Intended Use

The Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump is intended for use by qualified surgeons to provide irrigation/aspiration to the surgical site during arthroscopic surgical procedures in the joints of the shoulder, knees, ankle, elbow, wrist, and hip.

Device Description

The Karl Storz ARTHROPUMP POWER is a combined irrigation and suction pump for dilation and irrigation of joint capsules with a fluid. Both roller pumps are software controlled and automatically manage fluid and joint pressure based on settings chosen by the user. If desired, both flow and manage hald and joint proodividually adjusted. The device is used as part of a system that includes the unit, a footswitch and tubing, and can support integrated use of a shaver.

AI/ML Overview

The provided document is a 510(k) summary for the KARL STORZ ARTHROPUMP POWER Irrigation/Suction Pump. This type of document typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed acceptance criteria and performance metrics for a novel AI/software device.

Based on the information available in the document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states: "Test results show that all specifications have been met." However, it does not provide a table of
specific acceptance criteria or quantitative performance metrics (e.g., accuracy, precision, etc.) for the device. The focus is on functional, performance, and safety testing to demonstrate that the device meets its design specifications, but these specifications are not detailed in the summary.

2. Sample size used for the test set and the data provenance:

Not applicable. The document does not describe a test set in the context of an AI/software device evaluation. Performance testing mentioned is for the physical device's function, performance, and safety, not for evaluating an algorithm's output on a dataset.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This information is relevant for AI/software devices where human experts establish ground truth for testing.

4. Adjudication method for the test set:

Not applicable. This is typically used in the context of expert review of AI/software outputs.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, an MRMC study was not done. The device is an irrigation/suction pump, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI/software algorithm. The device itself is a physical pump with software control. The document mentions "microprocessor controlled dual peristaltic pump technology" and that "both roller pumps are software controlled," but this refers to the operational control of the pump, not an AI algorithm performing a diagnostic or analytical task.

7. The type of ground truth used:

Not applicable. The "ground truth" concept as applied to AI/software performance (e.g., pathology, outcomes data) is not relevant for this device. The "ground truth" for this device would be its correct physical function, pressure maintenance, and flow rates according to engineering specifications.

8. The sample size for the training set:

Not applicable. This device is not an AI/software algorithm that requires a training set.

9. How the ground truth for the training set was established:

Not applicable.

Summary:

The KARL STORZ ARTHROPUMP POWER is a medical device (an irrigation/suction pump) that utilizes microprocessor-controlled software to manage its mechanical functions. The provided 510(k) summary focuses on demonstrating its substantial equivalence to existing predicate devices (FMS Duo and Karl Storz Arthropump Plus).

The "study" or testing described within this document is typical for a hardware device with embedded software:

  • Non-Clinical Performance Data: "The KARL STORZ ARTHROPUMP POWER has been tested for function, performance and safety. Test results show that all specifications have been met."

This statement indicates that the device underwent engineering and functional testing to verify that it meets its pre-defined design specifications for parameters like pressure control, flow rates, and safety mechanisms. However, the exact specifications and the detailed results of these tests are not provided in this summary. The acceptance criteria would be those specific engineering and functional specifications, and the "study" is the internal design verification and validation testing conducted by the manufacturer.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "KARL STORZ - ENDOSKOPE" in a bold, sans-serif font. The text is white, and the background is dark and textured. The text appears to be a logo or title, possibly for a medical device company or product.

KARL STORZ Endoscopy-America, Inc. 2151 E. Grand Avenue El Segundo, California 90245

Toll Free 800 421 0837 Fax 800 321 1304

OCT 18 2010
ko9347)

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document are accurate and complete to the best of KSEA's knowledge.

Submitter:

Karl Storz Endoscopy-America, Inc. 2151 E. Grand Avenue El Segundo, CA 90245-5017 Phone: (424) 218-8381 (424) 218-8519 Fax:

Contact Person:

Regulatory Affairs Specialist Email: cdizol@ksea.com

Crystal Dizol

Date Prepared: September 16, 2010

Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump Device Trade Name:

Common Name: . Arthroscopic pump

Classification Name: Arthroscope and accessories

Regulation Number: 21 CFR 884.1730

HRX Product Code:

Predicate Device(s):

Future Medical Systems, Inc .: FMS Duo (K954465) Karl Storz: Arthropump Plus (K971649)

Device Description:

The Karl Storz ARTHROPUMP POWER is a combined irrigation and suction pump for dilation and irrigation of joint capsules with a fluid. Both roller pumps are software controlled and automatically manage fluid and joint pressure based on settings chosen by the user. If desired, both flow and manage hald and joint proodividually adjusted. The device is used as part of a system that includes the unit, a footswitch and tubing, and can support integrated use of a shaver.

Intended Use:

The Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump is intended for use by qualified The Ran Glorz An 11 in For only 1 Sweet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . the joints of the shoulder, knees, ankle, elbow, wrist, and hip.

Pape 3 of 3

{1}------------------------------------------------

Technological Characteristics:

The KARL STORZ ARTHROPUMP POWER Irrigation Pump utilizes microprocessor controlled dual peristaltic pump technology to maintain user-set pressure and flow, and to also perform lavage and hemostasis as necessary.

Non-Clinical Performance Data:

The KARL STORZ ARTHROPUMP POWER has been tested for function, performance and safety Test results show that all specifications have been met.

Determination of Substantial Equivalence:

The Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump is substantially equivalent to the predicate devices since the basic features, design, and intended use are similar. The minor differences between the Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump and the predicate devices raise no new issues of safety and effectiveness, as these design differences have no effect on the function or intended use of the devices. Where differences in performance or technology exist, it has been demonstrated that they do not adversely impact safety or effectiveness.

Conclusions:

The Karl Storz ARTHROPUMP POWER is substantially equivalent to the identified predicate devices and does not raise any new issues of safety and efficacy.

Att: Substantial Equivalence Table for Karl Storz ARTHROPUMP POWEF

2023 02 15
.
.
.
.

P. 2/2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Karl Storz Endoscopy-America, Inc. % Ms. Crystal Dizol Regulatory Affairs Specialist 2151 East Grand Avenue El Segundo, California 90245-5017

OCT 1 8 2010

Re: K093471

Trade/Device Name: Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: September 16, 2010 Received: September 20, 2010

Dear Ms. Dizol:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Ms. Crystal Dizol

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

for Peter D
Mark N. Melkerson
DEP Dir

Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

OCT 1 8 2010

510(k) Number (if known): K093471

Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump Device Name:

Indications for Use: The Karl Storz ARTHROPUMP POWER Irrigation/Suction Pump is intended for use by qualified surgeons to provide irrigation/aspiration to the surgical site during arthroscopic surgical procedures in the joints of the shoulder, knees, ankle, elbow, wrist, and hip.

× Prescription Use: (21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use: _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nittededen for nxa

sion Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of

510(k) Number K093471

§ 888.1100 Arthroscope.

(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.