of a risk management process, applicable FDA guidance documents and performance standards (21 CFR §882.4310
Surgify Halo is a sterile surgical burr indicated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and general surgical procedures.
The Surgify Halo is a sterile-packaged, single-use rotary cutting device (a burr) made from high-grade metallic materials, designed to cut bone and other hard tissues selectively while minimizing chattering. The device is similar to conventional surgical cutting burrs but includes a ring mechanism in the burr head to reduce chattering. Like traditional burrs, the device is designed to be compatible with commonly marketed, high-speed, surgical drill systems.
The Surgify Halo comprises:
- A head part, which is shaped to have cutting edge(s) to cut hard tissue.
- Dynamically positioned ring surrounding the burr head that controls the exposure of the cutting edge(s) decreasing chattering and enabling the tissue selective feature.
- A shaft, which transfers rotary movement from the drill motor to the head.
- An end-attachment feature, which allows the Surgify Halo to be locked to commonly marketed high-speed, surgical drill systems.
N/A
Ask a specific question about this device
(100 days)
Classification Name:** 21 CFR 882.4370 - Pneumatic cranial drill motor (Primary)
21 CFR 882.4310 -
882.4370; Pneumatic cranial drill motor | 21 CFR 882.4370; Pneumatic cranial drill motor21 CFR 882.4310
The Velocity Alpha MR High Speed Surgical Drill System is indicated for trephination, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone and biomaterials in or near a magnetic field of 3.0 Tesla or less for use in Neurosurgery.
The Velocity Alpha MR High Speed Surgical Drill System is a pneumatic powered surgical motor drill system. The system supplies pressurized air to motor which powers rotary surgical instruments through the attachments. Rotational speed of the system is controlled by the foot pedal. The MR System consists of a motor, motor hose, two attachments, and 4 cutting instruments. Additionally, the system is compatible with the Velocity Alpha Pneumatic High Speed Surgical Drill System foot pedal and wall hoses. The Velocity Alpha MR High Speed Surgical Drill System, including both system-specific and system-compatible components, can be used in or near a magnetic field of 3.0 Tesla level or less.
N/A
Ask a specific question about this device
(100 days)
Craniomaxillofacial Surgery" CMF Titanium Guides; CMF Plastic Models
Regulation Number: 21 CFR 882.4310
code: PPT, Secondary product code: LLZ |
| Classification Panel | Neurology |
| Device class | 21 CFR 882.4310
Planning System |
| 510(k) number | K201052 |
| Decision date | 08/31/2020 |
| Product code | PPT (21 CFR 882.4310
---------------------|
| Product code | PPT, LLZ | DZJ, LLZ | HAW | PPT |
| Classification | 21 CFR 882.4310
, Class II | 21 CFR 872.4120, Class II | 21 CFR 882.4560, Class II | 21 CFR 882.4310, Class II |
| Indications
CMF Titanium Guides
CMF Titanium Guides are intended to guide the marking of bone and/or guide surgical instruments in craniofacial surgery.
CMF Titanium Guides are used during bone repositioning/reconstruction surgical operations for orthognathic and reconstruction (including bone harvesting) indications.
CMF Titanium Guides are intended for children, adolescents and adults.
CMF Titanium Guides are intended for single use only.
CMF Titanium Guides are to be used by a physician trained in the performance of craniomaxillofacial surgery.
CMF Plastic Models
CMF Plastic Models are intended for visualization of the patient's anatomy, preparation of surgical interventions and fitting or adjustment of implants or other medical devices such as osteosynthesis plates or distractors, in craniofacial surgical procedures.
CMF Plastic Models are intended for infants, children, adolescents and adults.
CMF Plastic Models are intended for single use only.
CMF Plastic Models are to be used by a physician trained in the performance of craniomaxillofacial surgery.
Materialise Personalized Guides and Models for Craniomaxillofacial Surgery combines the use of 3D pre-operative planning software with patient-matched guides and models to improve and simplify the performance of surgical interventions by transferring the pre-operative plan to surgery. Materialise Personalized Guides and Models for Craniomaxillofacial Surgery are used in the craniofacial skeleton or in craniomaxillofacial surgeries.
The surgical planning is based on medical images of the patient that are segmented in order to create a 3D representation of the patient's anatomy. The surgical treatment of the patient is simulated based on instructions provided by the surgeon and the patient-matched devices are tailored to the treatment and the patient's needs. The patient-matched devices are manufactured from commercially pure Titanium, polyamide, or clear acrylic by means of additive manufacturing technologies. The patient-matched devices are provided non-sterile.
Materialise Personalized Guides and Models for Craniomaxillofacial Surgery include CMF Titanium Guides and CMF Plastic Models.
N/A
Ask a specific question about this device
(86 days)
K161704; respectively |
| Predicate Device Classification & Name | (1) 21 CFR 882.4300; (2) 21 CFR 882.4310
now Phasor Health, LLC) | N/A |
| Regulation Number | 21 CFR 882.4300 | 21 CFR 882.4300 | 21 CFR 882.4310
The Phasor LEGACY drill is a sterile, single-use, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
Phasor LEGACY drill is indicated for drilling of cranial bone by licensed and credentialed medical professionals. The LEGACY drill is manually powered by rotating the handle in a clockwise manner, and is a standalone drill driver without bits or any other items included. It is compatible via an AO quick-connect mechanism with similar AO-interface drill bits, ≤ 6.35 mm diameter and ≤ 70 mm usable length (protruding from the housing). The device is gamma-sterilized and composed of plastic housing with stainless steel shaft.
N/A
Ask a specific question about this device
(234 days)
60532
Re: K250815
Trade/Device Name: Hubly Drill (H100)
Regulation Number: 21 CFR 882.4310
60532
Re: K250815
Trade/Device Name: Hubly Drill (H100)
Regulation Number: 21 CFR 882.4310
4010 |
| Device Common Name: | Powered Cranial Drill |
| Device Regulation & Name: | 21 CFR 882.4310
---------------|
| Company | Hubly, Inc. | Hubly, Inc. | N/A |
| Regulation Number | 21 CFR 882.4310
| 21 CFR 882.4310 | Identical |
| Product Code | HBE – Class II | HBE – Class II | Identical |
|
The Hubly Auto-Stop Drill (H100) is a single-use, sterile, disposable device intended for use on adult patients during laminectomy and laminotomy procedures for drilling holes through the lamina.
The Hubly Drill is a battery powered drill. It is single-use disposable and provided ethylene oxide sterilized. The drill is designed to create optimal holes in bone. For trephination procedures the Hubly Drill is intended to drill cranial bone and create burr holes in the emergency room, at the bedside, or in the operating room. For laminectomy and laminotomy procedures the drill is intended to be used in the operating room to drill holes through the lamina to facilitate excision of the lamina.
The drill can be used one-handed using either hand. It is trigger-activated after removal of the battery pull tab. The drill has a single speed and turns off when the trigger is released. The drill bit has depth indicators at 5 and 10mm depth, which the physician may use to visually gauge depth of penetration while drilling. The drill features mechanical plunge prevention with a tapered stainless steel drill bit. The drill may be reactivated any number of times using the trigger if the physician desires.
The drill also has an auto-stop feature which detects when the bit breaks through the bone and immediately stops the drill. The device has an LED indicator which indicates to the user (green) when they are applying enough force and (red) when the drill stops.
N/A
Ask a specific question about this device
(122 days)
Houston, Texas 77054
Re: K252060
Trade/Device Name: Green
Regulation Number: 21 CFR 882.4310
Houston, Texas 77054
Re: K252060
Trade/Device Name: Green
Regulation Number: 21 CFR 882.4310
simple cranial drills, burrs, trephines, and their accessories |
| Device Regulation & Name | 21 CFR 882.4310
Company** | Phasor Health, LLC | Phasor Health, LLC | Stryker | N/A |
| Regulation Number | 21 CFR 882.4310
| 21 CFR 882.4310 | 21 CFR 870.2700 | No impact |
| Product Code | HBE (Class II) | HBE (Class II
Phasor GREEN Drill is a re-sterilizable drill driver (for use on one ore more patients, up to 10 total holes), with separately packaged single-use sterilized drill bit assembly -- for use on adult patients during neurosurgical procedures for drilling of cranial bone.
The Phasor™ GREEN Drill is composed of 2 items: (1) a re-sterilizable, non-rechargeable battery-operated Phasor Green Driver (made of plastic housing, capable of drilling up to 10 total holes in one or more patients) in conjunction with (2) a separately packaged, single-use Phasor Green Drill Bit Assembly comprised of a steel bit (of 6.35-mm or less), plastic adapter, and polybag secured using latex-free bands. The device is for drilling cranial or orthopedic bone, by prescription only and used by qualified users, with (1) Drill Driver (provided non-sterile) for sterilization using vaporized hydrogen peroxide (VHP)prior to use at user facility and (2) Drill Bit Assembly provided gamma-sterilized for single-use respectively.
N/A
Ask a specific question about this device
(28 days)
Saint Paul, MN 55114
Re: K252662
Trade/Device Name: UniBur
Regulation Number: 21 CFR 882.4310
Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories
Primary Classification: HBE 882.4310
UniBur is for single use only. This device is designed to be used with electric surgical instruments manufactured by Nakanishi INC. This device is intended for: cutting, drilling, removal, and shaping of bones in the fields of Neuro, Spine and ENT surgery.
The UniBur is a single-use, sterile medical device designed for cutting, drilling, removal, and shaping of bones, used with the Primado2 Total Surgical System (K132264) and used in the fields of neuro, spine, and ENT surgery.
The UniBur is designed to be connected to the slim motor handpiece of the Primado2 Total Surgical System.
The UniBur has 22 types of product variations based on differences in the length (3 types) and bending angle of the bur guard (3 types), the diameter of the bur for bone cutting (from φ0.6 mm to φ4.5 mm), and the grit size of the diamond embedded in the bur.
The UniBur consists of Diamond, Nickel, Stainless steel, Phosphor bronze, Stainless steel, FKM, Polyamide and PTFE.
N/A
Ask a specific question about this device
(138 days)
Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.
Phasor EZ-FIDUCIALS™ provide fixed reference points during neurosurgical procedures. The device is composed of 3 items, all single-use, provided gamma-sterilized, and placed within the same primary packaging (sealed Tyvek tray within a shelf carton):
-
Screws: Each of the four provided titanium screws is identical, with specific dimensions of each EZ-Fiducials screw as follows: screwhead circular shape screwhead with square-shaped engagement, non-threaded shaft diameter 2.9mm and length 15mm, threaded shaft diameter (major threads) 1.98mm, thread length 5mm; no protective caps are provided, and the screws should exclusively be utilized in conjunction with the screwdrivers (manual or EZ-Driver) provided together in the same tray (and with no other screwdriver);
-
Manual Screwdriver: handheld, to be solely used with the screws provided, not intended for use with any other screws;
-
EZ-driver: (electric, handheld, non-rechargeable battery-operated screwdriver, without software, single-use for solely tightening or loosening the provided screws exclusively, not intended for use with any other screws).
The provided FDA 510(k) clearance letter and summary for EZ-Fiducials focuses primarily on substantial equivalence to predicate devices based on technological characteristics and bench testing, rather than a clinical study with acceptance criteria based on human-in-the-loop performance or algorithm-only performance against a defined ground truth.
Therefore, for aspects related to "device performance," "acceptance criteria," "sample size," "expert ground truth," "adjudication methods," "MRMC studies," "standalone performance," and "ground truth for training/test sets," the available document does not provide this information. The submission relies on bench testing to demonstrate equivalence.
Here's an analysis of what information is provided and what explicitly is not provided based on your request:
Acceptance Criteria and Device Performance (Based on Available Information)
The document does not present a table of acceptance criteria and reported device performance in the manner typically seen for clinical or AI/algorithm performance studies (e.g., sensitivity, specificity, accuracy). Instead, it relies on demonstrating compliance with an ASTM standard and general performance adequacy, often by comparison to predicate devices' known characteristics.
Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):
| Acceptance Criteria Category (Inferred) | Specific Standard/Requirement | Reported Device Performance (EZ-Fiducials) | Met? |
|---|---|---|---|
| Material Biocompatibility | Biocompatible materials | Yes | Yes |
| Sterility | Gamma-sterilized | Yes, single-use, gamma-sterilized | Yes |
| Single-use | Disposable | Yes | Yes |
| Fiducial Screw Purchase/Deflection | Firm purchase into bone simulant; compliant with ASTM F543-23 | Verified by testing for deflection using CMM; compliant with ASTM F543-23 | Yes |
| Electric Screwdriver Functionality | Ability to drive/remove screws; adequate performance (per report "02670-019011-1" including Table 6) | Performs adequately, speed 30rpm (different from reference 120rpm, but supported by performance testing) | Yes |
| Packaging Integrity | Maintain sterility | Sealed Tyvek tray within a shelf carton | Yes |
| Safety Considerations | Safe operation | No negative impact on safety compared to predicates | Yes |
Study Proving Device Meets Acceptance Criteria
The study proving the device meets acceptance criteria is primarily bench testing. No clinical study data is presented.
1. A table of acceptance criteria and the reported device performance:
- See table above. This is inferred from the discussion of performance testing. Explicit quantitative acceptance criteria (e.g., "deflection must be less than X mm") are not detailed, but compliance with ASTM F543-23 and "adequate performance" are stated.
2. Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated for bench testing. The testing refers to "bone simulant" but the quantity of fiducials, screws, or instances tested is not specified.
- Data Provenance: Not applicable in terms of patient data. The testing is bench testing using "bone simulant." There is no indication of retrospective or prospective data or country of origin, as it's not a clinical study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This was a bench study, not a study requiring expert ground truth for clinical cases. The "ground truth" for the bench testing would be physical measurements and compliance with engineering standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. There was no human interpretation or subjective assessment of clinical data that would require adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study was not done. The device is a physical fiducial system and an electric screwdriver, not an AI or software device that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an AI algorithm. The device performance here refers to the physical characteristics and functionality of the fiducials and screwdriver.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Bench Testing Data/Engineering Standards. The ground truth for this submission's performance assessment is the physical performance and measurements of the device (deflection) and compliance with industry standards (ASTM F543-23).
8. The sample size for the training set:
- Not applicable. This device does not use machine learning or AI, so there is no training set.
9. How the ground truth for the training set was established:
- Not applicable. As above, no training set.
Summary of the Study:
The "study" referenced in the 510(k) for EZ-Fiducials is a series of bench tests focusing on:
- Biocompatibility of materials.
- Sterility validation.
- Packaging integrity.
- Performance of the screws (e.g., deflection characteristics, compliance with ASTM F543-23 when inserted into bone simulant).
- Performance of the electric screwdriver (its ability to drive screws adequately).
The clearance is based on the argument that these bench tests demonstrate the device's substantial equivalence in terms of safety and effectiveness to the predicate device (Medtronic Navigus Unibody Fiducial Marker System) and a reference device (OsteoMed Pinnacle Driver), even with some technological differences (e.g., screwdriver speed, sterilization method). The document explicitly states: "No clinical testing was needed or performed otherwise." The predicate device was also "cleared based upon bench testing alone."
Ask a specific question about this device
(27 days)
.NVG.H1); Surgify Halo (40.125.NVG.H1); Surgify Halo (40.000.SEE.H1)
Regulation Number: 21 CFR 882.4310
.NVG.H1); Surgify Halo (40.125.NVG.H1); Surgify Halo (40.000.SEE.H1)
Regulation Number: 21 CFR 882.4310
simple cranial drills, burrs, trephines, and their accessories. |
| Regulation Number: | 21CFR 882.4310
simple cranial drills, burrs, trephines, and their accessories. |
| Regulation Number: | 21CFR 882.4310
of a risk management process, applicable FDA guidance documents and performance standards (21 CFR §882.4310
Surgify Halo is a sterile surgical burr indicated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and General Surgical procedures.
The Surgify Halo is a sterile-packaged, single-use rotary cutting device (a burr) made from high-grade metallic materials, designed to cut bone and other hard tissues selectively while minimizing chattering. The device is similar to conventional surgical cutting burrs but includes a ring mechanism in the burr head to reduce chattering. Like traditional burrs, the device is designed to be compatible with commonly marketed, high-speed, surgical drill systems. The Surgify Halo comprises: Shank, Locking Features, Burr Head, and Safety Mechanism (a self-centering ring and spring mechanism installed in the burr head). The moving ring, larger than the burr head, can adopt two positions: First Position (Hard Tissue Mode) where the ring protrudes, preventing cutting edges from engaging, and Second Position (Soft Tissue Mode) where the ring extends beyond the cutting edge, shielding it to minimize accidental soft tissues and an exposed position for cutting hard surfaces like bone.
The Surgify Halo is a sterile surgical burr used for shaping and removing hard tissue and bone in neurosurgical, spinal, ENT, and general surgical procedures. The company performed performance testing and design validation/summative evaluation to demonstrate substantial equivalence to its predicate device (K250380).
Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria (Design Input Requirements) | Reported Device Performance (Results) |
|---|---|---|
| Functional Testing | ||
| Durability | Device functions as intended, all design and functional specifications met | All tests fulfilled Design input requirements |
| Cutting Effectiveness | Device functions as intended, all design and functional specifications met | All tests fulfilled Design input requirements |
| Chatter Rate | Device functions as intended, all design and functional specifications met | All tests fulfilled Design input requirements |
| Noise | Device functions as intended, all design and functional specifications met | All tests fulfilled Design input requirements |
| Thermal | Device functions as intended, all design and functional specifications met | All tests fulfilled Design input requirements |
| Design Validation/Summative Evaluation | ||
| Summative Usability Testing | User requirements met, consistent with predicate evaluation approach, addressed use errors and hazard-related use scenarios, in accordance with IEC 62366-1:2015 and FDA's 2000 Guidance "Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management." | All tests fulfilled Design input and user requirements |
Note: The provided text states that "All tests fulfilled Design input requirements" and "All tests fulfilled Design input and user requirements." It does not explicitly state numerical acceptance criteria for each parameter (e.g., "chatter rate < X dB"). The performance is reported as meeting these general design input requirements.
2. Sample Size and Data Provenance for the Test Set
- Sample Size:
- For Functional Testing: Not explicitly stated as a number of devices or trials. The "Results" column simply indicates that "All tests fulfilled Design input requirements."
- For Design Validation/Summative Usability Testing: Not explicitly stated as a number of participants or cases, only that "production equivalent Surgify Halo (4 mm)" was used.
- Data Provenance (Country of Origin and Retrospective/Prospective Data): Not specified in the provided document.
3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)
- Number of Experts: Not explicitly stated.
- Qualifications of Experts: For the Summative Usability Testing, it is mentioned that "Surgeons Summative Testing" was performed. Specific qualifications (e.g., years of experience, specialty beyond "Surgeons") are not provided.
4. Adjudication Method (Test Set)
- Adjudication Method: Not specified in the document. The usability testing was conducted as "Surgeons Summative Testing," implying direct user evaluation against scenarios, rather than an independent adjudication of results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done? No, the document does not mention an MRMC study or any comparison of human readers with vs. without AI assistance. The device is a surgical burr, not an AI-assisted diagnostic tool.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
- Standalone Performance Done? N/A. The device is a physical surgical tool, not an algorithm. Its performance is intrinsically linked to its use by a human operator in a surgical setting.
7. Type of Ground Truth Used (Test Set)
- Ground Truth Type:
- For Functional Testing: The "ground truth" was likely defined by engineering specifications and design input requirements for physical properties and operational performance (e.g., durability, cutting effectiveness, chatter, noise, thermal output).
- For Summative Usability Testing: The "ground truth" was established by user requirements and validated through hazard-related use scenarios, assessing whether "user requirements were met" and "user errors" were minimized. This is a form of expert consensus/observation of appropriate use based on human factors principles.
8. Sample Size for the Training Set
- Sample Size: N/A. As a physical medical device (surgical burr) that is not an AI/ML algorithm requiring training data, the concept of a "training set" in the context of machine learning does not apply. The development would involve engineering design, material science, and iterative physical prototyping/testing.
9. How Ground Truth for the Training Set Was Established
- How Ground Truth Was Established: N/A, for the same reasons as #8.
Ask a specific question about this device
(245 days)
Phasor EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma.
Phasor EVAC™ subdural evacuation system is indicated when access to the subdural space and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system consists of surgical instruments and accessories used for draining subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir without touching the brain. Utilizing a minimally invasive technique, the EVAC system components are designed to promote gradual brain re-expansion by creating a low homogeneous negative pressure throughout the subdural space as fluid is drained to an external suction reservoir.
This 510(k) clearance letter for the EVAC device (K243205) indicates that it is a central nervous system fluid shunt and components and it's regulated under 21 CFR 882.5550. While the letter confirms the device's substantial equivalence to predicate devices and states that various tests were conducted, it does not provide detailed information about specific acceptance criteria, study methodologies (like sample size, ground truth establishment, expert qualifications, or MRMC studies), or actual performance data in the format requested.
The document primarily focuses on explaining why the device is considered substantially equivalent to existing devices based on its intended use, indications for use, and technological characteristics. It mentions that "Various tests including for biocompatibility, packaging, sterility, safety, and performance were conducted and passed successfully," and that "testing for a closed system with firm purchase into bone was established and verified by testing." However, it does not disclose the specific data or criteria for these tests. It explicitly states, "No clinical testing was needed or performed otherwise."
Therefore, based on the provided FDA 510(k) clearance letter, I cannot populate the detailed table and answer all the specific questions about acceptance criteria and study proving device performance as requested, because this information is typically contained in the actual 510(k) submission, not the clearance letter itself. The clearance letter summarizes the FDA's decision, but not the detailed technical data from the submission.
Here's what can be inferred or explicitly stated from the provided document, with notes where information is not present:
Acceptance Criteria and Device Performance (Based on Inferred Safety and Effectiveness)
The acceptance criteria for a 510(k) cleared device are implicitly that it demonstrates substantial equivalence to a predicate device, meaning it is as safe and effective while having the same intended use. For this device, the "performance" described is its ability to meet the design intent and demonstrate equivalence through bench testing, rather than clinical efficacy.
Table 1: Inferred Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Inferred from 510(k) Basis) | Reported Device Performance (from document) |
|---|---|
| Biocompatibility | "Various tests including for biocompatibility... were conducted and passed successfully." (Specific criteria and results not provided) |
| Packaging Integrity | "Various tests including for... packaging... were conducted and passed successfully." (Specific criteria and results not provided) |
| Sterility | "Various tests including for... sterility... were conducted and passed successfully." (Specific criteria and results not provided) |
| Safety | "Various tests including for... safety... were conducted and passed successfully." (Specific criteria and results not provided) |
| Performance (Functional Equivalence) | "Various tests including for... performance were conducted and passed successfully." "Specifically, testing for a closed system with firm purchase into bone was established and verified by testing." "The performance and design validation testing conducted on the EVAC™ device on the bench demonstrated that it performs equivalent to the stated predicates..." (Specific criteria and results not provided) |
| Manufacturing Quality Systems | Compliance with Quality System (QS) regulation (21 CFR Part 820) and other general controls is required for marketing. (No specific test results provided in letter) |
Study Details (Based on Available Information)
-
Sample size used for the test set and the data provenance:
- The document states "No clinical testing was needed or performed otherwise." This indicates that the "test set" for demonstrating substantial equivalence was primarily based on bench testing and engineering evaluations of the device components and system.
- Specific sample sizes for these bench tests are not provided in the clearance letter.
- Data provenance (country of origin, retrospective/prospective): Not applicable or specified for bench testing.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Given that "No clinical testing was needed or performed," there's no mention of expert panels establishing ground truth in a clinical context. The "ground truth" for the engineering tests would be derived from accepted engineering standards and specifications.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable, as no clinical study or human reader evaluation requiring adjudication was performed.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The device is a surgical system (shunt and components), not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This question is more pertinent to software/AI devices. The EVAC system is a mechanical device. Its "performance" refers to its physical functioning (e.g., maintaining a closed system, firm purchase into bone), not an algorithmic output. The "Phasor Drill" component is explicitly stated to have been previously cleared (K161704), indicating its standalone performance was established previously.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For this device, the "ground truth" for its functional performance was based on engineering specifications, material science standards, and mechanical testing requirements (e.g., ability to form a "closed system with firm purchase into bone"). There's no mention of clinical ground truth (like pathology or outcomes data) because no clinical study was performed.
-
The sample size for the training set:
- Not applicable. This is a hardware device, not an AI/machine learning algorithm requiring a "training set."
-
How the ground truth for the training set was established:
- Not applicable for the same reason as above.