K Number
K250505

Validate with FDA (Live)

Device Name
EZ-FIDUCIALS
Manufacturer
Date Cleared
2025-07-09

(138 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.

Device Description

Phasor EZ-FIDUCIALS™ provide fixed reference points during neurosurgical procedures. The device is composed of 3 items, all single-use, provided gamma-sterilized, and placed within the same primary packaging (sealed Tyvek tray within a shelf carton):

  1. Screws: Each of the four provided titanium screws is identical, with specific dimensions of each EZ-Fiducials screw as follows: screwhead circular shape screwhead with square-shaped engagement, non-threaded shaft diameter 2.9mm and length 15mm, threaded shaft diameter (major threads) 1.98mm, thread length 5mm; no protective caps are provided, and the screws should exclusively be utilized in conjunction with the screwdrivers (manual or EZ-Driver) provided together in the same tray (and with no other screwdriver);

  2. Manual Screwdriver: handheld, to be solely used with the screws provided, not intended for use with any other screws;

  3. EZ-driver: (electric, handheld, non-rechargeable battery-operated screwdriver, without software, single-use for solely tightening or loosening the provided screws exclusively, not intended for use with any other screws).

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for EZ-Fiducials focuses primarily on substantial equivalence to predicate devices based on technological characteristics and bench testing, rather than a clinical study with acceptance criteria based on human-in-the-loop performance or algorithm-only performance against a defined ground truth.

Therefore, for aspects related to "device performance," "acceptance criteria," "sample size," "expert ground truth," "adjudication methods," "MRMC studies," "standalone performance," and "ground truth for training/test sets," the available document does not provide this information. The submission relies on bench testing to demonstrate equivalence.

Here's an analysis of what information is provided and what explicitly is not provided based on your request:


Acceptance Criteria and Device Performance (Based on Available Information)

The document does not present a table of acceptance criteria and reported device performance in the manner typically seen for clinical or AI/algorithm performance studies (e.g., sensitivity, specificity, accuracy). Instead, it relies on demonstrating compliance with an ASTM standard and general performance adequacy, often by comparison to predicate devices' known characteristics.

Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):

Acceptance Criteria Category (Inferred)Specific Standard/RequirementReported Device Performance (EZ-Fiducials)Met?
Material BiocompatibilityBiocompatible materialsYesYes
SterilityGamma-sterilizedYes, single-use, gamma-sterilizedYes
Single-useDisposableYesYes
Fiducial Screw Purchase/DeflectionFirm purchase into bone simulant; compliant with ASTM F543-23Verified by testing for deflection using CMM; compliant with ASTM F543-23Yes
Electric Screwdriver FunctionalityAbility to drive/remove screws; adequate performance (per report "02670-019011-1" including Table 6)Performs adequately, speed 30rpm (different from reference 120rpm, but supported by performance testing)Yes
Packaging IntegrityMaintain sterilitySealed Tyvek tray within a shelf cartonYes
Safety ConsiderationsSafe operationNo negative impact on safety compared to predicatesYes

Study Proving Device Meets Acceptance Criteria

The study proving the device meets acceptance criteria is primarily bench testing. No clinical study data is presented.

1. A table of acceptance criteria and the reported device performance:

  • See table above. This is inferred from the discussion of performance testing. Explicit quantitative acceptance criteria (e.g., "deflection must be less than X mm") are not detailed, but compliance with ASTM F543-23 and "adequate performance" are stated.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for bench testing. The testing refers to "bone simulant" but the quantity of fiducials, screws, or instances tested is not specified.
  • Data Provenance: Not applicable in terms of patient data. The testing is bench testing using "bone simulant." There is no indication of retrospective or prospective data or country of origin, as it's not a clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This was a bench study, not a study requiring expert ground truth for clinical cases. The "ground truth" for the bench testing would be physical measurements and compliance with engineering standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. There was no human interpretation or subjective assessment of clinical data that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study was not done. The device is a physical fiducial system and an electric screwdriver, not an AI or software device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI algorithm. The device performance here refers to the physical characteristics and functionality of the fiducials and screwdriver.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Bench Testing Data/Engineering Standards. The ground truth for this submission's performance assessment is the physical performance and measurements of the device (deflection) and compliance with industry standards (ASTM F543-23).

8. The sample size for the training set:

  • Not applicable. This device does not use machine learning or AI, so there is no training set.

9. How the ground truth for the training set was established:

  • Not applicable. As above, no training set.

Summary of the Study:

The "study" referenced in the 510(k) for EZ-Fiducials is a series of bench tests focusing on:

  • Biocompatibility of materials.
  • Sterility validation.
  • Packaging integrity.
  • Performance of the screws (e.g., deflection characteristics, compliance with ASTM F543-23 when inserted into bone simulant).
  • Performance of the electric screwdriver (its ability to drive screws adequately).

The clearance is based on the argument that these bench tests demonstrate the device's substantial equivalence in terms of safety and effectiveness to the predicate device (Medtronic Navigus Unibody Fiducial Marker System) and a reference device (OsteoMed Pinnacle Driver), even with some technological differences (e.g., screwdriver speed, sterilization method). The document explicitly states: "No clinical testing was needed or performed otherwise." The predicate device was also "cleared based upon bench testing alone."

FDA 510(k) Clearance Letter - EZ-Fiducials

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 9, 2025

Phasor Health, LLC
Ray King
Chief Executive Officer
8944 Kirby Drive
Houston, Texas 77054

Re: K250505
Trade/Device Name: EZ-Fiducials
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: HAW
Dated: June 9, 2025
Received: June 9, 2025

Dear Ray King:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250505 - Ray King Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250505 - Ray King Page 3

Sincerely,

Adam D. Pierce -S Digitally signed by Adam D. Pierce -S Date: 2025.07.09 09:43:44 -04'00'

Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250505

Device Name: EZ-Fiducials

Indications for Use (Describe)

Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary - Phasor Health, LLC

EZ-FIDUCIALS™

A. Device Description

CategoryComments
SponsorPhasor Health, LLC, Ray King, 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234
Correspondent Contact InformationRay King, 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234
Device Common Namesubdural fluid drainage kit
Device Regulation & Name21 CFR 882.4560
Classification & Product CodeClass II, HAW
Device Proprietary NameEZ-FIDUCIALS

Predicate Device(s) Information:

| Predicate Device(s)/ Reference Devices(s) | Navigus Unibody Fiducial Marker System/ OsteoMed Pinnacle Driver |
| Predicate Device/ Reference device Manufacturer(s) | (1) Medtronic, Inc., (2) OsteoMed; respectively |
| Predicate Device Common Name | (1) Navigus Unibody Fiducial Marker System; (2) Battery-operated drill/driver; respectively |
| Predicate Device Notification # | (1) K033619; (2) K162542 |
| Predicate Device Classification & Name | (1) 21 CFR 882.4560; (2) 21 CFR 882.4310; respectively |
| Predicate Device Classification & Product Code: | Class II for each of (1) HAW & (2) HBE, respectively |

B. Date Summary Prepared: July 8, 2025

C. Description of Device

Phasor EZ-FIDUCIALS™ provide fixed reference points during neurosurgical procedures. The device is composed of 3 items, all single-use, provided gamma-sterilized, and placed within the same primary packaging (sealed Tyvek tray within a shelf carton):

  1. Screws: Each of the four provided titanium screws is identical, with specific dimensions of each EZ-Fiducials screw as follows: screwhead circular shape screwhead with square-shaped engagement, non-threaded shaft diameter 2.9mm and length 15mm, threaded shaft diameter (major threads) 1.98mm, thread length 5mm; no protective caps are provided, and the screws should exclusively be utilized in conjunction with the screwdrivers (manual or EZ-Driver) provided together in the same tray (and with no other screwdriver);

  2. Manual Screwdriver: handheld, to be solely used with the screws provided, not intended for use with any other screws;

  3. EZ-driver: (electric, handheld, non-rechargeable battery-operated screwdriver, without software, single-use for solely tightening or loosening the provided screws exclusively, not intended for use with any other screws).

D. Indications for Use

Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.

E. Comparison of Technological Characteristics

Application DevicePredicate DeviceReference DeviceImpact on Substantial Equivalence
CompanyPhasor Health, LLCMedtronic, Inc.Osteomed Pinnacle

K250505, Page 1 of 3

Page 6

510(k)PendingK033619K162542Comparable
Regulation Number21 CFR 882.456021 CFR 882.456021 CFR 882.4310screws comparable to predicate device; accessory screwdriver similar to reference device
Product CodeHAWHAWHBEcomparable for predicate device
Device ClassIIIIIIComparable
Product CodeHAW (Class II)HAW (Class II)HBE (Class II)comparable for predicate device
Intended Use & Indications for UseIntended Use: Phasor EZ FIDUCIALS™ are intended for use as fixed reference points for presurgery CT imaging for patients requiring stereotactic surgery.Indications for Use: Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.Intended Use: The Medtronic Unibody Models FM-4007, FM-4010, and FM-4013 Bone Fiducials are intended for use as fixed reference points for presurgery CT imaging for patients requiring stereotactic surgery.Indications for Use: Navigus Unibody Fiducial Marker System is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging.Intended Use (and Indications for Use): The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with cranial surgical procedures.taken together, comparable to both devices
Technology Screwdriver Power SourceSingle-use Lithium non-rechargeable Batteries (for EZ-Driver); Manually operated for "Manual Screwdriver" includedManualDisposable, lithium, single use, sterile batterycomparable when considered together
Performance Characteristicsscrews into skull at threaded endscrews into skull at threaded endN/Acomparable screws to predicate device
Screwscrews into skull at threaded endscrews into skull at threaded endN/Acomparable screws to predicate device
Activation buttonsForward/Reverse button for EZ-Driver; N/A for manual screwdriver which can be turned in forward or reverse directionN/A for manual screwdriver which can be turned in forward or reverse directionForward/Reverse buttons for Drivercomparable to reference device
Insertion materialboneboneboneSame
Screw Sizes/Insertion2.0mm1.6mm1.2mm, 1.6mm, 2.0mmComparable
BiocompatibilityYesYes
SterileYesYes

K250505, Page 2 of 3

Page 7

Single-useYesYes
Speed of Electric Driver30rpmN/A120rpmdifferent, but ability to drive screw supported by performance testing
Performance Testing Standard(s) Met for ScrewsASTM F543-23No standard referencedNo standard referencedMet testing to noted standard for Application Device

F. Summary of Supporting Data

EZ-FIDUCIALS™ has the same intended use, patient population, and classification as the predicate. Accessory screwdrivers are comparable to the reference device noted. The subject and predicate devices are based upon the following technological elements, when the predicate and reference devices are taken together:

  • Sterile packaged for single use, and disposable.
  • OsteoMed Pinnacle Driver operates on single-use battery power, as per the reference predicate (lithium-ion battery);
  • Bone screws purchase into the skull in similar manner.

The following technical characteristics of the subject device differ from the predicate device:

  • Navigus Unibody bone fiducials are provided without an electric or manual screwdriver (although one would be needed) for insertion;
  • Navigus Unibody supplies "protective cap" and "screwdriver guide" (which are optional items, not required to the indications or intended use);
  • Osteomed Pinnacle driver is a high-speed driver with screwdriving and drilling capabilities, but otherwise comparable to EZ-Driver in that either can be used to drive screws (or remove them). Moreover, while Osteomed Pinnacle screwdriver can be used electrically or manually (in the still setting with power off to turn screws), Phasor EZ-Driver as part of EZ-Fiducials may be used similarly in power-on or power-off manual settings, but the EZ-Fiducials manual screwdriver is provided additionally to serve this same purpose of manual screw turning.
  • Navigus Unibody bone fiducials is sterilized using EO, whereas EZ-FIDUCIALS™ is sterilized with gamma with the manual screwdriver, electric screwdriver and bone fiducials/screws sealed together in a tray packaged in a shelf carton.

The proposed device does not appear to differ from the predicates taken together in any manner which would negatively impact the safety or effectiveness of the device.

G. Discussion of Performance Testing

Various tests including biocompatibility, packaging, sterility, safety, and performance have been validated with justifications as comparable and/or supported by documentation with adequate performance. Specifically, testing for a deflection of bone screws using a Coordinate Measuring Machine (CMM) with firm purchase into bone simulant was established and verified by testing. Further testing was performed to ensure the EZ-Fiducials screws was compliant with the ASTM F543-23 standard, with the electric screwdriver also tested and performing adequately (per report "02670-019011-1" including Table 6). No clinical testing was needed or performed otherwise.

H. Conclusion

The predicate device was cleared based upon bench testing alone. The performance and design validation testing conducted on the EZ-FIDUCIALS™ device on the bench demonstrated that it performs equivalent to the stated predicate and reference device that are currently marketed individually for the same intended use. The proposed device, comprised of the elements noted above, demonstrates that it should perform as safely as effectively as the predicate device.

K250505, Page 3 of 3

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).