K Number
K023593

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-02-06

(104 days)

Product Code
Regulation Number
866.5870
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the quantitative detection of IqG antibodies to thyroid peroxidase in human serum by enzyme immunoassay. The SeraQuest Anti-thyroid peroxidase test can be used with other serological tests and clinical findings, to aid in the diagnosis of thyroid diseases such as Autoimmune Thyroiditis and Graves' disease. For in vitro diagnostic use only.

Device Description

The SeraQuest Anti-Thyroid Peroxidase test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minutes incubations. It has been developed to detect IgG antibodies which are directed against Thyroid Peroxidase, in human serum. The Calibrators in the SeraQuest Anti-Thyroid Peroxidase test kit have been assigned values based on the NIBSC standard. Test results are reported as international units per milliliter (IU/mL).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the SeraQuest® Anti-Thyroid Peroxidase device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" but rather presents a clinical study comparing the new device (SeraQuest) against a predicate device (Varelisa TPO Antibodies test). The performance metrics of the SeraQuest device against the predicate are what would typically be used to demonstrate equivalence for regulatory acceptance.

MetricAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (SeraQuest vs. Varelisa)
Relative SensitivityHigh (e.g., comparable to predicate)100% (95% CI: 99.9-100%)
Relative SpecificityHigh (e.g., comparable to predicate)99.1% (95% CI: 97.7-100%)
Overall AgreementHigh (e.g., comparable to predicate)99.2% (95% CI: 97.3-100%)

Note: The document states, "The SeraQuest Anti-Thyroid Peroxidase test is substantially equivalent in intended use and performance, to the Pharmcia Varelisa TPO Antibodies, Freiburg, Germany." This indicates that the performance of the predicate device serves as the implicit acceptance criteria for the new device.


2. Sample Size and Data Provenance for the Test Set

  • Sample Size: 141 serum specimens.
  • Data Provenance: The study was conducted at Quest International, Inc., implying the data was collected or processed within the United States. The retrospective or prospective nature of the sample collection is not explicitly stated, but the specimens were "tested concurrently" by both devices, suggesting a prospective application to existing samples or a concurrent collection setup.

3. Number and Qualifications of Experts for Ground Truth

  • The document does not mention the use of experts to establish a "ground truth" in the traditional sense (e.g., a panel of radiologists for imaging).
  • Instead, the "ground truth" for the performance evaluation was established by comparing the SeraQuest device's results against a legally marketed predicate device, the Pharmacia Varelisa TPO Antibodies test. The predicate device itself acts as the reference standard.

4. Adjudication Method for the Test Set

  • No explicit adjudication method (like 2+1, 3+1) is described for the test set.
  • The comparison was a direct concurrent testing of each specimen by both the SeraQuest device and the Varelisa TPO Antibodies test, with results compared qualitatively (positive, equivocal, negative). For discordant results, one specimen was re-tested using a second legally marketed device (Scimedix Anti-Thyroid Peroxidase Test) to help resolve the discrepancy, but this wasn't a systematic adjudication process for all disagreements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. This device is an in-vitro diagnostic (IVD) test, not an imaging AI tool requiring human reader interpretation or assistance.

6. Standalone Performance Study

  • Yes, a standalone performance study was done in the sense that the SeraQuest device's results were independently generated for each specimen and then compared to the predicate. The performance metrics (sensitivity, specificity, agreement) directly reflect the algorithm's (or device's) standalone diagnostic capabilities relative to the predicate. There is no human-in-the-loop component for these specific performance metrics.

7. Type of Ground Truth Used

  • The ground truth used was the results obtained from a legally marketed predicate device (Pharmacia Varelisa TPO Antibodies test). For one discordant case, a second legally marketed device (Scimedix Anti-Thyroid Peroxidase Test) was used as an additional reference. This is a form of comparative ground truth against an established, similar diagnostic method.

8. Sample Size for the Training Set

  • The document does not provide information on a training set or its sample size. This is typical for IVD assay submissions where the focus is on analytical and clinical performance against a reference method rather than AI algorithm training.

9. How Ground Truth for the Training Set Was Established

  • Since no training set is mentioned, information on how its ground truth was established is not provided.

{0}------------------------------------------------

KO23593 510(k) SUMMARY

Applicant:Quest International, Inc.1938 N.E. 148th TerraceNorth Miami, FL 33181FEB 06 2003
Registration No.1061839
Contact Person:Robert A. Cort, V.P., Quality Assurance
Telephone:(305) 948-8788
Telefax:(305) 948-4876
Manufacturing Site:Same as above
Device:SeraQuest® Anti-Thyroid Peroxidase
Device Name:Anti-Thyroid Peroxidase, Multiple autoantibodies immunological tessystem (21CFR § 866.5660)5870
Device Classification:Class II (Performance Standards)

Description:

The SeraQuest Anti-Thyroid Peroxidase test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minutes incubations. It has been developed to detect IgG antibodies which are directed against Thyroid Peroxidase, in human serum.

The Calibrators in the SeraQuest Anti-Thyroid Peroxidase test kit have been assigned values based on the NIBSC standard. Test results are reported as international units per milliliter (IU/mL).

Principle:

Diluted samples are incubated in wells coated with thyroid peroxidase antigen. Antibodies directed against this antigen (if present) are immobilized in the wells. Residual sample is eliminated by washing and draining, and conjugate (enzyme-labeled antibodies to human IgG) is added and incubated. If IgG antibodies to thyroglobulin antigen are present, the conjugate will be immobilized in the wells. Residual coniugate is eliminated by washing, and the enzyme substrate is added and incubated. In the presence of the enzyme, the substrate is converted to a yellow end-product which is read photometrically at 405 nm.

Intended Use:

iended Use: The Anti-Thyroid Peroxidase test is intended for the quantative detection of human IgG antibodies to thyroid peroxidase antigen, in human serum by enzyme immunoassay. The presence of anti-thyroid peroxidase antibodies can be used with other serological tests and clinical findings to aid

{1}------------------------------------------------

diagnosing individuals with Autoimmune Thyroditis and Grave's Disease. For In Vitro Diagnostic Use Only.

Predicate Device:

The SeraQuest Anti-Thyroid Peroxidase test is substantially equivalent in intended use and performance, to the Pharmcia Varelisa TPO Antibodies, Freiburg, Germany.

Summary of Technological Characteristics:

CharacteristicSeraQuestAnti-Thyroid PeroxidaseTestPharmaciaVarelisaTPO Antibodies Test
Description:Enzyme ImmunoassayEnzyme Immunoassay
Intended Use:The detection of IgGantibodies againstthyroid peroxidasein human serum.The detection of IgGantibodies againstthyroid peroxidasein human serum.
Solid Phase:Polystyrene MicrowellPolystyrene Microwell
Antigen:Purifiedthyroid peroxidase(human thyroid)Recombinantthyroid peroxidase
Number of Incubation Periods:ThreeThree
Sample Dilution:1:511:101
Sample IncubationDuration:30 minutes30 minutes
Incubation Temperature:Room temperatureRoom temperature
Ezyme-labeled Conjugate:
AntibodyGoat anti-human IgGGoat anti-human IgG
EnzymeAlkaline phosphataseHorse Radish Peroxidase
Conjugate Volume:100 μl100 μl
Conjugate IncubationDuration:30 minutes30 minutes
Substrate:p-NitrophenylphosphateTMB
Subtrate Volume:100 μl100 μl
Substrate IncubationDuration:30 minutes10 minutes
Stop Reagent:0.5 M Trisodiumphosphate0.34 MSulfuric acid
Stop Reagent Volume:100 μl50 μl
Readout:SpectrophotometricSpectrophotometric
Wavelength:Reference Wavelength:405 nm620 nm450 nm
Normalization:Standard CurveStandard Curve
Reporting Unit:IU / mLIU / mL

{2}------------------------------------------------

APPENDIX 3.

Summary of Clinical Testing:

Experimental Procedure

To challenge the cutoff values, 141 serum specimens were tested at Quest International, Inc., concurrently by the SeraQuest Anti-Thyroid Peroxidase test, and the Varelisa TPO Antibodies test (Pharmacia & Upjohn). The assays were performed and interpreted according to the instructions in the manufacturer's package inserts.

Results and Conclusion

The qualitative agreement between the SeraQuest and the Pharmacia tests is shown in Table 1.

Of the 141 specimens tested, 22 were positive, 107 were negative and 1 was equivocal in both the SeraQuest and Varclisa tests (please see Table C-3). Of the 11 specimens remaining, 1 specimen which was negative by the Varelisa test, was positive by the SeraQuest test; 1 specimen which was negative in the Serclivest test, was equivocal by the Varelisa test; 7 which were equivocal in the SeraQuest test, 2 vere positive and 5 negative in the Varelisa test; 2 which were equivocal in the Varelisa test were positive in the SeraQuest test.

Excluding the equiv cal results, the sensitivity of the SeraQuest Anti-Thyroid Peroxidase (TPO) test relative to the Varaint test was 100%, or 99.9% to 100% (95% C.I.); the specificity was 99.1%, or 97.7% to 100% (95% C.I.); respectively. The overall agreement was 99.2%, or 97.3% to 100% (95% C.I.) (please see Talla C-3).

{3}------------------------------------------------

ABLE 1.

RESULTS OF SeraQuest ANTI-THYROGLOBULIN ASSAYS AND PHARMACIA ANTI-THYROGLOBULIN ASSAYS ON 141 SERUM SPECIMENS.

SeraQuest
PHARMACIAPOSEQUNEG%95% C.I.**
POS2220Relative Sensitivity*10099.9-100
EQU211
NEG15107Relative Specificity*99.197.7-100
Overall Agreement*99.297.3-100
  • Excluding equivocal results.

** 95%Confidence Interval calculated by the normal method.

The specimen that gave a discordant result was tested by a second legally marketed device, the Scimedix Anti-Thyroid Peroxidase Test, Scimedix Corp., Denville, New Jersey. The sample gave a ാsitive result with the Scimedix test.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for the Department of Human Services USA. The logo consists of a circular border with the text "HUMAN SERVICES USA DEPARTMENT" arranged around it. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked one behind the other, with flowing lines suggesting movement or connection.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FEB 0 6 2003

Mr. Robert A. Cort Vice President, Quality Assurance Quest International, Inc. 1938 N.E. 148th Terrace North Miami, FL 33181

Re: K023593

Trade/Device Name: SeraQuest Anti Thyroid Peroxidase (TPO) Regulation Number: 21 CFR 866.5870 Regulation Name: Thyroid Autoantibody Immunological Test System Regulatory Class: Class II Product Code: JZO Dated: December 23, 2002 Received: December 24, 2002

Dear Mr. Cort:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 –

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. . . - -

APPENDIX 6

Page

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: SeraQuest Anti-Thyroid Peroxidase (TPO)

Indications For Use:

  • For the quantitative detection of IqG antibodies to thyroid peroxidase in human ﺎ . serum by enzyme immunoassay.
  • The SeraQuest Anti-thyroid peroxidase test can be used with other serological 2. tests and clinical findings, to aid in the diagnosis of thyroid diseases such as Autoimmune Thyroiditis and Graves' disease.
    1. For in vitro diagnostic use only.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

A. Plewes for S. Bautista

Division of Clinical Labora 510/k) Number

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96)

§ 866.5870 Thyroid autoantibody immunological test system.

(a)
Identification. A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.(b)
Classification. Class II (performance standards).