K Number
K022800

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-11-25

(94 days)

Product Code
Regulation Number
866.5100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VisiQuant™ Anti-Nuclear Antibody Test System is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the in-vitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.

Device Description

The proposed VisiQuant™ ANA Test System uses the same VisiQuant™ ANA Test Kit as was previously cleared. Hyperion is now proposing to supply the VisiQuant™ Microscope as an additional reader for use with the VisiQuant™ ANA Test Kit. The reading of the samples remains the same except that the software will now generate the calibration curve and interpolate the titer. The prepared slide is placed onto the stage of the microscope. The microscope focuses on the images of each well and the CCD camera collects the images. The software captures and stores the images. The operator views the images to determine if the sample is positive and identifies the pattern. For the samples determined to be positive by the operator, the software calculates a fluorescence intensity unit (FIU) value. The proposed software will now generate a 4-parameter curve using the FIUs and ANA titers of Calibrators assayed in the same run, from which the VisiQuant titer of the test sample is interpolated. The proposed software now has the capability of generating calibration curve reports and laboratory reports.

AI/ML Overview

VisiQuant™ Anti-Nuclear Antibody Test System - Acceptance Criteria and Study Details

1. Acceptance Criteria and Reported Device Performance

The provided document describes a comparative evaluation of the VisiQuant™ ANA Test System against a previously cleared predicate device, the VisiQuant™ ANA Test Kit using a specially equipped fluorescent microscope. While explicit "acceptance criteria" are not listed as pass/fail thresholds, the study aimed to demonstrate "substantial equivalence." The performance metrics reported are agreement for qualitative results (negative/positive) and pattern agreement.

Performance MetricAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
Agreement for negative resultsHigh agreement with predicate device95.2%
Agreement for positive resultsHigh agreement with predicate device100%
Pattern agreement (for positive samples)High agreement with predicate device97.2%
Correlation coefficient (ANA titers)High correlation (e.g., >0.9 accepted in similar contexts)0.917

2. Sample Size and Data Provenance

  • Test Set Sample Size: 113 human serum samples were used for the qualitative comparison. Of these, 71 samples were positive by both methods and were further analyzed for pattern agreement and titer correlation.
  • Data Provenance: Not explicitly stated. However, given that the manufacturer is based in Miami, Florida, USA, and the device is intended for in-vitro diagnostic use in human serum samples, it is highly likely the data is from the United States. The study appears to be retrospective, comparing the new system's results to those obtained with a pre-existing predicate device.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly describe the use of experts to establish a "ground truth" for the test set in the traditional sense of independent expert review. The study's design is a method comparison against a predicate device. Therefore, the "ground truth" for the test set is effectively derived from the performance of the predicate device (VisiQuant™ ANA Test Kit using a specially equipped fluorescent microscope), whose accuracy and reliability would have been established in its own clearance process. The agreement percentages and correlation coefficients are then calculated against this predicate.

4. Adjudication Method

Not applicable. This study is a method comparison study between two devices, not an evaluation against a human expert consensus. Discrepancies were noted and analyzed (Table 4), but no formal adjudication process with experts is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed as described in the provided text. The study focuses on comparing the objective measurements and pattern identifications between a new automated system (VisiQuant™ ANA Test System) and a previously cleared manual-microscope-based system (VisiQuant™ ANA Test Kit with a specially equipped fluorescent microscope). The document mentions that the new software will "now generate the calibration curve and interpolate the titer" and that "the operator views the images to determine if the sample is positive and identifies the pattern" with the new system. However, the comparative study itself does not involve human readers evaluating images with and without AI assistance to determine an effect size.

6. Standalone Performance Study

Yes, a standalone study was implicitly performed, though it is framed as a "method comparison." The VisiQuant™ ANA Test System (which includes the VisiQuant™ Microscope and the new software) is evaluated in its entirety without human-in-the-loop performance being a variable in the comparison against the predicate device. The comparison directly assesses the output of the new system, including its software-generated calibration curves and interpolated titers, against the output of the predicate system. The software's role in pattern identification is limited to generating values after an operator identifies the pattern. The core automated function compared here seems to be the quantitative measurement and curve generation.

7. Type of Ground Truth Used

The "ground truth" used in this study is the performance of the predicate device, the VisiQuant™ ANA Test Kit using a specially equipped fluorescent microscope, which was previously cleared under Document Control No. K013213.

8. Sample Size for the Training Set

The document does not provide information regarding a specific training set size or its establishment. This is a 510(k) submission for a modification to an existing cleared device (adding a new reader/software system). The device itself (an ANA test system) does not necessarily involve a machine learning model that requires a "training set" in the typical sense. The software's function is primarily to process images, calculate fluorescence intensity units, and generate a calibration curve.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a training set for a machine learning model is not explicitly mentioned or implied for this device based on the provided text. The device's software appears to perform calculations and curve generation based on established immunoassay principles, not through a learned model requiring a labeled training dataset.

{0}------------------------------------------------

K022800

510(k) Summary of Safety and Effectiveness Information VisiQuant™ Anti-Nuclear Antibody Test System Revised September 20, 2002

NOV 2 5 2002 Hyperion, Inc. 14100 S.W. 136th Street Miami, FL. 33186 Contact Person: Radha Goolabsingh at 305-238-3020 x202 or Victor Rana at 305-238-3020 x208, or by facsimile at 305-232-7375.

Trade or Proprietary Name:VisiQuant™ Anti-Nuclear Antibody Test System
Common or Usual Name:VisiQuant™ ANA Test System
Classification Name:Anti nuclear antibody immunological test system
Registration Number:Manufacturer
Hyperion, Inc.14100 S.W. 136th StreetMiami, Florida 331861028110

The proposed VisiQuant™ ANA System is substantially equivalent to the VisiQuant™ ANA Test Kit using a specially equipped fluorescent microscope, previously cleared under Document Control No., K013213, on 6/04/2002. The proposed and the predicate device are both in-vitro diagnostic products, intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semi-quantitative measurement of ANA's in human serum sample with a single dilution as an aid in the diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren's Syndrome. There are no changes to the test kit and the VisiQuant Microscope supplied meets the minimum requirements specified for the test kit in K013213.

The proposed VisiQuant™ ANA Test System uses the same VisiQuant™ ANA Test Kit as was previously cleared. Hyperion is now proposing to supply the VisiQuant™ Microscope as an additional reader for use with the VisiQuant™ ANA Test Kit.

The reading of the samples remains the same except that the software will now generate the calibration curve and interpolate the titer. The prepared slide is placed onto the stage of the microscope. The microscope focuses on the images of each well and the CCD camera collects the images. The software captures and stores the images. The operator views the images to determine if the sample is positive and identifies the pattern. For the samples determined to be positive by the operator, the software calculates a fluorescence intensity unit (FIU) value. The proposed software will now generate a 4-parameter curve using the FIUs and ANA titers of Calibrators assayed in the same run, from which the VisiQuant titer of the test sample is interpolated. The proposed software now has the capability of generating calibration curve reports and laboratory reports.

{1}------------------------------------------------

Method Comparison:

A comparative evaluation of the proposed VisiQuant™ ANA Test System demonstrated substantial equivalence to the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope. The results are shown in the following tables and diagrams:

VisiQuant™ ANA Test KitNo. of samples with VisiQuant™ ANA Test System
NegativePositiveTotal
Negative40242
Positive07171
Total4073113

Agreement for negative results = 40/42 = 95.2% Agreement for positive results = 71/71 = 100%

Table3. Qualitative Results: The following table shows the pattern agreement for the 71 samples found to be positive by both the VisiQuant™ ANA Test Kit and the VisiQuant™ ANA Test System.

VisiQuant™ ANA Test KitNo. of samples with VisiQuant™ ANA Test System
HSNCH,SH,NTotal
H3131
S19120
N11
C22
H,S167
H,N1010
Total32191271071

Negative (Neg); Homogenous (H); Speckled (S); Nucleolar (N); Centromere (C)

Pattern agreement for samples positive by both tests = 69/71 = 97.2 %:

Table 4. Discrepant samples: Negative (Neg.), Not applicable (n.a); Speckled (S); Homogenous (H). Slight differences in reading could account for these discrepancies.

Sample IDVisiQuant™ ANA Test KitVisiQuant™ ANA Test System
TiterPatternTiterPattern
Y-SLE17406S314H,S
X-E4713H,S526H
Y-SS2Neg.n.a.130H
W-14Neg.n.a.97H,S

{2}------------------------------------------------

Seventy-one (71) serum samples, positive by both the proposed VisiQuant™ ANA Test System and the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope were compared for ANA titers. The following figure shows the correlation with all patterns (Homogenous, Speckled, Nucleolar, Centromere and mixed).

Image /page/2/Figure/3 description: This image is a scatter plot comparing ANA titers from two different VisiQuant ANA test systems. The x-axis represents the ANA titer of the VisiQuant ANA Test kit (VT), while the y-axis represents the ANA titer of the VisiQuant ANA Test System (VTM). The plot includes data points for various antibody patterns, such as Homogeneous, Speckled, Nucleolar, Centromere, Homog. & speckled, and Homog. & Nucleolar. A regression line and 95% confidence intervals are also displayed on the plot.

Fig. 1. Scatter diagram showing correlation in ANA titers for 71 positive results between VisiQuant™ ANA Test System and the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope, with a correlation coefficient of 0.917 and a regression equation of Log VTM = 0.0578 + 0.966 Log VT. The Linear Regression and its 95% confidence intervals are also shown.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The caduceus is rendered in black, and the text is also in black.

Food and Drug Administrat 2098 Gaither Road Rockville MD 20850

Ms. Radha Goolabsingh Director of Quality Assurance and Regulatory Affairs Hyperion, Inc. 14100 S.W. 136th Street Miami, FL 33186

Re:

K022800 Trade/Device Name: VisiQuant™ Anti-Nuclear Antibody Test System Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: DHN Dated: October 30, 2002 Received: October 31, 2002

Dear Ms. Goolabsingh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(K) NUMBER (IF KNOWN):

DEVICE NAME: VisiQuant™ Anti-Nuclear Antibody Test System

INDICATIONS FOR USE: The VisiQuant™ Anti-Nuclear Antibody Test System is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the in-vitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence Of CDRH, Office Of Device Evaluation (ODE)

Prescription Use A (Per 21 CFR 801.109)

OR

Over-The-Counter-Use

(Optional Format 1-2-96)

J. Reeves for J. Bautista

(Division Sign-Off) Division of Clinical Laboratory Devi-510(k) Number .

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).