K Number
K023693

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-03-12

(131 days)

Product Code
Regulation Number
880.5045
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The M-100 system is intended for use in filtering airborne particles from air for medical purposes.

Device Description

The M-100 is a portable HEPA-Filtered clean hood for use in providing a controlled environment to class 100 (ISO class 3,5) for medical applications that require a high degree of airborne particulate control. The system is controlled by embedded firmware and runs on standard 120volt, 5 amp power. Pushbutton Controls include airflow, light intensity, lift/lower. Manual controls include Filter tilt/angle, brakes.

AI/ML Overview

The provided text describes the Airex Inc. M-100 Mobile Medical Air Cleaner, a HEPA filtration system. However, it does not contain any information regarding acceptance criteria, device performance metrics, or any study details that would typically be associated with AI/ML medical devices.

The document is a 510(k) summary for a physical medical device (an air cleaner) from 2003, which precedes the widespread application of AI in medical devices and the associated regulatory requirements for studies proving AI performance.

Therefore, I cannot fulfill your request for the specific information points related to AI/ML device performance and study data (like sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment) because this information is not present in the provided text.

The closest relevant information provided is about the device's compliance with safety standards for physical medical equipment and air cleanliness:

  • UL 2601 Standard for Medical Electrical Equipment Part 1: General Requirements for Safety
  • ISO 14644-1 Classification of Air Cleanliness, Cleanrooms & Associated Controlled Environments, 1999.

These standards serve as "acceptance criteria" in a broad sense for the physical device's safety and air filtration capabilities, but they do not relate to the performance of an AI algorithm.

{0}------------------------------------------------

Airex Inc.

Image /page/0/Picture/2 description: The image shows a logo for AIREX. The word "AIREX" is written in white, bold, sans-serif font on a black rounded rectangle. The logo is surrounded by a grid of circles. The circles above the rectangle are filled in, while the circles below the rectangle are hollow.

MAR 1 2 2003

510(K) SUMMARY

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807,92, Content and Format of a 510(k) Summary.

  • Submitted By: 1. Airex Inc. 13704 SE 17th Street Bellevue, WA 98005
    1. Contact Person: William Haslebacher, President, CTO
  • Date Prepared: 3. January 26, 2003
    1. Proprietary Name; M-100 Mobile Medical Air Cleaner
    1. Common/ Usual Name: HEPA Filtration system

6. Classification Name: Sec. 880.5045 Medical Recirculating Air Cleaner. A device designed to remove particles from the air for medical purposes

7. Predicate Device:

The M-100 is substantially equivalent to the Micron 800M Air Purifier system by Biological Controls Inc (K974682, January23, 1998) in its design and its intended use.

8. Device Description:

The M-100 is a portable HEPA-Filtered clean hood for use in providing a controlled environment to class 100 (ISO class 3,5) for medical applications that require a high degree of airborne particulate control. The system is controlled by embedded firmware and runs on standard 120volt, 5 amp power. Pushbutton Controls include airflow, light intensity, lift/lower. Manual controls include Filter tilt/angle, brakes.

The M-100 has been designed to meet the following product safety standards;

  • UL 2601 Standard for Medical Electrical Equipment Part 1: General Requirements l for Safety
  • ISO 14644-1 Classification of Air Cleanliness, Cleanrooms & Associated Controlled Environments, 1999.

{1}------------------------------------------------

9. Intended Uses:

The M-100 system is intended for use in filtering airborne particles from air for medical purposes.

10, Technological Comparison to Predicate Device:

The M-100 is similar to the predicate device in that:

  • t Both system generate HEPA filtered air for medical purposes
  • . Both are portable, mobile
  • Both systems are UL approved .

The M-100 differs from the predicate device in that it can be used to direct highly filtered air to a specific location by means of available tilt and axial adjustments to the filter head assembly.

End of 510(k) Summary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design with three curved lines representing the bird's head and neck. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. The logo is black and white.

MAR 1 2 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. William Haslebacher President Airex Incorporated 13704 SE 17th Street Bellevue, Washington 98005

Re: K023693

Trade/Device Name: M-100 Mobile Medical Air Cleaner Regulation Number: 880.5045 Regulation Name: Medical Recirculating Air Cleaner Regulatory Class: II Product Code: FRF Dated: January 28, 2003 Received: January 29, 2003

Dear Mr. Haslebacher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Mr. Haslebacher

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Susan Roane

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Airex Inc.

Indications for Use

Page 1 of 1

510(k) Number (if known):

Device Name: M-100 Mobile Medical Air Cleaner

Indications for Use:

The M-100 system is intended for use in filtering airborne particles from air for medical purposes.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chiu S. Lim

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental D

510(k) Number. K023693

K023693addtofileA.doc

§ 880.5045 Medical recirculating air cleaner.

(a)
Identification. A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.(b)
Classification. Class II (performance standards).