the final tubing dimension.

{1}------------------------------------------------

K024048

Section 6

Summary of Safety and Effectiveness

The stent is formed by laser cutting the pattern from the tube, after which it is cleaned and electropolished to obtain smooth rounded struts.

The geometry is a continuous pattern consisting of large and small sinusoidal bands connected by axial struts. The deployed stent provides radial strength while conforming to the natural curvature of the anatomy. The stent provides a ghost like image using conventional radiographic imaging equipment.

The currently marketed Express Biliary LD Premounted Stent System is offered with stent diameters of 6 - 10 mm, in one mm increments. The 6 - 8 mm diameter stents has lengths of 17 mm, 37 mm, and 57 mm. The 27 mm length is proposed for these stent diameters and is the subject of this submission. These are considered to be the Small Lumen (SL) models. The 9 mm and 10 mm diameter stents have lengths of 25 mm. 37 mm, and 57 mm. These are considered to be the Large Lumen (LL) models.

Balloon Delivery Catheter

The balloon delivery catheter for the 27 mm Express Biliary LD Premounted Stent System will be identical to the one utilized on the currently marketed Express Biliary LD Premounted Stent Systems (K021630).

The Balloon Delivery Catheter is an over-the-wire catheter offered in a two lumen catheter shaft design. One lumen is used to pass the catheter over a guide wire. The proposed device is designed to be placed over guide wires which have outer diameters of 0.035" or smaller. This lumen can also be used for infusion of contrast medium.

The second lumen communicates with the balloon and is used to inflate and deflate the balloon during the stenting procedure. The guide wire lumen and the balloon lumen terminate at the proximal end of the catheter by means of a bifurcated hub with luer lock fittings.

A more detailed description of the devices is provided in Section 8.

6.4 Intended Use

The Express Biliary LD Premounted Stent System is indicated for the treatment of biliary strictures produced by malignant neoplasms.

{2}------------------------------------------------

K02 4043

Section 6

Summary of Safety and Effectiveness

6.5 Summary of Technological Characteristics

The Boston Scientific 27 mm Express Biliary LD Premounted Stent System will incorporate the identical design, method of deployment, fundamental technology, manufacturing, packaging, labeling, sterilization, and intended use as those in the currently marketed the Express Biliary LD Premounted Stent System (K021630).

Non-clinical Test Summary 6.6

Functional testing presented for the stent component consisted of tensile and elongation, bile corrosion resistance, dimensional verification (strut widths, stent lengths), foreshortening, recoil, expansion uniformity, compression resistance, over expansion, and deployment accuracy.

Functional testing presented for the balloon delivery catheter consisted of multiple balloon inflation within a stent, balloon burst within a stent, balloon inflation and deflation time, balloon proximal bond tensile, stent / balloon crossing profile, sheath insertion / withdrawal force, stent securement force, tracking, and hub to shaft tensile.

Biocompatibility and 2 year product shelf life have also been conducted. Test results verified that the Express Biliary LD Premounted Stent System is adequate for its intended use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 6 - 2003

Mr. Todd Kornmann Senior Regulatory Affairs Specialist Boston Scientific Scimed, Inc. One Scimed Place Maple Grove, Minnesota 55311-1566

Re: K024048

Trade/Device Name: Express™ Biliary LD Premounted Stent System (27 mm) Regulation Number: 21 CFR $876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: December 6, 2002 Received: December 9, 2002

Dear Mr. Kornmann:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{4}------------------------------------------------

Page 2 - Mr. Todd Kornmann

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification''(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sincerely yours,

Daniel C. Schultz, M.D.

Daniel G. Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K024048

Device Name: Express™ Biliary LD Premounted Stent System (27 mm)

FDA's Statement of the Indications For Use for device:

The Express™ Biliary LD Premounted Stent System is indicated for the palliation of malignant neoplasms in the biliary tree.

OR Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use_

Nancy C. Thornton

and 510(k) Nul

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.