K Number
K023036

Validate with FDA (Live)

Device Name
CARE ETS
Date Cleared
2003-06-19

(280 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Statement of Indication of Use (in EMG Mode):

  • Biofeedback
  • Relaxation Training
  • Muscle Re-education

Statement of Ladication of Use (in NMS Mode):
The Care ETS is recommended for use for the following conditions:

  • Relaxation of muscle spasms
  • Prevention or retardation of disuse atrophy
  • Increasing local blood circulation
  • Muscle re-education
  • Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
  • Maintaining or increasing range of motion.
  • Stroke Rehabilitation, using muscle re-education techniques
Device Description

A portable NMS device for muscle re-education.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the CARE ETS™ device:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific acceptance criteria in a numerical or measurable format for the CARE ETS™ device itself. Instead, it relies on a comparison to a predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence in technological characteristics"The CARE ETS™ has technological characteristics that are substantially equivalent to those of the predicate device, as determined by bench testing."
Substantial equivalence in labeling"The labeling of the CARE ETS™ is substantially equivalent to that of the predicate device."
Substantial equivalence in output characteristics (nonclinical)"Bench testing demonstrated that the output characteristics or CARE ETS™ are substantially equivalent to that of the predicate device."
No new questions of safety and effectiveness due to differences"The CARE ETS™ is substantially equivalent in electrical output to the predicate device and any differences between the devices do not pose new questions of safety and effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable. The document states "Clinical Testing: Not applicable." This indicates that no clinical studies with human participants were conducted as part of this 510(k) submission to demonstrate the device's performance or safety.
  • Data Provenance: Not applicable. As no clinical testing was performed, there is no clinical data provenance to report. The evaluation was based on bench testing comparison to a predicate device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable. No clinical testing or expert-adjudicated ground truth was established for the device itself.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There was no test set involving human data or expert review that would require an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. The document explicitly states "Clinical Testing: Not applicable," which means no human studies were conducted, including comparative effectiveness studies.
  • Effect size of human readers improvement with AI vs without AI assistance: Not applicable, as no such study was performed and the device is a neuromuscular stimulator, not an AI diagnostic tool for image interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was a standalone study done? No, not in the context of an "algorithm only" performance study typically seen with AI/ML devices. The device is a physical neuromuscular stimulator. The "standalone" performance shown was through bench testing, demonstrating its electrical output characteristics are equivalent to a predicate device.

7. Type of Ground Truth Used

  • Type of Ground Truth: For the purpose of the 510(k) submission, the "ground truth" was the performance and characteristics of the legally marketed predicate device (Automove AM800 K972997). The CARE ETS™ was deemed substantially equivalent based on its ability to match these characteristics through bench testing.

8. Sample Size for the Training Set

  • Sample Size: Not applicable. There is no mention of a training set as this is not an AI/ML device that requires data for model training.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment: Not applicable. As there was no training set, no ground truth needed to be established for it.

{0}------------------------------------------------

K023036

Appendix J: 510(k) Summary

510(k) Summary

Submitter's Information:Christian E. HuntCare Rehab®1124 Dominion CtMcLean, VA 22102Phone: 1-703-448-9644FAX: 1-703-356-2182.
Date of preparation:March 20, 2003
Proprietary Name:CARE ETS™
Common Name:Neuromuscular Stimulator ( EMG Triggered)
Classification Name:Powered Muscle Stimulator21 CFR 890.5850.
Panel Code:89
Product Code:IPF
Device Classification:Class II
Predicate Device:Automove AM800 (K972997)
Description of Device:A portable NMS device for muscle re-education.
Intended Use:The Care ETS is recommended for use for the following conditions:• Relaxation of muscle spasms• Prevention or retardation of disuse atrophy• Increasing local blood circulation• Muscle re-education• Immediate post-surgical simulation of calf muscles to prevent venousthrombosis• Stroke Rehabilitation , using muscle re-education techniques• Maintaining or increasing range of motion
Technological Comparison:The CARE ETS™ has technological characteristics that aresubstantially equivalent to those of the predicate device, as determinedby bench testing.
Labeling Comparison:The labeling of the CARE ETS™ is substantially equivalent to that of thepredicate device.
Nonclinical Testing:Bench testing demonstrated that the output characteristics or CAREETS™ are substantially equivalent to that of the predicate device.
Clinical Testing:Not applicable.
Conclusions from Testing:The CARE ETS™ is substantially equivalent in electrical output to thepredicate device and any differences between the devices do not pose newquestions of safety and effectiveness.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 9 2003

Mr. Christian Hunt President Care Rehab and Orthopedic Products, Inc. P.O. Box 580 McLean, Virginia 22101

Re: K023036

Trade/Device Name: Regulation Number: 21 CFR 890.5850, 21 CFR 882.5050 Regulation Name: Powered muscle stimulator, Biofeedback device Regulatory Class: II Product Code: IPF, HCC Dated: March 21, 2003 Received: March 21, 2003

Dear Mr. Hunt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Christian Hunt

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Mark H. Millken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use বা

510(k) Number:

Device Name: CARE ETS™

Statement of Indication of Use (in EMG Mode):

  • Biofcodback 1.
  • Relaxation Training 2.
    1. Muscle Rc-Education

Statement of Ladication of Use (in NMS Mode):

The Care ETS is recommended for use for the following conditions:

  • Relaxation of muscle spasms 】.
  • Prevention or retardation of disuse atrophy 2
  • ಗ Increasing local blood circulation
  • Muscle recordication র্ব
  • Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis . ਨ.
    • Maintaining or increasing range of motion. ర్
      1. Stroke Rehabilitation, using muscle re-education techniques

PLASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE AS NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark N. Milliken

(Divisio Jigi-Off) Division of General. Restorative and Neurological Devices

510" number K02309

A 20:51 E 0005 81 MUL

6

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).