Search Results
Found 170 results
510(k) Data Aggregation
(29 days)
Trade/Device Name: Optima Coil System (OptiOne Coil System)
Regulation Number: 21 CFR 882.5950
Review Panel:** Neurology; Cardiovascular
Product Code: HCG, KRD
Regulation Number: 21 CFR 882.5950
** | Neurology; Cardiovascular |
| Product Code: | HCG, KRD |
| Regulation Number: | 21 CFR 882.5950
The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.
N/A
Ask a specific question about this device
(107 days)
94538-2588
Re: K252694
Trade/Device Name: Target Detachable Coils
Regulation Number: 21 CFR 882.5950
Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950
Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950
Neurovascular Embolization Device | Same |
| Classification Regulation (21 CFR) | 870.3300, Class 2882.5950
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
- Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature
Stryker Neurovascular Target Detachable Coils are comprised of the following coil types: Target 360 Nano, Target 360 Ultra, Target 360 Soft, Target 360 Standard, Target Helical Nano, Target Helical Ultra, Target 3D, Target XXL 360, Target XL 360 Soft, Target XL 360 Standard, Target XL Helical, Target Tetra.
Target Detachable Coils are stretch resistant, electrolytically detachable coils consisting of a platinum-tungsten alloy coil attached to a stainless steel delivery wire.
Target Detachable Coils are specifically designed for use with Stryker Neurovascular's InZone® Detachment System (sold separately).
Target Detachable Coils are compatible with Stryker Neurovascular 2-tip marker microcatheters; refer to Instructions for Use (IFU) for the compatible microcatheter sizes.
N/A
Ask a specific question about this device
(123 days)
Trade/Device Name: InZone IST Detachment System; IZDS Connecting Cable
Regulation Number: 21 CFR 882.5950
Detachment System is a vascular and neurovascular embolization device under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950
The InZone IST Detachment System is intended for use with all versions of Stryker Neurovascular embolization devices in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.
Stryker Neurovascular's InZone IST Detachment System is a sterile, handheld, single-patient use device designed for use with Stryker Neurovascular embolization devices. The device consists of an enclosure with a detachment button, five LED indicator lamps, a funnel inset at its distal end, and a cable connection port. The device comes pre-loaded with two AAAA (1.5 VDC) batteries.
Stryker Neurovascular's IZDS Connecting Cable is a 180 cm cable intended for use with the InZone IST Detachment System in the detachment of monopolar embolization devices. The cable completes the connection between the InZone IST unit and a patient return electrode (a 20 or 22 gauge uncoated stainless-steel hypodermic needle) inserted subcutaneously at the patient's groin.
N/A
Ask a specific question about this device
(30 days)
Trade/Device Name:** Optima Coil System (Optima Packing Coil System)
Regulation Number: 21 CFR 882.5950
Review Panel:** Neurology; Cardiovascular
Product Code: HCG, KRD
Regulation Number: 21 CFR 882.5950
The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.
N/A
Ask a specific question about this device
(46 days)
| Classification** | |
|---|---|
| Class | II |
| Regulation Number | 21 CFR 870.330021 CFR 882.5950 |
Embosphere Microspheres are indicated for use in the embolization of:
- Hypervascular tumors, including symptomatic uterine fibroids
- Prostatic arteries for symptomatic Benign Prostatic Hyperplasia (BPH)
- Arteriovenous malformations
- Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature
Embosphere Microspheres are small, compressible, hydrophilic, biocompatible spheres made of acrylic polymer and porcine-derived gelatin. The microspheres are packaged in 0.9% saline and are provided sterile and non-pyrogenic in a vial or in a syringe.
The product is provided in seven size ranges to allow physicians to choose the appropriate size necessary for the vessel being embolized. The size ranges available are:
• 50-100 microns
• 40-120 microns
• 100-300 microns
• 300-500 microns
• 500-700 microns
• 700-900 microns
• 900-1200 microns
The principles of operation for the subject device Embosphere Microspheres are the same as the predicate device Embosphere Microspheres (K181300). Embosphere Microspheres are permanent implantable devices and are designed for controlled, targeted embolization. All indications for Embosphere Microspheres; uterine arteries, arteriovenous malformations, hypervascular tumors and prostate arteries all involve arterial embolization. The procedure of arterial embolization is similar for all arteries. Appropriately sized microspheres for target vessel occlusion are chosen by the trained interventional radiologist. The delivery procedure involves arterial access through an artery, using a guidewire and microcatheter under fluoroscopic guidance. Once the catheter tip is placed in the artery(ies) supplying the targeted tissue, Embosphere Microspheres mixed with a non-ionic contrast agent are delivered in a controlled manner under visualization to occlude the feeding vessel(s) to interrupt artery blood flow to the targeted area. The device is intended for single use.
I am sorry, but the provided text is an FDA 510(k) Clearance Letter for a medical device (Embosphere Microspheres). It details the regulatory clearance process, the device's intended use, and its equivalence to a predicate device.
This document does NOT contain information about any AI/ML model, its acceptance criteria, or a study proving that an AI/ML device meets those criteria.
Therefore, I cannot extract the information required to answer your request regarding the acceptance criteria and the study proving the device meets them, as it pertains to an AI model. The provided text describes a physical medical device and its regulatory review, not a software or AI-driven diagnostic tool.
Ask a specific question about this device
(69 days)
Trade/Device Name: Numen Coil Embolization System; NumenFR Detachment System Regulation Number: 21 CFR 882.5950
Promoting Embolization (KRD) |
| Regulation Number | 21 CFR 882.5950
RegulationDescription | 21 CFR § 870.3300, Device, Vascular,for Promoting Embolization21 CFR § 882.5950
Numen™ Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Numen™ Coil Embolization System is indicated for endovascular embolization of:
- . Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and . arteriovenous fistulae
- . Arterial and venous embolizations in the peripheral vasculature
NumenFR™ Detachment System is intended for use with MicroPort NeuroTech Numen™ Coil Embolization System in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature.
The Numen™ Coil Embolization System is composed of two parts as described below:
- An introducer sheath: The function of the introducer sheath is to facilitate introduction . of the coil into the microcatheter.
- . The coil system: The coil system is composed of a pusher and coil implant. The coil is a permanent implant intended to occlude blood flow in vascular abnormalities. The pusher is used to deliver the coil implant to the target lesion.
The MicroPort NeuroTech NumenFR™ Detachment System is a sterile, handheld, singlepatient use device designed for use with the MicroPort NeuroTech Numen™ Coil Embolization System. The device is operated by two pre-loaded batteries.
The provided text describes a 510(k) premarket notification for a medical device called the "Numen Coil Embolization System" and "NumenFR Detachment System." This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving enhanced performance or efficacy through extensive clinical studies. Therefore, much of the information typically sought in a study proving a device meets acceptance criteria (like MRMC studies, effect sizes, detailed ground truth establishment for AI models, and training set details) will not be present.
Based on the document, here's what can be extracted regarding acceptance criteria and performance:
The acceptance criteria are implicitly defined by the "Pass" result for each non-clinical bench test. The study proving the device meets these criteria is the Performance Testing section (Section 4).
1. A table of acceptance criteria and the reported device performance
| Test | Acceptance Criteria Summary (Implicit from Test Method and "Pass" result) | Reported Device Performance |
|---|---|---|
| Visual Inspection of Pusher | Pass visual inspection under specific magnification (e.g., no defects, meeting dimensional/surface quality). | Pass |
| Simulated Use | Perform as intended in a representative tortuous anatomical model (e.g., successful delivery, deployment, and retrieval). | Pass |
| Fatigue Testing | Durability after repeating simulated use six times, including coil retraction and re-deployment (e.g., no breakage, no significant functional degradation). | Pass |
| Detachment Time and Detachment Reliability | Reliable intentional detachment and reliable coil attachment after fatigue testing in a representative tortuous anatomical model (e.g., detaches within specified time, no premature detachment). | Pass |
| Delivery and Retraction Friction in Introducer Sheath | Max friction force when advancing or retracting the coil system in introducer sheath meets pre-specified limits. | Pass |
| Delivery, Deployment and Retraction Friction in Microcatheter | Max friction force when advancing, deploying or retracting the coil system through microcatheter in a relevant, tortuous, anatomical model meets pre-specified limits. | Pass |
| Kink Resistance | Resistance to kinking meets pre-specified acceptance criteria and can withstand bending forces encountered in clinical usage. | Pass |
| Torque Strength | Torque strength meets pre-specified criteria after rotating the proximal end of the device for 8 turns (e.g., no damage, no functional impairment). | Pass |
| Flexing Test (Per ISO 11070, Annex G) | Resistance to damage by flexing meets the requirements of ISO 11070, Annex G. | Pass |
| Fracture Test (Per ISO 11070, Annex F) | Resistance to fracture meets the requirements of ISO 11070, Annex F. | Pass |
Note: The document states "pre-specified acceptance criteria" for some tests, but the specific numerical values or detailed parameters of these criteria are not provided within the scope of this FDA letter.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify exact sample sizes (e.g., number of coils or devices) used for each bench test. It refers to "test units representative of final finished devices." Given that this is a 510(k) submission primarily relying on bench testing to demonstrate substantial equivalence for a modification (new coil type), detailed clinical test set provenance (country of origin, retrospective/prospective) is not applicable here, as it's not a clinical study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This document describes non-clinical bench testing, not a study involving human experts establishing ground truth for performance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As described above, this is bench testing, not an expert-adjudicated clinical study.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (coil embolization system) and not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the bench tests, the "ground truth" or reference for evaluating performance would be the predefined engineering specifications and performance targets for each test, as derived from relevant standards (e.g., ISO 11070 for Flexing and Fracture Tests) and internal product design requirements.
8. The sample size for the training set
Not applicable. This document describes testing for a physical medical device. The concept of a "training set" is relevant for machine learning models, which are not involved here.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for a machine learning model.
Ask a specific question about this device
(28 days)
Irvine, California 92618
Re: K242582
Trade/Device Name: Optima Coil System Regulation Number: 21 CFR 882.5950
|
| Regulation Number: | 21 CFR 882.5950
The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.
The provided text describes a 510(k) premarket notification for a medical device called the Optima Coil System (OptiBlock Line Extension). This submission focuses on demonstrating substantial equivalence to a predicate device (Optima Coil System, K223386) for which it is a modification or line extension.
Here's an analysis of the acceptance criteria and study information provided:
1. Table of acceptance criteria and the reported device performance:
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Visual Inspection, Dimensional Inspection, and Resistance Check | The test samples shall meet established test acceptance criteria for visual physical damage, secondary diameter and length, and resistance. | Pass |
| Simulated Use | The test samples shall be prepared in accordance with the instructions for use and meet established test acceptance criteria for device performance in a clinically relevant model. | Pass |
2. Sample size used for the test set and the data provenance:
The document explicitly states "The following non-clinical bench testing was performed to evaluate the device changes and to demonstrate substantial equivalence...". It mentions "test samples" but does not specify the sample size for either of the tests performed. The data provenance is from bench testing for a medical device line extension, not involving human subjects or clinical data in this submission. Therefore, country of origin is not applicable in the context of clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. Given that the studies are non-clinical bench tests, the "ground truth" would be established by engineering specifications and objective measurements rather than expert consensus on clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided and is generally not applicable to non-clinical bench testing. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable as the described device is a neurovascular embolization coil system, not an AI software or a device that assists human readers in interpreting medical images. There is no mention of AI in the document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable for the same reasons as point 5. The device is a physical medical implant, not an algorithm.
7. The type of ground truth used:
For the "Visual Inspection, Dimensional Inspection, and Resistance Check," the ground truth would be engineering specifications and manufacturing tolerances. For "Simulated Use," the ground truth would be pre-defined performance criteria within a clinically relevant model, also based on engineering and functional specifications for the device's intended operation. There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for these non-clinical tests.
8. The sample size for the training set:
This information is not provided and is not applicable given that this is not an AI/machine learning device. The concepts of "training set" and "ground truth for the training set" are relevant to machine learning model development, not for a physical medical device.
9. How the ground truth for the training set was established:
This information is not provided and is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(29 days)
California 94538
Re: K242243
Trade/Device Name: Target Detachable Coils Regulation Number: 21 CFR 882.5950
Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950
| Same |
| Regulation (21 CFR) | 882.5950
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
- Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature
Stryker Neurovascular Target Detachable Coils are comprised of the following coil types: Target 360 Nano, Target Helical Nano, Target 360 Ultra, Target Helical Ultra, Target 360 Soft, Target 3D, Target 360 Standard, Target XXL 360, Target XL 360 Soft, Target XL 360 Standard, Target XL Helical, Target Tetra. All Target Coils are stretch resistant coils. Target Coils incorporate a length of monofilament stretch resistant fiber or in the case of the Target Tetra, multi-strand material through the center of the coil designed to help resist stretching. Target Coils are designed for use with the Stryker Neurovascular InZone® Detachment System (sold separately). Each Target Coil type consists of a platinum-tungsten alloy coil attached to a stainless-steel delivery wire.
I am sorry, but based on the provided text, there is no information about acceptance criteria, device performance, sampling sizes, ground truth establishment, or clinical study results for the Target Detachable Coils.
The document appears to be a 510(k) premarket notification for the consolidation of instructions for use and patient information leaflets, along with other minor changes, for existing Target Detachable Coils. It asserts that the modifications do not alter the intended use, indications for use, or fundamental scientific technology of the predicate devices. Therefore, a comparative clinical study is not detailed here.
The document states:
- "This 510(k) requests clearance for the consolidation of three separate Target Instructions for Use (IFUs) into a single shared IFU and consolidates two separate Target Patient Information Leaflet (PIL) with Patient Information Card (PIC), making it a single PIL with PIC for all Target Detachable Coils, in addition to other changes made since the last 510(k) clearance resulting in a cumulative assessment that a new 510(k) is required."
- "The subject devices utilize the same design and materials of existing Target Coils, and the same manufacturing, packaging and sterilization processes."
- "The modified Stryker Neurovascular Target Detachable Coils have the same intended use/indications for use as the predicate Target Detachable Coils. The modifications do not alter the intended use, indications for use, or the fundamental scientific technology of the Predicate Devices."
- "Because the subject modifications do not alter the intended use or indications for use of the predicate devices, or the fundamental scientific technology of the predicate devices; and because the risk assessment of the modifications raise no new questions of safety and effectiveness, Stryker Neurovascular has determined the modified Target Detachable Coils to be substantially equivalent to the predicate device."
Therefore, the requested information regarding acceptance criteria and a study proving device performance is not present in the provided text.
Ask a specific question about this device
(111 days)
Trade/Device Name: Numen Coil Embolization System; NumenFR Detachment System Regulation Number: 21 CFR 882.5950
, for Promoting Embolization (KRD) |
| Regulation Number | 21 CFR 882.5950
RegulationDescription | 21 CFR §870.3300, Device, Vascular, forPromoting Embolization21 CFR §882.5950
Numen™ Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Numen™ Coil Embolization System is indicated for endovascular embolization of:
-Intracranial aneurysms
-Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae -Arterial and venous embolizations in the peripheral vasculature
NumenFR™ Detachment System is intended for use with MicroPort NeuroTech Numen™ Coil Embolization System in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature.
MicroPort NeuroTech has developed the Numen™ Coil Embolization System and NumenFR™ Detachment System. The Numen™ Coil Embolization System is designed to be used in conjunction with the NumenFRTM Detachment System (sold separately) for endovascular embolization of vascular abnormalities described in the intended use.
The Numen™ Coil Embolization System is composed of two parts as described below:
- . An introducer sheath: The function of the introducer sheath is to facilitate introduction of the coil into the microcatheter.
- . The coil system: The coil system is composed of a pusher and coil implant. The coil is a permanent implant intended to occlude blood flow in vascular abnormalities. The pusher is used to deliver the coil implant to the target lesion.
The MicroPort NeuroTech NumenFRTM Detachment System is a sterile, handheld, single-patient use device designed for use with the MicroPort NeuroTech Numen™ Coil Embolization System. The device is operated by two pre-loaded batteries.
The provided text describes a 510(k) premarket notification for a medical device, the Numen™ Coil Embolization System and NumenFR™ Detachment System. The document details the device, its intended use, comparison to a predicate device, and performance testing. However, it does not describe acceptance criteria for a study proving the device meets those criteria in the context of an Artificial Intelligence/Machine Learning (AI/ML) powered medical device.
The "Performance Testing" section (page 6) outlines bench testing conducted to evaluate device changes and demonstrate substantial equivalence to a predicate device. This is typical for traditional (non-AI/ML) medical devices, focusing on physical and functional properties, rather than AI performance metrics like sensitivity, specificity, or reader agreement.
Therefore, I cannot fulfill your request to describe acceptance criteria and a study demonstrating an AI/ML device's performance based on the provided text, as the text pertains to a different type of medical device assessment.
If you can provide a document that discusses the evaluation of an AI/ML medical device, I would be happy to analyze it according to your requested criteria.
Ask a specific question about this device
(28 days)
Trade/Device Name: Axium Detachable Coil; Axium Prime Detachable Coil Regulation Number: 21 CFR 882.5950
KRD |
| Regulation Number: | 21 CFR 882.5950
Axium™ Detachable Coil:
Axium™ Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. Axium™ Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: APB-X-Y-3D-ES, APB-X-Y-3D-SS, APB-X-Y-HX-ES, APB-X-Y-HX-SS)
The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: FC-X-Y-3D)
The Axium™ Prime Detachable Col is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil consist of a platinum embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a hand-held Instant Detacher (I.D.) which when activated detaches the coil from the delivery pusher tip. The I.D. is sold separately.
The document provided focuses on the substantial equivalence of the "Axium™ Detachable Coil" and "Axium™ Prime Detachable Coil" with the addition of fluorosafe markers. The primary purpose of the submission is to demonstrate that these changes do not raise new questions of safety and effectiveness compared to the predicate devices. Therefore, the "device" in question is not an AI/ML powered diagnostic device, but rather a neurovascular embolization device with a minor design change.
As such, many of the typical acceptance criteria and study details relevant to AI/ML powered devices, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable or provided in this document. The provided text describes bench testing and biocompatibility testing for the physical device itself.
However, I can extract the relevant "acceptance criteria" and "reported device performance" as presented for the specific tests conducted for this submission.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implicit from "Results" column) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ||
| Cytotoxicity | Not induce cytotoxicity | Did not induce cytotoxicity. |
| Sensitization | Not be considered a sensitizer | Were not considered a sensitizer. |
| Irritation | Be considered non-irritant | Are considered non-irritant. |
| Acute Systemic Toxicity | Show no mortality or evidence of acute systemic toxicity | Showed no mortality or evidence of acute systemic toxicity. |
| Indirect (extract) Hemolysis | Be considered non-hemolytic | Are considered non-hemolytic. |
| Material-Mediated Pyrogenicity | Meet USP 151 requirements and be non-pyrogenic | Met the requirements and were found to be non-pyrogenic. |
| Bench Testing | ||
| Visual Inspection | Darkness of fluorosafe etch mark and 360° etch mark around delivery pusher meet specifications | Met the acceptance criteria for visual inspection. |
| Marker Dimensional (Marker Position and Total Marker Length) Inspection | Total length of fluorosafe markers and their position on the delivery pusher meet specifications | Met the acceptance criteria for marker dimensional inspection. |
| Corrosion Resistance | No corrosion on the delivery pusher after soaking in saline and immersion in boiling water per ISO 10555-1, Annex A | Met the acceptance criteria for corrosion resistance. |
| Fluorosafe Marker Visibility | Meet user needs for visibility under simulated use conditions | Met the user needs for which it was designed and tested. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated for each test, but standard for device verification and validation. For biocompatibility, animal models were used (guinea pig, rabbit, mice). For bench tests, "devices" (plural) were evaluated, implying a sample size greater than one but not specified numerically.
- Data Provenance: Not applicable in the context of AI/ML data sets. These are laboratory tests on physical devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. The "ground truth" for this device modification is established through physical and objective measures against predefined specifications and recognized international standards (e.g., ISO 10993, USP 151) and user needs (for marker visibility). The "clinical users" who evaluated fluorosafe marker visibility are mentioned, but their number and specific qualifications are not detailed.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are typically for subjective assessments (e.g., image interpretation). These tests involve objective measurements and evaluations against specified criteria and standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This submission is for a physical medical device (embolization coil) with a minor design change (adding fluorosafe markers), not an AI-powered diagnostic tool. No MRMC study was conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" here is based on objective measurements against established engineering specifications, chemical/biological compatibility standards (ISO 10993, USP 151), and documented user needs. For example, meeting the criteria for "non-cytotoxic" or demonstrating "no corrosion."
8. The sample size for the training set
- Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
- Not applicable. There is no "training set."
Ask a specific question about this device
Page 1 of 17