    510k Summary Text (Full-text Search) :

    Trade/Device Name: Velocity Alpha® MR High Speed Surgical Drill System
    Regulation Number: 21 CFR 882.4370
    Trade/Device Name: Velocity Alpha® MR High Speed Surgical Drill System
    Regulation Number: 21 CFR 882.4370
    Classification Name:** 21 CFR 882.4370 - Pneumatic cranial drill motor (Primary)
    21 CFR 882.4310 -
    Manufacturer | Stryker Corporation | adeor medical AG |
    | K# | K061725 | K253772 |
    | Regulation | 21 CFR 882.4370
    ; Pneumatic cranial drill motor | 21 CFR 882.4370; Pneumatic cranial drill motor21 CFR 882.4310;

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Velocity Alpha MR High Speed Surgical Drill System is indicated for trephination, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone and biomaterials in or near a magnetic field of 3.0 Tesla or less for use in Neurosurgery.

    Device Description

    The Velocity Alpha MR High Speed Surgical Drill System is a pneumatic powered surgical motor drill system. The system supplies pressurized air to motor which powers rotary surgical instruments through the attachments. Rotational speed of the system is controlled by the foot pedal. The MR System consists of a motor, motor hose, two attachments, and 4 cutting instruments. Additionally, the system is compatible with the Velocity Alpha Pneumatic High Speed Surgical Drill System foot pedal and wall hoses. The Velocity Alpha MR High Speed Surgical Drill System, including both system-specific and system-compatible components, can be used in or near a magnetic field of 3.0 Tesla level or less.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233958

    Validate with FDA (Live)

    Device Name
    hekaDrill
    Manufacturer
    Date Cleared
    2024-03-14

    (90 days)

    Product Code
    Regulation Number
    874.4250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Handpiece (21 CFR 874.4250) Motor, Drill, Electric (21 CFR 882.4360) Motor, Drill, Pneumatic (21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hekaDrill system is intended for use in surgical procedures involving incision / cutting, removal, drilling and sawing of soft and hard tissue, bone and biomaterials. The system is specifically intended for use in Neurosurgical (Cranial including craniotomy); Ear, Nose and Throat (ENT), Maxillofacial, Orthopedic, Arthroscopic, Spinal, Stemotomy, and General Surgical Procedures.

    Device Description

    The hekaDrill System consists of electric and pneumatic drill handpieces, a power console, footswitches, attachments, connection cables, irrigation tube kits, cutting tools, and system accessories. The handpieces, attachments, and system accessories are provided non-sterile and are reusable. The console and footswitches are reusable and are not intended to be sterilised. The cutting tools and irrigation tube kits are provided sterile and are single use.

    AI/ML Overview

    The hekaDrill device primarily consists of various components for surgical procedures involving cutting, drilling, and sawing of soft and hard tissues. The performance assessment focused on comparing the device's cutting performance, electrical safety, electromagnetic compatibility, and biocompatibility with predicate devices and relevant standards.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Acceptance Criteria / Standard MetReported Device Performance
    General PerformanceEquivalent or better cutting performance compared to predicate drill system (vibration, noise, control, and performance)Cutting performance was equivalent or better to that of the predicate device.
    Electrical PerformanceElectrical safety according to IEC 60601-1:2005Instruments conform to IEC 60601-1:2005 for electrical safety.
    Electromagnetic compatibility according to IEC 60601-1-2:2014Instruments conform to IEC 60601-1-2:2014 for electromagnetic compatibility.
    BiocompatibilityNon-cytotoxic (L929 MEM Elution)No biological reactivity (Grade 0) of cells exposed to test article extract.
    Non-sensitizer (Kligman Maximization)Test article extracts elicited no reaction at challenge (0% sensitization).
    Non-irritant (Intracutaneous Injection)Test article sites did not show a significantly greater biological reaction than the control article.
    Non-toxic (Systemic Toxicity)Test article did not induce a significantly greater biological reaction than the control extracts.
    Non-pyrogenic (Pyrogenicity)Test article did not induce a pyrogenic response.
    Non-hemolytic (Indirect Hemolysis)Test article led to a hemolysis index above the negative control of 0.14%.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each performance test (e.g., number of cutting trials, number of biological samples). It also does not specify the country of origin for the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not mention the use of experts to establish a "ground truth" in the traditional sense for the performance tests described. The assessment relied on objective measurements and comparisons against established standards and predicate device performance. For the biocompatibility tests, the "ground truth" is determined by the specific reaction observed (e.g., cytotoxicity, sensitization), not expert consensus on an image or diagnosis.

    4. Adjudication Method for the Test Set:

    Not applicable in this context. The performance tests involve objective measurements and comparisons to standards, not subjective interpretations requiring adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The hekaDrill is a surgical drill system, not an AI-powered diagnostic or assistive device that would involve human readers or image interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable, as this is a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used:

    For the performance tests:

    • General Performance: The "ground truth" was established by comparing direct measurements of performance characteristics (vibration, noise, control, cutting effectiveness) against those same characteristics of a predicate device.
    • Electrical Performance: The "ground truth" was the adherence to established international safety standards (IEC 60601-1:2005 and IEC 60601-1-2:2014).
    • Biocompatibility: The "ground truth" was dictated by the specific biological responses observed in standardized tests, such as the absence of biological reactivity in cytotoxicity assays or no significant reaction in sensitization tests.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device. The concept of a "training set" is typically associated with machine learning or artificial intelligence models.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193630

    Validate with FDA (Live)

    Device Name
    hekaDrill
    Manufacturer
    Date Cleared
    2021-04-09

    (470 days)

    Product Code
    Regulation Number
    882.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification Name: Motor, Drill, Electric (21 CFR 882.4360) Motor, Drill, Pneumatic (21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hekaDrill system is intended for use in surgical procedures involving incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone. The system is specifically intended for use in Neurologic and general surgical procedures.

    Device Description

    The hekaDrill System consists of electric and pneumatic drill handpieces, a power console, footswitches, attachments, connection cables, irrigation tube kits, cutting tools, and system accessories. The handpieces, attachments, and system accessories are provided non-sterile and are reusable. The console and footswitches are reusable and are not intended to be sterilised. The cutting tools and irrigation tube kits are provided sterile and are single use.

    AI/ML Overview

    The provided text describes a 510(k) submission for the hekaDrill System. It includes details about performance testing, but it does not provide acceptance criteria in a quantitative format or a detailed study section with specific sample sizes, ground truth procedures, or expert involvement as requested.

    However, based on the limited information available in the "Discussion of the Performance Testing" table, I can extract and infer some aspects related to acceptance criteria and reported device performance.

    Here's an attempt to structure the answer based on the given information, with caveats about missing details:

    Acceptance Criteria and Supporting Study for hekaDrill System

    The K193630 submission for the hekaDrill System included performance testing to demonstrate the functionality and safety of the device compared to predicate devices. The following table summarizes the reported performance relative to general criteria.

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategorySpecific Criteria (Inferred from "Conclusion" column in source)Reported Device Performance
    General PerformanceCutting performance (vibration, noise, control) equivalent or superior to predicate.Cutting performance was equivalent or better to that of predicate device.
    Electrical PerformanceAdherence to IEC 60601-1:2005 for electrical safety.Instruments conform to IEC 60601-1:2005 for electrical safety.
    Electromagnetic CompatibilityAdherence to IEC 60601-1-2:2014 for electromagnetic compatibility.Instruments conform to IEC 60601-1-2:2014 for electromagnetic compatibility.
    BiocompatibilityNon-cytotoxic.Non-cytotoxic (No biological reactivity - Grade 0).
    Non-sensitizer (0% sensitization).Non-sensitizer (elicited no reaction at challenge).
    Non-irritant.Non-irritant (Test article sites did not show significantly greater biological reaction than control).
    Non-toxic (systemic).Non-toxic (Test article did not induce a significantly greater biological reaction than control extracts).
    Non-pyrogenic.Non-pyrogenic (Test article did not induce a Pyrogenic response).
    Non-hemolytic.Non-hemolytic (Test article led to a hemolysis index above the negative control of 0.14%).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the sample sizes used for any of the performance tests. For instance, for "Cutting performance," it does not state how many devices were tested or how many cutting tasks were performed. For biocompatibility tests, the number of samples or specimens analyzed is not provided.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that these are bench and lab-based performance tests for a new device submission, they would typically be prospective studies conducted in a controlled lab environment by Zethon Ltd (a UK company) or its contracted testing facilities.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The performance tests described (cutting performance comparison, electrical safety, EMC, biocompatibility) are primarily objective, quantitative measurements against established standards or predicate device performance, rather than requiring expert interpretation for ground truth.

    4. Adjudication method for the test set:

    • This information is not applicable/provided. The tests are objective measurements against standards or predicate performance, not subjective readings requiring adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable. The device, hekaDrill System, is an electric cranial drill motor, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study or assessment of AI assistance would not be relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable. As stated above, the hekaDrill System is a surgical instrument, not an algorithm.

    7. The type of ground truth used:

    • For the General Performance (Cutting performance), the ground truth was based on a comparison to a predicate drill system (Midas Rex Drill Systems), implying the predicate's performance served as a benchmark for "equivalent or better."
    • For Electrical Performance and Electromagnetic Compatibility, the ground truth was adherence to international consensus standards (IEC 60601-1:2005 and IEC 60601-1-2:2014, respectively).
    • For Biocompatibility, the ground truth was established by recognized biocompatibility testing methods (e.g., L929 MEM Elution for cytotoxicity, Kligman Maximization for sensitization, etc.) where specific biological responses or lack thereof (e.g., "Non-cytotoxic," "Non-sensitizer") define compliance.

    8. The sample size for the training set:

    • This information is not applicable/provided. The device is a surgical instrument, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • This information is not applicable. As there is no training set for a machine learning algorithm, there is no ground truth to establish for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202552

    Validate with FDA (Live)

    Date Cleared
    2020-11-02

    (60 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | 21 CFR 882.4360 (Motor,Drill, Electric)21CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midas Rex™ attachments and dissecting tools for Mazor are indicated for the incision/cutting, drilling, burring, and removal of hard tissue and bone in open and minimally invasive spine procedures.

    Computer-assisted surgery and its associated applications are intended as an aid for locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition for which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a long bone or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.

    Device Description

    Midas Rex™ attachments and surgical dissecting tools for Mazor are intended for the incision/cutting, drilling, burring, and removal of hard tissue and bone. The subject instruments are designed to be utilized through a cannula, for use with the Mazor X Stealth Edition™ system in open and minimally invasive spine procedures.

    The subject devices perform the intended function as part of the existing Medtronic Midas Rex TM surgical drill systems which consist of the subject attachments and dissecting tools, electric and pneumatic drill handpieces and system accessories.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study for the "Midas Rex Attachments and Dissecting Tools" which are intended for surgical navigation.

    Here's the detailed information based on your request:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mean Positional Error (mm): ≤ 2.0 mmMean Positional Error (mm): 1.12 mm
    Mean Trajectory Error (degrees): ≤ 2.0 degreesMean Trajectory Error (degrees): 0.37 degrees

    2. Sample size used for the test set and the data provenance

    The document indicates that "Navigational Accuracy Analysis" was performed, which "Provides confirmation that the subject devices satisfy the necessary navigational accuracy requirements." However, the specific sample size used for this test set is not explicitly mentioned in the provided text.

    Regarding data provenance, the study appears to be an internal performance verification study conducted by Medtronic Powered Surgical Solutions (MPSS) to demonstrate the device's adherence to specified accuracy metrics. The data would therefore be prospective in the context of this regulatory submission, generated specifically for the purpose of demonstrating substantial equivalence. The document doesn't specify the country of origin of the data beyond the manufacturer's location in Fort Worth, Texas.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the navigational accuracy analysis. The "Navigational Accuracy Analysis" is likely a technical measurement against a known reference, rather than an expert-adjudicated ground truth like in image interpretation studies.

    4. Adjudication method for the test set

    No adjudication method (e.g., 2+1, 3+1, none) is mentioned or implied for the navigational accuracy analysis. This type of test typically involves direct measurement against a known physical standard or simulated environment, rather than expert consensus on subjective interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No multi-reader multi-case (MRMC) comparative effectiveness study was done, as this device is a surgical tool with navigation capabilities, not an AI-assisted diagnostic or interpretive system involving human readers. The device itself provides assistance to the surgeon in locating anatomical structures.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation was conducted for the device's navigational accuracy. The "Navigational Accuracy Analysis" measured the device's performance (Mean Positional Error and Mean Trajectory Error) against predefined acceptance criteria, independent of a human operator actively engaged in a clinical task. This assessment is purely about the device's technical specifications.

    7. The type of ground truth used

    For the "Navigational Accuracy Analysis," the ground truth is based on technical specifications and precision measurements. The comparison is against a known, presumably highly accurate, reference or measurement standard that determines what constitutes "0 mm error" and "0 degrees error." It is not based on expert consensus, pathology, or outcomes data in the typical sense of a diagnostic device.

    8. The sample size for the training set

    The provided document describes a medical device (surgical tools with navigation) and its substantial equivalence submission, not a machine learning or AI model that requires a training set. Therefore, there is no mention of a "training set sample size."

    9. How the ground truth for the training set was established

    As mentioned above, this device is not an AI/ML model and therefore does not have a "training set" or a corresponding "ground truth for the training set" in the context of typical AI/ML development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191597

    Validate with FDA (Live)

    Date Cleared
    2019-11-01

    (137 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Depth Stop System: Electric cranial drill motor (21 CFR 882.4360) Pneumatic cranial drill motor (21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stealth Autoguide™ System is a positioning and guidance system intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments, based on a pre-operative plan and feedback from an image-guided navigation system with three-dimensional imaging software.
    The Stealth Autoguide™ System is a remotely-operated positioning and guidance system, indicated for any neurological condition in which the use of stereotactic surgery may be appropriate (for example, stereotactic EEG, laser tissue ablation, etc.).
    The Midas Rex™ Legend™ depth stop attachment and tools are incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.

    Device Description

    Stealth Autoguide™ System: The Stealth Autoguide System is a robotic positioning and guidance system intended to interpret navigation tracker coordinates and surgical plan coordinates from the StealthStation to robotically position and orient instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments to pre-defined plans.
    Midas Rex™ Legend™ Depth Stop System: The Midas Rex™ Legend™ Depth Stop System consists of a Depth Stop Attachment and specific surgical dissecting tools that will be used in conjunction with the Stealth Autoguide System to create cranial access holes for neurosurgical procedures.

    AI/ML Overview

    The provided text describes the Medtronic Stealth Autoguide System and Midas Rex Legend Depth Stop System. It includes information on performance testing for the Stealth Autoguide System, but lacks specific details on acceptance criteria and a study to prove the device meets all acceptance criteria in a comprehensive format. It also doesn't contain the requested information about training sets, expert ground truth development, MRMC studies, or standalone performance.

    However, based on the provided text, I can extract the following information concerning the performance testing for the Stealth Autoguide System's accuracy:

    Acceptance Criteria and Reported Device Performance for Stealth Autoguide™ System

    Acceptance CriterionReported Device Performance (Mean)Standard Deviation99% CI* Upper
    3D Positional Accuracy: Mean error ≤ 2.0 mm
        Biopsy Needle Accuracy Validation - StealthStation S70.92 mm0.47 mm3.03 mm
        Biopsy Needle Accuracy Validation - StealthStation S80.97 mm0.26 mm1.70 mm
        sEEG bolts/Visualase Accuracy Validation - StealthStation S71.50 mm0.68 mm3.08 mm
        sEEG bolts/Visualase Accuracy Validation - StealthStation S81.48 mm0.48 mm2.60 mm
    Trajectory Angle Accuracy: Mean error ≤ 2.0 degrees
        Biopsy Needle Accuracy Validation - StealthStation S71.22 degrees0.51 degrees2.41 degrees
        Biopsy Needle Accuracy Validation - StealthStation S80.59 degrees0.23 degrees1.11 degrees
        sEEG bolts/Visualase Accuracy Validation - StealthStation S71.04 degrees0.76 degrees2.81 degrees
        sEEG bolts/Visualase Accuracy Validation - StealthStation S80.42 degrees0.17 degrees0.82 degrees

    Details of the Accuracy Study:

    1. Sample size used for the test set and the data provenance: The document states that performance was determined using "overall end-to-end worst-case system level accuracy testing which incorporated clinically relevant anatomical phantoms." Further specifics about the sample size (e.g., number of phantoms, number of measurement points per phantom) and data provenance (e.g., country of origin, retrospective or prospective) are not provided in this document.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided in the document. The accuracy testing seems to be based on direct physical measurements against defined targets on phantoms rather than expert interpretation of images.

    3. Adjudication method for the test set: This information is not provided. Given the nature of the accuracy testing (physical measurements), traditional adjudication methods for image interpretation would likely not apply.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This information is not provided. The assessment described is a technical accuracy validation of the device's navigation and positioning capabilities, not a study involving human readers or AI assistance.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The "Stealth Autoguide™ System" is described as a "robotic positioning and guidance system" and the accuracy validation focuses on its "performance in 3D positional accuracy" and "trajectory angle accuracy." This implies standalone technical performance testing of the system's ability to achieve planned trajectories, before a human surgeon uses it to guide instruments. The system is designed to "robotically position and orient instrument holders or tool guides," suggesting its core function is algorithm-driven positioning. However, the evaluation here focuses on the accuracy of the guidance provided, which would then be utilized by a surgeon.

    6. The type of ground truth used: The ground truth for the accuracy study was established by defining "clinically relevant anatomical phantoms" and measuring the device's "performance in 3D positional accuracy" and "trajectory angle accuracy" against the known positions and trajectories on these phantoms. This is a phantom-based measurement ground truth.

    7. The sample size for the training set: This information is not provided. The document describes an accuracy validation study, not the development or training of an AI algorithm.

    8. How the ground truth for the training set was established: This information is not provided, as details about a training set are not included in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190172

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-06-14

    (133 days)

    Product Code
    Regulation Number
    882.4310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CFR 874.4250, Product code ERL) |
    | | Pneumatic cranial drill motor(21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otothinolaryngology; Craniofacial; Dental; and Endoscopic applications.

    The specific applications include Craniotomy, Laminotomy, Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.

    When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, splitting, fragmenting, extracting, removing, drilling, and reaming. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, smoothing or shaping of metal, and removing/rounding sharp edges.

    Device Description

    The Stryker Elite and HD Attachments are used within a system consisting of a variety of devices, including a console, powered motors, and cutting accessories. The attachments connect to the motors and the cutting accessories to complete the system for physician use. The Stryker Elite and HD Attachments are offered for prescription use only. The Elite and HD Attachments are intended to serve as interfaces between powered motors and cutting accessories for the purposes of:

    • Cutting bone, bone cement, and teeth;
    • Placing or cutting screws, metal, wires, pins, and other fixation devices; and
    • Providing a location for the user to hold and grip the device system. ●

    The Elite and HD Attachments are provided in straight and angled configurations. The subject Elite Attachments are offered in the following lengths: 7cm, 12cm, 17cm, and 20cm. The Stryker HD Attachments are offered in the following lengths: 9cm and 14cm. The Elite and HD Attachments are powered by, and compatible with, the Stryker electric and pneumatic motors. The attachments are also used with the cutting accessories (burs). All the attachments demonstrate a rotary mode of action by transmitting torque.

    The Elite and HD Attachments are made of stainless steel (SST). The attachments display a color band on the outer surface. The color bands serve to enhance the distinction of attachment and cutting accessory compatibility.

    The purpose of this submission is to gain clearance for colorant band modification for the Subject Device Attachments color band that exceeds the threshold as per FDA Guidance, "Deciding When to Submit a 510(k) for a Change to an Existing Device".

    AI/ML Overview

    The provided text is a 510(k) summary for the Stryker Elite and Heavy Duty (HD) Attachments, focusing on a colorant band modification. This document does not contain the detailed acceptance criteria or the study results for the original device's performance, but rather for the modification.

    Therefore, I cannot populate all sections of your request comprehensively. However, I can provide information based on what is available regarding the modification to the device.

    Here's a breakdown of the available information and what is missing:


    Description of Acceptance Criteria and Study for Colorant Band Modification

    The submission concerns a modification to the colorant band on the Stryker Elite and Heavy Duty (HD) Attachments. The primary purpose of this 510(k) is to gain clearance for this specific modification which "exceeds the threshold as per FDA Guidance, 'Deciding When to Submit a 510(k) for a Change to an Existing Device'". The underlying assumption is that the original device, without this specific color band modification, has already met its acceptance criteria through prior submissions (e.g., K143320).

    The study performed here specifically addresses the safety and effectiveness of the new colorant band material and colorant.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for Color Band Modification)Reported Device Performance (as stated in document)
    Safety: No new questions of safety raised"No new questions of safety and effectiveness were raised."
    Effectiveness: No new questions of effectiveness raised"No new questions of safety and effectiveness were raised."
    Performance Bench Testing: Met all pre-defined acceptance criteria"The Subject Device met all pre-defined acceptance criteria."
    Biocompatibility: Followed ISO 10993-1:2018 and FDA Guidance for limited exposure (< 24 hrs) externally communicating, tissue/bone/dentin devices"A biocompatibility evaluation was performed following the recommendations of ISO 10993-1: 2018 and FDA Guidance... as appropriate."
    Biocompatibility - Chemical Characterization: ICP-MSPerformed.
    Biocompatibility - Toxicological Risk AssessmentPerformed.
    ColorfastnessTested, results support substantial equivalence.
    DurabilityTested, results support substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated for bench testing, colorfastness, or durability tests for the specific modification. Biocompatibility testing was "adopted from K143320," implying reliance on prior data. The context suggests that the testing was performed on representative samples of the modified attachments.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, it is a prospective laboratory-based study for the modification.

    3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications

    • Not Applicable. This submission focuses on engineering bench testing and biocompatibility for a material change, not on diagnostic performance where expert ground truth would typically be established.

    4. Adjudication Method for the Test Set

    • Not Applicable. As above, this is for engineering and material testing, not clinical diagnosis.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This is not a diagnostic device or an AI-assisted device, so MRMC studies are not relevant. This is for a surgical instrument attachment.

    6. Standalone Performance (Algorithm Only without Human-in-the-loop)

    • Not Applicable. This device is a mechanical surgical attachment, not an algorithm.

    7. Type of Ground Truth Used

    • Bench Testing Results and Biocompatibility Standards. The "ground truth" for this modification's safety and effectiveness is established by meeting predefined engineering performance specifications and biological safety standards (ISO 10993-1, FDA Guidance) through laboratory testing.

    8. Sample Size for the Training Set

    • Not Applicable. This is not a machine learning device that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As above, no training set is relevant here.

    Summary of what the document confirms about the modification:

    • The change is related to the color band material and colorant.
    • The indications for use, intended use, operating principles, energy source, and functional outputs remain unchanged from the predicate device (K143320).
    • Bench testing (colorfastness, durability) was performed and met acceptance criteria.
    • Biocompatibility evaluation was performed, including chemical characterization (ICP-MS) and toxicological risk assessment, following established standards (ISO 10993-1, FDA Guidance). This biocompatibility testing was adopted from the predicate K143320, suggesting the new material passed similar or identical tests as the previous material.
    • No animal or clinical testing was required for this specific modification.
    • The conclusion is that the device, with the modification, is as safe and effective as the predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183644

    Validate with FDA (Live)

    Date Cleared
    2019-05-22

    (147 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | 21 CFR 882.4360 (Motor, Drill,Electric)21CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stealth-Midas MR8 System is indicated for the drilling, burring and removal of hard tissue and bone in spinal and cranial surgical procedures. Computer-assisted surgery and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.

    Device Description

    The Stealth-Midas MR8 consists of electric and pneumatic drill handpieces that feature an optical navigation tracker, enabling navigation in conjunction with the StealthStation S8 System and the Integrated Power Console (IPC). The system allows the navigation of a selection of currently available surgical dissecting tools. The navigated handpieces are provided non-sterile and are reusable. The associated attachments are provided non-sterile and are reusable. The associated surgical dissecting tools and MR8 ClearView burs are provided sterile and are single-use.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Medtronic Stealth-Midas MR8 System. The information provided outlines the device, its intended use, and comparative testing to establish substantial equivalence to predicate devices. However, the document does NOT contain a detailed study proving the device meets specific acceptance criteria with reported device performance metrics in the format requested. The document primarily focuses on establishing substantial equivalence based on technological characteristics and functional testing, not a formal clinical efficacy or diagnostic performance study with specific quantitative acceptance criteria typically seen for AI/ML-driven devices.

    Therefore, many of the requested items (e.g., sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, effect sizes, standalone performance, type of ground truth) are not explicitly present in the provided text. The device described is a surgical instrument with navigation capabilities, not an AI/ML-driven diagnostic or prognostic tool.

    Despite this, I will describe what is present and indicate what is missing based on the prompt's requirements.


    Device Name: Stealth-Midas MR8 System
    Device Type: Stereotaxic Instrument (Surgical Navigation System with Drill)

    The provided document describes the Stealth-Midas MR8 System, which combines an electric and pneumatic drill handpiece with an optical navigation tracker for computer-assisted surgery. The purpose of the 510(k) submission is to demonstrate substantial equivalence to legally marketed predicate devices, not necessarily to prove the device meets specific quantitative performance acceptance criteria in the context of an efficacy study as might be conducted for an AI/ML diagnostic algorithm.

    Here's a breakdown of the requested information based on the provided document:


    1. A table of acceptance criteria and the reported device performance

    The document focuses on "Navigational Accuracy Analysis" as a key performance test.

    Acceptance Criteria (from predicate)Reported Device Performance (Stealth-Midas MR8 System)
    StealthStation System Accuracy Requirements (Predicate Stealth-Midas with StealthStation S7):Stealth-Midas MR8 with StealthStation S8:
    3D Accuracy Testing: < 2.27 mm3D Accuracy Testing: < 2.00 mm
    2D Accuracy Testing: < 2.98 mm2D Accuracy Testing: < 3.00 mm
    Conclusion: The document states, "Accuracy validation testing conducted on the subject device confirms the Stealth-Midas MR8 when used with StealthStation S8 is equivalent in accuracy as the predicate Stealth-Midas when used with StealthStation S7."

    Other "performance testing" mentioned includes:

    • CAD Model Evaluation: Verified that CAD models are accurately reflected in the application software. (No quantitative acceptance criteria or reported performance given beyond "verified").
    • Formative Usability: Confirmed users can follow the navigated workflow and assemble the device. (No quantitative acceptance criteria or reported performance given beyond "confirmed").
    • Summative Testing: Confirmed users can safely and effectively use the Stealth-Midas MR8 navigated drill handpiece with the StealthStation S8 software. (No quantitative acceptance criteria or reported performance given beyond "confirmed").

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided document. The document refers to "Accuracy validation testing" but does not detail the number of tests performed or the nature of the samples (e.g., phantom, cadaver, pre-clinical/clinical data).
    • Data Provenance: Not specified. It can be inferred that testing was conducted internally by Medtronic, but no country of origin or whether it was retrospective/prospective is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the document. The accuracy testing refers to comparing the device's measured accuracy against a predicate device's established accuracy requirements. It's likely that "ground truth" for navigational accuracy pertains to physical measurements using precision instruments, not expert-derived labels.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not provided and is generally not applicable to the type of engineering/physical accuracy testing described for a surgical navigation system. Adjudication methods are typically used for establishing ground truth in image interpretation or clinical outcomes, which is not the primary focus of the performance tests detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done or mentioned. This type of study is relevant for AI-assisted diagnostic tools, not for a surgical drill and navigation system like the Stealth-Midas MR8. The device assists surgeons in locating anatomical structures, but it does not analyze images or provide diagnoses that human "readers" would interpret.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The concept of "standalone performance" for an algorithm is not applicable in the context of this device. The Stealth-Midas MR8 System is an instrument that functions in conjunction with a human surgeon and the StealthStation S8 System. Its "performance" is inherently linked to its ability to accurately track and display the position of surgical tools relative to patient anatomy, as used by a human, rather than an independent algorithmic output. The "Navigational Accuracy Analysis" seems to be a form of standalone technical performance assessment of the navigation component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth for the "Navigational Accuracy Analysis" would likely be physical measurements from a highly accurate measurement system (e.g., optical tracking system with known fiducials, coordinate measuring machine) in a controlled laboratory setting, rather than expert consensus, pathology, or outcomes data. The document implies a technical performance measurement against established benchmarks, not a clinical outcome.

    8. The sample size for the training set

    • Not applicable / Not provided. The Stealth-Midas MR8 System is a mechanical and optical medical device, not an AI/ML system that requires a "training set" in the computational sense. The device's design and operating parameters are based on engineering principles and validated through testing, not machine learning.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. As the device is not an AI/ML system, there is no "training set" or corresponding ground truth establishment process in the way it's understood for AI/ML.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183515

    Validate with FDA (Live)

    Date Cleared
    2019-05-12

    (145 days)

    Product Code
    Regulation Number
    882.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification Name:

    Motor, Drill, Electric (21 CFR 882.4360) Motor, Drill, Pneumatic (21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic MR8 Drill System is incision/cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials in Neurosurgical (Craniofacial including craniotomy); Ear, Nose and Throat (ENT), Maxillofacial, Orthopedic, Arthroscopic, Spinal, Sternotomy, and General Surgical Procedures.

    Additionally, the MR8 Drill System is indicated for the incision/cutting, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:

    • Lumbar Microdiscectomy
    • Lumbar Stenosis Decompression
    • Posterior Lumbar Interbody Fusion (PLIF)
    • Transforaminal Lumbar Interbody Fusion (TLIF)
    • Anterior Lumbar Interbody Fusion (ALIF)
    • Direct Lateral Interbody Fusion (DLIF)

    The Midas Rex MR8 ClearView Tools are used only in conjunction with the MR8 Drill System to perform as intended. Please refer to the Midas Rex MR8 Drill System and associated User's Guides for the Indications of Use.

    Device Description

    The Medtronic MR8TM Drill System is comprised of both Electric and Pneumatic powered, rotary cutting handpieces, attachments, surgical dissecting tools, and accessories designed to remove soft and hard tissue, bone, and biomaterials during various surgical procedures. The surgical dissecting tools are provided sterile and are single use, while the rest of the system components are provided non-sterile and are reusable.

    The Midas Rex™ MR8TM ClearView™ Tools are designed to interface with Midas Rex™ MR8 Drill System motor to support bone and tissue removal during surgical procedures. The Midas Rex™ MR8™ ClearView™ Tools are part of a larger portfolio of tools and accessories designed to be used with the Midas Rex™ MR8 System/Platform.

    AI/ML Overview

    This document describes the Medtronic MR8 Drill System and Midas Rex MR8 ClearView Tools. The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than independent performance claims against specific acceptance criteria. Therefore, the information provided primarily compares the device to existing predicate devices.

    1. Table of Acceptance Criteria & Reported Device Performance:

    The document does not explicitly present a table of "acceptance criteria" for the overall device in the typical sense of a clinical trial or performance study with defined thresholds. Instead, it details performance testing conducted for the Midas Rex MR8 ClearView Tools to ensure functionality with the MR8 Drill system and comparability to predicate devices. The "acceptance criteria" for this testing appear to be qualitative (e.g., "similar and/or better," "same or more," "below the burn threshold").

    TestAcceptance Criteria (Implied)Reported Device Performance
    Tool Chatter and Hand VibrationSimilar and/or better than equivalent predicatesScored similar and/or better than the equivalent Predicates
    Irrigation Rate vs IPC SettingSame or more than the rate displayed on the IPCDelivered the same or more than the one displayed on the IPC
    Thermal PerformanceCompleted respective duty cycles intact; Max temperature below burn thresholdCompleted duty cycles intact; Maximum temperature below burn threshold

    No additional testing was performed on the MR8 Drill System itself as there were no design changes to it for this submission. The Midas Rex MR8 ClearView Tools are the new elements being evaluated for their compatibility and performance within the existing MR8 Drill System.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample size for the individual performance tests (Tool Chatter and Hand Vibration, Irrigation Rate vs IPC Setting, Thermal Performance)conducted on the Midas Rex MR8 ClearView Tools.

    The provenance of the data is not explicitly stated (e.g., country of origin, retrospective/prospective). However, given that this is a 510(k) submission for a medical device by Medtronic, a US-based company, it is highly likely that the testing was conducted under standard quality systems and engineering practices, likely within a controlled laboratory environment. The tests appear to be engineering/bench testing rather than clinical studies with human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable to the type of testing described. The tests are engineering performance tests, not clinical evaluations requiring expert interpretation of ground truth (e.g., diagnosis from medical images).

    4. Adjudication Method for the Test Set:

    This information is not applicable. The tests are objective, quantitative measurements or qualitative observations during engineering performance testing, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This information is not applicable. The device described is a surgical drill system and associated tools, not an AI-assisted diagnostic or interpretative system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    This information is not applicable. This is not an AI-driven device or algorithm. The performance evaluation focuses on the mechanical and operational characteristics of the surgical tools.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance tests effectively refers to the physical and functional parameters of the device as designed and expected, as well as established safety thresholds (e.g., burn threshold for thermal performance). It's based on engineering specifications and safety standards relevant to surgical instruments.

    8. The Sample Size for the Training Set:

    This information is not applicable. No "training set" is mentioned as this is not a machine learning or AI-driven device.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable. As there is no training set mentioned, there is no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172907

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-09

    (410 days)

    Product Code
    Regulation Number
    882.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Pennsylvania 18034

    Re: K172907

    Trade/Device Name: ELAN 4 Air Motor System Regulation Number: 21 CFR 882.4370
    |
    | REGULATION NUMBER: | 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ELAN 4 Air Motor System is intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

    Device Description

    The ELAN 4 Air Motor System is a pneumatic motor system that can be used with any sterile, pressurized gas (usually nitrogen). The system consists of several components such as wall adaptors, foot pedals, and handpieces. The foot pedal has the ability to connect various handpieces with integrated motors that contain attachments such as burrs, saw blades, and drills. The system allows for high-speed dissection at up to 80.000 RPM while also allowing low speed cutting between 1.000 and 20.000 RPM. The speed and rotation direction of the handpieces can be controlled via the foot pedal.

    AI/ML Overview

    The provided text is a 510(k) summary for the ELAN 4 Air Motor System, a pneumatic cranial drill motor. It does not describe a study involving an AI or algorithm-based device, human readers, or a test set with expert ground truth establishment. Instead, it describes a medical device (a surgical drill) and its performance validation through design verification and biocompatibility testing.

    Therefore, I cannot answer your request as the information you've asked for (acceptance criteria for an AI/algorithm, sample size for AI test and training sets, expert consensus for ground truth, MRMC study, etc.) is not contained within the provided document.

    The document focuses on demonstrating substantial equivalence to predicate devices (other surgical drills) based on:

    • Indications for Use: The ELAN 4 Air Motor System is intended for high-speed cutting, sawing, and drilling of bone in various surgical fields.
    • Technological Characteristics: Comparison of components, power source (pneumatic vs. electric), operating speeds, rotation, materials, and control mechanisms.
    • Performance Data (Design Verification): Functional testing to ensure the device performs as intended and safely meets specifications over time. These tests are explicitly listed as "Pass: All requirements met."
    • Biocompatibility: Confirmation that materials used are safe for patient contact.

    To reiterate, the provided text does not describe the kind of study you are asking about, which typically pertains to diagnostic or assistive AI software rather than mechanical surgical instruments.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163565

    Validate with FDA (Live)

    Device Name
    MR8 Drill System
    Date Cleared
    2017-03-28

    (99 days)

    Product Code
    Regulation Number
    882.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification Name: Motor, Drill, Electric (21 CFR 882.4360) Motor, Drill, Pneumatic (21 CFR 882.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic MR8 Drill System is indicated for the incision/cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials in Neurosurgical (Craniofacial including craniotomy); Ear, Nose and Throat (ENT), Maxillofacial, Orthopedic, Arthroscopic, Spinal, Sternotomy, and General Surgical Procedures.

    Device Description

    The MR8 Drill System consists of electric and pneumatic drill handpieces, attachments, surgical dissecting tools, and system accessories. The handpieces, attachments, and system accessories are provided non-sterile and are reusable. The surgical dissecting tools are provided sterile and are single use.

    AI/ML Overview

    The document provided is a 510(k) summary for the Medtronic MR8 Drill System. It describes the device, its intended use, and comparison to predicate devices, along with a summary of performance testing. However, it does not provide the detailed acceptance criteria and study information typically found in a clinical study report for an AI/ML medical device. The MR8 Drill System is a surgical drill, not an AI/ML device.

    Therefore, many of the requested sections below, such as details on AI performance, sample sizes for training/test sets, expert ground truth establishment, adjudication methods, and MRMC studies, are not applicable to this submission. The information provided primarily focuses on the substantial equivalence of a physical surgical device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for the performance tests in the same way typical for AI/ML performance, but rather states the criteria implicitly by comparing to predicate devices or conformity to standards.

    TestAcceptance Criteria (Implicit from document)Reported Device Performance
    Motor Speed/Torque AnalysisSpeed/torque profile similar to that of predicate Legend deviceSpeed/torque profile is similar to that of predicate Legend device
    Drill System Cutting PerformanceCutting performance equivalent or better to that of predicate device in terms of tool chatter and hand vibrationCutting performance was equivalent or better to that of predicate device
    Electrical SafetyConformance to IEC 60601-1:2005Instruments conform to IEC 60601-1:2005 for electrical safety.
    Electromagnetic CompatibilityConformance to IEC 60601-1-2:2014Instruments conform to IEC 60601-1-2:2014 for electromagnetic compatibility.
    Cadaveric Simulated Use (Usability)Acceptable for its intended use in various surgical procedures by usersMR8 Drill System acceptable for its intended use in various surgical procedures (as evaluated by users on cadavers)

    Regarding AI/ML Specific Information (Many are Not Applicable to this Device):

    This device is a surgical drill system, a physical medical device, and not an AI/ML-driven diagnostic or therapeutic device. Therefore, the following sections are largely not applicable based on the provided FDA submission summary.

    2. Sample size used for the test set and the data provenance: Not applicable. The "test set" for this device involved physical product testing and cadaveric use. Specific sample sizes for the cadaveric study are not provided in this summary.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable in the context of AI/ML ground truth. For the cadaveric study, professional users (surgeons/medics) would have evaluated the device, but specific numbers and qualifications are not detailed in this summary.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of device and study.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the cadaveric use, the "ground truth" would be the subjective and objective assessment of the device's performance and acceptability by the users during surgical simulation. For other tests, it's conformance to technical specifications and safety standards.
    8. The sample size for the training set: Not applicable, as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.