K Number
K022889

Validate with FDA (Live)

Date Cleared
2002-09-25

(26 days)

Product Code
Regulation Number
870.1025
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INFINITY Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to a Siemens R50 Bedside recorder, either directly or via the INFINITY Network.

The INFINITY Modular monitors are capable of monitoring:

  • Heart rate
  • Respiration rate
  • Invasive pressure
  • Non-invasive pressure
  • Arrhythmia
  • Temperature
  • Cardiac output
  • Arterial oxygen saturation
  • Pulse rate
  • Apnca
  • ST Segment Analysis
  • 12-Lead ST Segment Analysis
  • tcp02/tcpC02
  • EEG signals
  • FiO2

With the MultiGas and MultiGas+ modules the monitors are capable of measuring respiration rate, Inspired and expired Carbon Dioxide (CO2), inspired and expired Oxygen (MultiGas+ only), average inspired Oxygen (MultiGas only), inspired and expired gas concentrations of Enflurane, Halothane, Isoflurane, Desflurane, Sevoflurane, and Nitrous Oxide.

With etCO2 the monitors can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored.

The monitors can interface with specific third party devices via an MIB protocol converter.

Device Description

The modifications implemented with the release of software version VF2 have not altered the basic fundamental technology of the INFINITY Modular Monitors. Testing with VF2 software indicates no new issues relative to safety and efficacy. The release of software version VF2 includes the following primary modifications:

    1. Wireless capability
  • Non-invasive blood pressure with step deflation 2.
    1. MICRO2+(K012770) interface
  • Support for Masimo sensors 4.

The INFINITY Modular Monitors (SC 7000 / SC 9000XL) with Wireless Option (a password protected option) allows the monitor to operate in a wireless network configuration. A wireless ethernet adapter installed in the PCMCIA port communicates with the INFINITY Network (K955059) through installed access points hardwired to the network. All patient related and network data is transmitted to the pre-installed access points allowing the transmitted data from the monitor to be viewed at an assigned MultiView WorkStation (K955059). This configuration provides the network with the same functionality as a hardwired system.

The Infinity Monitoring System utilizes the oscillometric method to measure noninvasive blood pressure (NIBP). With the SC7000/8000/9000XL VF2 release, the NIBP algorithm has changed from a linear to stepped deflation system.

The INFINITY Modular Monitor (SC 7000 / SC 8000 / SC 9000XL) VF2 release supports two simultaneous SpO2 measurements. Using a serial port already available on the monitors, the MicrO2+ pulse oximeter (K012770) can communicate with the monitors using a RS232 serial communication and power supply cable connected to the serial port on the monitor. The stand-alone MicrO2+ pulse oximeter transmits SpO2 and Pulse values to the monitor.

The monitor supports the use of Masimo SpO2 sensors via a password protected option and a specific adapter cable.

AI/ML Overview

The provided text describes modifications to the Siemens INFINITY Modular Monitors and makes references to equivalence assessments and clinical accuracy studies. However, it does not contain the detailed information necessary to fully answer your request regarding acceptance criteria, specific study designs, sample sizes, expert qualifications, or ground truth establishment.

The document states:

  • "Assessment of non-clinical performance data for equivalence: See Section J"
  • "Assessment of clinical performance data for equivalence: See Section J"
  • "Clinical accuracy studies referenced to cooximetry (see Section J) were conducted by Masimo using the following sensors in conjunction with the INFINITY Modular monitors"

Since "Section J" is not provided in the input, the specific details of the acceptance criteria and the studies proving the device meets them are unavailable.

Based on the available text, here's what can be extracted and what cannot:

1. Table of acceptance criteria and reported device performance:

  • Acceptance Criteria: Not provided in the given text.
  • Reported Device Performance: Not provided in the given text, beyond the general statement that "Testing with VF2 software indicates no new issues relative to safety and efficacy."

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not provided.
  • Data Provenance: Not provided (e.g., country of origin, retrospective/prospective). The text only states "clinical accuracy studies referenced to cooximetry... were conducted by Masimo."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not provided. The text mentions "clinical accuracy studies referenced to cooximetry," implying cooximetry serves as a reference, but it doesn't specify if human experts were involved in establishing ground truth or their qualifications.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not provided.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. The described device is a physiological monitor, not an AI-assisted diagnostic tool that human readers would use in a comparative effectiveness study. The text focuses on the device's ability to measure physiological parameters and its new features (wireless, NIBP, SpO2 interface, Masimo sensor support).

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • The text implies standalone testing of the device's physiological measurement capabilities through "clinical accuracy studies referenced to cooximetry." However, details of these studies are not provided.

7. The type of ground truth used:

  • For SpO2 measurements, the text explicitly states "clinical accuracy studies referenced to cooximetry." Cooximetry is a gold standard for measuring oxygen saturation in blood. For other parameters (Heart rate, Respiration rate, NIBP, etc.), the type of ground truth is not specified in this document.

8. The sample size for the training set:

  • Not provided. The document describes modifications to an existing device and refers to "accuracy studies" but does not detail a training set for an AI/algorithm.

9. How the ground truth for the training set was established:

  • Not provided, as details of a training set are not mentioned.

In summary, for a complete answer to your questions, "Section J" of the original 510(k) submission would be required. The provided excerpt only gives a high-level overview of the device modifications and mentions that performance data exists, but not the specifics of that data or the studies.

{0}------------------------------------------------

510(k) SUMMARY

as required per 807.92(c)

Submitters Name. Address:

SEP 2 5 2002

Siemens Medical Solutions USA, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Official Correspondent: Connie Hertel, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Penelope H. Greco Date submission was prepared: August 29, 2002

Trade Name, Common Name and Classification Name:

A. Trade Name:

Siemens INFINITY Modular Monitors (SC 7000 / SC 9000XL / SC 8000)

  • B. Common Name, Classification Name, Class and Regulation Number:
Common NameProduct CodeClassRegulation Number
Monitor, Physiological, Patient (witharrhythmia detection or alarms)MHXIII21 CFR 870.1025
Arrhythmia detector & Alarm74DSIIII21 CFR 870.1025

Legally Marketed Device Identification:

INFINITY SC 8000 Monitor, 510(k) K983632 / K990563 INFINITY SC 7000 / SC 9000XL Modular Monitors, 510(k) K003243/K982730/ K980882 INFINITY Explorer, 510(k) K013515 MICRO2+, 510(k) K012770

Description of Modification:

The modifications implemented with the release of software version VF2 have not altered the basic fundamental technology of the INFINITY Modular Monitors. Testing with VF2 software indicates no new issues relative to safety and efficacy. The release of software version VF2 includes the following primary modifications:

    1. Wireless capability
  • Non-invasive blood pressure with step deflation 2.
    1. MICRO2+(K012770) interface
  • Support for Masimo sensors 4.

1 of 3

Siemens Medical Solutions USA, Inc.

Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923

Tel: (978) 907-7500 Fax: (978) 750-6879

{1}------------------------------------------------

The INFINITY Modular Monitors (SC 7000 / SC 9000XL) with Wireless Option (a password protected option) allows the monitor to operate in a wireless network configuration. A wireless ethernet adapter installed in the PCMCIA port communicates with the INFINITY Network (K955059) through installed access points hardwired to the network. All patient related and network data is transmitted to the pre-installed access points allowing the transmitted data from the monitor to be viewed at an assigned MultiView WorkStation (K955059). This configuration provides the network with the same functionality as a hardwired system.

NIBP Step Deflation

The Infinity Monitoring System utilizes the oscillometric method to measure noninvasive blood pressure (NIBP). With the SC7000/8000/9000XL VF2 release, the NIBP algorithm has changed from a linear to stepped deflation system.

Second SpO2 - MICRO2+ Interface

The INFINITY Modular Monitor (SC 7000 / SC 8000 / SC 9000XL) VF2 release supports two simultaneous SpO2 measurements. Using a serial port already available on the monitors, the MicrO2+ pulse oximeter (K012770) can communicate with the monitors using a RS232 serial communication and power supply cable connected to the serial port on the monitor. The stand-alone MicrO2+ pulse oximeter transmits SpO2 and Pulse values to the monitor.

Support for Masimo Sensors

The monitor supports the use of Masimo SpO2 sensors via a password protected option and a specific adapter cable. Clinical accuracy studies referenced to cooximetry (see Section J) were conducted by Masimo using the following sensors in conjunction with the INFINITY Modular monitors:

  • . LNOP-Adt
  • . LNOP-Pdt
  • LNOP-Neo ●
  • LNOP-NeoPt ●
  • LNOP-DCI .
  • LNOP-DCIP .
  • . LNOP-YI
  • LNOP-Ear .
  • NR 125 ●

2 of 3

Siemens Medical Solutions USA, Inc.

Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923

Tel: (978) 907-7500 Fax: (978) 750-6879

{2}------------------------------------------------

Intended Use:

The INFINITY Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to a Siemens R50 Bedside recorder, either directly or via the INFINITY Network.

Assessment of non-clinical performance data for equivalence: See Section J

Assessment of clinical performance data for equivalence: See Section J

Biocompatability: Not applicable

Sterilization: Not applicable

See Section J Standards and Guidances:

3 of 3

Siemens Medical Solutions USA, Inc.

Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923

Tel: (978) 907-7500 Fax: (978) 750-6879

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three human profiles facing to the right, with three bars above them.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 25 2002

Siemens Medical Solutions USA, Inc c/o Ms. Penelope H. Greco Regulatory Submissions Manager Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923

Re: K022889

Trade Name: Siemens Medical INFINITY Modular Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: MHX Dated: August 29, 2002 Received: August 30, 2002

Dear Ms. Greco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Penelope H. Greco

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements; including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Elias Mallis

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

KO22889 510(k) Number (if known):

.Device Name: ________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The INFINITY Modular monitors are capable of monitoring:

  • Heart rate
  • Respiration rate
  • Invasive pressure
  • Non-invasive pressure
  • Arrhythmia
  • Temperature
  • Cardiac output
  • Arterial oxygen saturation
  • Pulse rate
  • Apnca
  • ST Segment Analysis
  • 12-Lead ST Segment Analysis
  • tcp02/tcpC02
  • EEG signals
  • FiO2

With the MultiGas and MultiGas+ modules the monitors are capable of measuring respiration rate, Inspired and expired Carbon Dioxide (CO2), inspired and expired Oxygen (MultiGas+ only), average inspired Oxygen (MultiGas only), inspired and expired gas concentrations of Enflurane, Halothane, Isoflurane, Desflurane, Sevoflurane, and Nitrous Oxide.

With etCO2 the monitors can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored.

The monitors can interface with specific third party devices via an MIB protocol converter.

The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Qutput, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only: etCO2 sidestream which is not available in neonatal mode; and tcpO2 which is to be used in the neonatal population only when the patient is not under gas anesthesia.

MRI Compatibility Statement:

The INFINITY Modular Monitors are not compatible for use in a MRI magnetic field.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseX
(Per 21 CFR 801.109)

OR

Over-The-Counter Use
(Optional Format 1-2-96)

Division of Cardiovascular & Respiratory Devices

510(k) NumberElico Malles
-----------------------------

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.