K Number
K014041

Validate with FDA (Live)

Date Cleared
2002-03-07

(90 days)

Product Code
Regulation Number
882.5950
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TRUFILL® DCS Detachable Coil System is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be: very high risk for management by traditional operative techniques, or, a) b) inoperable, and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature. The TRUFILL® DCS Detachable Coil System is also intended for arterial and venous embolizations in the peripheral vasculature. The TRUFILL® DCS Detachable Syringe is indicated for use with the TRUFILL® DCS TRUFILL® DCS Syringe Intended Detachable Coil. Use

Device Description

The TRUFILL® DCS Detachable Coil System is comprised of the TRUFILL® DCS Detachable Coil and TRUFILL® DCS Syringe.

  • The TRUFILL® DCS Detachable Coil consists of a delivery system (delivery tube and . coil introducer) and an embolic coil. The delivery tube is comprised of a hub, a strain relief, a stiff proximal section, and a floppy distal section. The coil introducer is a tube designed to protect the detachable embolic coil in the packaging dispenser and provide support for introducing the embolic coil into the microcatheter. The embolic coil is the implantable segment of the device. It is comprised of a vasocclusion coil wound from a platinum alloy wire (92% Platinum / 8% Tungsten) into a primary coil and then formed into a secondary helical or complex shape.
  • The TRUFILL®DCS Syringe consists of a 25-cc barrel with a pressure gauge, a threaded . plunger assembly with a locking wing mechanism, and a flexible high-pressure extension tube with a male luer connector.
AI/ML Overview

The provided text is a 510(k) Summary of Safety and Effectiveness for a medical device (TRUFILL® DCS Detachable Coil System). It focuses on demonstrating substantial equivalence to a predicate device rather than establishing novel acceptance criteria or presenting a standalone study for the device's performance against specific metrics.

Therefore, many of the requested sections about acceptance criteria, detailed study design, sample sizes, expert involvement, and ground truth establishment are not present in this document.

Here's a breakdown of what can be extracted and what is not available based on the provided text:

1. A table of acceptance criteria and the reported device performance

This information is not explicitly provided in terms of numerical acceptance criteria or performance metrics. The document instead focuses on comparative characteristics and "comparable safety and effectiveness" to a predicate device.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

This information is not provided. The document broadly mentions "clinical testing" but does not detail the size or nature of the test set, nor its provenance or study design (retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not provided. The concept of "ground truth" establishment by experts for a test set is not discussed in the context of this 510(k) summary.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not provided. Adjudication methods are not discussed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study is not mentioned and is not relevant for this type of device (an embolization coil, not an AI diagnostic tool).

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical implant, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

This information is not explicitly stated as the framework of "ground truth" against specific criteria is not detailed. The study's "clinical testing" likely refers to patient outcomes or clinical observations, but specifics are missing.

8. The sample size for the training set

This information is not provided. A "training set" in the context of machine learning is not applicable here. While there were "animal studies" and "in-vitro" testing, their sample sizes are not specified.

9. How the ground truth for the training set was established

This information is not applicable/provided. The concept of "ground truth for a training set" is for AI models, not for a physical medical device's testing directly.


Summary of Study and Device Performance (Based on provided text):

The study described is a comparative study aiming to demonstrate "substantial equivalence" of the TRUFILL® DCS Detachable Coil System to a predicate device, Target's Guglielmi Detachable Coil (GDC).

Acceptance Criteria (Implied by Substantial Equivalence):

The acceptance criteria are implicitly that the TRUFILL® DCS Detachable Coil System performs as safely and effectively as the predicate device (Target's GDC) across various characteristics and biological interactions. This is assessed through:

  • Comparative Characteristics: Matching or being comparable in design, materials, and functional aspects to the predicate device.
  • Non-Clinical Performance Data: Showing acceptable performance in in-vitro and animal tests.
  • Clinical Data: Demonstrating comparable safety and effectiveness in human use.

Table of Performance (Based on Comparative Characteristics):

CharacteristicsCordis Neurovascular, Inc. TRUFILL® DCS Detachable Coil and TRUFILL® DCS SyringeTarget's Guglielmi Detachable Coil GDCNotes
Anatomical SitesPeripheral and NeurovasculaturePeripheral and NeurovasculaturePerformance/Equivalence: Identical
Intended UseAneurysms, AVMs, AVFs, Long-Term UseAneurysms, AVMs, AVFs, Long-Term UsePerformance/Equivalence: Identical
Method of Coil AttachmentMechanical Elastomeric Press FitGold/Tin Solder JointPerformance/Equivalence: Different method, but implicitly considered equivalent in function.
Method of Coil DetachmentHydraulically expandable gripper materialElectrolytically dissolvable solder jointPerformance/Equivalence: Different method, but implicitly considered equivalent in function.
Detachment FeedbackPressure Gauge with detachment indicatorVoltmeter and ammeter with LED detachment indicatorPerformance/Equivalence: Different feedback mechanisms, but implicitly considered equivalent in providing necessary feedback for detachment.
Coil Shape ConfigurationsHelical and ComplexHelicalPerformance/Equivalence: TRUFILL® offers "Complex" in addition to "Helical," which is broadening the options but is not presented as a performance inferiority.
Coil Wire Outer Diameter (in)0.0015-0.0040.00175-0.004Performance/Equivalence: Very similar range, with TRUFILL® covering a slightly smaller minimum.
Primary Coil Diameter (in)0.010-0.0160.0095-0.015Performance/Equivalence: Very similar range, with TRUFILL® covering a slightly larger maximum and GDC a slightly smaller minimum.
Secondary Coil Diameter (mm)2 - 202 - 20Performance/Equivalence: Identical range.
Coil Length (cm)2 - 302 - 30Performance/Equivalence: Identical range.
Coil MaterialPlatinum/TungstenPlatinum/TungstenPerformance/Equivalence: Identical.
Delivery System Usable Length (cm)155-210175 - 195Performance/Equivalence: TRUFILL® offers a broader range of usable lengths.
Delivery System Body DesignMicrocatheter designGuidewire designPerformance/Equivalence: Different designs, but implicitly considered equivalent in delivering the coil.
Radiopaque Marker BandsPlatinum/Tungsten coils, Window designPlatinum/Tungsten coil with Gold/Tin solder attachment joints. T designPerformance/Equivalence: Different designs and materials for attachment, but implicitly considered equivalent in providing radiopacity for visualization.

The Study:

The study proving the device meets the acceptance criteria (of substantial equivalence) consisted of:

  • In-vitro testing:
    • Embolic coil radiopacity
    • Markerband radiopacity
    • Embolic coil softness
    • Force exerted on the wall of a simulated aneurysm
    • Force required to push the device through a microcatheter
    • Migration in a simulated worst-case fistula model
    • Distal tip softness
    • Distal tip angular displacement
  • Animal studies
  • Biocompatibility tests
  • Clinical testing: Demonstrated "comparable safety and effectiveness to the predicate device."

Conclusion: Results of in-vitro, animal, and clinical testing demonstrated that the TRUFILL® DCS Detachable Coil System is substantially equivalent to the predicate device, Target's GDC.

{0}------------------------------------------------

7 2002 MAR

Summary of Safety and Effectiveness

510(k) Summary of Safety and Effectiveness

Trade nameThe trade name of this device is: TRUFILL® DCS Detachable Coil and TRUFILL® DCSSyringe, also known as the TRUFILL® Detachable Coil System (DCS).
Common NameThe common name of this device is: Artificial Embolization Device.
Applicant'sNameCordis Neurovascular, Inc.14000 NW 57 CourtMiami, Florida 33014
Contact PersonAlina CaraballoRegulatory Affairs Manager(305) 512-6518(305) 512-6480 fax
Summary DateDecember 5, 2001
Continued on next page

{1}------------------------------------------------

K014041

510(k) Summary of Safety and Effectiveness, Continued

The TRUFILL® DCS Detachable Coil System is intended for embolizing certain intracranial TRUFILL® DCS aneurysms that - because of their morphology, their location, or the patient's general medical Detachable Coil condition - are considered by the treating neurosurgical team to be: Intended Use very high risk for management by traditional operative techniques, or, a) b) inoperable, and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature. The TRUFILL® DCS Detachable Coil System is also intended for arterial and venous embolizations in the peripheral vasculature. 2017年07月25日 11:54:00 AM R 11:50 PM R 11:00 AM 10:00 PM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 10:00 AM 1 The TRUFILL® DCS Detachable Syringe is indicated for use with the TRUFILL® DCS TRUFILL® DCS Syringe Intended Detachable Coil. Use The table below compares the TRUFILL® DCS Detachable Coil System with the currently

Comparative Characteristics

marketed device Target's Guglielmi Detachable Coil (GDC).

CharacteristicsCordis Neurovascular, Inc.TRUFILL® DCS Detachable Coiland TRUFILL® DCS SyringeTarget's Guglielmi Detachable CoilGDC
Anatomical SitesPeripheral and NeurovasculaturePeripheral and Neurovasculature
Intended UseAneurysms, AVMs, AVFs, Long-TermUseAneurysms, AVMs, AVFs, Long-TermUse
Method of Coil AttachmentMechanical Elastomeric Press FitGold/Tin Solder Joint
Method of Coil DetachmentHydraulically expandable grippermaterialElectrolytically dissolvable solder joint
Detachment FeedbackPressure Gauge with detachmentindicatorVoltmeter and ammeter with LEDdetachment indicator
Coil Shape ConfigurationsHelical and ComplexHelical
Coil Wire Outer Diameter (in)0.0015-0.0040.00175-0.004
Primary Coil Diameter (in)0.010-0.0160.0095-0.015
Secondary Coil Diameter (mm)2 - 202-20
Coil Length (cm)2 - 302 - 30
Coil MaterialPlatinum/TungstenPlatinum/Tungsten
Delivery System Usable Length (cm)155-210175 - 195
Delivery System Body DesignMicrocatheter designGuidewire design
Radiopaque Marker BandsPlatinum/Tungsten coils, Window designPlatinum/Tungsten coil with Gold/Tinsolder attachment joints. T design

Continued on next page

{2}------------------------------------------------

510(k) Summary of Safety and Effectiveness, Continued

DeviceCompany510 (k) Number/Concurrence DateProductCodePredicate for:
Guglielmi DetachableCoilBoston Scientific,TargetK951256HCGEmbolic coil
K960705Delivery system
K962503Detachment mechanism
K971395Sterilization
K991139
K993415
K993417
K993418
K001083
K002181

The predicate device is listed in the table below. Predicate Devices

Device Description

The TRUFILL® DCS Detachable Coil System is comprised of the TRUFILL® DCS Detachable Coil and TRUFILL® DCS Syringe.

  • The TRUFILL® DCS Detachable Coil consists of a delivery system (delivery tube and . coil introducer) and an embolic coil. The delivery tube is comprised of a hub, a strain relief, a stiff proximal section, and a floppy distal section. The coil introducer is a tube designed to protect the detachable embolic coil in the packaging dispenser and provide support for introducing the embolic coil into the microcatheter. The embolic coil is the implantable segment of the device. It is comprised of a vasocclusion coil wound from a platinum alloy wire (92% Platinum / 8% Tungsten) into a primary coil and then formed into a secondary helical or complex shape.
  • The TRUFILL®DCS Syringe consists of a 25-cc barrel with a pressure gauge, a threaded . plunger assembly with a locking wing mechanism, and a flexible high-pressure extension tube with a male luer connector.

Continued on next page

{3}------------------------------------------------

014041

510(k) Summary of Safety and Effectiveness, Continued

Non-Clinical Performance Data

AMA

,

The following in-vitro and animal testing was presented to support substantial equivalence to the predicate device.

Comparative Testing
Embolic coil radiopacity
Markerband radiopacity
Embolic coil softness
Force exerted on the wall of a simulated aneurysm
Force required to push the device through a microcatheter
Migration in a simulated worst case fistula model
Distal tip softness
Distal tip angular displacement
Animal studies

All appropriate biocompatibility tests were successfully performed on the materials used to Biocompatibility manufacture the TRUFILL® DCS Detachable Coil System.

Clinical Data Results of clinical testing demonstrated comparable safety and effectiveness to the predicate device.

2017年 - 2012 - 10:50 - 10:50 - 10:50 - 10:50 - 10:50 - 10:50 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00 - 10:00

Conclusion

Results of in-vitro, animal, and clinical testing demonstrated that the TRUFILL® DCS Detachable Coil System is substantially equivalent to the predicate device, Target's GDC.

】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 】 【 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 【 】 】 【 】 】 【 】 【 】 】 【 】 】 【 】 【 】 】 【 】 】 【 】 【 】 】 【 】 【 】 】 【 】 】 【 】

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff and a serpent winding around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

7 2002 MAR

Ms. Alina Caraballo Manager, Regulatory Affairs Cordis Neurovascular, Inc. 14000 NW 57th Court Miami Lakes, Florida 33014

Re: K014041

Trade/Device Name: TRUFILL® DCS Detachable Coil System Regulation Number: 882.5950 and 870.3300 Regulation Name: Artificial embolization device, Arterial embolization device Regulatory Class: III Product Code: HCG and KRD Dated: December 5, 2001 Received: December 7, 2001

Dear Ms. Caraballo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Ms. Alina Caraballo

This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally prematicated predicated on "The classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of Compliance at (301) 594-4639 Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Oiner general mistmances, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): K O1 YOU (

Device Name: TRUFILL® DCS Detachable Coil System comprised of the TRUFILL® DCS Detachable Coil and TRUFILL® DCS Syringe

Indications for Use Statement

The TRUFILL® DCS Detachable Coil System is intended for embolizing certain intracranial aneurysms that because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:

very high risk for management by traditional operative techniques, or, a)

inoperable, b)

and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature.

The TRUFILL® DCS Detachable Coil System is also intended for arterial and venous embolizations in the peripheral vasculature.

The TRUFILL® DCS Detachable Syringe is indicated for use with the TRUFILL® DCS Detachable Coil.

Miriam C. Provost

Division of General, Restorative and Neurological Devices

510(k) Number K014041

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X

OR

Over-The-Counter Use

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).