K Number
K211367

Validate with FDA (Live)

Device Name
Evo sEEG System
Date Cleared
2021-09-01

(120 days)

Product Code
Regulation Number
882.1330
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Evo sEEG System is intended for temporary (less than 24 hours) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

Device Description

The NeuroOne Evo sEEG System is comprised of the Evo sEEG Electrodes (which includes Anchor Bolt Cap and Stylet) accompanying Cable Assemblies, and the Evo Anchor Bolts. The Evo sEEG System components are provided sterile and are single use only. The Evo sEEG Electrodes are temporarily placed (less than 24 hours) at the subsurface level of the brain. The system is designed to be used in stereoelectroencephalography (sEG) procedures. The sEEG electrodes, by way of the accompanying cable assemblies, are connected to recording, monitoring, or stimulation/response instrumentation currently commercially available for use with other sEEG electrodes, including the predicate device. The sEEG electrodes are a 0.8 mm diameter polyimide electrode with platinum contacts and are available in varying numbers of contacts: 5 to 16 contacts, with an exploration length (recording depth) from 16 to 80 mm. The contact height is 2.0 mm and spacing is 1.5-3.2 mm apart. The Evo Anchor Bolts are available in 20 mm, 25 mm, 30 mm, and 35 mm lengths, can only be used and placed through a small 2.1 mm burr hole drilled in the skull, and should be used only when sEEG depth electrodes are warranted.

AI/ML Overview

The NeuroOne Evo sEEG System, a depth electrode for temporary brain electrical signal recording, monitoring, and stimulation, has been deemed substantially equivalent to its predicate device, the DIXI Medical Microdeep Depth Electrode. This determination is based on a comprehensive set of performance tests.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list specific quantitative acceptance criteria for each test. Instead, for all "Non-Clinical Performance Tests," "Biocompatibility" tests, and a "Simulated Use and Implantation Accuracy Study," the reported performance is a concise "Pass – All samples passed the acceptance criteria" or similar statements like "Pass - Non-cytotoxic," "Pass - Did not elicit a sensitization response," etc. This indicates that predefined criteria were met for each test, but the exact numerical thresholds are not provided in this summary.

Summary of Performance Testing and Results:

Test CategoryTest NameReported Device Performance
Non-Clinical Performance Tests
DimensionalElectrode Outer DiameterPass – All samples passed the acceptance criteria
Electrode Working LengthPass – All samples passed the acceptance criteria
Electrode Recording DepthPass – All samples passed the acceptance criteria
Electrode Contact SpacingPass – All samples passed the acceptance criteria
Electrode Contact SizePass – All samples passed the acceptance criteria
Electrode TailPass – All samples passed the acceptance criteria
Anchor Bolt Compatibility – Outer DiameterPass – All samples passed the acceptance criteria
Mechanical Performance - ImplantationAnchor Bolt Placement TorquePass – All samples passed the acceptance criteria
Anchor Bolt Removal TorquePass – All samples passed the acceptance criteria
Electrode Depth SettingPass – All samples passed the acceptance criteria
Electrode Depth Setting - Sliding ForcePass – All samples passed the acceptance criteria
Electrode Through Anchor BoltPass – All samples passed the acceptance criteria
Electrode Stylet RemovalPass – All samples passed the acceptance criteria
Anchor Bolt Cap TorquePass – All samples passed the acceptance criteria
Anchor Bolt Cap Torque MovementPass – All samples passed the acceptance criteria
Electrode MigrationPass – All samples passed the acceptance criteria
Electrochemical PerformanceImpedancePass – All samples passed the acceptance criteria
ReliabilityPass – All samples passed the acceptance criteria
StimulationPass – All samples passed the acceptance criteria
DetectionPass – All samples passed the acceptance criteria
Kink ResistancePass – All samples passed the acceptance criteria
Mechanical Performance – MonitoringElectrode RigidityPass – All samples passed the acceptance criteria
Anchor Bolt Cap TorquePass – All samples passed the acceptance criteria
Anchor Bolt Cap SealingPass – All samples passed the acceptance criteria
Electrode FlexibilityPass – All samples passed the acceptance criteria
Mechanical IntegrityAnchor Bolt Compatibility – TorquePass – All samples passed the acceptance criteria
Anchor Bolt RetentionPass – All samples passed the acceptance criteria
Electrode IntegrityPass – All samples passed the acceptance criteria
Anchor Bolt Cap IntegrityPass – All samples passed the acceptance criteria
Guiding Stylet IntegrityPass – All samples passed the acceptance criteria
Electrical SafetyIEC 60601-1 requirementsPass – All samples passed the acceptance criteria
PackagingISTA 3A and ASTM D-4169; DC13; AL1Pass – All samples passed the acceptance criteria
Shelf-LifeSimulated storage conditionsPass – All samples passed the acceptance criteria
SterilizationMinimum SAL of 10-6 (Ethylene Oxide)Pass – All samples passed the acceptance criteria
BiocompatibilityCytotoxicity (ISO MEM Elution)Pass - Non-cytotoxic
Sensitization (ISO Guinea Pig Maximization)Pass - Did not elicit a sensitization response
Irritation (ISO Intracutaneous Irritation)Pass - Requirements of the ISO Intracutaneous Reactivity Test have been met
Acute Systemic Toxicity (ISO Acute Systemic)Pass - The ISO Acute Systemic Injection Test have been met
Material Mediated Pyrogenicity (ISO Rabbit Pyrogen)Pass - Non-pyrogenetic
Implantation (Ovine brain, Rabbit tibia)Pass – Minimal or no reaction
Hemolysis (ASTM Hemolysis)Pass - Non-hemolytic
Simulated Use & Implantation AccuracyHuman cadaver studyNo significant difference between subject and predicate implantation accuracy

2. Sample Size Used for the Test Set and Data Provenance:

  • Non-Clinical Performance, Biocompatibility, Sterility, Shelf-Life, and Packaging Tests: The document consistently states "All samples passed the acceptance criteria" for these tests, implying multiple samples were tested for each. However, the exact sample sizes for each individual test are not specified.
  • Simulated Use and Implantation Accuracy Study: The sample size is not explicitly stated, but it was a "human cadaver study." The data provenance is implied to be from a controlled laboratory setting (conducted by NeuroOne Medical Technologies Corp.) rather than clinical data from a specific country, and it is a prospective study (testing the device).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

  • Non-Clinical Performance, Biocompatibility, Sterility, Shelf-Life, and Packaging Tests: These types of tests typically rely on established scientific and engineering standards and validated methodologies rather than expert consensus for ground truth. The "ground truth" would be the fulfillment of the specified technical criteria within the standards. Therefore, an explicit number of experts and their qualifications for establishing ground truth as one might see in a clinical algorithm study is not applicable here.
  • Simulated Use and Implantation Accuracy Study: The document does not specify the number or qualifications of experts used to establish the ground truth for implantation accuracy in the human cadaver study. The comparison was made against the predicate device, implying that accuracy was measured objectively according to predefined metrics.

4. Adjudication Method for the Test Set:

  • For the non-clinical tests, the adjudication method is via comparison to predefined acceptance criteria within established standards (e.g., ISO, ASTM, IEC). This is typically an objective pass/fail assessment rather than a human adjudication process.
  • For the "Simulated Use and Implantation Accuracy Study," the adjudication method is not explicitly stated. The conclusion that "no significant difference" was found implies a statistical comparison of measured implantation accuracy between the subject and predicate devices. No human adjudication process (like 2+1 or 3+1) is mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was mentioned. The study focused on demonstrating substantial equivalence through technical and performance characteristics, and a simulated use study comparing the subject device to a predicate, not on human reader performance with or without AI assistance. The device itself is an implantable electrode, not an AI-powered diagnostic tool.

6. Standalone (Algorithm Only) Performance Study:

This concept is not relevant to the Evo sEEG System, as it is a physical medical device (electrode) and not an algorithm or software-only device. Performance studies were conducted on the device's physical and electrical characteristics.

7. Type of Ground Truth Used:

  • Non-Clinical Performance, Biocompatibility, Sterility, Shelf-Life, and Packaging Tests: The ground truth for these tests is based on adherence to established international and national standards (e.g., ISO, ASTM, IEC) and internal specifications. This is objective and measurable data compared against predefined thresholds.
  • Simulated Use and Implantation Accuracy Study: The ground truth for implantation accuracy would be derived from objective measurements taken within the cadaver study, comparing the planned trajectory/placement with the actual placement, likely using imaging modalities.

8. Sample Size for the Training Set:

The concept of a "training set" is not applicable as this is a physical medical device, not a machine learning algorithm that requires training data.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable, as there is no training set for this device.

{0}------------------------------------------------

September 1, 2021

Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

NeuroOne Medical Technologies Corp. % Joseph Ostendorf Manager and Principal Consultant Ostendorf Consulting, LLC 23879 Blue Spruce Road Sauk Centre, Minnesota 56378

Re: K211367

Trade/Device Name: Evo sEEG System Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: July 30, 2021 Received: August 2, 2021

Dear Joseph Ostendorf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211367

Device Name Evo sEEG System

Indications for Use (Describe)

The Evo sEEG System is intended for temporary (less than 24 hours) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

NeuroOne

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number:K211367
Date Prepared:September 01, 2021
Applicant:NeuroOne Medical Technologies Corp.7599 Anagram DriveEden Prairie, MN 55344Phone: (952) 426-1383E-mail: info@n1mtc.com
Contact Person:Joseph OstendorfRegulatory Affairs Consultant7599 Anagram DriveEden Prairie, MN 55344

SUBJECT DEVICE

Trade/Device Name:Evo sEEG System
Device Regulation Number:21 CFR§882.1330
Device / Regulation Name:Electrode, Depth
Regulation Description /Common Name:Depth electrode
Product Code:GZL
Device Class / RegulationClassification:Class II

{4}------------------------------------------------

DEVICE DESCRIPTION (For the Device Subject to this 510(k) Premarket Notification)

The NeuroOne Evo sEEG System is comprised of the Evo sEEG Electrodes (which includes Anchor Bolt Cap and Stylet) accompanying Cable Assemblies, and the Evo Anchor Bolts. The Evo sEEG System components are provided sterile and are single use only. The Evo sEEG Electrodes are temporarily placed (less than 24 hours) at the subsurface level of the brain. The system is designed to be used in stereoelectroencephalography (sEG) procedures.

The sEEG electrodes, by way of the accompanying cable assemblies, are connected to recording, monitoring, or stimulation/response instrumentation currently commercially available for use with other sEEG electrodes, including the predicate device.

The sEEG electrodes are a 0.8 mm diameter polyimide electrode with platinum contacts and are available in varying numbers of contacts: 5 to 16 contacts, with an exploration length (recording depth) from 16 to 80 mm. The contact height is 2.0 mm and spacing is 1.5-3.2 mm apart.

The Evo Anchor Bolts are available in 20 mm, 25 mm, 30 mm, and 35 mm lengths, can only be used and placed through a small 2.1 mm burr hole drilled in the skull, and should be used only when sEEG depth electrodes are warranted.

INDICATIONS FOR USE (For the Device Subject to this 510(k) Premarket Notification)

The Evo sEEG System is intended for temporary (less than 24 hours) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

DEVICE CLASSIFICATION, INTENDED USE/INDICATIONS FOR USE, AND TECHNOLOGICAL AND PERFORMANCE CHARACTERISTICS COMPARISONS

The following tables provide a side-by-side comparison of the Evo sEEG System to the predicate device to support this pre-market notification.

Device Classification Comparison
Subject DeviceK211367Predicate DeviceK170959
Trade/Device Name:Evo sEEG SystemDIXI Medical Microdeep DepthElectrode
Device Regulation Number:21 CFR§882.133021 CFR§882.1330
Device / Regulation Name:Electrode, DepthElectrode, Depth
Regulation Description /Common Name:Depth electrodeDepth electrode
Product Code:GZLGZL
Device Class / RegulationClassification:Class IIClass II

{5}------------------------------------------------

Intended Use/Indications for Use Comparison
Subject DeviceK211367Predicate DeviceK170959
IntendedUseA depth electrode is an electrode used fortemporary stimulation of, or recordingelectrical signals at, subsurface levels of thebrain.A depth electrode is an electrode used fortemporary stimulation of, or recordingelectrical signals at, subsurface levels of thebrain.
Indicationsfor UseThe Evo sEEG System is intended fortemporary (less than 24 hours) use withrecording, monitoring, and stimulationequipment for the recording, monitoring, andstimulation of electrical signals at thesubsurface level of the brain.The DIXI Medical Microdeep DepthElectrodes are intended for temporary (<30day) use with recording, monitoring andstimulation equipment for the recording,monitoring and stimulation of electricalsignals at the subsurface level of the brain.
Electrode Technological and Performance Characteristics Comparison
Subject DeviceK211367Predicate DeviceK170959
ManufacturerNeuroOne Medical TechnologiesCorp.DIXI Medical
Electrode Diameter0.8 mm0.8 mm
Electrode MaterialPolyimidePolyamide
Number of Contacts5 - 165 - 18
Contact Height/Length2 mm2 mm
Contact Spacing1.5 – 3.2 mm1.5 - 11 mm
Contact Material(s)PlatinumPlatinum/iridium
Exploration Length(Recording Depth)16 - 80 mm16 - 80.5 mm
Single patient use, DisposableYesYes
Provided SterileYesYes
Environment of UseIntraoperative and Neurologicalmonitoring locations at subsurfacelevels of the brainIntraoperative and Neurologicalmonitoring locations at subsurfacelevels of the brain
Duration of UseLess than 24 hoursLess than 30 days
Principles of OperationDepth electrodes are temporarilyplaced at subsurface levels of thebrain to record, monitor, andstimulate electrical signalsDepth electrodes are temporarilyplaced at subsurface levels of thebrain to record, monitor, andstimulate electrical signals

{6}------------------------------------------------

Anchor Bolt Technological and Performance Characteristics Comparison
Subject DeviceK211367Predicate DeviceK170959
ManufacturerNeuroOne Medical TechnologiesCorp.DIXI Medical
Depth Electrode DiameterCompatibility0.8 mm0.8 mm
Material(s)TitaniumTitanium
Lengths Offered20 - 35 mm15 - 35 mm
Anchor Bolt ID0.9 mm0.9 mm
Single patient use, DisposableYesYes
Provided SterileYesYes
Environment of UseIntraoperative and Neurologicalmonitoring locationsIntraoperative and Neurologicalmonitoring locations
Duration of UseLess than 24 hoursLess than 30 days
Principles of OperationThreaded into a pre-drilled hole inthe skull.Threaded into a pre-drilled hole inthe skull.

{7}------------------------------------------------

SIMILIARITIES AND DIFFERENCES TO THE PREDICATE (For the Device Subject to this 510(k) Premarket Notification)

There are similarities and differences between the subject and predicate devices. The NeuroOne Evo sEEG System and the DIXI Medical Microdeep Depth Electrode devices have the same Device Regulation Numbers, Device / Regulation Names, Regulation / Common Names, Product Codes, and Device Class / Regulation Classifications.

Both devices have identical Intended Uses and similar Indications for Use.

The devices further possess the same Electrode Diameters. The NeuroOne Evo sEEG System has a similar material as the DIXI Medical Microdeep Depth Electrode as both utilize polymer materials for the electrode body and for their insulative properties. The NeuroOne Evo sEEG System has a similar Number of Contacts and is within the range of contacts offered for the DIXI Medical Microdeep Depth Electrodes. Both devices share the same electrode Contact Height/Length. The NeuroOne Evo sEEG System has similar Contact Spacing and utilizes variable contact spacing across the device size configurations whereas the DIXI Medical Microdeep Depth Electrodes utilizes the same Contact Spacing across all device size configurations. The devices are similar in both utilizing platinum contacts, although the DIXI Medical Microdeep Depth Electrodes have platinum/iridium contacts. The NeuroOne Evo sEEG System has a similar Exploration Length (Recording Depth) and is within the range of Exploration Length (Recording Depth) offered for the DIXI Medical Microdeep Depth Electrodes. Both devices are Single Patient Use, Disposable devices, which are provided Sterile. They both are used in the same Environment of Use and share similar Durations of Use. And finally, they both have the very same principles of operation.

The NeuroOne Evo Anchor Bolts and the DIXI Medical Microdeep Depth Electrode anchor bolts have the same Depth Electrode Diameter Compatibility and Material(s). The anchor bolts Lengths Offered with the NeuroOne Evo Anchor Bolts are similar to the DIXI Medical Microdeen Depth Electrode anchor bolts and are within the range offered by the DIXI Medical Microdeep Depth Electrode anchor bolts. The Evo Anchor Bolts have the same Anchor Bolt IDs as the DIXI Medical Microdeep Depth Electrode anchor bolts. Both devices are Single Patient use, Disposable devices, which are provided Sterile. They both are used in the same Environment of Use and share similar Durations of Use. And finally, they both have the very same Principles of Operation.

Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device. A comparison was made against the predicate where data was available. The Evo sEEG System has undergone bench, electrical, biocompatibility, packaging, and sterilization testing to demonstrate the differences in the sizes and materials do not raise questions of safety or efficacy.

{8}------------------------------------------------

SUMMARY OF PERFORMANCE TESTING AND STANDARDS

Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device. A comparison was made against the predicate where data was available. The Evo sEEG System has undergone non-clinical performance, biocompatibility, sterility, shelf-life, and packaging to demonstrate that the results do not raise questions of safety or efficacy.

Non-Clinical Performance Tests
TestTest Method SummaryResults and Conclusions
DimensionalElectrode Outer DiameterPass – All samples passed theacceptance criteria
Electrode Working LengthPass – All samples passed theacceptance criteria
Electrode Recording DepthPass – All samples passed theacceptance criteria
Electrode Contact SpacingPass – All samples passed theacceptance criteria
Electrode Contact SizePass – All samples passed theacceptance criteria
Electrode TailPass - All samples passed theacceptance criteria
Anchor Bolt Compatibility –Outer DiameterPass – All samples passed theacceptance criteria
Mechanical Performance -ImplantationAnchor Bolt Placement TorquePass – All samples passed theacceptance criteria
Anchor Bolt Removal TorquePass – All samples passed theacceptance criteria
Electrode Depth SettingPass – All samples passed theacceptance criteria
Electrode Depth Setting - SlidingForcePass – All samples passed theacceptance criteria
Electrode Through Anchor BoltPass – All samples passed theacceptance criteria
Electrode Stylet RemovalPass – All samples passed theacceptance criteria
Anchor Bolt Cap TorquePass – All samples passed theacceptance criteria
Anchor Bolt Cap TorqueMovementPass – All samples passed theacceptance criteria
Electrode MigrationPass – All samples passed theacceptance criteria
Electrochemical PerformanceImpedancePass – All samples passed theacceptance criteria
ReliabilityPass - All samples passed theacceptance criteria
StimulationPass - All samples passed theacceptance criteria
DetectionPass - All samples passed theacceptance criteria
Kink ResistancePass - All samples passed theacceptance criteria
Mechanical Performance –MonitoringElectrode RigidityPass – All samples passed theacceptance criteria
Anchor Bolt Cap TorquePass – All samples passed theacceptance criteria
Anchor Bolt Cap SealingPass – All samples passed theacceptance criteria
Electrode FlexibilityPass – All samples passed theacceptance criteria
Mechanical IntegrityAnchor Bolt Compatibility –TorquePass – All samples passed theacceptance criteria
Anchor Bolt RetentionPass – All samples passed theacceptance criteria
Electrode IntegrityPass – All samples passed theacceptance criteria
Anchor Bolt Cap IntegrityPass – All samples passed theacceptance criteria
Guiding Stylet IntegrityPass – All samples passed theacceptance criteria
Electrical SafetyProduct shall meet the applicablerequirements of the IEC 60601-1standard.Pass – All samples passed theacceptance criteria
PackagingThe packaged device and labelingshall withstand the conditions ofISTA 3A and ASTM D-4169;DC13; AL1 without loss offunction, sterility, or legibility.Pass – All samples passed theacceptance criteria
Shelf-LifeThe packaged device and labelingshall withstand simulated storageconditions without loss offunction, sterility, or legibility.Pass – All samples passed theacceptance criteria
SterilizationThe sterilization process shall bevalidated to demonstrate aminimum of SAL of 10-6 for theproduct using Ethylene OxidePass – All samples passed theacceptance criteria

{9}------------------------------------------------

Biocompatibility
TestTest SummaryConclusions
CytotoxicityISO MEM Elution Using MouseFibroblast Cells - 72 HourExtraction (GLP)Pass - Non-cytotoxic
SensitizationISO Guinea Pig MaximizationSensitization(GLP - 2 Extracts)Pass - Did not elicit a sensitizationresponse
IrritationISO Intracutaneous Irritation(GLP - 2 Extracts)Pass - The requirements of theISO Intracutaneous ReactivityTest have been met
Acute Systemic ToxicityISO Acute Systemic Toxicity(GLP - 2 Extracts)Pass - The ISO Acute SystemicInjection Test have been met
Material Mediated PyrogenicityISO Materials Mediated RabbitPyrogen(GLP)Pass - Non-pyrogenetic

{10}------------------------------------------------

ImplantationProvide general information onthe health hazards likely to arisefrom continuous exposure of thedevice for local tissue/boneresponse to implantation sites inboth an ovine brain implantationstudy (including assessment forneurotoxicity) and rabbit tibiastudy (Evo sEEG anchor boltonly).Pass – Minimal or no reaction
HemolysisASTM Hemolysis - DirectContact and Extract Method(GLP)Pass - Non-hemolytic

Simulated Use and Implantation Accuracy Study

Results and Conclusions

A human cadaver study was conducted by NeuroOne Medical Technologies Corp. with the purpose of demonstrating the simulated use of the Evo sEEG System, and its implantation accuracy, when implanted according to its Instructions For Use (IFU). The implantation accuracy of the Evo sEEG System was compared to the implantation accuracy of the predicate DIXI Medical Microdeep Depth Electrode (K170959). The results demonstrated no significant difference between the implantation accuracy of the subject and predicate devices and therefore these devices are substantially equivalent when evaluating implantation accuracy.

BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The Intended Use, Fundamental Scientific Technology, and Principles of Operation for the Evo sEEG System are the same as those described for the predicate device. The Evo sEEG System does not raise new questions regarding safety and effectiveness when compared to the predicate device and has been determined by NeuroOne Medical Technologies Corp. to be substantially equivalent.

In summary, the Evo sEEG System has the following similarities to the predicate device, which has previously received 510(k) clearance:

  • . Has the same device classification
  • Has the same intended use ●
  • Has the similar indications for use
  • Has the same environment of use ●
  • Uses similar technological characteristics
  • . Uses the same principles of operation
  • . Uses the same sterilization methodology
  • Is biocompatible for its intended use

Therefore, the conclusions drawn from the non-clinical tests (biocompatibility) demonstrate the device is as safe, as effective, and performs as well as the legally marketed predicate device, per 21 CFR 807.92(b)(3). The Evo sEEG System is substantially equivalent to the predicate device.

SUBSTANTIAL EQUIVALENCE CONCLUSIONS

Through the thorough comparison of technological and performance characteristics, the subject device is determined to be substantially equivalent to the predicate device.

§ 882.1330 Depth electrode.

(a)
Identification. A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.(b)
Classification. Class II (performance standards).