K Number
K020894

Validate with FDA (Live)

Date Cleared
2002-04-29

(41 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The products encompassed by this request are intended for in-vitro diagnostics use and are intended for r ne products chooling and flushing the sample flow path of the equivalent OEM Analyzers. Mission reagents are intended to serve as direct replacements to like named products manufactured by the OEM. The products encompassed are to be handled using normal laboratory precautions.

Device Description

The Calibrators for the OEM Instruments are aqueous reagents with salts added to obtain desired analyte levels to provide calibration of the electrodes and rinse the sample path. The reagents are intended for use on equivalent OEM Instruments. The original equipment manufacturer (OEM) of the instruments and the predicate reagents which are necessary for the continued operation and use of the instruments. Mission uses a similar composition, description and packaging as that used by the OEM in its products, as shown in the packaging section of this submission.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Mission Diagnostic Reagents, based on the provided document.

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a numerical or categorical format. However, the performance is demonstrated through precision coefficent of variation (%CV) and correlation (slope and R^2) against OEM reagents. The implicit acceptance criterion for this type of device (calibrators/reagents intended as direct replacements) would be that the Mission reagents perform comparably to the OEM reagents, demonstrating "substantial equivalence."

A reasonable interpretation of the acceptance criteria, based on the presented data, would be:

  • Precision (%CV): Low %CV values, indicating good repeatability. While no specific threshold is given, the reported values are generally very low (often < 1%, with a few instances up to ~10%) which are typical for clinical chemistry assays.
  • Correlation (Slope and R^2): Slope near 1 and R^2 near 1, indicating a strong linear relationship and good agreement between Mission and OEM reagents. This is explicitly stated in the document: "Correlations demonstrated slopes of 1 and R^2's of 0.99, which support a claim of substantial equivalence."

Table: Acceptance Criteria (Implicit) and Reported Device Performance

Performance MetricImplicit Acceptance CriterionReported Device Performance (Summary)
Precision (%CV)Low %CV, comparable to established assay performance.Generally very low, ranging from 0.0% to 10.4% across various analytes and instruments. Most are < 5%.
Correlation (Slope for Y=mX+b)Slope (m) close to 1 (indicating proportional agreement).Ranging from 0.95 to 1.04 across various analytes and instruments (for both human serum and QC results).
Correlation (R^2 for Y=mX+b)R^2 close to 1 (indicating strong linear correlation).Ranging from 0.99 to 1.00 across various analytes and instruments (for both human serum and QC results).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Sizes (Test Set):

    • Precision Studies:
      • Ciba-Corning 238: N=9 per QC level for pH, pO2, pCO2.
      • Ciba-Corning 248: N=10 per QC level for pH, pO2, pCO2.
      • Ciba-Corning 348: N=20-21 per QC level for Na, K, Ca, pH.
      • Ciba-Corning 664: N=16 per QC level for Na, K, Cl, tCO2.
      • Ciba-Corning 288: N=18-25 per QC level for Na, K, Ca, pH.
    • Correlation Studies (Regression Analysis):
      • Human Serum:
        • Na (348, 288, 664): N=19-44.
        • K (348, 288, 664): N=44-49.
        • Ca (348, 288): N=35-44.
      • QC Results:
        • pH (238, 248, 348, 288): N=21-30.
  • Data Provenance: The document does not explicitly state the country of origin. The studies appear to be retrospective in the sense that they are conducted on existing instruments and control/serum samples. However, the data collection, where Mission reagents are compared against OEM reagents on the same analyzer and same test days, indicates a controlled study design. The human serum samples are implicitly "prospective" for the correlation study, as they were "spiked with Human Serum Albumin to yield varying concentrations."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the ground truth for these medical device assays (calibrators) is established by the reference values of certified control materials or reference methods, not by expert consensus.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication typically refers to expert review processes for subjective assessments (e.g., image interpretation). For analytical performance studies of calibrators, the "ground truth" or reference values are derived from quantitative chemical analysis or certified reference materials, not expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This document describes an in-vitro diagnostic (IVD) calibrator, not an AI-assisted diagnostic device or a system involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is not applicable. This document describes an in-vitro diagnostic (IVD) calibrator, not an algorithm or AI system. The "performance" refers to the analytical performance of the chemical reagents.

7. The Type of Ground Truth Used

  • Precision Data: The ground truth for precision is the consistency of results obtained from the Mission reagents themselves when repeatedly measured against known control material levels. The "ground truth" or target values for the control materials (QC1, QC2, QC3) would have been established by their manufacturer.
  • Correlation Data:
    • Human Serum Samples for Na, K, Cl, Ca: The ground truth for the comparison is the results obtained using the OEM reagents on the same analyzer. The values obtained with the OEM reagents serve as the reference for determining the equivalence of the Mission reagents. These samples were also "spiked with Human Serum Albumin to yield varying concentrations," indicating prepared samples with expected values.
    • QC Samples for pH: Similar to human serum, the ground truth for comparison is the results obtained using the OEM reagents on the same analyzer for the three levels of QC.

8. The Sample Size for the Training Set

This information is not applicable. This device is a chemical reagent, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above.

{0}------------------------------------------------

APR 2 9 2002

Page 49 of 70

K020894

510(k) Submission for Mission Diagnostic Reagents on pH/Blood Gas &/or Electrolyte Analyzers

1. Submitter's Name & AddressMission Diagnostics
331 Fiske St
Holliston MA 01746
FAX: 508-429-0452
Establishment Registration Number:In Process
Date of Preparation:March 18, 2002
Contact Person:Linda Stundtner
QA/RA Manager
508-429-0450
2. Identification of the Device:
Proprietary/Trade name:Calibrating Material, Cal-Pak, Buffers
Common or usual nameCalibrators for ISE and/or pH/Blood Gas automated systems
Classification name:Calibrator, secondary
Device ClassificationII
Regulation Number:21 CFR § 862.1150
Panel:Chemistry (75)
  • . Mission manufactures calibrators intended to serve as direct replacements to like named products manufactured by Original Equipment Manufactures (OEM)

Predicate Device: 3.

Product Code:

  • Mission claims substantial equivalence to the OEM Calibrating Materials listed below: .
    JIT
Mission ProductOEM Equivalent
AV-BP0136D Type 1 Buffer 7.383BP0136 Type 1 Buffer 7.383 AVL
AV-BP0137D Type 2 Buffer 6.841BP0137 Type 2 Buffer 6.841 AVL
CD-473691AD Cal-Pak for Ciba-Corning 238473691 pH/Blood Gas pH 7.38 Cal
CD-473496AD Cal-Pak for Ciba-Corning 248473496 6.8/7.3 Buffer for 248
CD-104227AD Cal-Pak for Ciba-Corning 348104227 6.8/7.3 Buffer for 348
CD-478855D Ciba-Corning 200 Series 7.3 Buffer478855D Ciba-Corning 200 Series 7.3
CD-478856D Ciba-Corning 200 Series 6.8 Buffer478856D Ciba-Corning 200 Series 6.8
CD-478864D Ciba-Corning 664 Cal/Wash478864D Ciba-Corning 664 Cal/Wash
CD-478865D Ciba-Corning 664 Slope478865D Ciba-Corning 664 Slope
CD-473385D Bayer 800 Series 7.3/Co-ox Zero473385D Bayer 800 Series 7.3/Co-ox Zero
CD-473386D Bayer 800 Series 6.8 Buffer473386D Bayer 800 Series 6.8 Buffer
CD-473387D Bayer 800 Series Wash/Cal G/L Zero473387D Bayer 800 Series Wash/Cal G/L Zero
CD-570096D Bayer 800 Series Cal G/L570096D Bayer 800 Series Cal G/L
  • . The OEM products for Ciba-Corning Instruments (238, 248, 348, 664, & 200 Series) were originally released under the Ciba-Corning name. The company has under gone several owner and name changes: starting out as Corning Glass, Ciba-Corning, Chiron, and Iastly

{1}------------------------------------------------

Bayer. For the purposes of this 510(k) the OEM for theses products will be referred to as Ciba-Corning.

  • The OEM products for Bayer Instrument (800 Series (aka RapidLab) were originally released . under the Chiron name. The company has under gone several owner and name changes: starting out as Corning Glass, Ciba-Corning, Chiron, and lastly Bayer. For the purposes of this 510(k) the OEM for theses products will be referred to as Bayer.

4. Device Description:

  • The Calibrators for the OEM Instruments are aqueous reagents with salts added to obtain . desired analyte levels to provide calibration of the electrodes and rinse the sample path.
  • Intended Use: .
  • The reagents are intended for use on equivalent OEM Instruments. .
  • The original equipment manufacturer (OEM) of the instruments and the predicate reagents . which are necessary for the continued operation and use of the instruments.
  • Mission uses a similar composition, description and packaging as that used by the OEM in its . products, as shown in the packaging section of this submission.
  • Performance equivalence is shown in the following manner: ●
    • Through a method comparison where results obtained on an equivalent OEM analyzer, . calibrated with Mission calibrating material are compared with results obtained on the same analyzer calibrated with OEM calibrating material.
    • A summary of the results of these studies follows: .

5. Performance Characteristics:

Precision Data

Precision data are collected from the analysis of three levels of control materials tested over a minimum of 7 days on each OEM analyzer. Analyzers are calibrated with Mission reagents and alternately with the OEM reagents for testing.

{2}------------------------------------------------

Precision Data Table - 238 Ciba-Corning pH/BG Analyzer

238 Instr

pH

Mission
LevelNMeanStdevMinMax% CV
QC197.150.017.157.160.1%
QC297.410.017.407.440.2%
QC397.640.017.637.650.1%
of performer I it in continue
Instr238

pO2

Mission
LevelNMeanStdevMinMax% CV
QC1963162641.2%
QC291012981031.6%
QC3915861511693.7%

238 Instr

pCO2

Mission
LevelNMeanStdevMinMax% CV
QC1981277832.7%
QC2945144461.6%
QC3918018192.4%

Precision Data Table - 248 Ciba-Corning pH/BG Analyzer

Instr 248

pH

Mission
LevelNMeanStdevMinMax% CV
QC1107.1560.0037.1517.1600.0%
QC2107.4030.0037.3977.4070.0%
QC3107.6300.0057.6227.6390.1%

Instr 248 pO2

Mission
LevelNMeanStdevMinMax% CV
QC11053.92.551.659.74.6%
QC21094.91.193.096.91.1%
QC310160.83.3157.2167.72.1%

Instr 248 pCO2

LevelNMeanStdevMinMax% CV
QC11077.31.076.379.71.3%
QC21047.00.446.647.80.8%
QC31021.80.221.522.00.7%

{3}------------------------------------------------

Precision Data Table - 248 Ciba-Corning pH/BG & Electrolyte Analyzer

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------IT THE FORT COLL A LANSING MANUFACTURES POR LANDInstrSeptember 19, 2011348

Na

Mission
LevelNMeanStdDevMinMax%CV
QC1211132.01101181.8%
QC2211392.71351471.9%
QC3201622.41581691.5%

K

Instr348
------------
LevelMission
NMeanStdDevMinMax%CV
QC1212.100.022.062.130.8%
QC2214.380.064.164.431.3%
QC3216.880.086.747.011.2%

Ca

LevelMission
NMeanStdDevMinMax%CV
QC1212.060.081.752.134.1%
QC2211.160.060.941.255.4%
QC3210.580.060.440.6510.4%
ALL PROPERTY OF COCOLLEGE

pH

Mission
LevelNMeanStdevMinMax%CV
QC1107.1660.0067.1547.1770.1%
QC2107.4420.0197.4187.4760.3%
QC3107.6690.0277.6347.7170.4%

{4}------------------------------------------------

Precision Data Table - 664 Ciba-Corning Analyzer

Instr664
------------

Na

Mission
LevelNMEANstdevMinMax% CV
QC116112.01.4110.0114.91.3%
QC216137.60.9136.8139.40.6%
QC316162.31.0161.1164.40.6%
The former of the state of the comments of the comments of the county of the county ofs C
Instr1

K

Mission
LevelNMeanStdevMinMax% CV
QC1161.980.031.952.041.3%
QC2164.380.044.344.460.9%
QC3166.990.076.927.141.0%
Instrନିର୍ବିଧ001
---------------------------

Cl

Mission
LevelNMeanStdevMinMax% CV
QC116834.677995.6%
QC216997.9911278.0%
QC3161229.4881277.7%

Instr 664 tCO2

Mission
LevelNMeanStdevMinMax% CV
QC11628.81.326.331.14.6%
QC21630.41.328.432.34.2%
QC31623.30.821.924.83.6%

{5}------------------------------------------------

Precision Data Table - 288 Ciba-Corning pH/BG & Electrolyte Analyzer

AND
str288

Na

LevelMission
NMeanStdevMinMax% CV
QC119108.00.8107.0110.00.7%
QC219134.30.7132.7135.70.5%
QC319159.21.0157.3160.70.6%
Instr288
------------

K

Mission
LevelNMeanStdevMinMax% CV
QC1191.660.021.631.691.1%
QC2194.250.024.224.290.6%
QC3197.070.047.027.140.6%

Instr 288 Ca

Mission
LevelNMeanStdevMinMax% CV
QC1192.130.032.032.181.6%
QC2191.120.011.091.141.0%
QC3180.480.010.450.512.8%

Instr 288 pH

Mission
LevelNMeanStdevMinMax% CV
QC1257.1820.0127.1667.2100.2%
QC2257.4260.0127.4037.4550.2%
QC3257.6510.0057.6457.6610.1%

Correlation with OEM Reagents

Correlation data for electrolytes on the 348 were obtained from human serum samples for Na, K, Cl, Ca. Samples are spiked with Human Serum Albumin to yield varying concentrations of each of the measuring analytes. The serum samples were measured on the 348 analyzer calibrated with Mission reanents for 1 run and 1 comparison run with OEM reagents each test day

Correlation for pH were obtained by comparing 3 levels of QC.

Linear regression analysis was performed using Mission data as the independent X variable and Corning as the dependent Y variable in the equation Y = mX +b

Correlations demonstrated slopes of 1 and R 's of 0.99, which support a claim of substantial equivalence. Graphs of Correlation are in Attachment Section.

{6}------------------------------------------------

Regression Analysis of Mission vs OEM; Human Serum Results

Na
NSlopeInterceptR2Range
348441.02-0.960.99103 - 185
288190.958.881.00103 - 171
664250.956.571.0093 - 173

Y = mX + b where Y = OEM results, X = Mission results, m = Slope, b = Intercept

NSlopeInterceptR2Range
348491.000.051.002.39 - 5.87
288441.04-0.061.001.54 - 6.59
664441.02-0.041.001.82 - 6.56

..

Ca
NSlopeInterceptR2Range
348350.960.070.990.94 - 2.11
288441.00-0.050.990.94 - 2.48
1.1
LALE MELLE BOOK BOOK******A MIND AND A1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------11Company of Children Co. Co.--CAnd Advance-----

Regression Analysis of Mission vs OEM; QC Results

Y = mX + b where Y = OEM results, X = Mssion results, m = Slope, b = Intercept

pH
NNSlopeInterceptR2Range
238301.000.031.007.15 - 7.65
248301.01-0.061.007.151 - 7.639
348301.00-0.010.997.151 - 7.717
288210.980.171.007.163 - 7.661

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular pattern around the eagle. The logo is black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms Linda Stundtner OA/RA Manager Diamond Diagnostics Inc. Mission Diagnostics 333 Fiske Street Holliston, MA 01746

APR 2 9 2002

Re: K020894

Trade/Device Name: Mission Diagnostic Reagents for pH/BG &/or Electrolyte Analyzers Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II

Product Code: JIT Dated: March 18, 2002 Received: March 19, 2002

Dear Ms. Stundtner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely vours.

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

510(k( Number

020894

Device Name: Mission Diagnostic Reagents for pH/BG &/or Electrolyte Analyzers

Indication For Use:

The products encompassed by this request are intended for in-vitro diagnostics use and are intended for r ne products chooling and flushing the sample flow path of the equivalent OEM Analyzers.

Mission Product PNOEM Instrument(s) Used on
AV-BP0136D Type 1 Buffer 7.383AVL 99X and Compact Analyzers
AV-BP0137D Type 2 Buffer 6.841AVL 99X and Compact Analyzers
CD-473691AD Cal-Pak for Ciba-Corning 238Ciba-Corning 238 pH/BG Analyzer
CD-473496AD Cal-Pak for Ciba-Corning 248Ciba-Corning 248 pH/BG Analyzer
CD-104227AD Cal-Pak for Ciba-Corning 348Ciba-Corning 348 pH/BG & Electrolyte Analyzer
CD-478855D Ciba-Corning 200 Series 7.3 BufferCiba-Corning 200Series pH/BG Analyzer
CD-478856D Ciba-Corning 200 Series 6.8 BufferCiba-Corning 200 Series pH/BG Analyzer
CD-478864D Ciba-Corning 664 Cal/WashCiba-Corning 664 Analyzer
CD-478865D Ciba-Corning 664 SlopeCiba-Corning 664 Analyzer
CD-473385D Bayer 800 Series 7.3/Co-ox ZeroBayer 800Series Analyzers
CD-473386D Bayer 800 Series 6.8 BufferBayer 800Series Analyzers
CD-473387D Bayer 800 Series Wash/Cal G/L ZeroBayer 800 Series Analyzers
CD-570096D Bayer 800 Series Cal G/LBayer 800Series Analyzers

Mission reagents are intended to serve as direct replacements to like named products manufactured by the OEM.

The products encompassed are to be handled using normal laboratory precautions.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrance of CDRH, Office of the Device Evaluation (ODE)

T. Cox
(Division Sign-Off)

510(k) Number

Prescription Use X
(per 21 CFR 801.109)

(Optional format 3-10-98)

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.