K Number
K021923

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-10-24

(135 days)

Product Code
Regulation Number
864.7750
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for is intended for quantitative prothrombin time testing of fresh, capillary whole blood for monitoring of oral anticoagulation therapy by trained patients or their caregivers, on the prescription or other order of a treating physician.

Device Description

The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage prothrombin time test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control tests are required for the INRatio system. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result. The INRatio is the same as the INRatio, except that the labeling has been modified for readability by the lay user.

AI/ML Overview

Here's an analysis of the provided text regarding the INRatio Prothrombin Time Monitoring System's acceptance criteria and related study information:

Note: The provided text is a 510(k) summary and associated FDA correspondence. It does not include detailed study reports or comprehensive acceptance criteria. Therefore, several sections below will indicate "Not explicitly stated in the provided text." or "Not applicable/Not explicitly detailed in the provided text."


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Anticipated)Reported Device Performance (Summary)
Equivalence to established reference method for Prothrombin Time and INR measurements.The INRatio System was found to perform equivalently when used by trained lay users and healthcare professionals. Furthermore, both user populations generated results found to be equivalent to an established reference method.
Performance of internal quality control mechanisms.The device includes internal control channels. If control results are not within a set range, an error message is reported instead of a result, indicating a problem with the test.
Readability/Usability for lay users for self-testing.Labeling has been modified for readability by the lay user. The device is intended for use by trained patients or their caregivers.

Explanation: The document strongly emphasizes "equivalence" as the primary performance benchmark against an "established reference method." Specific quantitative acceptance criteria (e.g., specific bias limits, precision thresholds) are not detailed in this 510(k) summary. The performance data section provides a general summary statement about equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in the provided text.
  • Data Provenance: Not explicitly stated in the provided text regarding country of origin or whether it was retrospective or prospective. The statement "The INRatio System was found to perform equivalently when used by trained lay users and healthcare professionals" suggests a clinical study, which would typically be prospective, but this is an inference.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not explicitly stated in the provided text.
  • Qualifications of Experts: Not explicitly stated in the provided text.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly stated in the provided text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? Not applicable in the traditional sense for this device. This is an In Vitro Diagnostic (IVD) device that provides a quantitative measurement (PT/INR), not an imaging device requiring human reader interpretation. The study involved comparing results from different user types (lay users vs. healthcare professionals) to a reference method, which is a different type of comparison.
  • Effect size of human readers improvement with AI vs without AI assistance: Not applicable/Not explicitly detailed in the provided text. The device itself is not described as an "AI" device as per current definitions, nor does it assist human readers in interpretation; it provides the result.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? Yes, in a way. The device itself calculates the PT and INR. The comparison involved the device's output being assessed against a reference method, performed by both lay users and healthcare professionals. The device, by its nature, functions in a "standalone" capacity to produce a result, with human input being the blood sample and operation of the machine. The key test was the agreement of the device's output (regardless of who operated it) with the reference standard.

7. Type of Ground Truth Used

  • Type of Ground Truth: An "established reference method." The specific nature of this reference method (e.g., a laboratory-based gold standard PT/INR test) is not detailed but is implied. For PT/INR, this usually refers to a central lab-based method that is considered the clinical standard.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not explicitly stated in the provided text. IVD devices like this typically undergo extensive calibration and verification during development (which could be considered a form of "training" data for the device's internal algorithms), but precise details are not in this summary.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not explicitly stated in the provided text. During the development of such a device, ground truth for calibration and internal algorithm optimization would typically be established using a combination of certified reference materials, spiked samples, and clinical samples analyzed by established laboratory reference methods.

{0}------------------------------------------------

KO21923

OCT 2 4 2002

Appendix F 510(k) Summary

: 上

Introduction:According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1. Submitter's name address, contactHemoSense600 Valley WayMiilpitas, CA 95035(408) 719-1393 (phone)(408) 719-1184 (fax)
Contact person: Judith Blunt
Date prepared: June 10, 2002
2. Device nameCommon or Usual Name: Prothrombin Time Test
Classification Name: Prothrombin Time Test
Trade or Proprietary Name: INRatio
3. Predicate device:The INRatio System: device for testing Prothrombin Time and INR in whole blood.
4. Device description:The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage prothrombin time test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control tests are required for the INRatio system. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result. The INRatio is the same as the INRatio, except that the labeling has been modified for readability by the lay user.

{1}------------------------------------------------

    1. Intended use:
      .

The INRatio is intended for quantitative prothrombin time testing of fresh, capillary whole blood for monitoring of oral anticoagulation therapy by trained patients or their caregivers, on the prescription or other order of a treating physician.

    1. Comparison to The INRatio is substantially equivalent in materials, design and intended use to other products predicate device that measure Prothrombin Time INR in human blood. Most notably, it is substantially equivalent to the INRatio, manufactured by HemoSense. In fact, it is identical in materials, design and function to the INRAtio, but the labeling has been changed for physician directed, self-test use.
      1. Summary of The INRatio System was found to perform equivalently when used by trained lay users and performance data healthcare professionals. Furthermore, both user populations generated results found to be equivalent to an established reference method.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A.". The text is arranged around the circle's perimeter.

ockville MD 20850

Ms. Judith Blunt Director of Clinical Affairs HemoSense, Inc. 600 Valley Way Milpitas, California 95035

Re: K021923

Trade/Device Name: HemoSense INRatio Self-Test PT Monitoring System Regulation Number: 21 CFR § 864.7750 Regulation Name: Prothrombin Time Test Regulatory Class: II Product Code: GJS Dated: August 29, 2002 Received: August 30, 2002

Dear Ms. Blunt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Under Section 522(a) of the act, manufacturers of certain Class II or Class III devices that meet certain criteria may be ordered to conduct Postmarket Surveillance (PS) of that device. We are considering ordering PS for the HemoSense INRation Self-Test System. Please contact Valerie Dada at 301-594-1243 within 15 days of receipt of this letter to arrange a meeting to discuss the objective and design of a PS plan for your device."

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Ms. Judith Blunt

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Appendix G. Indications for Use Statement

June 10, 2002

PREMARKET NOTIFICATION INDICATIONS FOR USE STATEME (As required by 21 CFR 807.87(i))

510(K) Number: KO21923

Device Name: HemoSense INRatio PT Monitoring System

The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for is intended for quantitative prothrombin time testing of fresh, capillary whole blood for monitoring of oral anticoagulation therapy by trained patients or their caregivers, on the prescription or other order of a treating physician.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Signed: Judith Blunt

Name: Judith Blunt

Position: Director of Clinical Affairs and Product Support

Date: June 10, 2002

Prescription

Home-use

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).