K Number
K011561

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-11-20

(183 days)

Product Code
Regulation Number
886.5928
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sereine Extra Strength Daily Cleaner is indicated for use as a daily cleaner for removal of film, deposits and other debris which accumulate on soft (hydrophilic) and hard (PMMA) contact lenses during wear. Sereine Extra Strength Daily Cleaner may be used with either thermal (heat) or chemical (cold) disinfection systems. Follow the recommendation of your eye care practitioner when disinfecting your lenses.

Device Description

Sereine Extra Strength Daily Cleaner is a sterile, viscous, aqueous solution with an alkaline pH. The cleaner contains an amphoteric surfactant to aid in the removal of proteins, lipids and other debris which accumulate on contact lenses during wear. The cleaner is not meant for use in the eye and must be rinsed from the lenses. Sereine Extra Strength Daily Cleaner is packaged in a 1 FL OZ plastic bottle (polyethylene) with red dropper tip (polyolefin) and over-cap (polypropylene). It has an outside cardboard carton. It has an expiration date of two years from the date of bottling.

AI/ML Overview

This document describes the Sereine Extra Strength Daily Cleaner, a contact lens cleaning solution, and outlines its similarity to a predicate device, MiraFlow Extra Strength Daily Cleaner. The submission focuses on demonstrating substantial equivalence rather than conducting a separate clinical study with distinct acceptance criteria for novel performance.

Here's an breakdown of the information provided, tailored to your request:

Acceptance Criteria and Device Performance (Demonstrated Equivalence)

The acceptance criteria for Sereine Extra Strength Daily Cleaner appear to be based on demonstrating close similarity in key physical and chemical properties to the legally marketed predicate device, MiraFlow Extra Strength Daily Cleaner. The reported performance demonstrates that Sereine meets these criteria by matching the predicate's values.

Acceptance Criteria (Based on Predicate Match)Reported Device Performance (Sereine Extra Strength Daily Cleaner)
pH: 8.3 (matching MiraFlow)8.3
Viscosity: 74 cp (matching MiraFlow)74 cp
Total solids: 20.6% (matching MiraFlow)20.8%
Sterility: Sterile (matching MiraFlow)Sterile
UV-visible spectrum: Match MiraFlowMatch MiraFlow

The "study" that proves the device meets (or, more accurately, is substantially equivalent to) the acceptance criteria is a series of laboratory tests comparing the Sereine Extra Strength Daily Cleaner to the MiraFlow Extra Strength Daily Cleaner.


Study Details

Due to the nature of this 510(k) submission for a contact lens cleaning solution, the provided document does not describe a clinical study with human participants, expert ground truth, or an MRMC comparative effectiveness study in the context of diagnostic AI tools. Instead, it focuses on laboratory bench testing and non-clinical performance.

  1. Sample size used for the test set and the data provenance:

    • pH: Three different lots of Sereine Extra Strength Daily Cleaner were tested. Data provenance is internal laboratory testing.
    • Viscosity: Three lots of Sereine Extra Strength Daily Cleaner were tested. Data provenance is internal laboratory testing.
    • Total Solids: Three samples of Sereine Extra Strength Daily Cleaner were tested. Data provenance is internal laboratory testing.
    • Sterility: Sereine Extra Strength Daily Cleaner was tested for sterility by randomly sampling bottles from the lot. The exact number of samples is not specified beyond "randomly sampling bottles." The testing was performed by Nelson Laboratories, Inc., Salt Lake City, UT. (Provenance: US, prospective testing for the submission).
    • UV-visible spectrum: One sample of Sereine Extra Strength Daily Cleaner was compared. Data provenance is internal laboratory testing.
    • Stability Study (pH, Viscosity, Active Ingredients, Physical Appearance, Sterility, Preservative Effectiveness): Three lots of Sereine Extra Strength Daily Cleaner were stored at ambient temperature for two years and tested periodically. (Provenance: US, prospective testing for the submission).
    • Solution Compatibility:
      • 10 Metrotint (hydrophilic) lenses
      • 10 Horizon 55 (hydrophilic) lenses
      • 10 PMMA (hard) lenses
        (Provenance: US, prospective testing for the submission).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the study involved physical and chemical characterization of a cleaning solution, not a diagnostic task requiring expert interpretation. The ground truth was established by laboratory measurements and adherence to specified testing procedures (e.g., USP 24 for sterility, FDA Guidelines for preservative effectiveness).
  3. Adjudication method for the test set:

    • Not applicable. The tests involved objective measurements (pH, viscosity, total solids, UV-Vis spectra, sterility) and direct observation for physical appearance and solution compatibility. There was no human interpretation or adjudication method described in the context of expert consensus.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a 510(k) submission for a contact lens cleaning solution, not a diagnostic AI device. Therefore, no MRMC study with human readers assisting AI was conducted or is relevant.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable, as this is not an AI algorithm. The performance described is that of the cleaning solution itself.
  6. The type of ground truth used:

    • Objective Laboratory Measurements: For pH, viscosity, total solids, and UV-visible spectrum, the "ground truth" was established by direct measurement using calibrated equipment (Piccolo pH meter, Gilmont Falling Ball Viscometer, Shimadzu UV-visible Spectrophotometer).
    • Sterility: The ground truth for sterility was established by culturing methods as per USP 24, performed by a third-party laboratory (Nelson Laboratories, Inc.).
    • Preservative Effectiveness: Ground truth was established by microbiological challenge testing according to FDA Guidelines documented in the "Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products," performed by a third-party laboratory (Northview Pacific Laboratories, Inc.).
    • Solution Compatibility: Ground truth was based on objective measurements of optical parameters (before and after 30 cycles) and visual inspection for physical changes (e.g., discoloration, cracks, precipitates) by laboratory personnel.
  7. The sample size for the training set:

    • Not applicable. This is not an AI device that requires a training set. The term "training set" is typically used in machine learning.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K011561 NOV 20 2001

Optikem International, Incorporated 2172 South Jason Street Denver, Colorado 80223 USA

510(K) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS

  1. Submitter's Name

Optikem International, Inc. 2172 South Jason Street Denver, CO 80223

Telephone: 303-936-1137 FAX: 303-936-7320

  1. Contact Person

Sally Cook President

  1. Date Summary was Prepared:

May 8, 2001

Sereine
Contact Lens
Solutions

Optisoap
Hand Soap

{1}------------------------------------------------

4. Name of Device

Trade Name: Sereine Extra Strength Daily Cleaner Common Name: Daily Cleaner for Contact Lenses -Classification Name: Contact Lens Cleaner

    1. Legally Marketed Device to Which Equivalence is Claimed:
      MiraFlow Extra Strength Daily Cleaner, CIBA Vision
      Cleaner, Carength Ca., 00155 Corporation, Duluth, GA 30155.

Description of Device: 6.

Sereine Extra Strength Daily Cleaner is a sterile, viscous, aqueous Sereine Extra Strength Dany Cloaner contains no added solution with an alkaline pri. The citains an amphoteric surfactant to aid in
preservatives. The cleaner contains an amphoteric surfactant to aid in preservatives. The cleaner contains and other debris which accumulate on the removal of proteins, lipids and other dealer is not meant for use in the eye
contact lenses during wear. The cleaner is not meant for use in the eye contact lenses during weak ed from the lenses.

Sereine Extra Strength Daily Cleaner is packaged in a 1 FL OZ plastic Sereine Extra Strength Daily Citcano is passessorse on white bottle (polyethylene) with red uropper tip (poryon) (and carton. It has an
over-cap (polypropylene). It has an outside cardboard carton. It has an
the exp (polypropylene) over-cap (polypropylene). It som the date of bottling.

{2}------------------------------------------------

Intended Use of the Device: 7.

Sereine Extra Strength Daily Cleaner is a daily cleaner for use Sereine Extra Daving in ids and other accumulated debris which collects on contact lenses during wear.

Sereine Extra Strength Daily Cleaner is intended for use with all soft (hydrophilic) lenses and hard (PMMA) lenses on a an soft (1) arope cleaner is for use on lenses immediately following removal from the eye and prior to disinfection.

Sereine Extra Strength Daily Cleaner is intended for use each time the lenses are removed from the eye.

Sereine Extra Strength Daily Cleaner is not intended for use in Screme Entra Buche insed from the lens after cleaning with either the Cyc and mast cophilic lenses) or water (hydrophobic lenses).

{3}------------------------------------------------

    1. Technological Characteristics of the Device as Compared to the Predicate Device:
MiraFlowSereine
pH8.38.3
Viscosity74 cp74 cp
Total solids20.6%20.8%
SterilitySterileSterile
UV-visible spectrumMatch Miraflow

{4}------------------------------------------------

DESCRIPTION OF TESTS CONTAINED IN TABLE ON PREVIOUS PAGE:

1. pH

The pH of three different lots of MiraFlow Extra Strength Daily File pri of there different lots of Sereine Extra Strength Daily Cleaner was obtained by using a calibrated Piccolo pH meter. The results of the three lots were averaged.

2. Viscosity

The viscosity of the MiraFlow Extra Strength Daily Cleaner and the Sereine Extra Strength Daily Cleaner was determined using a Gilmont Falling Ball Viscometer. Three lots of each cleaner were tested and the results averaged. The Gilmont Falling Ball Viscometer was Size 2 (2-100 cps) with a stainless steel I alling Dur . The viscometer was placed in a constant temperature water bath of 25C. Approximately 5 ml of solution was placed in the viscometer and the stainless steel ball was dropped. The time for the stainless steel ball to descend through the liquid was timed with a calibrated Cole-Parmer Model #94410-20 Stop Watch. The viscosity was calculated by averaging the times for the lots of MiraFlow Extra Strength Daily Cleaner. The procedure was repeated for the lots of Sereine Extra Strength Daily Cleaner.

{5}------------------------------------------------

DESCRIPTION OF TESTS CONTAINED IN TABLE (CONT.)

3. Total Solids

The total non-volatile solids were determined by weighing an The total non-voluene vished product of MiraFlow Extra Strength Daily Cleaner in a porcelain evaporating dish. Streign Dany Cloater in been previously weighed (empty). The evaporating dish had boing oven at 105C. The dish was The dish was placed in a diffing obtained. The procedure was Three uned using Sereine Extra Strength Daily Cleaner. Three samples of each product were tested and the results averaged. Sumples or courts were comparable as to total non-volatile solids.

4. Sterility

MiraFlow Extra Strength Daily Cleaner is labeled as a sterile product. Sereine Extra Strength Daily Cleaner was tested for sterility by Nelson Laboratories, Inc., Salt Lake City, UT storing of Norther contained in USP 24. Sereine Extra Strength Daily Cleaner was tested for sterility by randomly sampling Daily Cleaner wers from the lot. Sereine Extra Strength Daily Cleaner was found to be sterile.

{6}------------------------------------------------

DESCRIPTION OF TESTS CONTAINED IN TABLE (CONT).

5. UV-Visible Spectra Comparison

MiraFlow Extra Strength Daily Cleaner and Sereine Extra Strength Milal 10W Examer were compared using a Shimadzu UV-visible Dany Cloudiol were Sompl UV-160. The cleaner was placed neat in a 10 mm rectangular silica absorption cell. The reference was ill a 10 min receanganas orbance range was 0-2.5 absorbance units and the spectral range was 200-1000 nm. The spectra of the two cleaners were comparable indicating like components.

{7}------------------------------------------------

9. Summary of Non-clinical tests

The following non-clinical tests were performed on Sereine The following non emileaner. These tests were not performed Latra Strength Daily Cleaner as Optikem On Mind Tow Encra Ses not have access to materials and data necessary to conduct these tests.

Although the non-clinical tests were not performed on MiraFlow Anthough the non cinner for comparison, it is presumed that Extra Strengan Daily Cleaccessfully in order to obtain pre-market approval.

A. Stability Study

The stability or shelf-life study for Sereine Extra Strength Daily Cleaner was conducted using lots stored at ambient temperature for two years. The ambient temperature was between 60 and 80 F and was recorded on a daily basis as part of the quality assurance program. Three lots of Sereine Extra Strength Daily Cleaner were tested for stability. These lots were bottled in 1 FL OZ (30 ml) plastic bottles. Bottles were capped and the caps sealed with tamper-resistant seals. Bottles were not subjected to any specific conditions as far as humidity or light. The relative humidity is recorded on a daily basis as part of the quality assurance program.

{8}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Summary of Non-clinical tests (cont). 9.

A. Stability Study (cont).

The following characteristics of Sereine Extra Strength Daily Cleaner were monitored:

  1. pH

The pH of each of three lots was taken using a calibrated The pH of each of thee lots was compared to the original
Piccolo pH meter. The result was compared to the original Piccolo pH meter. The finished device acceptance pH obtained as part of the humaneng date. The pH was not
standards from the original bottling and storage standards from the original soming and storage.

  1. Viscosity

The viscosity of each of three lots was measured using a The viscosity of each of the City of Ciscometer. The viscosity Gilmont Size 2 Falling Ball Type Tibecausessity as determined at the time of bottling.

  1. Active Ingredients

The ingredients were determined using a Shimadzu UV-visible The ingredients were determinou uses acception silica absorption Spectrophotometer, Model 100 will a scan of 200-800 mm.
cells. Three stored lots were measured in a scan of 200-800 mm. cells. Three stored lots were was not performed on the lots as part of the original finished device acceptance, the scans were as part of the original imislied on the use. The ingredients were
compared to a recently formulated batch. The ingredied as indicated compared to a recently formations of the storage period as indicated by the UV-visible spectra.

{9}------------------------------------------------

K 0 11 56 1

510(k) SUMMARY

SUMMARY OF SAFETY AND EFFECTIVENESS

Summary of Non-clinical tests (cont.) g.

A. Stability Study (cont.)

    1. Physical Appearance
      Physical appearance of the bottle, tip and cap was observed. Physical appearance of the occases, inc. are discoloration, cracks, Packaging components were Extra Strength Daily Cleaner itself was observed for any precipitates, turbidity or color change. was observed for any precipitates, turelows of the chemical components were noted.
    1. Sterility
      The sterility of the stored solution was tested using USP The Stermly of the Stored Stated States, Inc., Salt Lake City, UT. All three stored lots tested were sterile.
  1. Preservative Effectiveness with 14 day re-challenge

Preservative effectiveness testing of three stored lots was performed by Northview Pacific Laboratories, Inc. of Hercules, CA. The testing was conducted according to the FDA Guidelines CA. The testing was conceeded ation (510(k)) Guidance Document . For Contact Lens Care Products". The stored lots passed the requirements for this test.

Conclusion:

Shelf-life studies indicated that the Sereine Extra Strength Daily Cleaner is safe and effective for the period of the proposed shelf-life of two years.

{10}------------------------------------------------

9. Summary of Non-clinical tests (cont.)

B. Chemistry

1. Solution Compatibility

The compatibility of Sereine Extra Strength Daily Cleaner with soft (hydrophilic) lenses and hard (PMMA) lenses with soft (figuropining) tofnen ing the "Solution Compatibility was unlined in the "Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products". Ten lenses of Metrotint (enhancing tints of polymacon, 38% water) and ten lenses of Horizon 55 (visibility tints of 55% water) were used to test solution compatibility of hydrophilic lenses. Metrotint lenses are manufactured by Metro Optics, Austin, TX. Horizon 55 lenses are by Medio Opace, Westcon Contact Lenses, Grand Junction, CO. Ten PMMA lenses were also tested for compatibility with Sereine Extra Strength Daily Cleaner. These were blue lenses manufactured by Lens Dynamics, Lakewood, CO. The lens blanks were manufactured by Optikem Polymers of Englewood, CO.

The optical parameters were recorded before the test. Lenses were inspected for any physical defects. For hydrophilic lenses, each lens was flexed and then cleaned with Sereine Extra Strength Daily Cleaner as per product directions. After rinsing with sterile saline, the lenses were placed in a lens case and disinfected with Opti-One Solution (Alcon Laboratories, Ft Worth, TX) as per label instructions. After disinfection the lenses were removed from the lens case and the cycle repeated. The procedure was repeated for 30 cycles for each lens. After the 30 proces, the lens parameters and physical appearance of the lenses were again recorded.

{11}------------------------------------------------

Summary of Non-clinical tests (cont.) ல்

B. Chemistry (cont.)

1. Solution Compatibility

For PMMA lenses, the lenses were cleaned with Sereine Extra For PMMA lenses, the lenses were cicance in The lenses were
Strength Daily Cleaner as per label directions. The lenses were Strength Daily Cleaner as per iaber as per interes were
rinsed with water and placed in lens cases. The lenses were rinsed with water and placed in left soaking Solution as per disinfected with Sereine wetting or boulenges were removed
label directions. After disinfection, the lenses were removed label directions. And the cycle repeated. After 30 cycles,
from the lens cases and the cycle repeated. After 30 cycles, from the lens cases and the Cycle reposee of the lenses were the parameters and physical appearantee on

No significant changes in optical parameters or physical No significant changes no opine Extra Strength Daily appearance was noted. Serence DAGE Burenge

  1. Cleaning Effectiveness

The Determination of the critical micelle concentration of the The Determination of the circles tension of the solution done according
surfactant and the surface tension of the solution (510(k)) surfactant and the surface tension of also to the bottfication (510(k))
to the procedure contained in "Premarket North Corp. Products" to the procedure contained in "re Lens Care Products", Guidance Document for Someon Donator Donator Book Testing Appendix B. Testing Company, Inc. of Passaic, NJ.

{12}------------------------------------------------

9. SUMMARY OF NON-CLINICAL TESTS (CONT.)

C. Microbiological Tests

  1. Sterility

Sterility of Sereine Extra Strength Daily Cleaner was confirmed using USP methodology. Sterility testing of the product was done by Nelson Laboratories, Salt Lake City, UT. All lots tested were sterile. The manufacturing process used by Optikem International, Inc. to produce Sereine Extra Strength Daily Cleaner was adequate to ensure that the solution was sterile.

  1. Preservative Effectiveness with 14 day rechallenge

Preservative effectiveness testing was conducted according to the procedure outlined in "Pre-market Notification (510(k)) Guidance Document for Contact Lens Care products". The testing was conducted by Northview Pacific Laboratories, Hercules, CA.

Sereine Extra Strength Daily Cleaner passed the test for preservative effectiveness and is safe for use in a multi-dose container.

{13}------------------------------------------------

10. Conclusions

Based on the data outlined on the preceding pages, Sereine Extra
tion of the outlined offered officative for use as a daily clear Based on the data outlined on the preceding pagos, our as a daily cleaner for
Strength Daily Cleaner is safe and effective for use as a daily cleaner for Strength Daily Cleaner is safe un deposits and debris from the removal of accumulated (PMMA) lenses. The expiration date of
soft (hydrophilic) and bard (PMMA) lensessmiste based on data outlined soft (hydrophilic) and hard (PMMA) tenses. "The expropriate based on data outlined on the preceding pages.

{14}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV Z 0 2001

Ms. Sally Cook President, Optikem International, Inc. Wilsa, Incorporated 2172 South Jason Street Denver, CO 80223

Re: K011561

K011501
Trade/Device Name: Sereine Extra Strength Daily Cleaner Regulation Number: 21 CFR 886.5928 Regulation Number: 21 Of St Irophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN; HPX Dated: October 1, 2001 Received: October 2, 2001

Dear Ms. Cook:

We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Section 310(ts) notheration of the indications for the indications for above and we have determined the cence is substainly of commence prior to May 28, 1976, 1976, 1976, 1976, 1976, 1976, 1976, use stated in the enclosure) to device interestments, or to devices that have been the enactment date of the Medical Device runes of Food, Drug, and Cosmetic Act
reclassified in accordance with the provisions of the Federal controls provisions of reclassified in accordance with the provisions on the general controls provisions of the (Act). You may, therefore, market the actives seauirements for annual registration,
Act. The general controls provisions of the Act include requirements mishranding Act. The general controls provisious of the Act include requiremisms against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III
n major If your device is classified (see above) into essues additional controls. Existing major (Premarket Approval), it may be subject to such and in the Code of Federal Regulations, Title 21,
regulations affecting your device can be found in the Code of Federal parth regulations affecting your device can or in the management on assumes compliance with the Parts 800 to 895. A substantially cquivalents, as set forth in the Quality System
Current Good Manufacturing Practice requirements, as set forth in the Rest, 800 and that, Current Good Manufacturing Fractic requiremently is over CFR Part 820) and that, through
Regulation (QS) for Medical Devices: General regulation (21 CFR Parily such Regulation (QS) for Medical Devices. Ocher a voginistration (FDA) will verify such periodic QS inspections, the Food and Drag regulation may result in regulatory action. In the Federal assumptions. Tandre to comply with the Other toge oncerning your device in the Federal addition, FDA may publish lurther announcements on notification submission does not affect
Register. Please note: this response to your premarket not for devices under Register. Flease note: this response to your f 542 of the Act for devices under any obligation you might have under sociologio sportions, or other Federal laws or regulations. I

{15}------------------------------------------------

Page 2 - Ms. Sally Cook

This letter will allow you to begin marketing your device as described in your Section 510(k)
. The subscription for a Tim FDA freding of sybstantial equivalence of your dev This letter will allow you to begin marketing your device of your device to a would and the research I its retect will and of the FDA finding of substantial equivalence of your device to trigure of the first of the first of the for proceed to the market.

procedure specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your device on our labeling regulation on earther Office of
additionally 21 CFR Part 809.10 for in vited diagnosic devices, please contact t If you attended in the most of on the diagnostic devices); pressoons on the promotion and advertising of
Compliance at (301) 594-4613. Additionally, for questions on the pro Compliance at (301) 594-4613. Additionally, for questions on and the promotes and Ocuped on the Office of Compliation of Compliance an (101). "C/C/R Part 807.97).
your device, please contact the Office of Compliation "(21CFR Part 807.97).
regulation entitl regulation entitled, "Misbranding by responsibilities may be obtained from the Division
Other general information on your responsibilities under the Act may be obtained from rogeneral information on your responsibilities under the Act may 10 of the many of the manager (800)
of Small Manufacturers, International and Consumer Assistance at its tol of Small Manufacturers, International its Internet address .
638-2041 or (301) 443-6597 or at its Internain html 638-2041 of (501) 175-604) 175-607 14-15-2041 11-1

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{16}------------------------------------------------

K011561

NOV 2 0 2001

INDICATIONS FOR USE STATEMENT

Sereine Extra Strength Daily Cleaner is indicated for use as a daily
our and other that the first be acits and other debris which accums Sereine Extra Strength Daily Cleaner is Indicator for which accumulate
cleaner for removal of film, deposits and other debris which accumulate
s and of the county in and DJ A cleaner for removal of film, deposits and other were are and

Sereine Extra Strength Dailly Cleaner may be used with either thermal
s and the success to belief of ation systems . Follow the recommends Sereine Extra Strength Dailly Cleaner may be used with exammendation
(heat) or chemical (cold) disinfection systems. Follow the recommendation
Services of chemical (cold) di (heat) or chemical (cold) disinicollor systems
of your eye care practitioner when disinfecting your lenses.

Eis O Ph.D.

(Division Sign-Off) Division of Ophthalmic Devices Koll561 510(k) Number_

Over-the-Counter Use

§ 886.5928 Soft (hydrophilic) contact lens care products.

(a)
Identification. A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”