K Number
K023971

Validate with FDA (Live)

Date Cleared
2003-01-27

(59 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Greiner VACUETTE® Trace Elements Evacuated Blood Collection Tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory.

Device Description

VACUETTE® Tubes are used to collect, transport and process blood for testing serum, plasma or whole blood in the clinical laboratory. The VACUETTE® Trace Elements tube with sodium heparin may be used to collect a whole blood/plasma sample. The VACUETTE® Trace Elements tube with no additive may be used to collect a serum sample. The tubes are composed of clear plastic. The caps are roval blue and made of plastic and rubber: the inner cap rings are black and made of plastic. The tubes' size is 13 x 75 mm. 6mL draw. The tubes are equipped with a vacuum tube holder to assist in positioning the product when obtaining blood samples. The vacuum tube holder is composed of plastic.

AI/ML Overview

The Greiner VACUETTE® Trace Elements Tubes were evaluated through two studies to demonstrate substantial equivalence to a predicate device.

1. Table of Acceptance Criteria & Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance
No background presence of trace elements in the tubes."Test results for trace elements to be claimed showed no background presence of trace elements in the tubes."
Equivalent performance with blood samples compared to the predicate device."equivalent performance with blood samples."

2. Sample Size and Data Provenance

  • Sample Size (Test Set): 40 blood donors for the comparison testing with blood samples. The sample size for the deionized water testing is not specified, but it implies multiple tubes were tested.
  • Data Provenance: Not explicitly stated, but given the submission is to the FDA, it is highly likely the studies were conducted in the US. The studies were likely prospective, as they involved collecting blood from donors specifically for the purpose of testing the new device.

3. Number of Experts and Qualifications

  • This information is not provided in the document. The studies focused on analytical performance (presence of trace elements, comparison of results) rather than diagnostic interpretation requiring expert consensus.

4. Adjudication Method

  • Not applicable. The studies involved analytical testing of trace element levels, not subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This type of study (comparing human reader performance with and without AI assistance) is not relevant for this device, which is a blood collection tube and not an AI diagnostic tool.

6. Standalone Performance Study

  • Yes, implicitly. The study testing for the "background presence of trace elements in the tubes using deionized water" functions as a standalone performance assessment to ensure the tubes themselves do not contribute contaminants. The comparison testing with blood samples also assesses the device's performance in its intended use, albeit in comparison to a predicate.

7. Type of Ground Truth Used

  • The ground truth for the deionized water study was the absence of trace elements in the deionized water itself. The device's performance was measured against this known clean baseline.
  • The ground truth for the comparison testing with blood samples was the predicate device's performance. The results obtained from the Greiner VACUETTE® tubes were compared to those obtained using the Becton Dickinson Vacutainer® Trace Elements tube. The assumption is that the predicate device provides accurate (or substantially equivalent) measurements of trace elements in blood.

8. Sample Size for the Training Set

  • Not applicable. This device is a blood collection tube, not an algorithm that requires a training set.

9. How Ground Truth for Training Set was Established

  • Not applicable. As stated above, there is no training set for this device.

{0}------------------------------------------------

JAN 2 7 2003

November 27, 2002

510(K) SUMMARY

Greiner VACUETTE® Trace Elements Tubes

Greiner VACUETTE® North America. Inc. P.O Box 1026 Monroe, NC 28111

For information regarding this 510(k) Summary, please contact Greiner VACUETTE® North America, Douglas L. Harris.

Device Names:

Proprietary Name: VACUETTE® Trace Elements Tubes

Common Name: Blood Collection Tubes

Classification Name: Tubes, Vials, Systems, Serum Separators, Blood Collection

Device Description:

VACUETTE® Tubes are used to collect, transport and process blood for testing serum, plasma or whole blood in the clinical laboratory. The VACUETTE® Trace Elements tube with sodium heparin may be used to collect a whole blood/plasma sample. The VACUETTE® Trace Elements tube with no additive may be used to collect a serum sample. The tubes are composed of clear plastic. The caps are roval blue and made of plastic and rubber: the inner cap rings are black and made of plastic. The tubes' size is 13 x 75 mm. 6mL draw. The tubes are equipped with a vacuum tube holder to assist in positioning the product when obtaining blood samples. The vacuum tube holder is composed of plastic.

Intended Use:

VACUETTE® Tubes, Holders and Needles are used together as a system for the collection of venous blood. The Greiner VACUETTE® Trace Elements tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory.

Substantial Equivalence:

The Greiner VACUETTE® Trace Elements tubes are substantially equivalent to the Becton Dickinson Vacutainer® Trace Elements tube with sodium heparin (pre-amendment). The blood collection tubes have the same intended use and contain the same tube material and stopper material; the VACUETTE® Trace Elements tube containing sodium heparin contains the same anticoagulant.

Two studies were conducted on the tubes – testing for presence of trace elements in the tubes using deionized water and comparison testing using blood collected from 40 blood donors. Test results for trace elements to be claimed showed no background presence of trace elements in the tubes and equivalent performance with blood samples.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Greiner Bio-One Vacuette® North America c/o Ms. Judi Smith Principal Sienna Partners, L.L.C. P.O. Box 103 Baldwin, MD 21013

JAN 2 7 2003

K023971 Re:

Trade/Device Name: Greiner VACUETTE® Trace Elements Tubes Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: November 27, 2002 Received: November 29, 2002

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Ko2397/ 510(k) Number (if known):

Device Name: Greiner VACUETTE® Trace Elements Evacuated Blood Collection Tubes with sodium heparin or no additive

Indications For Use:

Greiner VACUETTE® Trace Elements Evacuated Blood Collection Tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lan V.A. La De Cooper

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K093471

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96)

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.