K Number
K993415

Validate with FDA (Live)

Date Cleared
2000-01-21

(112 days)

Product Code
Regulation Number
882.5950
Age Range
0 - 65
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Guglielmi Detachable Coil (GDC) is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:

a)very high risk for management by traditional operative techniques, or,

b) inoperable,

and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature.

The GDC is also intended for arterial and venous embolizations in the peripheral vasculature.

The indications for use for Stretch Resistant GDC are identical to those for the predicate GDC device cleared under K962503.

Device Description

The GDC system consists of:

  • · GDC power supply
  • · GDC occlusion coil attached to a delivery wire
  • · set of GDC connecting cables
  • patient return electrode
  • · two 9-volt batteries

each of which is sold separately.

The occlusion coil is detached by electrolytically dissolving a small portion of the delivery wire upon desired placement of the coil in the anatomy.

GDC occlusion coils are manufactured from platinum wire which is first wound into a primary coil and then formed into a secondary helical shape.

Coils are attached to a delivery wire, which consists of a ground stainlesssteel core wire with a stainless-steel coil welded at the distal end and a Teflon® outer jacket. The delivery wire is the same as that employed for the predicate GDC cleared under K962503.

The GDC Power Supply is a self-contained, battery-operated unit designed to initiate and control detachment of a GDC coil. It has an on/off button, connections for the GDC cables, and the following controls and displays:

  • · Current setting switch and display Allows for selection of the different current settings. The selected current (mA) is briefly displayed after which the unit displays the current flowing through the system.
  • · Voltage diplay and indicator Displays the DC output voltage.
  • · Time display Displays the elapsed time that the current has been flowing through the GDC system.
  • Check indicator Flashes on and off at the beginning of a detachment procedure if there is a poor connection to the patient.
  • · Detach indicator Flashes on and off when the power supply has detected detachment. Detachment is accompanied by five beeps.
  • · Battery indicator Illuminates when the voltage of the internal 9-volt batteries falls below specification.
AI/ML Overview

This document describes the safety and effectiveness of the Guglielmi Detachable Coil (GDC), a Class III artificial embolization device. The information provided focuses on the "Stretch Resistant GDC" and its comparison to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

The document summarizes a "Verification Test Summary Table: Comparison of Stretch Resistant GDC Devices to the Predicate Device". The acceptance criteria for most tests were established by the predicate device (K962503). For the "Main Coil Stretch Test," acceptance criteria were established for the modified device.

Test or Point of ComparisonAcceptance Criteria (explicitly stated or inferred from predicate comparison)Reported Device Performance (Stretch Resistant GDC)
FrictionMeets acceptance criteria established for predicate device.Meets acceptance criteria established for predicate device.
Tensile Strength, Main Coil WeldMeets acceptance criteria established for predicate device.Meets acceptance criteria established for predicate device.
Detachment TimeMeets acceptance criteria established for predicate device.Meets acceptance criteria established for predicate device.
Detachment in Saline w/ particulate analysisNo change made that would result in particulate generation.No change was made which would result in the generation of particulate during detachment.
Heating Effect of ElectrolysisNo change made that would influence or increase heating effect.No change made which would influence or increase heating effect.
Heating Effect of MRINo change made that would increase heating effect of MRI.No change made which would increase heating effect of MRI.
Tip Ball StrengthMeets acceptance criteria established for strength of tip ball.Meets acceptance criteria established for strength of tip ball.
Main Coil Stretch TestMeets acceptance criteria for the modified device.Meets acceptance criteria for the modified device.

2. Sample Sizes Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for each specific test mentioned in the "Verification Test Summary Table". It is a descriptive summary of test outcomes rather than a detailed study report with sample sizes.

The data provenance is not specified. However, given that this is a 510(k) submission for a medical device in the US, the testing and data generation would typically follow US regulatory guidelines and standards. It is retrospective in the sense that these tests were conducted to demonstrate substantial equivalence to a previously cleared device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

This document is a technical summary for regulatory submission, focusing on material and functional comparisons of a medical device (GDC coils). It does not involve "ground truth" derived from expert consensus on diagnostic interpretations (like for an AI algorithm). Instead, the performance is assessed against established engineering and safety standards, benchmarks set by the predicate device, or specific acceptance criteria defined for the modified device. Therefore, information about the number or qualifications of experts used to establish "ground truth" in this context is not applicable.

4. Adjudication Method for the Test Set:

Not applicable. The tests described are objective engineering and material performance assessments, not subjective diagnostic interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance:

Not applicable. This document pertains to a physical medical device (embolization coil), not an AI algorithm for diagnostic assistance. Therefore, an MRMC study or the effect size of AI assistance on human readers is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This is not an AI algorithm.

7. The Type of Ground Truth Used:

The "ground truth" in this context refers to established engineering specifications, performance benchmarks derived from the predicate device (K962503), or specific design requirements for the new "Stretch Resistant GDC." For example:

  • Friction: Measured against a specific frictional force threshold.
  • Tensile Strength, Main Coil Weld: Compared against a minimum tensile strength value.
  • Detachment Time: Evaluated against a specified temporal range for detachment.
  • Particulate Analysis: Absence of particulate generation.
  • Heating Effects (Electrolysis, MRI): No increase beyond acceptable thresholds or predicate device performance.
  • Tip Ball Strength: Compared against a minimum strength requirement.
  • Main Coil Stretch Test: Evaluated against specific criteria established for the elasticity and resistance of the modified device.

These benchmarks and criteria are based on established medical device safety and efficacy standards and the performance history of the predicate device.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a handwritten string of alphanumeric characters. The string appears to be "K993415". The characters are written in a cursive style, with some connections between the letters and numbers. The image is in black and white.

Section 2 - 510 (k) Summary of Safety and Effectiveness and Class III Certification and Summary

a. Summary Of Safety And Effectiveness

Contact Person

Roxane Baxter Regulatory Affairs Manager Boston Scientific / Target 47900 Bayside Parkway Fremont, CA. 94538

Trade Name

Guglielmi Detachable Coil (GDC), Class III

Common Name

Occlusion Coil

Classification Name

Artificial Embolization Device (21 CFR Section 882.5950)

{1}------------------------------------------------

Predicate Devices

NumberDescriptionPredicate forClearanceDate
K962503(Boston Scientific /Target)Guglielmi Detachable CoilGDC-10 and GDC-10(2D) Stretch ResistantGuglielmiDetachable CoilsandGDC-18 and GDC-18(2D) Stretch ResistantGuglielmiDetachable Coils20 Sept. 1996
K930738 (Deknatel,Division ofHowmedica, Inc.)Implantable SurgicalSuturesSuture material26 July 1994
K971395(Boston Scientific /Target)Shelf Life, Various Class IIand Class III Devices forFor shelf-lifevalidation methodand subsequentplacement of shelf-life information ondevice label14 July 1997

Intended Use

The Guglielmi Detachable Coil (GDC) is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:

a)very high risk for management by traditional operative techniques, or,

b) inoperable,

and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature.

The GDC is also intended for arterial and venous embolizations in the peripheral vasculature.

The indications for use for Stretch Resistant GDC are identical to those for the predicate GDC device cleared under K962503.

{2}------------------------------------------------

Device Description

The GDC system consists of:

  • · GDC power supply
  • · GDC occlusion coil attached to a delivery wire
  • · set of GDC connecting cables
  • patient return electrode
  • · two 9-volt batteries

each of which is sold separately.

The occlusion coil is detached by electrolytically dissolving a small portion of the delivery wire upon desired placement of the coil in the anatomy.

GDC occlusion coils are manufactured from platinum wire which is first wound into a primary coil and then formed into a secondary helical shape.

Coils are attached to a delivery wire, which consists of a ground stainlesssteel core wire with a stainless-steel coil welded at the distal end and a Teflon® outer jacket. The delivery wire is the same as that employed for the predicate GDC cleared under K962503.

{3}------------------------------------------------

The GDC Power Supply is a self-contained, battery-operated unit designed to initiate and control detachment of a GDC coil. It has an on/off button, connections for the GDC cables, and the following controls and displays:

  • · Current setting switch and display Allows for selection of the different current settings. The selected current (mA) is briefly displayed after which the unit displays the current flowing through the system.
  • · Voltage diplay and indicator Displays the DC output voltage.
  • · Time display Displays the elapsed time that the current has been flowing through the GDC system.
  • Check indicator Flashes on and off at the beginning of a detachment procedure if there is a poor connection to the patient.
  • · Detach indicator Flashes on and off when the power supply has detected detachment. Detachment is accompanied by five beeps.
  • · Battery indicator Illuminates when the voltage of the internal 9-volt batteries falls below specification.

{4}------------------------------------------------

Technological Characteristics Comparison

Coil Dimensional Attributes

GDC(Predicate device clearedunder K962503)Stretch Resistant GDC
Coil Primary Wind OD0.010" - 0.015"0.0095", 0.010", 0.0135" or0.015"
Secondary Coil ODGDC-10: 2 mm - 30 mmGDC-10: 2 mm - 30 mm
GDC-18: 2 mm - 30 mmGDC-18: 2 mm - 30 mm
Coil Wire ODRange: 0.001" - 0.005"Current wire sizes used in theproduction of GDC devicesare:GDC-10: 0.00175", 0.002"GDC-18: 0.00225", 0.003",0.004"GDC-10: 0.0175", 0.002"GDC-18: 0.00225", 0.003"
Delivery Wire Length50 - 200 cm175 cm
Delivery WireProximal ODGDC-10: 0.010"GDC-18: 0.010"GDC-10: 0.010"GDC-18: 0.010"
Delivery WireDistal ODGDC-10: 0.0028"GDC-18: 0.0028"

{5}------------------------------------------------

.

Technological Characteristics Comparison (cont.)

Materials

Stretch Resistant GDC
Main CoilSame as predicate device
Stretch ResistantFilamentPolymer
Delivery Wire:
Core wirew/coatingSame as predicate device
Proximal CoilSame as predicate device
Proximal MarkerCoilSame as predicate device
Sheath, DeliveryWire (heat shrinktubing)Same as predicate device
Proximal TubingSame as predicate device
BushingSame as predicate device
HypotubeN/A
Anchor CoilPlatinum

{6}------------------------------------------------

Technological Characteristics Comparison (cont.)

Power Supply

GDC Power Supply,Stretch Resistant GDC
PowerSame as for predicate device.
BatteriesSame as for predicate device.
Expected BatteryLifeSame as for predicate device.
Red CableSame as for predicate device.
Black CableSame as for predicate device.
Current SettingsSame as for predicate device.
CurrentSame as for predicate device.
VoltageSame as for predicate device.
Operating Temp.Same as for predicate device.
Storage Temp.Same as for predicate device.
Relative HumiditySame as for predicate device.
Unit SizeSame as for predicate device.
Unit WeightSame as for predicate device.

{7}------------------------------------------------

Verification Test Summary Table: Comparison of Stretch Resistant GDC Devices to the Predicate Device

Test or Point of ComparisonStretch Resistant GDC
FrictionMeets acceptance criteria established forpredicate device.
Tensile Strength, Main Coil WeldMeets acceptance criteria established forpredicate device.
Detachment TimeMeets acceptance criteria established forpredicate device.
Detachment in Saline w/ particulateanalysisNo change was made which would resultin the generation of particulate duringdetachment.
Heating Effect of ElectrolysisNo change made which would influenceor increase heating effect.
Heating Effect of MRINo change made which would increaseheating effect of MRI.
Tip Ball StrengthMeets acceptance criteria established forstrength of tip ball.
Main Coil Stretch TestMeets acceptance criteria for themodified device.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol consisting of three stylized human figures connected by flowing lines, representing health and human services.

Public Health Service

JAN 2 1 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Roxane K. Baxter, RAC Manager, Regulatory Affairs Boston Scientific/Target 47900 Bayside Parkway Freemont, California 94538

Re: K993415

Trade Name: GDC-10 Stretch Resistant Guglielmi Detachable Coil GDC-10 (2D) Stretch Resistant Guglielmi Detachable Coil GDC-18 Stretch Resistant Guglielmi Detachable Coil GDC-18 (2D) Stretch Resistant Geglielmi Detachable Coil Regulatory Class: III Product Code: HCG and KRD Dated: September 29, 1999 Received: October 1, 1999

Dear Ms. Baxter:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{9}------------------------------------------------

Page 2 - Ms. Roxane K. Baxter, RAC

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

James E. Dillard III

James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains a sequence of handwritten alphanumeric characters. The characters appear to be "K993415". The characters are written in a cursive style, with some connections between the letters and numbers. The image is in black and white.

Image /page/10/Picture/1 description: The image shows the logo for Boston Scientific TARGET. The words "Boston Scientific" are stacked on top of each other in a large, bold font. Below that, the word "TARGET" is written in a larger, bolder font, and it is underlined. The logo is simple and professional.

INDICATIONS FOR USE STATEMENT

510(k) Number: __

Device Name: GDC-10 Stretch Resistant Guglielmi Detachable Coil GDC-10 (2D) Stretch Resistant Guglielmi Detachable Coil and GDC-18 Stretch Resistant Guglielmi Detachable Coil GDC-18 (2D) Stretch Resistant Guglielmi Detachable Coil

Indications for Use:

The Guglielmi Detachable Coil (GDC) is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:

a) very high risk for management by traditional operative techniques, or,

b) inoperable,

and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature.

The GDC is also intended for arterial and yenous embolizations in the peripheral vasculature.

(Division Sign-Off)
Division of General Restorative Devices K993415
510(k) Number.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseX
(Per 21 CFR 801.109)

510(k) Notification, Boston Scientific/Target, GDC-10 and GDC-18 Stretch Resistant GDC CONFIDENTIAL Section 3, Page 2 September 1999

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).