K Number
K023362

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-12-09

(63 days)

Product Code
Regulation Number
864.5400
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fisher Diagnostics ThromboScreen® 1000 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma specimens in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), and Fibrinogen.

The Pacific Hemostasis® Fibrinogen Reagent plus Kaolin is intended to be used on the Fisher Diagnostics ThromboScreen® 1000 Coagulation Instrument for the quantitative determination of fibrinogen in plasma.

Device Description

The ThromboScreen® 1000 (TS1000) is a photo-optical instrument used for the performance of in-vitro diagnostic clotting procedures in the clinical laboratory. The instrument utilizes photo-optical principles to measure and record the time required for subject plasma specimens to clot. The TS1000 light source is provided by a 660 nm LED. The incubator block is temperature regulated to 36.5 - 37.5°C and contains six measuring positions and six reagent positions.

The Pacific Hemostasis® Fibrinogen Reagent plus Kaolin is identical to the Pacific Hemostasis® Thrombin for Fibrinogen Kit, except that the thrombin is reconstituted with water containing kaolin rather than water. Kaolin is added to increase the visibility of the clot in the stirred reaction cell.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Fisher Diagnostics ThromboScreen® 1000 and Pacific Hemostasis® Fibrinogen Reagent plus Kaolin, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative manner (e.g., "The correlation coefficient must be >0.95"). Instead, it presents comparative performance data against predicate devices to demonstrate substantial equivalence. The implication is that the performance must be comparable to the legally marketed predicate devices.

Test (Reagent, Unit)Performance MetricPredicate Device Performance (MLA 900C/1600C) (Implied acceptance range)ThromboScreen® 1000 Performance
Method Comparison Studies (Correlation Coefficient, r)
Prothrombin Time (PT) (Thromboplastin DS, seconds) - General Clinical SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.98, 0.99, 0.97
Prothrombin Time (Thromboplastin DS, INR) - General Clinical SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.99, 0.98, 0.98
Prothrombin Time (PT) (Thromboplastin DS, seconds) - Coumadin SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.96, 0.97, 0.96
Prothrombin Time (Thromboplastin DS, INR) - Coumadin SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.96, 0.97, 0.96
Activated Partial Thromboplastin Time (APTT-LS reagent, seconds) - General Clinical SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.99, 0.98, 0.98
Activated Partial Thromboplastin Time (APTT-LS reagent, seconds) - Heparin SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.94, 0.97, 0.94
Fibrinogen Concentration (mg/dL) - General Clinical SamplesCorrelation Coefficient (r)Close to 1.0 (as shown by predicate)0.97, 0.95, 0.96
Precision Studies (%CV)
PT - Within-Run PrecisionRange of %CV(Comparable to MLA 900C/1600C)1.4% - 5.8%
APTT - Within-Run PrecisionRange of %CV(Comparable to MLA 900C/1600C)1.2% - 3.4%
Fibrinogen Concentration - Within-Run PrecisionRange of %CV(Comparable to MLA 900C/1600C)1.2% - 3.4%
PT - Between-Run Precision (Total %CV)Range of %CV(Comparable to MLA 900C/1600C)1.8% - 9.4% (Across sites, within/run/day)
APTT - Between-Run Precision (Total %CV)Range of %CV(Comparable to MLA 900C/1600C)2.4% - 7.5% (Across sites, within/run/day)
Fibrinogen - Between-Run Precision (Total %CV)Range of %CV(Comparable to MLA 900C/1600C)3.3% - 8.5% (Across sites, specific concentration)

Important Note: The acceptance criteria are implicitly met by demonstrating "substantial equivalence" to the predicate devices. The correlation coefficients are all very high (close to 1.0), and the precision data (CV%) also appears comparable between the ThromboScreen 1000 and the predicate MLA instruments.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Sizes:

    • Method Comparison (Correlation Studies):
      • Prothrombin Time (General Clinical Samples): 60 samples per site (across 3 sites = 180 total)
      • Prothrombin Time (Coumadin Samples): 100 samples per site (Site 1 & 2), 92 samples (Site 3) = 292 total
      • Activated Partial Thromboplastin Time (General Clinical Samples): 58 (Site 1), 60 (Site 2), 60 (Site 3) = 178 total
      • Activated Partial Thromboplastin Time (Heparin Samples): 60 samples per site (across 3 sites = 180 total)
      • Fibrinogen Concentration (General Clinical Samples): 28 (Site 1), 30 (Site 2), 30 (Site 3) = 88 total
    • Precision Studies: The specific number of samples for precision studies is not explicitly stated, but it involves "Coag 1, Coag 2, Coag 3" which likely represent different control levels or concentrations, and were performed "within-run," "run-to-run," and "day-to-day" at three sites.
  • Data Provenance: The studies were conducted "in-house and at two external testing laboratories." While specific countries are not mentioned, the context of the FDA submission (U.S. Department of Health & Human Services) strongly suggests the data was generated within the United States or from sites compliant with U.S. regulatory standards. The data is prospective as it involves comparison testing and precision measurements on collected specimens. The document indicates specimens were collected from "apparently healthy individuals and from subjects with different pathological conditions which are expected to affect the results for a particular assay."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

The description does not mention the use of "experts" in the sense of human readers adjudicating results or establishing ground truth for the test results. This is an in-vitro diagnostic device (IVD) where the "ground truth" is typically established by comparing the device's quantitative measurements against a legally marketed predicate device (MLA 900C/1600C) using the same patient samples and/or established reference methods for precision. The performance is assessed by correlation coefficients and precision metrics against these established methods, not by expert consensus on interpretations.

4. Adjudication Method for the Test Set:

Not applicable. As described above, this is an IVD device measuring quantitative coagulation parameters. The "ground truth" is based on the results from predicate devices and internal validation, not human adjudication of subjective interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for imaging or diagnostic interpretation devices where multiple human readers interpret cases. The ThromboScreen® 1000 is an automated instrument for quantitative laboratory tests (Prothrombin Time, APTT, Fibrinogen), not an interpretation aid for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, a standalone performance evaluation was done. The entire study focuses on the performance of the ThromboScreen® 1000 instrument and the Pacific Hemostasis® Fibrinogen Reagent plus Kaolin. Its measurements are compared directly to those of established predicate instruments (MLA 900C/1600C). There is no "human-in-the-loop" aspect to its measurement process; it's an automated photo-optical instrument.

7. The Type of Ground Truth Used:

The ground truth for the device's performance was established by:

  • Predicate Device Comparison: The results obtained from the ThromboScreen® 1000 were compared against results from legally marketed predicate devices, the MLA 900C and MLA 1600C, using the same "citrated plasma specimens."
  • Internal Validation/Reference Methods: For precision studies, it implies that control materials or repeated measurements on samples were used to establish the variability (CV%) of the device itself and in comparison to the predicate devices.

Therefore, the ground truth is based on comparative measurements to predicate devices and established laboratory standards for precision and accuracy, rather than pathology, expert consensus, or outcomes data in the typical sense for imaging or pathology devices.

8. The Sample Size for the Training Set:

The document does not contain information about a "training set" or "training data." This type of terminology is usually associated with machine learning or artificial intelligence models. The ThromboScreen® 1000 is a photo-optical instrument. Its design and operational parameters would be developed through engineering and calibration processes, not a "training set" in the AI sense.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no mention of a "training set" in the context of this device's development or evaluation.

{0}------------------------------------------------

11.0PREMARKET NOTIFICATION 510(K) SUMMARY
---------------------------------------------
DEC 09 2002
Applicant:Jerald Steiner
Fisher Diagnostics
8365 Valley Pike
Middletown, Virginia 22645
Telephone: 540 869 8382
Fax: 540 869 8108
Contact:Jerald Steiner
Telephone: 540 869 8382
Fax: 540 869 8108
Date:10/4/02
Trade Name:Fisher Diagnostics ThromboScreen® 1000
Pacific Hemostasis® Fibrinogen Reagent
plus Kaolin
Common Name:Automatic Coagulation Instrument
Fibrinogen Reagent
Classification Name:Coagulation Instrument
21 CFR 864.5400
Fibrinogen Test
21 CFR 864.7340
Equivalent Devices:MLA 900C (K884863)
MLA 1600C (K931206)
Fibrinogen Reagent (K781880)
K023362
-----------

Description of the Devices

The ThromboScreen® 1000 (TS1000) is a photo-optical instrument used for the performance of in-vitro diagnostic clotting procedures in the clinical laboratory. The instrument utilizes photo-optical principles to measure and record the time required for subject plasma specimens to clot. The TS1000 light source is provided by a 660 nm LED. The incubator block is temperature regulated to 36.5 - 37.5°C and contains six measuring positions and six reagent positions. A detailed description of the device, including an explanation of how it functions, is described in the TS 1000 Operator's Manual, Section l , Introduction.

The Pacific Hemostasis® Fibrinogen Reagent plus Kaolin is identical to the Pacific Hemostasis® Thrombin for Fibrinogen Kit, except that the thrombin is reconstituted with water containing kaolin rather than water. Kaolin is added to increase the visibility of the clot in the stirred reaction cell.

{1}------------------------------------------------

Intended Use of the Devices

The Fisher Diagnostics ThromboScreen® 1000 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma specimens in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT). Activated Partial Thromboplastin Time (APTT), and Fibrinogen.

The Pacific Hemostasis® Fibrinogen Reagent plus Kaolin is intended to be used on the Fisher Diagnostics ThromboScreen® 1000 Coagulation Instrument for the quantitative determination of fibrinogen in plasma.

Summary of Substantial Equivalence Comparisons

The ThromboScreen® 1000 (TS1000) was compared to the MLA 900C and the MLA 1600C (K884863 & K931206, respectively). All three instruments have a similar intended use: for in-vitro diagnostic coagulation testing in the clinical laboratory. Further, the proposed device and the predicate devices have the same measurement system for clotting assays: photo-optical clot detection systems.

The TS1000 is an automated coagulation instrument. In contrast, the MLA 900C is semiautomated and the MLA 1600C is a fully-automated instrument. The MLA 900C requires manual save an automatic pipetting system, which adds both sample and test reagent. The light source for the MLA instruments is a Halogen lamp and the wavelength is set at 550 nm (for clotting assays). In contrast, the TS1000 utilizes an LED optic at 660 nm. Although differences in light source and wavelength exist, all instruments have been optimized for their light source/filter combinations. The performance data generated support this statement (Tables 1-3).

Comparison testing was performed in-house and at two external testing laboratories using Pacific Hemostasis (PH) brand reagents. As part of this submission, kaolin was added to the fibrinogen reagent used on the TS1000 to optimize detection. Specimens were evaluated from apparently healthy individuals and from subjects with different pathological conditions which are expected to affect the results for a particular assay. Table 1 summarizes the results of the comparison studies between the proposed and the predicate devices.

{2}------------------------------------------------

Test(Reagent, Unit)Site & Sample #CorrelationCoefficient, rRegressionEquation
Prothrombin Time(PT)(Thromboplastin DS,seconds)General ClinicalSamplesSite 1 - 60Site 2 - 60Site 3 - 600.980.990.97y = 1.475x - 6.66y = 1.045x - 1.26y = 1.126x - 2.51
Prothrombin Time(Thromboplastin DS,INR)Site 1Site 2Site 30.990.980.98y = 1.437x - 0.41y = 1.037x - 0.11y = 1.031x - 0.02
Prothrombin Time(PT)(Thromboplastin DS,seconds)CoumadinSamplesSite 1 - 100Site 2 - 100Site 3 - 920.960.970.96y = 1.142x - 1.16y = 1.142x - 1.27y = 0.924x + 6.04
Prothrombin Time(Thromboplastin DS,INR)Site 1Site 2Site 30.960.970.96y = 1.066x + 0.14y = 0.984x + 0.13y = 0.786x + 0.68
Activated PartialThromboplastin Time*(APTT-LS reagent,seconds)General ClinicalSamplesSite 1 - 58Site 2 - 60Site 3 - 600.990.980.98y = 1.189x - 2.99y = 1.205x - 4.24y = 1.161x - 2.93
Heparin SamplesSite 1 - 60Site 2 - 60Site 3 - 600.940.970.94y = 1.219x - 6.16y = 0.896x + 6.87y = 1.108x - 3.23
FibrinogenConcentration(mg/dL)General ClinicalSamplesSite 1 - 28Site 2 - 30Site 3 - 300.970.950.96y = 0.735x + 43.3y = 0.696x + 56.09y = 0.818x - 22.07

Table 1 Summary of Method Comparison Studies Between the ThromboScreen® 1000 and the MLA 900C/1600C

Precision studies were also performed to assess the performance of the TS1000. The following coefficients of variation were obtained for within-run and between-run precision studies:

{3}------------------------------------------------

TS1000MLA 900C/1600C
TestCoag 1Coag 2Coag 3Coag 1Coag 2Coag 3
PT
Site 12.2%3.2%5.8%2.9%5.1%3.1%
Site 21.9%2.3%4.8%
Site 31.4%2.2%2.8%
APTT
Site 11.2%2.0%3.0%0.9%2.0%0.7%
Site 21.6%3.4%3.1%
Site 31.5%1.8%2.0%
FibrinogenConcentration
Site 11.2%2.0%3.0%0.9%2.0%0.7%
Site 21.6%3.4%3.1%
Site 31.5%1.8%2.0%

Table 2 Summary of Within-Run Precision Studies, %CV

{4}------------------------------------------------

Coag 1Coag 2Coag 3
TS1000
Site 1
Within-run2.3%3.0%4.9%
Run-to-run2.8%4.3%8.0%
Day-to-day3.6%4.5%9.4%
Total2.5%3.7%6.6%
Site 2
Within-run1.1%1.7%2.3%
Run-to-run2.1%2.6%6.9%
Day-to-day1.8%2.6%7.4%
Total1.8%2.2%5.1%
Site 3
Within-run1.8%1.8%2.9%
Run-to-run2.7%3.2%4.4%
Day-to-day3.4%3.5%4.2%
Total2.3%2.6%3.7%
MLA
Site 1
Within-run1.8%2.0%2.4%
Run-to-run2.5%8.8%8.8%
Day-to-day3.0%6.4%8.5%
Total2.2%6.4%6.4%
Site 2
Within-run2.1%2.7%2.6%
Run-to-run2.9%9.3%9.4%
Day-to-day3.2%10.3%11.1%
Total2.5%6.8%6.8%
Site 3
Within-run1.5%2.2%2.3%
Run-to-run3.9%6.9%8.2%
Day-to-day4.3%6.9%9.6%
Total3.0%5.1%6.0%
Coag 1Coag 2Coag 3
TS1000Site 1
Within-run1.7%2.1%2.7%
Run-to-run2.9%5.3%2.9%
Day-to-day2.2%7.2%3.3%
Total2.4%4.0%2.9%
Site 2
Within-run1.4%1.7%1.9%
Run-to-run4.9%4.9%4.8%
Day-to-day4.5%4.5%5.9%
Total3.8%3.8%4.4%
Site 3s
Within-run2.2%1.7%2.3%
Run-to-run6.0%7.1%10.4%
Day-to-day7.3%7.2%13.2%
Total4.5%5.1%7.5%
MLASite 1
Within-run1.5%0.4%0.7%
Run-to-run2.6%3.6%3.8%
Day-to-day2.9%4.4%3.6%
Total2.2%2.5%2.7%
Site 2
Within-run1.1%0.8%0.9%
Run-to-run3.1%3.3%3.4%
Day-to-day3.2%3.3%3.8%
Total2.3%3.7%2.5%
Site 3
Within-run0.9%0.9%0.8%
Run-to-run6.1%8.0%7.6%
Day-to-day5.8%10.7%8.9%
Total4.2%5.6%5.3%

Table 3a Summary of Between-Run Precision Testing - PT

{5}------------------------------------------------

Table 3b Summary of Between-Run Precision Testing - APTT

{6}------------------------------------------------

Low FibCoag 1High Fib
timeconctimeconctimeconc
TS1000Site 1
Within-run3.2%4.1%5.2%
Within-day5.4%5.3%6.4%
Run-to-run6.7%5.3%8.2%
Day-to-day8.0%8.2%4.2%3.5%9.1%7.7%
Total5.1%6.8%4.7%4.5%6.8%7.1%
Site 2
Within-run3.5%2.4%4.1%
Within-day3.6%3.8%6.5%
Run-to-run4.4%4.1%8.7%
Day-to-day5.2%5.4%3.7%3.4%8.9%7.6%
Total4.1%4.6%3.3%3.5%6.1%7.1%
Site 3
Within-run4.9%2.3%3.6%
Within-day8.7%5.5%7.5%
Run-to-run7.8%6.9%9.4%
Day-to-day7.3%8.1%7.6%6.5%10.6%9.0%
Total6.5%8.5%5.1%6.0%7.4%8.2%

Table 3b Summary of Between-Run Precision Testing - Fibrinogen

In conclusion, the similar intended use, technological characteristics and performance data support the claim that the ThromboScreen® 1000 is substantially equivalent to the MLA 900C and the MLA 1600C.

Furthermore, the TS1000 specific fibrinogen assay reagent, Fibrinogen Reagent plus Kaolin, is substantially equivalent the Pacific Hemostasis® Fibrinogen Reagent.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Jerald Steiner, Ph.D. Senior Applications Scientist Fisher Diagnostics 8365 Valley Pike P.O. Box 307 Middletown, Virginia 22645-0307

DEC 0 9 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Re: K023362

Trade/Device Name: Fisher Diagnostics ThromboScreen 1000; Pacific Hemostasis Fibrinogen Reagent plus Kaolin

Regulation Number: 21 CFR § 864.5400 Regulation Name: Coagulation Instrument Regulatory Class: II Product Code: GKP, GIS Dated: October 4, 2002 Received: October 7, 2002

Dear Dr. Steiner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{8}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

K023362

Device Name:

Pacific Hemostasis® Fibrinogen Reagent plus Kaolin

Sponsor Name:

Fisher Diagnostics

Indications for Use:

The Pacific Hemostasis® Fibrinogen Reagent plus Kaolin is intended to be used on the Fisher Diagnostics ThromboScreen® 1000 Coagulation Instrument for the quantitative determination of fibrinogen in plasma.

Do Not Write Below This Line - Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use Over-The-Counter Use

12

Jouchina Bautista

(Division Sign-Off) Division of Clinical Laboratory Devices 023362 510(k) Number -

{10}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

KO23362

Fisher Diagnostics ThromboScreen® 1000

Sponsor Name:

Device Name: :

Fisher Diagnostics

Indications for Use:

The Fisher Diagnostics ThromboScreen® 1000 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma specimens in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), and Fibrinogen.

Do Not Write Below This Line - Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use Over-The-Counter Use

17 0

Giuseppina Fantini

(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number .

§ 864.5400 Coagulation instrument.

(a)
Identification. A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.(b)
Classification. Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.