Re: K251802
Trade/Device Name: STAR RF Ablation System
Regulation Number: 21 CFR 882.4725
| Classification | |
|---|---|
| Class: | Class II |
| Regulation: | 21 CFR 882.4725 |
| FDA Product | |
| Regulation Name | Radiofrequency Lesion Probe |
| Regulation Number | 882.4725 |
| 882.4725 | |
| Product Code | GXI |
| Product Code Description | Probe, Radiofrequency Lesion |
The STAR RF Ablation System is intended for ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change).
The subject device (STAR RF Ablation System) is a bipolar, high frequency electrosurgical system comprising the SpineSTAR Ablation Instrument, MetaSTAR RF Generator, AE Cable, Hand Switch Cable, StabiliT Introducer and PowerCURVE Navigational Osteotome. The subject device is intended to produce lesions by direct application of radiofrequency currents for the relief of chronic low back pain by basivertebral nerve ablation. The MetaSTAR RF Generator applies radiofrequency (RF) energy into the SpineSTAR. StabiliT Introducer is used to create access to the vertebral body and PowerCURVE Navigational Osteotome is used to create a channel to the target site for ablation. During lesion creation, targeted tissue in vertebral body is exposed to RF energy. The application of RF energy causes a thermal reaction at the targeted tissue site to ablate the basivertebral nerve. When used as directed, the SpineSTAR Ablation Instrument is connected to the AE Cable, which is connected to MetaSTAR. A Hand Switch Cable is connected to MetaSTAR to control delivery of RF energy to the target tissue. MetaSTAR monitors the temperature at the periphery of the ablation zone to enable controlled ablation.
N/A
Ask a specific question about this device
(115 days)
K251243**
Trade/Device Name: OneRF Trigeminal Nerve Radiofrequency Probes
Regulation Number: 21 CFR 882.4725
NeuroOne OneRF® Trigeminal Nerve Radiofrequency (TN-RF) Probe
Device Regulation Number: 21 CFR§882.4725
Probe which received marketing authorization after FDA review and clearance of K050084 under 21 CFR 882.4725
Probe which received marketing authorization after FDA review and clearance of K231675) under 21 CFR 882.4725
| 21 CFR§882.4725 | 21 CFR§882.4725 | Same for All |
| Device / Regulation Name: | Radiofrequency lesion
The NeuroOne OneRF® Trigeminal Nerve Ablation System is indicated for use in procedures to create radiofrequency lesions for the treatment of pain, or for lesioning nerve tissue for functional neurosurgical procedures.
OneRF® Ablation System for Trigeminal Nerve (TN) Ablation uses radiofrequency (RF) ablation to create lesion(s) in an area of nerve tissue that the surgeon has identified for ablation. The ablation site may be identified by diagnostic stimulation of the trigeminal nerve using the Trigeminal Nerve Radiofrequency (TN-RF) Probe to accurately locate the target area for ablation.
The subject device is identical to the sEEG-RF Probe cleared after FDA review of K231675. In K231675, the FDA reviewed and cleared the sEEG-RF Probe when used with the NeuroOne OneRF® Generator to form the OneRF® Ablation System. The OneRF® Ablation System is indicated to create radiofrequency lesions in nervous tissue for functional neurosurgical procedures.
In this new 510(k), we propose to market the identical sEEG-RF Probes to create lesion(s) in the trigeminal nerve "for the treatment of pain." This use for treatment of facial pain requires new accessories (Cannula and Tuohy Borst Adapter) for insertion/placement of the sEEG-RF Probe. When the sEEG-RF Probe is used to create lesion(s) in the trigeminal nerve, it will be marketed as the Trigeminal Nerve Radiofrequency (TN-RF) Probe. The TN-RF Probe and new accessories, when combined with the cleared (K231675) NeuroOne OneRF® Generator and accessories, will be referred to as the OneRF® Trigeminal Nerve Ablation System.
The OneRF® for TN Ablation System components/accessories consist of the following:
- TN-RF Probe (with Stylet)
- 16mm x 5 Contact TN-RF Probe Kit
- 26.5mm x 8 Contact TN-RF Probe Kit
- Insertion Components
i. Insulated Cannula w/Stylet
ii. Tuohy Borst Adapter (Adapter) - Stimulation Components
i. Cable Assembly (5 or 8 connector pins) (CA) - Ablation Components
i. Stylet
ii. Temperature Accessory (TA)
iii. Spacer Tubes
iv. Radiofrequency Connector Box (RFCB)
v. Equipment- Generator
- Generator Interface Cable (GIC)
- Cart
- Foot Pedal (optional)
- Ground Pad
The provided document is a 510(k) Clearance Letter for a medical device. It does not describe a study that uses a test set, training set, or ground truth to evaluate algorithm performance. The clearance is based on the substantial equivalence of the "NeuroOne OneRF Trigeminal Nerve Radiofrequency Probes" to existing predicate devices, supported by non-clinical performance tests and biocompatibility assessments.
Therefore, I cannot provide the requested information in a table format as there are no relevant acceptance criteria or reported device performance metrics for an AI/algorithm-based study within the document.
Here's why each of your requested points cannot be answered from the provided text:
- A table of acceptance criteria and the reported device performance: The document focuses on performance testing related to mechanical, dimensional, packaging integrity, and sterilization aspects of the physical device, not an AI algorithm's performance. The "Lesion Size Testing" states "Lesion sizes were proportional to time and temperature. Lesion size is comparable to predicate," but it doesn't provide specific quantitative acceptance criteria or detailed numerical results beyond this qualitative statement.
- Sample size used for the test set and the data provenance: Not applicable. The studies are non-clinical, involving physical device testing, not data analysis on a test set.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no ground truth established by experts for a test set in these non-clinical tests.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic or interpretative device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
Summary of available information regarding performance testing:
The document outlines a series of non-clinical performance tests to support the substantial equivalence claim. These tests primarily verify the physical and functional integrity of the device and its components.
- Lesion Size Testing:
- Overview: Measured lesion size as a function of temperature and time, for monopolar and bipolar configurations, and temperature control/manual modes.
- Results: "Lesion sizes were proportional to time and temperature. Lesion size is comparable to predicate." (No specific numerical data or explicit acceptance criteria provided in the summary).
- Dimensional Verification of the Cannula:
- Overview: Evaluated dimensional characteristics and compatibility between components.
- Results: "Pass – The test results indicate that the Cannula meet the dimensional requirements."
- Mechanical Performance:
- Overview: Verified specifications related to mechanical interaction between the TN-RF Probe, Cannula, and Tuohy Borst Adapter.
- Results: "Pass – The test results indicate that the TN-RF Probe, Cannula, Tuohy Borst Adapter designs meet the mechanical performance requirements."
- Mechanical Integrity:
- Overview: Evaluated mechanical integrity of the TN-RF Probe, Cannula, and Tuohy Borst Adapter.
- Results: "Pass - The test results indicate that the TN-RF Probe, Cannula and Tuohy Borst Adapter designs meet the mechanical integrity requirements."
- TN-RF Probe Kit Package Integrity:
- Overview: Tested the packaged device and labeling against conditions of packaging, shelf life, and distribution per ISO 11607-1, ISTA 3A, ASTM D4169, ASTM F1980, ASTM F2096, ASTM F88.
- Results: "Pass - The test results indicate that the TN-RF Probe Kit (...) packaging designs meet the integrity requirements (i.e., seal strength, bubble leak, label inspection, and no damage that impacts device sterility)."
- Sterilization:
- Overview: Validated ethylene oxide sterilization process to achieve a minimum SAL of 10⁻⁶ per ISO 11135.
- Results: "Pass – All criteria passed and the new product/package configuration was adopted into the validated sterilization cycle."
- Usability – Summative Validation:
- Overview: Performed in accordance with FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices," February 3, 2016.
- Results: "Pass – The NeuroOne OneRF® TN-RF Ablation System has been found to be safe and effective for the intended users, uses, and use environments."
- Biocompatibility:
- Overview: Tested sEEG-RF Probe for prolonged contact and Cannula w/stylet, Tuohy Borst Adapter for limited contact.
- Results: "Passed" for all tested components. No testing for Temperature Accessory/Spacer Tubes/Stylet and Radio Frequency Connector Box as there is no direct or indirect patient contact.
The provided document specifically clarifies that the device is substantially equivalent, and the "conclusions drawn from the nonclinical testing demonstrate the device is as safe, as effective, and performs as well as the legally marketed device predicates, per 21 CFR 807.92(b)(3)." This is a traditional medical device clearance, not an AI/ML-driven software clearance.
Ask a specific question about this device
(111 days)
OptaBlate BVN Intraosseous Nerve Ablation System (10mm Probe Dual Kit)
Regulation Number: 21 CFR 882.4725
Intraosseous Nerve Ablation System |
|---|---|
| Regulation | Radiofrequency lesion probe (21 CFR 882.4725
The Stryker OptaBlate BVN Intraosseous Nerve Ablation System is intended for the ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least six months duration that has not responded to at least six months of conservative care, and is also accompanied by features consistent with Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change).
The subject device Stryker Optablate® BVN Intraosseous Nerve Ablation System is a bipolar, high frequency electrosurgical system comprising the BVN probe, Microinfuser, syringe, introducer handpiece, introducer conduit, 10 G access cannula with diamond tip stylet, and 10 G bevel tip stylet. The subject device is intended to be used in conjunction with the existing Optablate radiofrequency (RF) generator (K221074), MultiGen2 Splitter Cable (K170242), and Optablate Microinfuser (K221074) to produce lesions by the direct application of radiofrequency currents for the relief of chronic low back pain. The subject generator applies temperature-controlled, radiofrequency (RF) energy into the probe. During lesion creation, targeted tissue is exposed to RF energy using an active probe inserted into a conduit that is within an access cannula. The application of RF energy causes a thermal reaction at the targeted tissue site to ablate the basivertebral nerve. It is indicated for the L3 through S1 vertebrae. When used, the subject OptaBlate BVN probes are connected to the splitter cable, which is connected to the generator, to deliver RF energy to the target tissue. The Optablate Microinfuser is connected to the probe to deliver a small amount of saline to the ablation site. The saline exits the probe between the emitters and helps ensure there is good electrical connection between the emitters and the tissue. The quality of the connection is measured in units called impedance. Impedance goes up if the connection is poor. If the impedance goes to high, the generator will trigger an error and stop the ablation. The Microinfuser slowly introduces saline to prevent impedance rises.
The provided FDA 510(k) clearance letter and summary for the OptaBlate BVN Intraosseous Nerve Ablation System focuses on the **device's safety and effectiveness **through comparison with a predicate device and non-clinical performance testing. It explicitly states "No clinical testing was required to support this submission." This means there was no study proving the device's performance in a clinical setting against acceptance criteria that would typically involve human or AI reader performance metrics, effect sizes, or ground truth established by experts.
Therefore, I cannot fulfill the request for information related to:
- Table of acceptance criteria and reported device performance based on clinical outcomes.
- Sample size for a test set (as no clinical test set was used for effectiveness).
- Number of experts and their qualifications used to establish ground truth.
- Adjudication method for a test set.
- MRMC comparative effectiveness study, effect size of human reader improvement.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for training set (as no AI/machine learning component is described).
- How ground truth for training set was established.
However, I can extract information related to the non-clinical acceptance criteria and proof of mechanism/performance.
Non-Clinical Acceptance Criteria and Device Performance (Summary):
The device's acceptance criteria primarily revolve around its ability to perform its intended function safely and effectively in a technical/mechanical/biological sense, rather than a diagnostic or AI-assisted clinical performance sense.
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):
| Acceptance Criteria Category | Specific Test/Standard | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Sterilization | ISO 11135:2014 & ISO 10993-7:2008/Amd 1:2019 (Ethylene Oxide - Kits); ISO 11137-1:2006, ISO 11137-2:2013, ISO 11137-3:2017 (Radiation - Microinfuser) | Conformance with standards for effective sterilization and acceptable residuals. | Pass (Compliance demonstrated) |
| Biocompatibility | ISO 10993-5:2009, ISO 10993-12:2021 (Cytotoxicity); ISO 10993-23:2021, ISO 10993-12:2021 (Irritation); ISO 10993-11:2017, ISO 10993-12:2021 (Acute Systemic Toxicity); ISO 10993-10:2010, ISO 10993-12:2021 (Sensitization) | Test articles to be non-cytotoxic, show no greater biological reaction than controls, and cause no sensitization reaction. | Pass (All tests met criteria) |
| Electrical Safety & EMC | IEC 60601-1 Ed. 3.2 (General Safety); IEC 60601-1-2 Ed. 4.1 (EMC); IEC 60601-1-8 Ed. 2.2 (Alarms); IEC 60601-2-2 Ed. 6.1 (HF Surgical Equipment) | Compliance with essential performance and safety requirements of specified electrical and electromagnetic compatibility standards. | Pass (All applicable clauses tested and conform) |
| Physical/Mechanical Performance | ASTM D4169 (Transit); ASTM F2096 (Bubble Test); ASTM F88 (Seal Peel Test); Internal Specs (Mechanical Testing, Dimensional, Visual Inspection) | Device and packaging integrity maintained during transit; no gross leaks; adequate seal strength; specified mechanical properties (insertion force, retraction force, flow rate, tensile strength, twist cycle, torsional stiffness, impact force); adherence to dimensional specifications; absence of surface defects. | Pass (All tests met criteria) |
| Functional Performance (BVN Probe Specific) | Internal Test Method (Temperature Accuracy); Internal Test Method (Lesion Size) | Accuracy verified across full functional use range; measured RF lesion size in ex vivo tissue model. | Pass (Accuracy verified; ex vivo lesion size measured and acceptable) |
2. Sample Size Used for the Test Set and Data Provenance:
- Test Set Sample Size: Not applicable in the context of clinical performance; the testing described is non-clinical (e.g., in-vitro biocompatibility tests, mechanical tests of device components, ex vivo tissue models for lesion formation).
- Data Provenance: The data comes from the manufacturer's internal testing (Stryker Instruments) as part of their 510(k) submission. No specific country of origin for test data is mentioned beyond "Stryker Instruments" (headquartered in Portage, MI, USA). The studies are "non-clinical" performance evaluations rather than retrospective or prospective human subject studies.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable. The ground truth for this device's non-clinical performance is established by engineering specifications, international and national standards (e.g., ISO, IEC, ASTM), and the physical/chemical properties measured via established test methods. No human experts were used to establish "ground truth" in the diagnostic sense for this submission.
4. Adjudication Method for the Test Set:
- Not applicable. This concept applies primarily to clinical studies where human or AI interpretations are adjudicated. For non-clinical testing, results are typically binary (pass/fail) based on pre-defined quantitative or qualitative acceptance criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No. The submission explicitly states, "No clinical testing was required to support this submission." Therefore, no MRMC study to compare human reader performance with or without AI assistance was conducted or reported.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Not applicable. This device is an ablation system, not a diagnostic algorithm. Its performance is related to its mechanical, electrical, and biological safety and its ability to create a lesion, not to interpret medical images or data.
7. The Type of Ground Truth Used:
- For Biocompatibility: Established by standardized biological response measurements against control samples, as per ISO 10993 series.
- For Electrical Safety/EMC: Established by conformance to specified IEC standards for medical electrical equipment.
- For Mechanical and Functional Testing: Established by engineering specifications and measurements in controlled laboratory settings (e.g., ex vivo tissue models for lesion size) and validated against ASTM standards where applicable. This is not "expert consensus, pathology, or outcomes data" in the clinical sense.
8. The Sample Size for the Training Set:
- Not applicable. This device does not involve an AI/machine learning component that would require a "training set."
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as no training set was relevant for this device submission.
Ask a specific question about this device
(254 days)
201100
China
Re: K241367
Trade/Device Name: RF Cannula
Regulation Number: 21 CFR 882.4725
mztmedtech.com
Device Trade Name: RF Cannula
Device Class: Class II
Regulation Number: 21 CFR 882.4725
The RF cannula is intended for use in RF heat lesion procedures for the relief of pain.
The RF Cannula is composed of Cannula, Cannula hub, Stylet, Stylet hub, Insulating layer, Depth stop and Protection tube.
Materials being used include SUS 304 stainless steel, K-Resin, polyethylene, polyethylene terephthalate.
RF Cannula is supplied sterile and is only for single-use. It is used in conjunction with a RF Generator and RF Electrode and other accessories to create Radiofrequency (RF) Lesions of nerve tissue or for use in percutaneous nerve blocks. During the RF heat lesion procedure, the RF cannula bare tip is placed into the target tissue; preloaded stylet is withdrawn and replaced by the RF Electrode. The radio frequency is produced by the RF Generator and released by the cannula bare tip.
N/A
Ask a specific question about this device
(181 days)
-------------------------|
| Device Regulation Number: | 21 CFR§882.440021 CFR§882.4725
Probe which received marketing authorization after FDA review of K010202 and was cleared under 21 CFR 882.4725
| NA |
| Device RegulationNumber | 21 CFR§882.4725
| 21 CFR§882.4725
The NeuroOne OneRF Radiofrequency Ablation System is indicated for creation of radiofrequency lesions in nervous tissue for functional neurosurgical procedures.
The OneRF™ Ablation System components consist of the:
- Radiofrequency (RF) Generator and Accessories
- Generator Interface Cable (GIC), Cart and Foot Pedal (optional)
- sEEG-RF Probe (with Universal Cable Assembly) and Ablation Accessories
- Temperature Accessory, Spacer Tubes, Stylet and RF Connector Box
The OneRF™ Ablation System uses radiofrequency ablation to create lesion (s) in an area of nervous tissue that the neurosurgeon has identified for ablation. This is accomplished by diagnostically locating the area to ablate using the implanted Stereoelectroencephalography (sEEG). Once the area of ablation is located, the generator and ablation accessories are taken to the patient with implanted sEEG(s). The sEEG(s) now function as an sEEG-RF Probe for RF ablation.
To perform the ablation the Temperature Accessory (TA) is inserted into the sEEG-RF Probe to the predetermined electrode contact location using the specified spacer tube. A stylet may be used to ensure the lumen is patent prior to inserting the TA. The universal cable assembly remains attached to the sEEG-RF Probe while its cables are disconnected from the diagnostic head box. One (monopolar) or two (bipolar) cables from the universal cable assembly are inserted into the radiofrequency connector box (RFCB) depending on the contact area(s) to be ablated. The TA is also connected to the RFCB is attached to the Generator Interface Cable which is attached to the Generator. Temperature and time are used to create lesion (s).
The provided text describes the NeuroOne OneRF Ablation System, a radiofrequency lesion generator and probe indicated for creating radiofrequency lesions in nervous tissue for functional neurosurgical procedures.
Here's an analysis of the acceptance criteria and study proving the device meets them, based solely on the provided text:
Important Note: The provided text is a summary from an FDA 510(k) premarket notification. It describes the device, its intended use, comparison to predicate devices, and a high-level summary of performance testing. It does not contain detailed pass/fail acceptance criteria values for most tests or granular study design information (e.g., specific sample sizes for particular experiments, expert qualifications, or detailed ground truth methodologies beyond "lesion size was determined"). The information here is inferred from the "Results and Conclusions" section, which consistently states "Pass" or "Met applicable requirements."
1. Table of Acceptance Criteria and Reported Device Performance
Given the lack of specific quantitative acceptance criteria in the provided text, the table below will list the tests performed and their reported outcomes, implying that the device met the (unstated) acceptance criteria for each.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Lesion Size: Lesion sizes determined based on time and temperature should be comparable to the predicate. | Pass - Lesion sizes were determined based on time and temperature. Lesion size is comparable to predicate. |
| Dimensional Verification & RFCB Cable Durability: Dimensional characteristics and compatibility between components must be demonstrated; flexural durability of RFCB cables must be met. | Pass - The test results indicate that the sEEG-RF Probe / Temperature Accessory / Stylet / Spacer Tube and Radio Frequency Connector Box designs meet the dimensional and cable durability requirements. |
| Mechanical Performance: Specifications related to the mechanical interaction between the sEEG-RF Probe and Accessories must be verified. | Pass - The test results indicate that the sEEG-RF Probe / Temperature Accessory / Stylet / Spacer Tube and Radio Frequency Connector Box designs meet the mechanical performance requirements. |
| Mechanical Integrity: Mechanical durability of the sEEG-RF Probe Accessories must be demonstrated. | Pass - The test results indicate that the sEEG-RF Probe Accessories (Temperature Accessory and Radio Frequency Connector Box) designs meet the mechanical integrity requirements. |
| Ablation System Performance: Specifications related to energy delivery and temperature accuracy, including durability after use, must be met. | Pass - The test results indicate that the sEEG-RF Probe / Temperature Accessory / Stylet / Spacer Tube and Radio Frequency Connector Box designs meet the system performance requirements. |
| Generator System Performance: Specifications related to the Generator and UI Software must be met. | Pass - The test results indicate that the Generator and UI Software designs meet the system performance requirements. |
| Electrical Safety: Product must meet applicable requirements of Electromagnetic Compatibility and Electrical Safety standards (e.g., IEC 60601-1, -2, -6 and IEC 60601-2-2). | Pass - Met applicable requirements. |
| Temperature Accessory Kit and RFCB Package Integrity: Packaged device and labeling must withstand packaging, shelf life, and distribution testing (e.g., ISO 11607-1, ISTA 3A, ASTM D4169, ASTM F1980-16, ASTM 2096, ASTM F88) without loss of function, sterility, or legibility. | Pass - The test results indicate that the sEEG-RF Probe Accessories (Temperature Accessory / Stylet / Spacer Tube and Radio Frequency Connector Box) packaging designs meet the integrity requirements (i.e., seal strength, bubble leak, label inspection, and no damage that impacts device sterility). |
| Sterilization: Sterilization process must be validated to demonstrate a minimum SAL of 10-6 for the product using Ethylene Oxide per ISO 11135. | Pass - All criteria passed and the sterilization cycle was validated. |
| Usability - Summative Validation: Device must be found safe and effective for intended users, uses, and use environments, in accordance with FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices." | Pass – The NeuroOne OneRF™ Ablation System has been found to be safe and effective for the intended users, uses, and use environments. |
| Software: Software must address applicable requirements in accordance with FDA Guidance "Content of Premarket Submissions for Software Contained in Medical Devices" and IEC 62304. | Software analysis addressed applicable requirements. |
| Cybersecurity: Cybersecurity analysis must address applicable requirements in accordance with FDA guidance documents. | Cybersecurity analysis addressed applicable requirements. |
| Biocompatibility: Components with prolonged (>24 hours to 30 days) contact with tissue/bone and limited (<24 hours) contact with intact skin must pass biocompatibility tests. | Passed – Reference (K211367/K222404). (Note: Sterile components that have no direct or indirect patient contact did not require testing). |
Details of the Study Proving Device Meets Acceptance Criteria:
Given the nature of the device (radiofrequency ablation system) and the provided 510(k) summary, the "study" primarily refers to a series of non-clinical performance tests rather than a typical clinical trial involving medical imaging data analysis.
-
Sample sizes used for the test set and the data provenance:
- Test Set (Non-Clinical Performance Tests): The document does not specify exact sample sizes for each non-clinical test (e.g., number of probes tested for durability, number of lesions created for size testing). It generally refers to these as "testing" or "activities."
- Data Provenance: Not explicitly stated, however, the tests are laboratory-based ("ex-vivo tissue" mentioned for lesion size testing) rather than human or geographical. All studies are described as "design verification and validation activities."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. For engineering performance tests (e.g., electrical safety, mechanical integrity, sterilization), "ground truth" is typically established by objective measurement against engineering specifications rather than expert consensus. For "Lesion Size Testing," the document states "Lesion sizes were determined based on time and temperature," implying measured outcomes rather than expert visual assessment establishing "ground truth."
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not mentioned. For engineering performance tests, outcomes are typically definitive (pass/fail against a specification) and do not usually involve adjudication in the way that clinical imaging studies might.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study (MRMC for AI-assisted human readers) is not applicable to a radiofrequency ablation system as described. The device is a direct treatment tool, not an AI diagnostic aid that assists human readers in interpreting medical images.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The device functions as a system with a human operator (neurosurgeon) in control ("Operator controlled; RF delivered from RF generator to compatible sEEG - RF probes to create lesions in nervous tissue"). Performance tests were conducted on the system, including the generator, probes, and software, to ensure it operates according to specifications. While software verification was done, the "standalone" performance in the context of an AI-only diagnostic algorithm is not relevant here.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For Lesion Size Testing: "Lesion sizes were determined based on time and temperature." This suggests direct physical measurement or calculation based on set parameters and observed outcomes, likely in ex-vivo tissue as mentioned in the table.
- For other engineering tests: The "ground truth" is adherence to established engineering specifications, international standards (e.g., IEC, ISO, ASTM), and design requirements. This is verified through direct measurement, functional testing, and environmental challenges.
-
The sample size for the training set:
- Not applicable. This device is not an AI/machine learning algorithm that requires a training set in the conventional sense for diagnostic image analysis. Its software facilitates control and monitoring of the ablation process. Software verification was conducted in accordance with relevant FDA guidance and IEC 62304.
-
How the ground truth for the training set was established:
- Not applicable, as there is no "training set" in the context of an AI algorithm learning from data. The software within the device is deterministic and performs control functions based on predefined algorithms and parameters. Ground truth for software verification typically involves demonstrating that the software meets its specified functional and non-functional requirements through testing against those requirements.
Ask a specific question about this device
(421 days)
Cannula |
| Regulation Number: | 21 CFR 882.4725
Radiofrequency (RF) Electrode |
| Regulation Number: | 21 CFR 882.4725
Intended for use for lesioning of neural tissue and for pain management. It is indicated for use in the peripheral nervous system. The APEX 6 is to be used with LCCS electrodes and cannulae and Conmed Thermogard Dispersive Electrodes.
The RF Innovations APEX 6 is a desktop RF lesioning generator which is used for the lesioning of neural tissue. The APEX 6 is a multi-lesioning, 6 channel portable generator that can provide continuous or pulsed RF output at 460 kHz, and monopolar or dual electrode modes. The device includes sensory and motor stimulation functions to fine tune electrode placement for procedures. Based on performance testing the device is designed to connect to LCCS FDA cleared lesioning probes which are inserted into patients for lesioning of neural tissue during medical procedures. Device features include a touch screen monitor incorporating microprocessor and graphics display for user interface as well as self-diagnostics, calibration checks, and recordkeeping functions.
This document is a 510(k) summary for the APEX 6 Lesioning Generator. It outlines the device's characteristics, its intended use, and how it compares to a predicate device (NeuroTherm NT 2000 RF Lesioning Generator) to establish substantial equivalence.
Here's an analysis of the acceptance criteria and the supporting study, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Lesion Size (comparison to predicate device) | Affected area differences between APEX 6 and predicate device should be minimal (implied: insignificant for clinical equivalence). | "All affected area differences were less than 1 mm." The study confirms "the Apex 6 and the predicate device are substantially equivalent in regards to affected tissue size." |
| Comparison of Treatment Times (comparison to predicate device) | Time to ramp up to set temperature and treatment times for APEX 6 should be substantially equivalent to the predicate device under worst-case conditions. | "Results demonstrated that the APEX 6 was substantially equivalent in terms of treatment times compared to the predicate device." |
| Design Validation Review | All documented device requirements (Requirement Specification, Traceability Matrix, Software Specifications, Validation Plans, Instructions for Use) must be met. | "The finished unit design review verified that all documented device requirements were met." |
| Every Unit Functional Test | Each manufactured unit must pass a full functional test regimen (Voltage checks, Program and Impedance Testing, Software testing, Main GUI, Final testing: Impedance/Temperature measurement, Electrical safety testing) prior to shipment. | "Each unit must pass all tests prior to shipment." |
| Safety (Bench Testing) | Compliance with ANSI/AAMI 60601-1, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI 60601-2-2, and CAN/CSA-C22.2 No. 60601-2-2. | "Tested to ANSI/AAMI 60601-1:2005 + C1:2009 + A2:2010 + A1:2012, CAN/CSA-C22.2 & 60601-1:2014 AND ANSI/AAMI 60601-2-2:2017, CAN/CSA-C22.2 No. 60601-2-2:2019." The document states compliance implicitly by reporting "Tested to...". |
| Electromagnetic Compatibility (EMC) (Bench Testing) | Compliance with IEC 60601-1-2 / EN 60601-1-2. | "EMC was tested in accordance with IEC 60601-1-2:2014/ EN 60601-1-2:2015/IEC 60601-2-2." The document states compliance implicitly by reporting "Tested in accordance with...". |
| Software Performance (Validation to FDA Moderate Level of Concern) | Software performs as expected per validation to FDA guidance for Moderate Level of Concern. | "Software testing supports that the APEX 6 performs as expected. Validation was performed to the FDA Moderate Level of Concern per the FDA software guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." |
Study Details:
The provided document describes bench testing to demonstrate the substantial equivalence of the APEX 6 to its predicate device, primarily focusing on physical performance characteristics and safety standards. There is no clinical study (human-in-the-loop or standalone AI performance) described for this device, as it is a medical device generator and not an AI/Software as a Medical Device (SaMD).
Here's an breakdown of the specific points requested, based on the provided text:
-
A table of acceptance criteria and the reported device performance: (See table above)
-
Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify exact sample sizes for the "Lesion size" or "Comparison of treatment times" tests. It refers to "minimal, typical, and maximum energy delivery" for lesion size and "worse case conditions" for treatment times, implying a set of controlled experimental conditions rather than a large clinical "test set" in the common AI/SaMD sense. For "Every unit functional test," the sample size is "Every unit manufactured."
- Data Provenance: The tests are described as "Bench testing," indicating they were performed in a laboratory or controlled environment. There is no information regarding country of origin or whether it's retrospective or prospective, as it's not a clinical data study.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable to the type of testing described. The "ground truth" for a device like a radiofrequency lesion generator is based on physical measurements (e.g., lesion dimensions, electrical parameters, time) and established engineering/safety standards. It's not about expert interpretation of medical images or clinical outcomes that would require human expert consensus for ground truth.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This is not applicable. Adjudication methods are typically used in clinical studies where multiple human readers interpret data, and significant disagreements require a tie-breaking mechanism or consensus. The tests described are bench tests with objective physical measurements.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This type of study is relevant for AI/SaMD devices where the AI assists human readers in diagnostic or screening tasks. The
APEX 6is a physical medical device generator, not an AI software.
- No, an MRMC study was not done. This type of study is relevant for AI/SaMD devices where the AI assists human readers in diagnostic or screening tasks. The
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- While the
APEX 6itself performs functions autonomously (e.g., generating RF energy), the "standalone" concept usually applies to AI algorithms whose performance is evaluated independently of human input. No such "algorithm-only" performance study is described here, as the device's function is the generation and control of RF energy, which is evaluated through bench tests against physical parameters and predicate device performance. Software functions were validated, but this is different from a standalone AI performance study.
- While the
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the bench tests was based on physical measurements (e.g., linear dimensions in millimeters, time in seconds, electrical parameters like impedance, voltage, current, power) and compliance with established international and national standards (e.g., IEC 60601 series, ANSI/AAMI, CAN/CSA).
- For "Lesion size" and "Comparison of treatment times," the "ground truth" was derived from the performance of the legally marketed predicate device, used as a benchmark for substantial equivalence.
-
The sample size for the training set:
- This information is not applicable. The device is not an AI/machine learning model that requires a "training set" in the sense of data used to train a statistical or learning algorithm. Its design and validation rely on engineering principles, product specifications, and comparisons to predicate devices.
-
How the ground truth for the training set was established:
- This information is not applicable, as there is no training set for an AI/ML model for this device.
Ask a specific question about this device
(89 days)
Trade/Device Name: Intracept Intraosseous Nerve Ablation System Regulation Number: 21 CFR 882.4725
| Classification Name: | Radiofrequency lesion probe, 21 CFR 882.4725
The Intracept Intraosseous Nerve Ablation System is intended to be used in conjunction with radiofrequency (RF) generators for the ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least six months duration that has not responded to at least six months of conservative care, and is also accompanied by features consistent with Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change).
The Intracept Intraosseous Nerve Ablation System (Intracept System) is comprised of sterile, singleuse components: The Intracept Access Instruments include cannulas and stylets that provide access to the intended site of radiofrequency (RF) ablation. The Intracept RF Probe conducts RF energy to the target location. To obtain the energy needed for tissue ablation, the Intracept RF Probe is used with the Relievant Radiofrequency Generator (RFG). The Relievant RFG is a universal AC powered, microcontroller controlled, bipolar RF generator intended to deliver RF energy to a targeted site. During RF energy delivery, power is continuously monitored and controlled, based on temperature and impedance measurements at the treatment site, to ensure proper operation. Currently, the Relievant RF Generator (reference device (K171143)) is the only compatible RF generator for use with the Intracept System. The Intracept System uses RF energy to ablate the basivertebral nerve for relief of chronic low back pain and involves a two-step process. First, utilizing the Access Instruments, based on a minimally invasive, transpedicular or extrapedicular approach, a cannula and stylets are placed into the vertebral body to create a path or channel to the terminus of the basivertebral foramen. The RF Probe is then placed into this channel and controlled RF energy is delivered to ablate the basivertebral nerve. This nerve has been identified as a proprioceptive sensory nerve with enervation of the vertebral endplates.
The Relievant Medsystems, Inc. Intracept Intraosseous Nerve Ablation System (K222281) did not undergo a clinical study as substantial equivalence was determined based on non-clinical performance testing. The device is a modification of a previously cleared device (K213836). Modifications were made to the Access Instruments component of the system to improve usability and performance, but the RF Probe remained unchanged.
Here's a summary of the non-clinical testing:
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Biocompatibility Testing | Compliance to applicable ISO 10993 standards for tissue/bone/dentin contact <24 hours. Materials used in construction are equivalent to the Predicate Device. | All tests passed. |
| Cytotoxicity | ISO 10993-5 (MEM Elusion) | PASS |
| Sensitization | ISO 10993-10 (Guinea Pig Maximization Sensitization) | PASS |
| Acute Systemic Toxicity | ISO 10993-11 (Acute Systemic Injection Test in Mice) | PASS |
| Intracutaneous Reactivity | ISO 10993-10 (Intracutaneous Reactivity Irritation Test in Rabbits) | PASS |
| Dimensional and Functional Testing | Devices are equivalent in size, materials, and construction to the Predicate Device. | All tests passed. |
| Corrosion | ISO 10555-1, Annex A: No visible signs of corrosion. | PASS |
| Transit | ASTM D4169 - Standard Practice for Performance Testing of Shipping Containers and Systems (DC 13, assurance level II) | PASS |
| Gross Leaks | ASTM F2096 - Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) | PASS |
| Seal Strength | ASTM F88/F88M - Standard Test Method for Seal Strength of Flexible Barrier Materials (Seal peel ≥1.0 lbs/in) | PASS |
| Mechanical | Met all performance testing per Product Specifications. | PASS |
| Sterilization | ANSI/AAMI/ISO 11137-1 and 11137-2: Sterility assurance of 10-6. | PASS |
| Interface and Primary Operating Function Testing | Devices have equivalent performance to the Predicate Device. | All tests passed. |
| Simulated Use | Following exposure to 6 full deployments, devices remained functional without damage and met interface requirements. | PASS |
| Usability | IEC 62366-1: Application of Usability Engineering to Medical Devices. Usability testing simulated in sawbones. | PASS |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated for each bench test, but the usability testing involved "15 users."
- Data Provenance: Not applicable as these were non-clinical bench tests. The provenance is internal laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as this was non-clinical bench testing. The evaluation was against established standards and product specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this was non-clinical bench testing. Results were objectively measured against predefined criteria in relevant standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done, as this was a non-clinical submission for device modification, not an AI/software device requiring comparative effectiveness studies with human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is a physical medical device (nerve ablation system), not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For non-clinical bench testing, the "ground truth" is defined by the technical specifications and international standards (e.g., ISO, ASTM, AAMI, IEC) to which the device must conform.
8. The sample size for the training set
Not applicable, as this was not an AI/ML device requiring a training set.
9. How the ground truth for the training set was established
Not applicable, as this was not an AI/ML device requiring a training set.
Ask a specific question about this device
(92 days)
Trade/Device Name: Intracept Intraosseous Nerve Ablation System Regulation Number: 21 CFR 882.4725
|
| Classification Name: | Radiofrequency lesion probe, 21 CFR 882.4725
The Intracept Intraosseous Nerve Ablation System is intended to be used in conjunction with radiofrequency (RF) generators for the ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least six months duration that has not responded to at least six months of conservative care, and is also accompanied by features consistent with Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change).
The Intracept Intraosseous Nerve Ablation System is comprised of sterile, singleuse components:
- The Intracept Access Instruments include introducers, cannulas and stylets that provide access to the intended site of radiofrequency (RF) ablation.
- The Intracept RF Probe conducts RF energy to the target location.
To obtain the energy needed for tissue ablation, the Intracept RF Probe is used with the Relievant Radiofrequency Generator (RFG).
The Intracept System uses RF ablation of the basivertebral nerve for relief of chronic low back pain and involves a two-step process. First, utilizing the Access Instruments, based on a minimally invasive, transpedicular or extrapedicular approach, a cannula and stylets are placed into the vertebral body to create a path or channel to the terminus of the basivertebral foramen. The RF Probe is then placed into this channel at the terminus of the basivertebral foramen and controlled RF energy is delivered to ablate the basivertebral nerve (BVN). This nerve has been identified as a proprioceptive sensory nerve with enervation of the vertebral endplates.
The provided text describes the 510(k) premarket notification for the Intracept Intraosseous Nerve Ablation System. It focuses on demonstrating substantial equivalence to a predicate device, specifically regarding the introduction of an additional set of ablation parameters.
Based on the provided information, the device in question is a medical device for radiofrequency ablation, not an AI/ML powered device. Therefore, many of the requested elements for describing an AI/ML powered medical device, such as acceptance criteria based on metrics like sensitivity/specificity, sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and how ground truth for training was established, are not applicable and not present in this document.
The document primarily focuses on demonstrating that the new version of the device, with additional ablation parameters, is substantially equivalent to a previously cleared version. The "acceptance criteria" here are effectively the criteria for establishing substantial equivalence, which is shown through non-clinical performance testing.
Here's an interpretation based on the provided text, recognizing its focus on a hardware device rather than an AI/ML algorithm:
Acceptance Criteria and Study Proving Device Meets Acceptance Criteria
1. Table of Acceptance Criteria and Reported Device Performance:
Since this is a hardware device (RF ablation system) rather than an AI/ML diagnostic device, the "acceptance criteria" are not reported in terms of standard diagnostic performance metrics (e.g., sensitivity, specificity, AUC). Instead, they are framed around demonstrating safety and efficacy comparable to a predicate device, particularly with the introduction of new ablation parameters.
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance (as stated in the 510(k) Summary) |
|---|---|
| Device performs as intended. | "Subject Device (Intracept System) with additional ablation parameters performs as intended." |
| No new safety and/or efficacy concerns raised by changes. | "does not raise any new safety and/or efficacy concerns when compared to the legally marketed Predicate Device (Intracept System)." |
| Risk profile of additional ablation parameters is acceptable. | "Risk analysis of an additional set of ablation parameters at lower temperature, ramp, and time showed no increase in risk profile." |
| Functional equivalence to predicate device. | "The Subject and Predicate Devices have a functionally equivalent design, with the difference being the introduction of an additional set of ablation parameters." |
The key difference for the new device is the addition of ablation parameters:
- New Ablation Parameters: 75° C, 0.5°C/second ramp, 7 minutes (420 seconds) time. The predicate had only 85° C, 1°C/second ramp, 15 minutes (900 seconds).
2. Sample Size Used for the Test Set and Data Provenance:
This document does not describe a clinical study with a "test set" of patient data for evaluating an AI/ML algorithm. The "testing" mentioned is "Non-Clinical Performance Testing." This likely refers to in-vitro or bench testing, and potentially animal studies, to validate the device's physical performance, safety, and the effects of the new ablation parameters on tissue.
- Sample Size: Not explicitly stated, as it's non-clinical performance testing, not a clinical trial with a patient test set.
- Data Provenance: Not applicable as it refers to non-clinical testing, not patient data (e.g., retrospective/prospective).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. The ground truth for a hardware device's performance is typically established through engineering specifications, physical measurements, and biological response studies, not expert consensus on medical images or diagnoses.
4. Adjudication Method for the Test Set:
Not applicable, as there is no "test set" in the context of expert review of data for AI/ML performance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
Not applicable. This is a hardware device submission, not an AI/ML algorithm where MRMC studies are relevant for assessing human reader performance with and without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:
Not applicable. This is a standalone device in the sense that it performs a physical action (ablation), but not a standalone AI algorithm generating an output without human intervention for diagnostic or clinical decision-making purposes.
7. The Type of Ground Truth Used:
The "ground truth" for this device's performance is based on non-clinical performance testing. This would involve:
- Engineering specifications and measurements: Ensuring the device delivers RF energy as expected.
- Pre-clinical (in-vitro/ex-vivo/animal) studies: Demonstrating the desired biological effect (ablation leading to cellular necrosis) and safety profile in relevant tissues using the specified parameters.
- Risk analysis: Evaluating potential increases in risk profile with the new ablation parameters.
8. The Sample Size for the Training Set:
Not applicable. This is not an AI/ML model that undergoes a "training set" process.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as it's not an AI/ML model with a training set.
Ask a specific question about this device
(74 days)
Re: K203066
Trade/Device Name: COOLIEF Cooled Radiofrequency Kit Advanced Regulation Number: 21 CFR 882.4725
|
| Regulation Number | 21 CFR §882.4725
Classification Name | Probe, Radiofrequency Lesion |
| Regulation Section | 21 CFR 882.4725
Classification Name | Probe, Radiofrequency Lesion |
| Regulation Section | 21 CFR 882.4725
The COOLIEF* Cooled Radiofrequency Kit Advanced is to be used in conjunction with the Radiofrequency (RF) Pain Management Generator (PMG-ADVANCED) or the COOLIEF* Radiofrequency Generator (CRG-Advanced) to create lesions in nervous tissue. The device includes a fluid delivery system for commonly used fluid agents limited to contrast medium, saline, and/or anesthetic solution delivery at the target site. This device is also indicated for creating radiofrequency lesions of the genicular nerves for the management of moderate to severe knee pain of more than 6 months with conservative therapy, including medication, in patients with radiologically-confirmed ostecarthritis (grade 2-4) and a positive response (≥50% reduction in pain) to a diagnostic genicular nerve block.
The COOLIEF® SINERGY* Cooled Radiofrequency Kit Advanced is to be used in conjunction with the Radiofrequency (RF) Pain Management Generator (PMG-ADV ANCED) or the COOLIEF* Radiofrequency Generator (CRG-ADVAN CED) to create RF lesions in nervous tissue. The device includes a fluid delivery system for commonly used fluid agents limited to contrast medium, saline, and/or anesthetic solution delivery at the target site.
The COOLIEF* Cooled Radiofrequency Kit Advanced is a modification to the predicate COOLIEF* Cooled Radiofrequency Kit cleared under K163461. The subject device consists of the following:
- COOLIEF* Cooled Radiofrequency Probe Advanced
- COOLIEF* Cooled Radiofrequency Fluid Tubing Kit ●
- COOLIEF* Cooled Radiofrequency Fluid Delivery Introducer ●
The modifications to the subject device are specific to the probe and fluid tubing kit.
The subject probe and fluid tubing kit are also included in the COOLIEFSINERGY Cooled Radiofrequency Kit Advanced.
The subject device has several configurations based on the anatomic region of use, length of the introducer and probe, and the length of the active electrode. The COOLIEF* Cooled Radiofrequency Kit Advanced is composed of fluid delivery introducers, probes, and fluid tubing kit. The kit is sterile, non-pyrogenic, and single-use. The COOLIEF* Cooled Radiofrequency Kit Advanced is to be used in conjunction with the Radiofrequency (RF) Pain Management Generator (PMG-ADVANCED) or the COOLIEF* Radiofrequency Generator (CRG-Advanced) to create lesions in nervous tissue.
The COOLIEF* SINERGY* Cooled Radiofrequency Kit Advanced is to be used in conjunction with the Radiofrequency (RF) Pain Management Generator (PMG-ADVANCED) or the COOLIEF* Radiofrequency Generator (CRG-ADVANCED) to create RF lesions in nervous tissue. This device includes the same components as the COOLIEF* Cooled RF Kit Advanced, but also includes an Epsilon The Epsilon Ruler is a circular stainless-steel ruler that may be used during the COOLIEF Ruler. SINERGY* procedure to assist the user in providing a 'template' around the foramen, which is specific to performing procedure around the sacroiliac (SI) nerve.
This document is a 510(k) summary for the COOLIEF® Cooled Radiofrequency Kit Advanced, a medical device for creating lesions in nervous tissue. It details the device's characteristics, comparison to predicate devices, and non-clinical testing results.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Name / Description | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Sterilization and Shelf-Life (ANSI/AAMI/ISO 11135:2014) | Compliance with standard | Pass |
| Bacterial Endotoxins Testing (ST72:2019) | Compliance with standard | Pass |
| Human Factors (62366-1:2015) | Compliance with standard | Pass |
| Biocompatibility (10993-5: 2009, 10993-10: 2010, 10993-11: 2006) | Compliance with standards | Pass |
| Electromagnetic Compatibility and Electrical Safety (60601-1:2005+AM1 (2012), 60601-1-6: 2010+A1:2013, 60601-1-9:2013, 60601-2-2:2017) | Compliance with standards | Pass |
| Performance Verification: Lesion Size Testing | As intended | Pass |
| Performance Verification: Probe Thermocouple Accuracy Testing | As intended | Pass ($\pm$ 3°C as per Table 3) |
| Performance Verification: Probe Fluid Delivery Port Injection Force Testing | As intended | Pass |
| Performance Verification: Flow Rate Testing | As intended | Pass |
| Performance Verification: High Pressure Leak Testing | As intended | Pass |
| Performance Verification: Probe Mass Testing | As intended | Pass |
| Performance Verification: Structural Strength Testing | As intended | Pass |
| Performance Verification: Sterile Water and Saline Cooling Fluids Testing | As intended | Pass |
| Performance Verification: Packaging Verification (ANSI/AAMI/ISO 11607-1:2019) | Compliance with standard | Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states that clinical data was not applicable for the subject device and therefore was not provided. All testing mentioned is "non-clinical testing" or "bench performance testing." The sample sizes for these non-clinical tests are not specified in the provided text, nor is the provenance of the data beyond being "bench performance testing."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as there was no clinical study described that involved establishing a ground truth with experts. The stated tests are non-clinical, focusing on device performance and safety parameters.
4. Adjudication Method for the Test Set
This information is not applicable as there was no clinical study described. Adjudication methods are typically relevant for human-led assessments in clinical trials.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This information is not applicable. The device described is a radiofrequency kit, not an AI-powered diagnostic or assistive tool for human readers. There is no mention of AI or MRMC studies in the provided text.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This information is not applicable. The device is hardware (a radiofrequency kit) and not an algorithm.
7. The Type of Ground Truth Used
This information is not applicable as no clinical studies with a ground truth were conducted or described. The "ground truth" for the non-clinical tests would be the established engineering and safety standards and specifications to which the device was tested.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical medical device, not a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable. As stated above, the device does not involve a training set or ground truth in the context of an algorithm.
Ask a specific question about this device
(365 days)
Salt Lake City, Utah 84116
Re: K190259
Trade/Device Name: Vesta RF Cannula Regulation Number: 21 CFR 882.4725
Cannula, Radiofrequency Lesion Radiofrequency Lesion Probe Neurology Radiofrequency Lesion Probe 21 CFR 882.4725
|
| ClassificationRegulation | 21 CFR 882.4725
| 21 CFR 882.4725
The Vesta RF Cannula, in combination with an RF generator and probe, are intended for use in radiofrequency (RF) heat lesion procedures for relief of pain.
The Vesta RF Cannula consists of an insulated cannula with an active tip for use in radiofrequency heat lesion procedures for the relief of pain. It is designed to be used with compatible pain management generators and probes that have a maximum voltage rating less than or equal to 280 V. The Vesta RF Cannula is sterilized and intended for single use only.
The information provided is about the Biomerics Vesta RF Cannula, cleared under K190259, which is a medical device and not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device studies (such as sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training data) are not applicable.
However, I can provide the acceptance criteria and study information that was conducted for this device. The regulatory submission primarily relies on bench testing and biocompatibility testing to demonstrate substantial equivalence to a predicate device.
Acceptance Criteria and Device Performance (Not applicable for AI/ML device studies)
The device, the Vesta RF Cannula, is a non-AI medical device. Its performance is demonstrated through meeting various industry standards for physical characteristics, sterilization, and biocompatibility, rather than metrics like sensitivity, specificity, or AUC as would be seen in an AI study.
Table of Acceptance Criteria and Reported Device Performance:
| Standard/Test | Acceptance Criteria (Implied by standard compliance) | Reported Device Performance |
|---|---|---|
| Bench Testing | ||
| ISO 6009 (Needle Diameter) | Compliance with ISO 6009 standard | Passed |
| ISO 9626 (Cannula Verification) | Compliance with ISO 9626 standard | Passed |
| ISO 7864 (Needle Tip Inspection) | Compliance with ISO 7864 standard | Passed |
| ISO 80369-7 (Luer Acceptance) | Compliance with ISO 80369-7 standard | Passed |
| ISO 15223-1 (Label Proof Approval) | Compliance with ISO 15223-1 standard | Passed |
| ISO 11135, ISO 10993-7, ISO 11138-2 (Sterilization Assessment) | Compliance with relevant ISO standards for sterilization and residuals | Passed |
| ISO 11607-1 (Stability and Aging Assessment) | Compliance with ISO 11607-1 standard for packaging and shelf-life | Passed |
| Biocompatibility Testing | ||
| Irritation – Intracutaneous (Pooled Blue/Green Hub) | No significantly greater biological reaction than control | Non-irritant |
| Sensitization – Kligman Maximization (Pooled Blue/Green Hub) | No reaction at challenge (0% sensitization) | Non-sensitizer |
| Cytotoxicity – MEM Elution (Blue Hub) | No reactivity (Grade 0) | Non-cytotoxic |
| Cytotoxicity – MEM Elution (Green Hub) | No reactivity (Grade 0) | Non-cytotoxic |
| Systemic Toxicity (Blue Hub) | No clinical signs consistent with toxicity | Non-toxic |
| Systemic Toxicity (Green Hub) | No clinical signs consistent with toxicity | Non-toxic |
| EO Residuals | Below allowable limits established in ISO 10993-7 Section 4.3 for limited exposure device (≤24 hours) | EO sterilization is appropriate |
| Material Mediated Pyrogen | Pyrogenicity risk assessed and found acceptable | Non-pyrogenic |
Since this is a non-AI/ML device, the following points are largely inapplicable or have no information provided in the given document.
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- Test Set Sample Size: Not applicable for an AI/ML test set. For bench and biocompatibility testing, the sample sizes are determined by the respective ISO or test standards. These typically involve a small number of samples for physical and chemical tests (e.g., a few units per test).
- Data Provenance: Not applicable. The tests are bench and lab-based studies performed to evaluate device characteristics, not clinical data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable as this is a non-AI/ML device. Ground truth is established by objective measurements against engineering specifications and industry standards for physical and chemical properties.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable as this is a non-AI/ML device. Performance is determined by objective measurements and standard compliance.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an RF cannula, not an AI-powered diagnostic or assistive tool. No MRMC study was performed.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For this device, the "ground truth" for its safety and effectiveness is established by compliance with recognized international standards (ISO, IEC, ASTM) for material properties, sterilization, physical dimensions, and biocompatibility. This involves objective measurements and qualitative assessments (e.g., "non-irritant," "non-cytotoxic") based on established laboratory protocols.
-
The sample size for the training set:
- Not applicable. This is a non-AI/ML device, so there is no "training set."
-
How the ground truth for the training set was established:
- Not applicable. This is a non-AI/ML device.