K Number
K012899

Validate with FDA (Live)

Date Cleared
2001-09-18

(20 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Green Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner (21 CFR 880.6250).

Device Description

Nitrile rubber films form an excellent barrier to body fluids and bloodborne pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing fine movement necessary for treatment. The absence of natural rubber latex in the product yields no latex protein allergens. Nitrile rubber is known to create a superior barrier to bloodborne pathogens and body fluids.

AI/ML Overview

The provided document describes a 510(k) submission for a medical examination glove, which is a low-risk device. This type of submission relies on demonstrating substantial equivalence to a predicate device, rather than extensive clinical studies with complex acceptance criteria and ground truth establishment typically seen for higher-risk devices or AI/ML-driven diagnostics.

Therefore, many of the requested categories are not applicable or do not have direct equivalents in this type of submission.

Here's an analysis based on the provided text, indicating where information is not applicable:

1. Table of Acceptance Criteria and Reported Device Performance

SpecificationAcceptance Criteria (Predicate Performance)Reported Device Performance (Proposed Device)
Performance StandardsASTM standardsASTM standards (meets or exceeds)
WatertightnessASTM standardsASTM standards (meets or exceeds)
Safety (Non-clinical tests)
Rabbit IrritationPasses (Implied by predicate performance)Passes
Guinea Pig SensitizationPasses (Implied by predicate performance)Passes
Modified Draze TestPasses (Implied by predicate performance, though not performed on proposed device, similarity is key)N/A (Similarity to predicate device is the basis for not repeating this specific test)

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated as this is a non-clinical, performance-based comparison to a predicate device, not a human reader study or clinical trial with a defined "test set" in the context of diagnostic accuracy.
  • Data Provenance: The data refers to "Non-clinical tests (Human Study)" (referring to aspects like rabbit irritation and guinea pig sensitization, which use animal models, not human subjects for testing the glove's barrier properties or direct clinical performance). The provenance is best manufacturing company and not a country of origin. This is retrospective in the sense that the test results for the predicate are already established, and the proposed device is being evaluated against those established results.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This submission does not involve human expert interpretation of data to establish a "ground truth" for diagnostic purposes. The ground truth for glove performance is established by standardized ASTM tests and animal safety models.

4. Adjudication method for the test set

  • Not Applicable. There is no adjudication method in the context of expert review for diagnostic agreement as this is not a diagnostic device or a clinical study involving human judgment on cases.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a medical examination glove, not an AI-assisted diagnostic device. Therefore, no MRMC study with human readers and AI assistance was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical medical device, not an algorithm or software. Therefore, no standalone algorithm performance was assessed.

7. The type of ground truth used

  • Standardized Performance Metrics: The "ground truth" is defined by established ASTM (American Society for Testing and Materials) standards for watertightness and "Performance Standards," as well as results from standardized biocompatibility tests (Rabbit Irritation, Guinea Pig Sensitization, Modified Draze Test). These are objective, laboratory-based measurements rather than expert consensus, pathology, or outcomes data in the typical sense for diagnostic devices.

8. The sample size for the training set

  • Not Applicable. This is not an AI/ML device that requires a training set. The "training" for this device comes from the established manufacturing processes and material science, and its substantial equivalence is based on direct comparison to an already approved device.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no training set involved, the concept of establishing ground truth for it does not apply.

{0}------------------------------------------------

510(K) SUBMISSION FOR GREEN NITRILE POWDER-FREE MEDICAL EXAMINATION GLOVE (S SUBMISSION DATE: 2001-06-15

SUMMARY OF SAFETY AND EFFECTIVENESS

A. INFORMATION

  1. SUBMITTER'S

Name:

BEST MANUFACTURING COMPANY

Address:579 Edison StreetMenlo, GA 30731 USA
Telephone Number:706 862 2302
Contact Person:David C. Young
Date Summary Prepared:2001-06-15
2. NAME OF DEVICETrade or Proprietary Name:Nitrile Powder-Free MedicalExamination Glove (Green)
Common or Usual NameNon-Sterile Nitrile Powder-Free PatientExamination Glove
Classification Name:Patient Examination Glove
3. PREDICATE DEVICEIDENTIFICATION NAMEN-DEX Nitrile Powder-Free MedicalExamination Glove, K992170

4. DESCRIPTION OF DEVICE

NUMBER

a. How the device functions:

Nitrile rubber films form an excellent barrier to body fluids and bloodborne pathogens.

  • b. Scientific concepts that form the basis for the device: The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing fine movement necessary for treatment. The absence of natural rubber latex in the product vields no latex protein allergens.
  • c. Physical and performance characteristics such as design, materials, and physical properties:

Nitrile rubber is known to create a superior barrier to bloodborne pathogens and body fluids. PAGE I OF 3

{1}------------------------------------------------

510(K) SUBMISSION FOR GREEN NITRILE POWDER-FREE MEDICAL EXAMINATION GLOVE SUBMISSION DATE: 2001-06-15

    1. STATEMENT OF INTENDED USE, INCLUDING DESCRIPTION OF THE DISEASE OR CONDITIONS THAT THE DEVICE WILL ADDRESS This is a disposable device, intended for medical purposes, that is worn on the examiner's hand to prevent contamination between the patient and examiner. Powder-free examination gloves are suitable in situations where powder is not desirable.
    1. EXPLANATION OF SIMILARITIES OR DIFFERENCES TO PREDICATE DEVICE
    • · The proposed device is identical to the predicate device, except for the following:

The proposed device has been rendered green in color, while the predicate is blue.

B. IF SE DECISION BASED ON PERFORMANCE DATA: 1. DISCUSSION OF NON-CLINICAL TESTS

(Human Study)

SpecificationProposedPredicate
N-DEX Nitrile Powder-Free Medical Examination Glove (green)N-DEX Nitrile Powder-Free Medical Examination Glove (blue)
Performance StandardsASTMASTM
WatertightnessASTMASTM
2. DISCUSSION OF CLINICAL TESTS
SpecificationProposedPredicate
Safety
Rabbit IrritationPassesPasses
Guinea Pig SensitizationPassesPasses
Modified Draze TestN/APasses

PAGE 2 OF 3

{2}------------------------------------------------

510(K) SUBMISSION FOR GREEN NITRILE POWDER-FREF. MEDICAL EXAMINATION GLOVE SUBMISSION DATE: 2001-06-15

DISCUSSION OF SAFETY OR EFFECTIVENESS DATA OBTAINED (with specific reference to adverse effects and complications)

See Section M: Biocompatibility Testing.

    1. CONCLUSIONS DRAWN FROM NONCLINICAL AND CLINICAL TESTS THAT DEMONSTRATE SAFETY AND EFFECTIVENESS, AND Performance == > PREDICATE PRODUCT
      The Nitrile Powder-Free Medical Examination Glove (green) has been carefully compared to a legally marketed device in the 510(k). The data summaries indicate that the proposed product meets or exceeds accepted scores for the predicate product in both physical and nonclinical tests and satisfies the requirements for a safe and effective "powder-free" medical glove.

Pursuant to 21 C.F.R. 807.87 (j), I David C. Young, Director, Regulatory Affairs and Quality Assurance, certify that to the best of my knowledge and belief and based upon the data and information submitted to me in the course of my responsibilities as the Director, Regulatory Affairs and Quality Assurance, for the Best Manufacturing Company, and in reliance thereupon, the data and information submitted in this premarket notification are truthful and accurate and that no facts material to a review of the substantial equivalence of this device have been knowingly omitted from this submission.

David C. U

David C. Young, Director Regulatory Affairs & Quality Assurance

2001-06-15 DATE

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image contains the words "Public Health Service". The text is in a simple, sans-serif font. The words are arranged on a single line, with "Public Health" on the left and "Service" on the right.

Image /page/3/Picture/3 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol in the center consists of three stylized, curved lines that resemble a stylized human figure.

SEP 1 8 2001

Mr. David C. Young Director of Regulatory Affairs Best Manufacturing Company 579 Edison Street Menlo, Georgia 30731

Re: K012899

Trade/Device Name: Green Nitrile Powder-Free Medical Examination Glove Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA Dated: June 15, 2001 Received: August 29, 2001

Dear Mr. Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

{4}------------------------------------------------

Page 2 - Mr. Young

of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (section 534-542 of the Act; 21); CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Los A. Walberto

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(K) SUBMISSION FOR GREEN NITRILE POWDER-FREE MEDICAL EXAMINATION GLOVE SUBMISSION DATE: 2001-06-15

INDICATIONS FOR USE

Best Manufacturing Company Applicant: 510(k) Number (if known) KO12899 Device Name: Green Nitrile Powder-Free Medical Examination Glove

The Green Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner (21 CFR 880.6250).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

..............................................................................................................................................................................

Concurrence of CDRH Office of Device Evaluation (ODE) ાર Over-The-Counter______________________________________________________________________________________________________________________________________________________________ Prescription Use Per 21 CFR 801.109 * For a new submission, do NOT fill in the 510(k) number blank.

PAGE 3 OF 11

(Division Sign-Off) Division of Dental, Infection Control and General Hospital Devi 510(k) Number .

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.