K Number
K023217

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-10-25

(29 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

Device Description

The device description of the proposed SMART™ Control™ Nitinol Stent Transhepatic Biliary System is as follows.
• 6 French stent delivery system profile;
• Stent material - Nickel Titanium alloy and tantalum micromarkers;
• Expanded stent diameters 9 and 10 mm;
• Stent lengths: 80 mm;
• Stent delivery system usable length 80 and 120 cm; Guidewire lumen 0.035";
• Proximal Deployment Handle.

AI/ML Overview

The provided text is related to a 510(k) submission for a medical device (SMART™ Control™ Nitinol Stent Transhepatic Biliary System) and describes its substantial equivalence to predicate devices, rather than a study outlining acceptance criteria and detailed device performance metrics.

Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth for an AI/ML device is not applicable or cannot be extracted from the given document.

However, I can provide the following based on the available text:

Acceptance Criteria and Device Performance (Not Applicable - This document is a 510(k) summary for a substantial equivalence determination for a medical stent, not a performance study with explicit acceptance criteria for a new device's efficacy as per your request's format.)

The document focuses on establishing substantial equivalence to predicate devices through pre-clinical testing, rather than defining and proving specific performance metrics against pre-defined acceptance criteria for a novel AI/ML device.


Here's an attempt to answer the questions based on the type of document provided:

  1. A table of acceptance criteria and the reported device performance

    • Not applicable. This document is a 510(k) summary for a medical stent, asserting "substantial equivalence" to predicate devices, not reporting performance against explicit, quantifiable acceptance criteria in the manner typically seen for AI/ML diagnostic devices. The equivalence was confirmed through "pre-clinical testing," but no specific metrics or targets are provided in this summary.
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This document does not pertain to a study with a "test set" in the context of AI/ML or diagnostic performance. It refers to "pre-clinical testing" for a physical medical device. No sample size, data provenance, or study type (retrospective/prospective) is mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This document does not describe a study involving expert-established ground truth for a test set.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This document does not describe a study involving adjudication for a test set.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This document does not describe an MRMC study or AI assistance.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This document is not about an algorithm/AI device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. This document does not describe a study requiring ground truth in the context of an AI/ML device. The "pre-clinical testing" likely refers to bench testing, mechanical stress tests, biocompatibility assessments, and potentially animal studies, which establish performance against engineering specifications or biological responses, not a "ground truth" for diagnostic accuracy.
  8. The sample size for the training set

    • Not applicable. This document does not describe a training set as it is not about an AI/ML device.
  9. How the ground truth for the training set was established

    • Not applicable. This document does not describe a training set or its ground truth.

{0}------------------------------------------------

K023217

Page 112

OCT 2 5 2002

510(k) Summary of Safety and Effectiveness

GeneralProvisionsThe name of the device is:
Proprietary NameCommon or Usual Name
SMART™ Control™ Nitinol StentTranshepatic Biliary SystemBiliary Stent
Name ofPredicateDevicesThe device is substantially equivalent to:• Cordis SMART™ Control™ Nitinol Stent Transhepatic Biliary System(510(k) # K021898 - August 14, 2001) - Cordis Corporation.• S.M.A.R.T.™ Nitinol Stent and Delivery System (Improved Radiopacity)(510(k) # K001843 – July 18, 2000 -- Cordis Corporation.
ClassificationClass II.
PerformanceStandardsPerformance standards have not been established by the FDA under section514 of the Food, Drug and Cosmetic Act.
Indicationsfor UseThe SMART™ Control™ Nitinol Stent Transhepatic Biliary System isintended for use in the palliation of malignant neoplasms in the biliary tree.
DeviceDescriptionThe device description of the proposed SMART™ Control™ Nitinol StentTranshepatic Biliary System is as follows.• 6 French stent delivery system profile;• Stent material - Nickel Titanium alloy and tantalummicromarkers;• Expanded stent diameters 9 and 10 mm;• Stent lengths: 80 mm;• Stent delivery system usable length 80 and 120 cm; Guidewirelumen 0.035";• Proximal Deployment Handle.
BiocompatibilityAll materials used in the SMART™ Control™ Nitinol StentTranshepatic Biliary System are biocompatible.

{1}------------------------------------------------

K023217

c 2/2

The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is Summary of Substantial substantially equivalent to the predicate devices. The equivalence was Equivalence confirmed through pre-clinical testing.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three overlapping human figures, possibly representing the department's focus on people and well-being.

OCT 2 5 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Sam Mirza Manager, Regulatory Affairs Cordis, a Johnson & Johnson Company 14201 N.W. 60th Avenue MIAMI LAKES FL 33014

Re: K023217

Trade/Device Name: Cordis S.M.A.R.T.TM Control™ Nitinol Stent Transhepatic Biliary System Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: September 25, 2002 Received: September 26, 2002

Dear Mr. Mr. Mirza:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{3}------------------------------------------------

Page 2 - Mr. Sam Mirza

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow vou to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Daniel C. Schultz, M.

Daniel G. Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K023217

Device Name: Cordis S.M.A.R.T.TM Control™ Nitinol Stent Tranhepatic Biliary System

FDA's Statement of the Indications For Use for device:

The Cordis S.M.A.R.T.™ Control™ Nitinol Stent Tranhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.

Prescription Use OR (Per 21 CFR 801.109)

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

Urind A. Slymon

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.