K Number
K013213

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-06-04

(251 days)

Product Code
Regulation Number
866.5100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Hyperion VisiQuant™ ANA Test Kit is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the invitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.

Device Description

VisiQuant™ ANA is a unique ANA IFA. The test samples are assayed with a single dilution as in the traditional qualitative procedure to determine positive and pattern while obtaining a VisiQuant titer from the same image. The stained slides are read objectively with a digital camera through the FAN microscope to measure the fluorescence intensity of reacted substrate (Hip-2 calls). The Visiguant tive of the test sample is interpolated from a standard curve. VisiQuant ANA uses a unique fluorophore, La Jolla Blue (LJB), which has a longer fluorescence life (photostability) than fluorescent. LJJP has nearinfrared peak excitation and emission wavelengths to minimize autofluoresconce commonly present in biological substances.

AI/ML Overview

The Hyperion VisiQuant™ ANA Test Kit is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the in-vitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.

1. Table of acceptance criteria and the reported device performance:

Agreement in Test Results (Qualitative)

CriteriaAcceptance Criteria (Implicit)Reported Device Performance
Agreement for negative resultsHigh agreement, similar to predicate device98.92 %
Agreement for positive resultsHigh agreement, similar to predicate device94.12 %

Pattern Agreement (for positive samples)

CriteriaAcceptance Criteria (Implicit)Reported Device Performance
Pattern agreement (both positive, VisiQuant pattern matches Bion pattern)High agreement, similar to predicate device94.79 %
Pattern agreement (Bion positive, VisiQuant pattern matches Bion pattern)High agreement, similar to predicate device89.21 %

Quantitative Correlation (VisiQuant Titer vs. Bion IFA Titer)

CriteriaAcceptance Criteria (Implicit)Reported Device Performance (R-value)
Correlation for all positive patternsStrong positive correlation0.731 (N=96, p<0.0001)
Correlation for Homogenous pattern onlyVery strong positive correlation0.909 (N=36, p<0.0001)

The acceptance criteria are implicitly defined by the sponsor demonstrating "substantial equivalence" to the predicate device (BION Antinuclear Antibody (ANA) Test Kit/Control) through high agreement percentages and strong statistical correlations.

2. Sample size used for the test set and the data provenance:

Test Set Sample Size:

  • Qualitative and Pattern Agreement: 195 samples (93 negative, 102 positive by Bion Test)
  • Quantitative Correlation (All patterns): 96 serum samples (positive by both VisiQuant ANA and Bion ANA)
  • Quantitative Correlation (Homogenous pattern only): 36 serum samples (with homogenous ANA pattern, positive by both tests)
  • Titration equivalence: 11 serum samples

Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). However, it is an in-vitro diagnostic product, and the samples are human serum. The study is a comparative evaluation against a legally marketed predicate device, suggesting these were collected samples tested in parallel.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The ground truth for the test set was established using the predicate device, the BION Antinuclear Antibody (ANA) Test Kit/Control, which involves visual reading.

For the titration of 11 serum samples (Table 4) using the Bion IFA ANA, the document states: "Average of separate readings by two persons." This indicates that at least two persons were involved in establishing the Bion IFA ANA titers for these specific 11 samples.

The qualifications of these "two persons" are not specified (e.g., "radiologist with 10 years of experience"). For the broader qualitative and pattern assignments by the Bion test, the number of experts or readers and their qualifications are not mentioned. Given it's a visual read, it implies trained laboratory personnel.

4. Adjudication method for the test set:

For the Bion IFA ANA titers in Table 4, the adjudication method was the average of separate readings by two persons.

For the general qualitative and pattern results of the Bion test (predicate device), the adjudication method is not explicitly stated. It's implied that the results from the predicate device serve as the reference.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a multi-reader multi-case (MRMC) comparative effectiveness study specifically comparing human readers with AI assistance versus without AI assistance was not reported.

This submission describes the performance of the Hyperion VisiQuant™ ANA Test Kit as a standalone system (which includes a digital camera and software for objective reading) in comparison to a predicate manual visual reading method (Bion Test). While the VisiQuant™ uses digital imaging and software for quantitative measurements, the study does not evaluate human readers' improvement with this new technology versus without it in a structured MRMC setting.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance study was done for the VisiQuant™ ANA Test Kit. The device is described as reading slides "objectively with a digital camera through the FAN microscope" and then using "VisiQuant software" to calculate fluorescence intensity units (FIUs) and interpolate VisiQuant titers from a standard curve. The reported "Reported Device Performance" values in the table above represent the results achieved by this system.

While the "determination of the absence (negative) or presence (positive) of ANA pattern(s)" is mentioned as being performed "as in the commonly used qualitative (screen) procedure for ANA IFA," for the positive samples, the VisiQuant titer is obtained by saving the image and processing it with the software. This indicates that the VisiQuant system (digital camera + software) operates in a standalone manner to generate the quantitative results.

7. The type of ground truth used:

The ground truth used was the results from the predicate device, the BION Antinuclear Antibody (ANA) Test Kit/Control. This predicate device uses the indirect fluorescent antibody technique (IFA) and involves visual reading to determine the titer and pattern. Therefore, the ground truth is essentially expert visual interpretation (potentially consensus as noted for titration) from a legally marketed device.

8. The sample size for the training set:

The sample size for the training set is not explicitly stated. The document focuses on the comparative evaluation of the VisiQuant™ ANA Test Kit with a predicate device. It is mentioned that "The FIUs and ANA titers of Calibrators assayed in the same run are used to build a standard curve, from which the VisiQuant titer of the test sample is interpolated." This implies a calibration process, but a dedicated "training set" for an AI algorithm is not detailed.

9. How the ground truth for the training set was established:

As no specific "training set" for an AI algorithm is explicitly detailed, the method for establishing its ground truth is also not specified. The document describes the process for generating the "standard curve," which serves as the reference for interpreting VisiQuant FIUs into titers. This standard curve is built using "FIUs and ANA titers of Calibrators assayed in the same run." The ground truth for these calibrators would be their established analytical values, likely determined through a rigorous validation process by the manufacturer, but the specifics are not provided in this 510(k) summary.

{0}------------------------------------------------

6-4-02

Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string is "KO13213". The characters are written in a simple, slightly messy style, with varying stroke thicknesses. The image is in black and white.

510(k) Summary of Safety and Effectiveness Information Hyperion VisiQuant™ ANA Test Kit September 25, 2001 Revised: May 31, 2002

Hyperion, Inc. 14100 S.W. 136th Street Miami, FL. 33186 Contact Person: Radba Goolabsingh at 305-238-3020 x202 or Victor Rana at 305-238-3020 x208, or by facsimile at 305-232-7375.

Trade or Proprietary Name: Hyperion VisiQuant™ ANA Test Kit

Common or Usual Name:ANA IFA Kit
Classification Name:Anti nuclear Antibody immunological test system
Registration Number:Manufacturer
Hyperion, Inc.14100 S.W. 136th Street
1028110

Miami, Florida 33186

The proposed Hyperion VisiQuant™ ANA Test Kit is substantially equivalent to the BION Antinuelear Antibody (ANA) Test Kit/Control manufactured by BION Ent. I.d., previously cleared under Documnti control No., K872845, on 01/17/87. The proposed and the predicate device are in-viro diagnosic products, intended for use to determine the antinuelear antibody in human serum to aid in the diagnosis of certain autoimmune diseases by utilizing the indirect fluorescent antibody technique (FFA),

The proposed Hyperion VisiQuant™ ANA Test Kit differs from the BION Antinuclear Antibody (ANA) Test KirControl in the configuration of the individual components included in their respective kists. They also differ in the way the read. The BION slides are serially diluted and visually read to determine the titer. The VisiQuant™ ANA slides are read off a standard corve to provided and visually titer,

VisiQuant™ ANA is a unique ANA IFA. The test samples are assayed with a single dilution as in the traditional qualitative procedure to determine positive and pattern while obtaining a VisiQuant titer from the same image. The stained slides are read objectively with a digital camenthrough the FAN microscope to measure the fluorescence intensity of reacted substrate (Hip-2 calls). The Visiguant tive of the test sample is interpolated from a standard curve. VisiQuant ANA uses a unique fluorophore, La Jolla Blue (LJB), which has a longer fluorescence life (photostability) than fluorescent. LJJP has nearinfrared peak excitation and emission wavelengths to minimize autofluoresconce commonly present in biological substances.

{1}------------------------------------------------

510(k) Summary of Safety and Effectiveness Information Hyperion VisiQuant™ ANA Test Kit Atlachment D, Page 2

Header Revised on 7/24/02

The antigen of the ANA substrate is a human cpithelial cell (HEp-2) which are cultured and fixed in wells of a glass slide. Diluted test serum samples are placed in the wells and incubated and incubated. In ANA arc present, they will bind with the antigens expressed by the cells. A fter washing the wells 11 renove unbound antigen, the conjugate (La Jolla Blue labeled goat antibody to human InG) is add and incubated. The conjugate will bind immunologically to the bound ANA on the HEP-2 cell antigens. After another washing of the wells to remove unbound conjugate, the wells are read with a microscope equipped with a digital camera and computer to obtain the immununofluorescence image for the determination of the absence (negative) or presence (positive) of ANA patteres) as in the commonly used qualitative (screen) procedure for ANA IFA.

For the positive sample, a VisiQuant tiler may be obtained by saving the image to a tagged image file format (TTF) file for processing by the VisiQuant software. The sollware calculates a fluorescence intensity unit (FIU) value. The FIU is calculated by measuring the average thrightness of the image from the canera. It then calculates the mean brightness of the picture elements (preditioss on two groups, one and of the love the more below. The difference between the two group mean values is reported as the FIU value. The FIUs and ANA titers of Calibrators assayed in the same run are used to build a standard curve, from which the VisiQuant titer of the test sample is interpolated.

A counparative evalualion of the proposed Hyperion VisiQuant™ ANA Test Kit demonstrated substantial cquivalence to the BION Antinuelear Antibody (ANA) Test KillControl. The results are shown in the following tables and diagrams:

Table 1. Agreement in Test Results

VisiQuantANANo. samples with Bion Test results
NegativePositiveTotal
Negative92698
Positive19697
Total93102195

Agreement for negative results = 92/93 = 98.92 % Agreement for positive results = 96/102 = 94.12 %

Table2. Qualitative Results

Negative (Neg); Homogenous (H); Speckled (S); Nucleolar (N); Centromere (C)
-------------------------------------------------------------------------------------
VisiQuantpatternNegIISNCH,SH,NS,NTotal
Neg925198
H13637
S2727
N44
C44
H,S21012
H,N19212
S,N11
Total933634541123195

Pattern agreement for samples positive by both tests = 91/96 = 94.79 %; for Bion positive samples = 91/102= 89.21 %

{2}------------------------------------------------

510(k) Summary of Safety and Effectiveness Information Hyperion VisiQuant™ ANA Test Kit Attachment D, Page 3

Header Revised on 7/24/02

Table 3. Discrepant samples (Negative, --; positive, titer with pattern; Speckled (S); Homogenous (H) and Nucleolar (N); positive but not titrated, n.a.). Borderline reactivity accounts for 7 discrepancies as detailed below. Six (6) showed titers of 40 by the Bion ANA but negative by VisiQuant ANA test; one (1) was negative by the Bion ANA but positive low titer of 175 by the VisiQuant ANA test. Slight differences in assay detection and in reading could account for these discrepancies. Some normal subjects may demonstally tow level ANA's which are clinically insignificant,

Item no.Sample IDBion testVisiQuant ANA
TiterPatternTiterPattern
1W-27--175H
2Y-N1740S----
3Y-N2540S----
4W-2140S----
5W-240S----
6W-1440H,S----
7Y-SS2n.a.S----
8X-B15n.a.S128H,S
9X-B8160N196H,N
10W-2440S,N254H,N
11W-51280N; 160S227H,N
12W-41640S189H,S

Table 4. Titration of 11 serum samples with VisiQuant ANA. Because VisiQuant titer values could be underestimated for some test samples with contromere, nucledian or some speckled patterns, nine (9) such samples were titrated by the traditions, 2-fold serial dilution and assayed with the VisiQuant ANA. The results as summarized below revealed the equivalence of VisiQuant ANA titers from titration with Bion's IFA ANA titers. Two high positive controls (items 10 and 11) were included in the experiment.

Itemno.SampleIDVisiQuant ANABion IFA ANATiter & Pattern
Assay at 1/40PatternTiterTitrationTiter*
1W-5H, N227160II,N1280N; 160S
2X-C2C21512802560C
3X-C4C13512802560C
4Z8C60640640C
5Z6N95240640N
6Z11N72960640N
7X-B12S1049601280S
8W-42S1729601280S
9X-E1N,S4741280N,#2560N; 320S
1013-3-1H554038405120H
1113-3-11S291325605120S
  • Average of separate readings by two persons.

#This sample showed a speckled pattern at the initial dilution of 1:40. It was not tested at 1:80, however it was tested at 1:160 and did not show a speckled pattern.

{3}------------------------------------------------

510(k) Summary of Sufety and Effectiveness Information Hyperion VisiQuant™ ANA Test Kit Attachment D, Page 4

Header Revised on 7/24/02

Un25

Nincty-six (96) serum samples, positive by both the VisiQuant ANA and the Bion ANA, were compared for ANA titers. Figure 2 shows the correlation with all patterns (Homogenous, Speckled, Nucleolar, Centromere and mixed) and Figure 3 shows the correlation with the Homogenous pattern only.

Image /page/3/Figure/4 description: This image is a scatter plot comparing VisiQuant ANA titer and IIF ANA titer. The x-axis represents the IIF ANA titer, ranging from 10 to 10000, while the y-axis represents the VisiQuant ANA titer, also ranging from 10 to 10000. The plot includes a regression line with dotted lines indicating confidence intervals, and data points are categorized by VisiQuant ANA pattern, including Homogeneous, Speckled, Nucleolar, Centromere, Homogeneous & speckled, Homogeneous & nucleolar, and Speckled & nucleolar.

Figure 2. Scatter diagram showing correlation between VisiQuant titer (VT) and Bion II A titer (BT) for 96 samples positive by both test. R=0.731, N=96, p<0.0001, Log VT= 0.964+0.654 x Log BT. The Linear Regression (solid line) and its 95% confidence intervals (dotted lines) are also shown.

Image /page/3/Figure/6 description: This image is a scatter plot comparing "VisiQuant ANA Titer" on the y-axis and "Dion IFA ANA titer" on the x-axis. Both axes are on a logarithmic scale, ranging from 10 to 10000. The plot shows a positive correlation between the two measures, with data points clustered around a regression line and a legend indicating that the data points represent "Homogeneous" ANA patterns.

Figure 3. A scatter diagram showing correlation between VisiQuant titer (VT) and Bion IFA titer (BT) for 36 samples with homogenous ANA pattern. R=0.909, N=36, p<0.0001, Log VT-0.631+0.806 x I.og BT. The Linear Regression (solid line) and its 95% confidence intervals (dottod lines) are also shown.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of an eagle or bird with three heads, each head facing to the right.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 0 4 2002

Ms. Radha Goolabsingh Director of Quality Assurance and Regulatory Affairs Hyperion, Inc. 14100 S.W. 136th Street Miami. Florida 33186

Re: K013213

Trade/Device Name: Hyperion VisiQuant™ ANA Test Kit Regulation Number: 21 CFR & 866.5100 Regulation Name: Antinuclear Antibody Immunological Test System Regulatory Class: II Product Code: DHN Dated: May 24, 2002 Received: May 28, 2002

Dear Ms. Goolabsingh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(K) NUMBER (IF KNOWN): K013213

DEVICE NAME: Hyperion VisiQuant™ ANA Test Kit

INDICATIONS FOR USE: The Hyperion VisiQuant™ ANA Test Kit is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the invitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.

Sousnn S. Alta.e

(Division Sign-Off) Division of Clinical Laboratory Devices

619(k) Number K013213

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence Of CDRH, Office Of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) $\checkmark$

OR

Over-The-Counter-Use .

(Optional Format 1-2-96)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).