K Number
K014069

Validate with FDA (Live)

Date Cleared
2002-06-10

(182 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cut-off of 300 ng/mL in urine to assist in screening of abuse samples. For In vitro Diagnostic Use
Immunoassay for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cut-off of 300 ng/mL in urine to assist in screening of abuse at home. For In vitro Diagnostic Use

Device Description

LifeSign Home Drug Test (COC) is simple one step immunochromatographic test for the rapid, qualitative detection of cocaine.
Status Stik" COC is simple one step immunochromatographic test for the rapid, qualitative detection of cocaine.

AI/ML Overview

The provided text describes two separate 510(k) submissions, 14060 and 1601-1067, for devices intended for cocaine detection. Since this request specifically asks for an analysis of a "single study," and the text for 14060 contains a "Consumer Study" while 1601-1067 does not, I will focus on the information relevant to the study mentioned for LifeSign®Home Drug Test (COC) (associated with 510(k) number 14060).

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided input:

1. Table of Acceptance Criteria and Reported Device Performance:

The "acceptance criteria" appear to be implicit rather than explicitly stated as numerical thresholds. The study primarily focuses on demonstrating substantial equivalence to a predicate device and overall accuracy.

Acceptance Criterion (Implicit)Reported Device Performance
Substantial equivalence to predicate (K990786)"The tests demonstrated 100 % correlation when 100 specimens (50 negative and 50 positive) were compared." Note: This applies to the predicate comparison, not the consumer study directly.
Overall accuracy (for home use)"LifeSign® Home Drug Test (COC) showed over 96% overall accuracy."

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Substantial Equivalence Comparison: 100 specimens (50 negative and 50 positive).
  • Sample Size for Consumer Study: Not explicitly stated, beyond stating "over 96% overall accuracy."
  • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective, using existing "specimens" for the substantial equivalence comparison, and implying a user-based "consumer study," but details are scarce.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the given text. The text does not detail how the ground truth for the 100 specimens (used for substantial equivalence) or the consumer study samples was established, nor does it mention any experts involved in this process.

4. Adjudication Method for the Test Set:

This information is not provided in the given text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not mentioned. The studies described are focused on the device's performance against known samples or in consumer use, not on comparing human readers with and without AI assistance.

6. Standalone Performance Study:

Yes, a standalone study (algorithm/device only without human-in-the-loop performance) was conducted. The "Consumer Study" and the "Substantial Equivalence" comparison both assess the device's performance directly (e.g., "LifeSign® Home Drug Test (COC) showed over 96% overall accuracy"). While the "Consumer Study" involves users, it's evaluating the device's ability to produce an accurate result when used by consumers, rather than the consumers' interpretation with or without AI assistance.

7. Type of Ground Truth Used:

The type of ground truth is implicitly based on known positive and negative specimens. For the "Substantial Equivalence" comparison, 50 negative and 50 positive specimens were used, implying their true status was already established. For the "Consumer Study," the "overall accuracy" suggests a comparison against an established gold standard, but the nature of this standard is not explicitly detailed (e.g., laboratory confirmation, expert consensus).

8. Sample Size for the Training Set:

The text does not provide any information about a training set since these are immunoassay devices, which typically do not involve machine learning models that require explicit training sets in the same way AI algorithms do. The "Substantial Equivalence" study and "Consumer Study" are performance validation studies.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable/provided as there is no mention of a training set for an AI/algorithm-based device.

{0}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is: 14060

    1. Date of Summary: Nov.30, 2001
    1. Submitted by: Princeton BioMeditech Corporation 4242 U.S. Route 1. Monmouth Jct., NJ 08852 PHONE 732-274-1000 732-274-1010 FAX Contact Person: Jemo Kang
    1. Device Name: Trade Name: LifeSign®Home Drug Test (COC) Common or Usual Name: Immunoassay for detection of cocaine in human urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DIO for Enzyme Immunoassay)
    1. Identification of legally marketed device to which claims equivalence: K990786; StatusDS™DOA10(MET/OPI/COC/THC/PCP/BZO/BAR/MTD/TCA/AMP)
    1. Device Description: LifeSign Home Drug Test (COC) is simple one step immunochromatographic test for the rapid, qualitative detection of cocaine.
    1. Intended Use: LifeSign Home Drug Test (COC) is designed for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cutoff of 300 ng/mL in human urine to assist in screening of drugs of abuse samples at home.
    1. Substantial Equivalence: LifeSign®Home Drug Test (COC) is substantially equivalent to K990786; Status DS "DOA10. Both products use the same assay principle and immunochromatographic assay to detect cocaine metabolite qualitatively. The detection cutoff level is the same. The tests demonstrated 100 % correlation when 100 specimens (50 negative and 50 positive) were compared. The differences are that LifeSign®Home Drug Test (COC) detects cocaine use only, while StatusDS "DOA10 detects nine other drugs of abuse in addition to COC, and different antibody is used for cocaine detection.
    1. Consumer Study: In a consumer study, LifeSign® Home Drug Test (COC) showed over 96% overall accuracy.
  • Conclusion: The device is substantially equivalent to the legally marketed device K990786. Status DS""DOA10 (MET/OPI/COC/THC/PCP/BZO/ BAR/MTD/TCA/AMP).

{1}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is: 1601-1067

  1. Date of Summary: Nov.30, 2001

  2. Submitted by: Princeton BioMeditech Corperation 4242 U.S. Route 1, Monmouth Jct., NJ 08852 PHONE 732-274-1000 FAX 732-274-1010 Contact Person: Jemo Kang

  3. Device Name: Trade Names: Stick Device: Status Stik™ COC, AccuSign® Stik COC, AccuStik™ COC Card Device: AccuSign® COC, Status DS™ COC Strip Test: AccuStrip™ COC Common or Usual Name: Immunoassay for detection of cocaine in human urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DIO for Enzyme Immunoassay)

    1. Identification of legally marketed device to which claims equivalence: K990786; StatusDS™DOA10(MET/OPI/COC/THC/PCP/BZO/BAR/MTD/TCA/AMP)
    1. Device Description: Status Stik" COC is simple one step immunochromatographic test for the rapid, qualitative detection of cocaine.
    1. Intended Use: Status Stik™ COC is designed for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cutoff of 300 ng/mL in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use.
    1. Substantial Equivalence: Status Stik" COC is substantially equivalent to K990786; Status DS™DOA10. Both products use the same assay principle and immunochromatographic assay to detect cocaine metabolite qualitatively. The detection cutoff level is the same. The tests demonstrated 100 % correlation when 100 specimens (50 negative and 50 positive) were compared. The differences are that Status Stik™ COC detects cocaine use only, while StatusDS™DOA10 detects nine other drugs of abuse in addition to COC, and different antibody is used for cocaine detection.
  • Conclusion: The device is substantially equivalent to the legally marketed device K990786. Status DS "DOA10 (MET/OPI/COC/THC/PCP/BZO/ BAR/MTD/TCA/AMP).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Jemo Kang, Ph.D. Director Princeton BioMeditech Corporation 4242 U.S. Route 1 Monmouth Junction, NJ 08852-1905

JUN 1 0 2002

Re: K014069

Trade/Device Name: LifeSign®Home Drug Test (COC) Status Stik™ COC, AccuSign®Stik COC, AccuStik™ COC, AccuSign® COC, Status DSTM COC, Strip: AccuStrip™ COC Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: April 9, 2002 Received: April 10, 2002

Dear Dr. Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and i additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours.

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page of

KO14067 510(k) Number (if known):

Device Name: Stick: Status Stik™ COC, AccuSign®Stik COC, AccuStik™ COC Card: AccuSign® COC, Status DSTM COC Strip: AccuStrip™ COC

Indications For Use:

Immunoassay for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cut-off of 300 ng/mL in urine to assist in screening of abuse samples. For In vitro Diagnostic Use

Sean Cooges
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K014069

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Professional Use: Prescription Use:_X (Per 21 CFR 801.109)

OR

Over The Counter Use:

(Optional Format 1-2-96) Page of of of of a

{5}------------------------------------------------

Page of

510(k) Number (if known): [ 0 1406 }

Device Name: LifeSign®Home Drug Test (COC)

Indications For Use:

Immunoassay for the qualitative detection of cocaine metabolite, benzoylecgonine, at the cut-off of 300 ng/mL in urine to assist in screening of abuse at home. For In vitro Diagnostic Use

Jean Cooper

(Division -Off)
Division or inical Laboratory Devices
510(k) Num: K014069

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Professional Use:_

Prescription Use:_ (Per 21 CFR 801.109) OR

Over The Counter Use: X

(Optional Format 1-2-96)

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).