    Trade/Device Name:** FemVue® Controlled Saline-Air Device (FSA-300)
    Regulation Number: 21 CFR§ 884.4530
    Regulation Name:** Obstetric-gynecologic specialized manual instrument

    Regulation Number: 21 CFR 884.4530
    Device | |
    |---|---|---|---|
    | Submission Number | K253403 | K241693 | N/A |
    | Regulation | 21 CFR 884.4530
    | 21 CFR 884.4530 | Same |
    | Device Classification | Class II | Class II | Same |
    | Product Code | LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemVue® Controlled Saline-Air Device is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    The FemVue® Controlled Saline-Air Device is a sterile dual-barrel contrast media syringe that is connected to an intrauterine catheter to instill saline-air contrast medium as part of a sono-hysterosalpingogram (Sono-HSG) procedure. Ultrasound of the fallopian tubes can be performed with or without assessment of the uterine cavity.

    The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via ethylene oxide (EO) sterilization and is intended for single-use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252260

    Validate with FDA (Live)

    Date Cleared
    2025-11-26

    (128 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    TAIWAN

    Re: K252260
    Trade/Device Name: RELIEEV HSG Catheter (HSG7FA1)
    Regulation Number: 21 CFR 884.4530
    Name:** RELIEEV HSG Catheter (HSG7FA1)
    Common Name: HSG Catheter
    Regulation Number: 21 CFR 884.4530
    Regulation | Class II, 21 CFR | Class II, 21 CFR | Same |

    Page 7

    K252260
    Page 3 of 8

    | | 884.4530
    | 884.4530 | |
    |---|---|---|---|
    | Product code | LKF | LKF | Same |
    | Indications for Use | The RELIEEV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RELIEEV HSG Catheter (HSG7FA1) is for the delivery of contrast media or saline into the uterine cavity during hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes.

    Device Description

    The RELIEEV HSG Catheter (HSG7FA1) is a 7 French (Fr) dual lumen polyurethane catheter with a working length of 30 cm. RELIEEV HSG Catheter (HSG7FA1) is available with a closed-end tip with two side ports. The catheter consists of two lumens: an open lumen used to inflate the balloon and an open lumen used to inject diagnostic media into the uterus. The silicone balloon is designed with a balloon volume of 3.5 mL. When the balloon is inflated inside the uterus, the subject device seals the cervix. The catheter features a proximal fitting of a double-lumen polyurethane hub with a polycarbonate stopcock attached. Both the hub and stopcock accept a standard Luer lock. A stopcock is provided at the proximal end of the device to allow inflation of the balloon with a syringe. The center lumen is open through the device to the distal end, and the device has a luer connector at the proximal end for injecting fluids. An introducing sheath is placed over the tube as a guide and into the endocervix to the level of the internal os (opening of the cervix) with no dilatation required; therefore, allowing the catheter to pass easily through the cervix and into the uterine cavity.

    The RELIEEV HSG Catheter (HSG7FA1) consists of a balloon catheter and a 5 mL syringe. The subject device is supplied sterile and is single-use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242002

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-22

    (136 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Georgia 30024

    Re: K242002

    Trade/Device Name: FemVue MINI Saline-Air Device Regulation Number: 21 CFR 884.4530
    instrument |
    | Regulation Number: | 21 CFR 884.4530
    | N/A |
    | RegulationNumber | 884.4530
    | 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemVue MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    Fem Vue® MINI is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterilization and is intended for single-use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies mentioned for the FemVue MINI Saline-Air Device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it states that "all predetermined acceptance criteria were met" for various tests. Therefore, the table below is constructed by inferring the acceptance criteria from the type of test conducted.

    Acceptance Criterion (Inferred from Test)Reported Device Performance
    Sterility (achieves sterility after Ethylene Oxide sterilization)Met (per ISO 11135:2014, AAMI TIR 28:2016, ISO 10993-7: 2008)
    Package Integrity (maintains sterility barrier)Met (per Visual inspection, ASTM F2096-11, ASTM F88/ F88M-23, ASTM F1886/F1886-16)
    Withstand Transportation (maintains integrity after transport simulation)Met (per ASTM D4169-22)
    Biocompatibility (non-cytotoxic, non-sensitizing, non-irritating)Met (non-cytotoxic, non-sensitizing, and non-irritating per ISO 10993-5: 2009, ISO 10993-10: 2021)
    Visual Inspection (meets specified visual appearance)Met (before and after accelerated aging)
    Functional/Cycle Testing (operates as intended through its lifecycle)Met (before and after accelerated aging)
    Saline-Air Quantification (delivers consistent alternating pattern of saline and air)Met (before and after accelerated aging)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the individual bench performance tests. It only states that these tests were conducted. It also does not explicitly state the data provenance (e.g., country of origin) or if the studies were retrospective or prospective, though bench testing is inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are primarily physical, chemical, and functional bench tests, rather than clinical studies requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of the bench tests, it's unlikely a formal adjudication method (like 2+1, 3+1) would be applicable. The tests likely have objective pass/fail criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance data for a device, not a diagnostic algorithm that would typically require such a study to assess human reader improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study (algorithm only) was not done. This device is a manual instrument, not an AI or software algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical performance studies described, the "ground truth" refers to established engineering standards, material specifications, and functional requirements for a medical device. This is determined by the specific ASTM and ISO standards cited (e.g., ISO 11135 for sterility, ISO 10993 for biocompatibility, ASTM F1980-16 for accelerated aging). The concept of expert consensus, pathology, or outcomes data as ground truth is not applicable to these types of engineering and functional tests.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. This device is a hardware product, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240364

    Validate with FDA (Live)

    Date Cleared
    2024-10-16

    (253 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    TAIWAN

    Re: K240364

    Trade/Device Name: "RELIEEV" Uterine Manipulator Injector Regulation Number: 21 CFR 884.4530
    | Regulation Name: | Obstetric-gynecologic specialized manual instrument |
    | Regulation Number: | 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use of the RELIEEV Uterine Manipulator Injector is indicated in Diagnostic Laparoscopy, Minilaparotomy, Fertility Exams, and Salpingoplasty procedures where manipulation of the uterus is required.

    Device Description

    The "RELIEEV" Uterine Manipulator Injector is intended to be used during medical procedures where manipulation of the uterus or injection of fluids into the uterine lumen are needed. This device is a sterile, single use product consisting of a plastic (PVC) insertion tube that includes two lumens, one for inflation of a 10 mL intrauterine balloon (cuff) at the distal end of the insertion tube via the inflation valve using the provided 10 mL syringe and the other for injection of fluid through a distal end-port using a user-provided syringe via the Leur fitting. The device also includes a movable/removable rigid handle that allows alteration of the device insertion depth and includes a cervical stop, laser etched depth markings to aid in setting the insertion depth, and a pilot balloon to assess maintenance of cuff inflation. The device is curved to facilitate forward uterine manipulation. The device has an overall length of 36 cm and an outer diameter of the catheter component of 5.0 mm. When inflated with 10 mL of air, the balloon has a diameter of 23.0-24.9 mm.

    AI/ML Overview

    This document is an FDA 510(k) summary for the "RELIEEV" Uterine Manipulator Injector. It describes the device and claims substantial equivalence to a predicate device, the Panpac Uterine Manipulator Injector, Model Umi 4.5 (K092980).

    Based on the provided text, the device in question is a medical instrument, not an AI/software device. Therefore, the questions related to AI/software performance, ground truth, human readers, and training/test sets are not applicable to this document.

    The acceptance criteria and study proving the device meets them are related to the physical and biological safety and performance of the Uterine Manipulator Injector, rather than an AI algorithm.

    Here's the information extracted from the document, tailored to the type of device described:


    Acceptance Criteria and Device Performance for "RELIEEV" Uterine Manipulator Injector

    This device is a physical medical instrument, and its performance evaluation focuses on its physical characteristics, sterility, biocompatibility, and mechanical integrity, comparing it to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details various tests performed to ensure the device meets safety and performance standards for a physical medical device. The "acceptance criteria" are implied by the successful completion of these tests per the referenced standards and the statement that "all predetermined acceptance criteria were met."

    Area of PerformanceSpecific Test/EvaluationAcceptance Criteria (Implied by standard and successful outcome)Reported Device Performance
    SterilizationEO sterilization validationConformance to ISO 11135-1:2014 & ISO 10993-7:2008 for residual EtOMet (testing performed)
    PackagingVisual inspectionConformance to ASTM F1886/F1886M-16Met (testing performed)
    Seal Strength testingConformance to ASTM F88/F88M-23Met (testing performed)
    Dye Penetration testConformance to ASTM F1929-15Met (testing performed)
    TransportationTransportation Simulation testingConformance to ASTM D4169-23Met (testing performed)
    BiocompatibilityCytotoxicityNon-cytotoxic per ISO 10993-5:2009Non-cytotoxic
    SensitizationNon-sensitizing per ISO 10993-10:2021Non-sensitizing
    IrritationNon-irritating per ISO 10993-23:2021Non-irritating
    Acute Systemic ToxicityNon-systemically toxic per ISO 10993-11:2017Non-systemically toxic
    Material Mediated PyrogenicityNon-pyrogenic per USP <151>Non-pyrogenic
    Bench Performance (after accelerated aging)Dimensional specificationsMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Bend test (Force to tip deflection)Met pre-determined criteriaAll "predetermined acceptance criteria were met"
    Cuff Burst TestMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Tensile strength testing (all joints/connections)Met pre-determined criteriaAll "predetermined acceptance criteria were met"
    Repeated cuff inflationMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Prolonged inflationMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Insertion depth marker integrity testingMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Handle locking mechanism testing (force to move from set position)Met pre-determined criteriaAll "predetermined acceptance criteria were met"

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for the various tests. It refers to established testing standards (e.g., ISO, ASTM, USP), which would specify appropriate sample sizes for each test type. The data provenance is from bench testing and laboratory evaluations conducted by the manufacturer (LI Medical Corporation LTD.) to support regulatory submission. This is a prospective assessment of newly manufactured devices under controlled conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This concept is not applicable to the assessment of a physical medical device like a uterine manipulator. "Ground truth" in this context refers to the defined acceptable parameters for physical and biological performance, which are established by international standards and regulatory guidelines, not by individual experts evaluating images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable for a physical medical device. Performance is judged against specified engineering, material, and biological safety standards, not through adjudication of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable, as this is not an AI/software device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is not an AI/software device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on established industry standards and regulatory requirements for medical device safety and efficacy. This includes:

    • Physical and mechanical properties (e.g., dimensions, tensile strength, inflation capabilities).
    • Sterility assurance (e.g., EtO residuals).
    • Biocompatibility (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).
    • Packaging integrity.

    These are quantitative metrics or pass/fail criteria derived from validated test methods.

    8. The Sample Size for the Training Set

    Not applicable, as this is not an AI/software device. There is no "training set" in the context of a physical medical device's performance validation; rather, there are samples tested according to predefined standards.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241693

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-06

    (86 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K241693

    Trade/Device Name: FemChec Controlled Saline-Air Device (FCD-250) Regulation Number: 21 CFR 884.4530
    Name: | Obstetric-gynecologic specialized manual instrument |
    | Regulation Number: | 21 CFR 884.4530
    ----------------------------------------------------|
    | Classification | 21 CFR 884.4530
    | 21 CFR 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    FemChec® is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via EO sterilization and is intended for single-use.

    AI/ML Overview

    The provided text is a 510(k) summary for the FemChec Controlled Saline-Air Device (FCD-250), which is a medical device and not an AI or software device. Therefore, the questions related to AI/algorithm performance, training sets, test sets, ground truth establishment by experts, adjudication methods, and MRMC studies are not applicable to this document. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data for a physical medical device.

    However, I can extract the acceptance criteria and a summary of the studies performed from the provided text for the FemChec device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests/CriteriaReported Device Performance
    SterilizationEthylene Oxide Sterilization Validation per ISO 11135-1:2014, AAMI TIR 28:2016, ISO 10993-7:2008Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    PackagingVisual inspection, Bubble Leak test per ASTM F2096-11, Seal Strength testing per ASTM F88/F88M-15Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    TransportationTransportation Simulation testing per ASTM D4169-14Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    BiocompatibilityCytotoxicity per ISO 10993-5:2009, Sensitization ISO 10993-10:2021, Irritation per ISO 10993-10:2021Tested to be non-cytotoxic, non-sensitizing, and non-irritating.
    Bench PerformanceVisual Inspection, Functional/Cycle Testing, Saline-Air Quantification (before and after accelerated aging to 12 months real-time aging per ASTM F1980-16)All predetermined acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable as the studies conducted are non-clinical bench performance and material compatibility studies for a physical device, not a performance study involving a test set of data points or images.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable. Ground truth establishment by experts is relevant for AI/software devices requiring interpretation of data (e.g., medical images), which is not the case for this physical medical device.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods are relevant for studies where multiple readers/experts assess cases, which is not applicable to the non-clinical studies for this physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance studies for a physical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical medical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    This information is not applicable in the context of AI/software performance. For the described bench tests, the "ground truth" would be established by the specifications and standards (e.g., flow rates, material properties, sterility levels) defined in the test protocols themselves, which are based on recognized international and ASTM standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230849

    Validate with FDA (Live)

    Device Name
    ELLAVI UBT
    Date Cleared
    2023-11-13

    (230 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ave Crofton, Maryland 21114

    Re: K230849

    Trade/Device Name: ELLAVI UBT Regulation Number: 21 CFR 884.4530
    Name: | Obstetric-Gynecologic Specialized Manual Instrument |
    | Regulation Number: | 21 CFR § 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ELLAVI UBT device is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.

    Device Description

    The ELLAVI UBT is an 18 cm long intrauterine balloon tamponade device primarily comprised of thermoplastic elastomer. The balloon and internal tubing are made of thermoplastic elastomer, while the exterior tubing is made of polyvinyl chloride. A supply bag is the filled, connected, and hung at a pre-determined height based on the patient's systolic blood pressure. The device consists of a balloon that is inserted into the uterus transvaginally into the postpartum uterus after delivery and removal of the placenta. A valve is then opened, and this allows the fluid to inflate the balloon and apply pressure to achieve a hemostatic effect.

    AI/ML Overview

    The ELLAVI UBT device, intended for temporary control or reduction of postpartum uterine bleeding, underwent a series of non-clinical performance tests to demonstrate substantial equivalence to its predicate device, the Bakri Postpartum Balloon. The provided documentation primarily focuses on mechanical, biocompatibility, sterilization, and shelf-life testing. No clinical study data is provided in the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Mechanical Testing
    Visual AssessmentNo visual defectsVerified no visual defects
    Dimensional VerificationVerification of tube and balloon dimensionsVerified tube and balloon dimensions
    Stretch PercentageVerify the stretch percentage of the balloon when filled with 750 mlVerified stretch percentage
    Tube KinkingVerify that the tube is unable to kink when handled as per normal useVerified no kinking
    External Wall PressureThe external wall of the balloon must be able to exert 60 mm Hg to 120 mm Hg on the uterine wallVerified pressure exertion within range
    Pressure Maintenance (Balloon)Maintain a pressure of 12 kPa for 20 seconds (p-test)Verified pressure maintenance
    Pressure Maintenance (Supply Bag)Maintain a pressure of 12 kPa for 8 seconds (p-test)Verified pressure maintenance
    Balloon Burst VolumeVerify the burst volume of the balloonVerified burst volume
    Balloon Leak/RuptureNo leakage and/or rupture when subjected to 180 mm Hg for 24 hoursVerified no leakage/rupture
    Balloon Drop StrengthWithstand a drop from 1 m when filled to maximum fill levelVerified drop strength
    Balloon Fill TimeVerify the time it takes to fill the balloon to the normal fill level for different supply bag heightsVerified fill time
    Balloon Fill with Partial BlockageAble to fill to the normal fill level while inner tube is pressed against balloon wallVerified fill with blockage
    O-Ring Seal ForceMaintain a connection between the tube and the balloonVerified connection maintenance
    Force to Separate Balloon from TubingVerify the force required to break the O-ring seal in emergency while balloon is filledVerified force required
    Balloon Leak in Absence of O-RingNo leaks when filled to maximum operating fill level with no O-ringVerified no leaks
    Tube/Tap Removal ForceVerify the force required to remove the tube from the tapVerified force required
    Supply Bag/Tap Removal ForceVerify the force required to pop the tube off the tap connection for a closed tapVerified force required
    Balloon Draining TimeVerify the time it takes the balloon to drain back to the supply bag for different supply bag heights and tube orientationsVerified draining time
    Balloon Fill with a Kinked TubeCan fill and/or drain with a kinked tubeVerified fill/drain with kinked tube
    Biocompatibility Testing
    Cytotoxicity (ISO 10993-5:2009)Not cytotoxicNot cytotoxic
    Maximization Sensitization (ISO 10993-10:2010)Not sensitizingNon-sensitizing
    Intracutaneous Irritation (ISO 10993-10:2010)Not irritatingNon-irritating
    Acute Systemic Toxicity (ISO 10993-11:2017)Not acutely systemically toxicNot acute systemically toxic
    Material Mediated Pyrogenicity (ISO 10993-11:2017)Does not elicit a material-mediated pyrogenic responseDoes not elicit a material-mediated pyrogenic response
    Sterilization & Shelf-Life
    SterilizationSAL of 10⁻⁶Achieved SAL 10⁻⁶ (Ethylene Oxide)
    Shelf-Life3-year shelf-lifeEstablished based on accelerated aging (ASTM F1980-21)
    Simulated Shipping DistributionIn accordance with ASTM D4196-16Conducted in accordance with ASTM D4196-16
    Package IntegrityIn accordance with ASTM F1980-16, F88/F88M-15, and ASTM F1929-15Conducted in accordance with specified ASTM standards
    Endotoxin TestingLess than <20 EU/deviceDemonstrated less than <20 EU/device (USP <85>)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document details non-clinical performance data (mechanical, biocompatibility, sterilization, and shelf-life testing). It does not describe a "test set" in the context of clinical or image-based studies with human subjects. Therefore, information about sample size, country of origin, or retrospective/prospective nature of such a test set is not applicable from this document. The tests performed are laboratory-based evaluations of the device's physical and material properties.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The document describes non-clinical engineering and material science tests. "Ground truth" in the context of expert consensus for clinical diagnosis is not relevant to these types of tests.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "test set" in the context of clinical or diagnostic performance that would require an adjudication method. The reported performance is based on direct measurements and observations from laboratory testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or conducted, as this documentation pertains to non-clinical testing for substantial equivalence of a medical device.

    6. Standalone Performance Study

    The documented studies are standalone in the sense that they evaluate the device itself (its mechanical properties, biocompatibility, etc.) without human interaction as part of the performance evaluation. However, this is not a "standalone performance" in the context of algorithm-only performance for a diagnostic device, as the ELLAVI UBT is a physical medical instrument.

    7. Type of Ground Truth Used

    The "ground truth" for the non-clinical tests described is based on engineering specifications, material science standards, and established laboratory testing protocols. For instance:

    • Mechanical Testing: Physical measurements against design specifications, established parameters (e.g., pressure ranges, fill volumes, time limits).
    • Biocompatibility Testing: Compliance with ISO 10993 series standards, where reactivity or toxicity is evaluated against established biological models.
    • Sterilization and Shelf-Life: Compliance with ISO 11135 and ASTM standards for sterility assurance levels, aging profiles, and package integrity.

    8. Sample Size for the Training Set

    Not applicable. The ELLAVI UBT is a physical medical device. It does not involve machine learning algorithms or a "training set" in the context of artificial intelligence.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230730

    Validate with FDA (Live)

    Device Name
    Apyx
    Manufacturer
    Date Cleared
    2023-07-25

    (131 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NW Washington, DC 20004-1109

    Re: K230730

    Trade/Device Name: Apyx Regulation Number: 21 CFR§ 884.4530
    of Device: Apyx

    Common or Usual Name: Fixation device for the pelvic floor

    Classification Number: 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apyx is intended for attaching sutures to ligaments of the pelvic floor.

    Device Description

    The Apyx device is indicated for anchoring sutures to ligaments of the pelvic floor. The device consists of an implantable nitinol anchor with 4 prongs configured with either non-absorbable or resorbable suture. The anchor-suture assembly is contained within a cartridge, wherein the anchor / suture assembly is deployed to the target site from the cartridge with an applicator. An optional use retriever may be used to remove the anchor in the event of sub-optimal placement of an Anchor during the index procedure. The Apyx also includes an optional securement element which may be used as a suture retention device to distribute suture tension over a larger tissue area and aid in wound healing.

    The Apyx device is a single use device which is supplied sterile.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study proving the device meets those criteria in the typical format of a clinical or performance study with specific metrics like sensitivity, specificity, or AUC, as these would be associated with diagnostic or AI-powered devices.

    However, based on the information provided, here's a breakdown of what can be extracted regarding performance and testing:

    1. A table of acceptance criteria and the reported device performance

    The document mentions various tests conducted, generally implying that the device met the requirements of these tests to demonstrate safety and effectiveness. Since specific numerical acceptance criteria (e.g., "pull-out strength > X N") and the exact reported performance values are not explicitly stated, I will infer the acceptance as "Met requirements" based on the conclusion of substantial equivalence.

    Acceptance Criterion (Implied)Reported Device Performance
    BiocompatibilityMet requirements (ISO 10993 compliant)
    Chemical Characterization and Toxicological Risk AssessmentMet requirements (ISO 10993-18 compliant)
    Corrosion ResistanceMet requirements
    Packaging ValidationMet requirements
    Shelf-Life ValidationMet requirements
    Securement Element Pull-OutMet requirements
    Cadaver Model Testing (Delivery)All anchors accurately delivered into target ligament; no complications or unanticipated risks observed. Successfully performed.
    Cadaver Model Testing (Retrieval)Successfully performed. Successfully removed.
    Cadaver Model Testing (Securement)All anchors secured using the securement element and properly fixed at the desired position with ease.

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: The "Cadaver Testing" section does not specify the number of cadavers or the number of anchors tested. It uses general phrasing like "users," "all anchor delivery and retrieval procedures," and "All anchors."
    • Data provenance: Not explicitly stated, but cadaver testing implies a controlled laboratory environment. The country of origin of the data is not mentioned. It is prospective testing carried out for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided. The cadaver testing describes the actions of "users" (implying medical professionals) who delivered and retrieved anchors, but doesn't mention independent expert assessment or ground truth establishment in a formalized way for outcomes like accuracy or safety observed during the testing. The "ground truth" here is the observed outcome of the procedure itself (e.g., successful placement, no damage).

    4. Adjudication method for the test set

    • Not applicable/Not described. The cadaver testing describes qualitative observations of successful delivery, retrieval, and proper fixation. There's no mention of an adjudication process for resolving disagreements between multiple observers, as might be seen in diagnostic studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a surgical instrument (fixation device), not an imaging or AI-assisted diagnostic tool. Therefore, an MRMC study and AI-related effectiveness are not relevant or discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This device is a manual instrument for surgical use, not an algorithm.

    7. The type of ground truth used

    • For the cadaver testing, the ground truth was the direct observation of the physical outcome and success of the procedure (e.g., "anchors were accurately delivered," "no damage observed," "properly fixed"). This is akin to direct experimental observation and physical verification rather than consensus, pathology, or outcomes data in a clinical trial.

    8. The sample size for the training set

    • Not applicable. This device is a physical medical instrument, not a machine learning algorithm, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for a physical medical instrument.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221356

    Validate with FDA (Live)

    Device Name
    Wee Bell
    Manufacturer
    Date Cleared
    2023-02-02

    (268 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    GA 30339

    Re: K221356

    Trade/Device Name: wee BELL Circumcision Device Regulation Number: 21 CFR§ 884.4530
    -----|-----------------------------------------------------|
    | Classification Regulation: | 21 CFR §884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The wee BELL is a newborn and infant circumcision device. It is used on males with intact foreskin for the purpose of performing a circumcision (removal of the foreskin/prepuce). It consists of two components: wee Bell Circumcision device and ligature. The ligature is intended for use with the wee Bell Circumcision device only.

    Device Description

    The wee BELL is a newborn and infant circumcision device. It is used on males with intact foreskin for the purpose of performing a circumcision (removal of the foreskin/prepuce). It consists of two components: wee Bell Circumcision device and ligature is intended for use with the wee Bell Circumcision device only. The wee Bell is designed to assist in the performance of a circumferential crushing of the foreskin (prepuce) by tightening the ligature around the bell portion of the device. The remaining foreskin is then removed.

    The wee Bell is comprised of two components: a transparent plastic bell and a ligature. It is available in six (6) sizes to accommodate anatomy variation in the size of newborn and infant glands. The device is EtO sterilized and is a single-use device.

    AI/ML Overview

    The provided text is a 510(k) summary for the "wee BELL Circumcision Device" and describes its characteristics, intended use, and comparison to a predicate device. However, it does not contain information about acceptance criteria or a study proving performance against such criteria for a medical device that would involve a test set, ground truth, expert opinions, or MRMC studies.

    The document states:

    • "Nonclinical functional performance testing was performed on the subject device and included functional testing (visual inspection, tensile testing and smoothness testing), packaging testing per ASTM F88-21, ASTM D4169-22 and ISO 11607, sterilization per ISO 11135:2014 and biocompatibility according to ISO 10993-1 for a surface device in prolonged contact with intact skin. Additionally, material mediated pyrogenicity was conducted."

    This indicates engineering and material-based tests were performed to ensure the device's physical properties, sterility, and biocompatibility, but not clinical performance benchmarks in terms of diagnostic accuracy or effectiveness measured against a ground truth developed by experts. The device is a "manual instrument," not an AI/ML-driven diagnostic device, which is typically what the questions (sample size for test set, ground truth, experts, MRMC, etc.) pertain to.

    Therefore, I cannot provide the requested table and information based on the provided text, as the type of evaluation described is foundational for a medical device's physical integrity and safety, rather than a clinical performance study with statistical endpoints related to diagnostic accuracy or efficacy improvement with human-in-the-loop.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223866

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-25

    (33 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Switzerland

    Re: K223866

    Trade/Device Name: Carevix™ Suction Cervical Stabilizer Regulation Number: 21 CFR$ 884.4530
    |
    | Regulation Name &Number: | 21 CFR 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carevix™ Suction Cervical Stabilizer is indicated to snare, grasp, hold and manipulate cervical tissue.

    Device Description

    Carevix™ Suction Cervical Stabilizer is a sterile, single-use, two-piece device consisting of three main parts: a reloadable vacuum reserve embedded inside the main body, including a piston to generate the vacuum, an activation/deactivation push-button, and an anatomic-shaped suction head to put in contact with the cervix. The suction head includes a rod that connects the main body.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Carevix™ Suction Cervical Stabilizer. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a full study report for a novel medical device, especially an AI/ML-driven one.

    Therefore, many of the requested details, particularly those pertaining to AI/ML device performance (sample size for test set, data provenance, number of experts for ground truth, adjudication, MRMC studies, standalone performance, training set details), are not present in the provided text, as this is a traditional medical device submission.

    Here's what can be extracted based on the provided text:

    Acceptance Criteria and Device Performance:

    The document states that the device was evaluated against its design and user specifications, and that "All the results were in accordance with the test acceptance criteria." However, it does not provide a table of specific quantitative acceptance criteria or the reported device performance metrics for different parameters (e.g., specific force applied, vacuum pressure achieved, or a pass/fail rate for a clinical outcome).

    Instead, it lists the types of tests performed and their general outcomes:

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritating"demonstrated the subject device is non-cytotoxic non-sensitizing, and non-irritating."
    SterilizationMeets ANSI/AAMI/ISO 11137-2:2013 standards"Sterilization validation was performed... and was conducted following testing protocols used for the predicate device." (Implied successful validation)
    Performance & SafetyMeets design and user specifications, no new safety/effectiveness questions raised compared to predicate."All tests were conducted according to the protocols used to validate the predicate device ASPIVIX v1.1. All the results were in accordance with the test acceptance criteria."

    Missing Information (as per the provided text):

    • Sample size used for the test set and the data provenance: Not mentioned. The testing appears to be primarily bench testing (biocompatibility, sterilization, performance) rather than clinical studies with patient data.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/not mentioned. Ground truth as typically understood for AI/ML devices (e.g., disease presence) is not relevant to this device's testing.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/not mentioned.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No. This is not an AI/ML device nor a diagnostic imaging device that would typically involve human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, as this is not an AI/ML algorithm.
    • The type of ground truth used: For biocompatibility, the ground truth is established by standardized lab tests (e.g., cell viability, skin irritation). For performance, the ground truth is the device's design specifications. No "expert consensus" or "pathology" ground truth is mentioned.
    • The sample size for the training set: Not applicable, as this is not an AI/ML device.
    • How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

    Summary of what the document focuses on instead:

    The document outlines a 510(k) submission for a modified medical device. The primary goal is to demonstrate "substantial equivalence" to a previously cleared predicate device (ASPIVIX v1.1 Cervical Suction Tenaculum, K203820). This is achieved by:

    • Comparing Indications for Use: Stating they are the same ("to snare, grasp, hold and manipulate cervical tissue").
    • Comparing Technological Characteristics: Identifying both similarities and differences (e.g., changes in rod length, inner diameter, trigger mechanism, vacuum generation, and two-piece vs. one-piece design).
    • Providing Safety and Performance Data for the Modifications: This includes:
      • Biocompatibility testing: Performed in accordance with ISO 10993-1, 10993-5, 10993-10, and ISO 14971:2019. Tests confirmed non-cytotoxic, non-sensitizing, and non-irritating results.
      • Sterilization validation: Performed via e-beam in accordance with ANSI/AAMI/ISO 11137-2:2013.
      • Performance and Safety testing: Conducted according to protocols used for the predicate device, with all results meeting acceptance criteria. The document states that newly identified risks (related to assembly) do not raise new questions of safety or effectiveness because they are identifiable before patient use.
    • Conclusion: The testing results support the determination of substantial equivalence, meaning the modified device is considered as safe and effective as the predicate device.

    In essence, this is a clearance for a physical medical instrument based on engineering and biological testing, not an AI/ML software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222798

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-16

    (91 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    MA 01730

    Re: K222798

    Trade/Device Name: Rejoni Intrauterine Catheter Regulation Number: 21 CFR§ 884.4530
    |
    | Regulation Number: | 21 CFR 884.4530

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rejoni Intrauterine Catheter is for the delivery of contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes.

    When used for HSG. the Rejoni Intrauterine Catheter can be used for evaluation of tubal patency.

    The Rejoni Intrauterine Catheter is used to access the uterine cavity for the delivery of saline for Sonohysterography (SHG).

    Device Description

    The Rejoni Intrauterine Catheter is a sterile, single lumen uterine catheter that facilitates access to the uterus to deliver fluids during Hysterosalpingography (HSG) or Sonohysterography (SHG). The Rejoni Intrauterine Catheter consists of an inner shaft and an additional support sheath over the proximal portion of the catheter shaft to provide additional stiffness. A repositionable stopper ("acom") is connected to the support sheath, which is placed on the catheter shaft, and can be repositioned from 0 cm up to 10 cm from the distal tip of the catheter. The stopper maintains placement of the catheter against the external cervical os during the procedure, and a luer lock-style adapter on the proximal end is available for compatibility with syringes. The distal end contains a straight through hole for delivery of fluids during the procedure.

    AI/ML Overview

    The Rejoni Intrauterine Catheter is a sterile, single-lumen uterine catheter intended for delivering contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes. It can also be used for evaluation of tubal patency during HSG.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategorySpecific Test / CriterionAcceptance CriteriaReported Device Performance
    Mechanical TestingPerformance - MechanicalIn accordance with device performance specificationsPass
    Performance - FunctionalIn accordance with device performance specificationsPass
    Dimensional VerificationMet design specificationsPass
    Visual AssessmentMet design specificationsPass
    Tensile Testing (luer to inner shaft; acorn to outer shaft)Met design specificationsPass
    Kink TestingMet design specificationsPass
    Leak TestingMet design specificationsPass
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Non-cytotoxicPass
    Sensitization (ISO 10993-10:2021)Non-sensitizingPass
    Intracutaneous Reactivity (ISO 10993-10:2021)Non-irritatingPass
    SterilitySterility Assurance Level (SAL)10$^{-6}$10$^{-6}$ (validated)
    Sterilization Process ValidationConforms to ISO 11137-1, -2, -3Validated
    Shelf LifeDevice performance over shelf lifeMaintains performance as per mechanical testingPass
    Sterile barrier integrity over shelf lifeMaintains sterile barrierPass
    Package integrity after accelerated aging (ASTM 1980-16)No gross leaks, adequate seal strength, visual integrityPass
    Package integrity after simulated shipping (ISTA-3A:2018, ASTM D4169-16)No gross leaks, adequate seal strength, visual integrityPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test (e.g., how many devices were used for tensile testing, biocompatibility, etc.). However, it indicates these were pre-clinical tests (lab-based) rather than clinical studies with patient data. Therefore, data provenance such as country of origin of data or retrospective/prospective characteristics is not applicable in this context.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The described studies are pre-clinical performance and safety tests (mechanical, biocompatibility, sterility, shelf life) rather than studies that require expert-established ground truth on diagnostic accuracy, for example. The "ground truth" for these tests is defined by established industry standards (ISO, ASTM) and the device's own design specifications.

    4. Adjudication Method:

    Not applicable. As noted above, these are pre-clinical tests following defined protocols and standards, not studies requiring expert adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The device is a physical catheter, not an AI-assisted diagnostic tool requiring evaluation of human reader improvement.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    No, a standalone (algorithm only) study was not done. This device is a manual instrument and does not involve an algorithm.

    7. Type of Ground Truth Used:

    The ground truth for these tests is implicitly or explicitly defined by:

    • Design Specifications: For mechanical and functional performance, the device's own engineering and design requirements serve as the ground truth.
    • International Standards (ISO, ASTM): For biocompatibility, sterility, and packaging/shelf life, the ground truth is established by the specified ISO and ASTM standards and their acceptance criteria (e.g., non-cytotoxic, SAL of 10$^{-6}$).

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.