Search Filters

Search Results

Found 235 results

510(k) Data Aggregation

    K Number
    K260533

    Validate with FDA (Live)

    Device Name
    ELISIO™-H
    Date Cleared
    2026-03-19

    (30 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , Florida 33172

    Re: K260533
    Trade/Device Name: Elisio™-h
    Regulation Number: 21 CFR 876.5860
    Name:** Dialyzer, High Permeability With Or Without Sealed Dialysate System
    Regulation Number: 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELISIO™-H dialyzers are intended for hemodialysis, hemodiafiltration, hemofiltration, and isolated ultrafiltration in patients with acute kidney injury or chronic kidney disease when conservative therapy is judged to be inadequate.

    Device Description

    The ELISIO™-H is a single-use, high-flux hemodialyzer intended for the extracorporeal treatment of patients with renal failure. The device is constructed of a plastic cylindrical housing comprised of two compartments, one for blood and one for dialysate, separated by a semi-permeable membrane. Blood is circulated through the blood compartment, while dialysate flows counter currently through the dialysate compartment. This flow pattern enables the removal of uremic toxins, excess fluid, and electrolytes via diffusion and/or convection, depending on the selected treatment modality.

    The dialyzer interfaces with the patient via an ISO 8637 compliant blood tubing set and must be used with dialysis machines equipped with an ultrafiltration controller or accurate fluid balancing system. The ELISIO™-H features the POLYNEPHRON™ membrane (Polyethersulfone), for high clearance efficiency and a polypropylene housing not made with BPA or DEHP.

    The device is available in eight sizes (surface area): 0.9, 1.1, 1.3, 1.5, 1.7, 1.9, 2.1, and 2.5 m². It is sterile (gamma radiation) and non-pyrogenic with a three-year shelf life. No integrated components or accessories are included.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252459

    Validate with FDA (Live)

    Date Cleared
    2026-02-10

    (189 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: multiFlux 130 (F00013123); multiFlux 160 (F00013124)
    Regulation Number: 21 CFR 876.5860
    Regulation Name:** High Permeability hemodialysis system
    Regulatory Class: Class II per 21 CFR § 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The multiFlux 130 and 160 filters are intended for single use only for extracorporeal blood purification during continuous veno-venous hemofiltration (CVVH), continuous veno-venous hemodialysis (CVVHD), and continuous veno-venous hemodiafiltration (CVVHDF) using pre-, post- or mixed-dilution modes for patients suffering from acute kidney injury.

    Device Description

    The multiFlux 130 and 160 hemofilters are high-flux, single-use, steam-sterilized filters. The hemofilters are provided blood pathway sterile and non-pyrogenic. The hemofilters allow for the transfer of water and solutes between blood and dialysate using semipermeable, hollow fiber membranes.

    The multiFlux 130 and 160 hemofilters are high-flux, sterile devices designed for single-use acute hemodialysis. The hemofilters are configured to connect to a bloodline set which connects to a patient's vascular access system when used with a hemodialysis machine equipped with ultrafiltration control. During hemodialysis, blood is pumped from the patient's body through an extracorporeal circuit, one component of which is the hemofilters. The hemofilters contain semi-permeable membranes that allow for diffusion and/or ultrafiltration to transport toxins and excess fluid from the blood compartment (fiber lumen) to the dialysate compartment. Hemofilters utilize a counter-current flow in which dialysate and blood flow in opposite directions in the hemofilter. The counter-current flow maintains the concentration gradient across the membrane for waste and fluid removal.

    The multiFlux 130 and 160 hemofilters' components are composed of the following materials:
    Housing: Polypropylene
    Potting Resin: Polyurethane
    Fiber Bundle: Polysulfone-polyvinylpyrrolidone blend, alpha-tocopherol (vitamin E)
    Sealing Ring: Silicone
    Flange: Polypropylene
    Blood Port Cap(s): Polypropylene
    Dialysate Port Cap(s): Styrol-Ethylen-Butylen-Styrol, Polypropylene

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253412

    Validate with FDA (Live)

    Date Cleared
    2026-01-26

    (118 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: Tablo® Hemodialysis System (PN-0008000, PN-0006000U)
    Regulation Number: 21 CFR§ 876.5860
    Hemodialysis delivery system and water purification system
    Classification Regulations: 21 CFR § 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tablo® Hemodialysis System and TabloCart™ are indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility, and/or in the home. Treatments must be administered under a physician's prescription and observed by a trained individual who is considered competent in the use of the device. Treatment types available include Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration.

    Device Description

    The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute, chronic care facility or in the home. The Tablo Hemodialysis System consists of:

    1. Tablo Console
    2. Tablo Cartridge

    The proposed modifications for Tablo Hemodialysis System (Console) does not affect the functionality of the Tablo cartridge, which was cleared under K190793 and K210782.

    The TabloCart with Prefiltration Drawer ("TabloCart") is an optional accessory to the Tablo Hemodialysis System that raises the height of the system, and features large, 360° rotating wheels to aid in system mobility. The Prefiltration Drawer features two replaceable cartridge filters (configurable with either two sediment (2S), two carbon (2C) or one sediment filter, and one carbon filter (SC)) that serve as "prefilters" to help remove sediment and/or chlorine/chloramines from supply water before it enters the Tablo's integrated water purification system.

    The Tablo Hemodialysis System is designed to operate and achieve its intended use with or without use of the optional TabloCart. The TabloCart consists of TabloCart Wheeled Platform, Fluidics Drawer and Software. The Tablo Hemodialysis System uses its integrated water purification system, consisting of carbon, sediment, RO, and ultra-filters, to generate AAMI-quality water for dialysate fluid, even where the water is prefiltered using the TabloCart.

    The proposed modifications for Tablo Hemodialysis System (Console) does not affect the functionality of the Tablo Cart, which was cleared under K232776.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253518

    Validate with FDA (Live)

    Date Cleared
    2025-12-18

    (36 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K253518
    Trade/Device Name: FX CorAL 40; FX CorAL 50
    Regulation Number: 21 CFR 876.5860
    Regulation Name:** High Permeability hemodialysis system
    Regulatory Class: Class II per 21 CFR § 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FX CorAL hemodialyzers are intended for single use only for extracorporeal blood purification during intermittent renal replacement therapies hemodialysis (HD), hemodiafiltration (HDF) using pre-, post or mixed-dilution modes, and isolated ultrafiltration for patients suffering from renal insufficiency, including pediatric patients.

    Consider body and dialyzer surface area, blood flow, body weight and extracorporeal blood volume when selecting dialyzers for use with pediatric patients.

    Device Description

    The FX CorAL dialyzers are high-flux, single-use, steam-sterilized hemodialyzers. The dialyzers are provided blood pathway sterile and non-pyrogenic. The dialyzers allow for the transfer of water and solutes between blood and dialysate using semipermeable, hollow fiber membranes.

    The FX CorAL dialyzers are high-flux, sterile devices designed for single-use acute and chronic hemodialysis. The dialyzers are configured to connect to a bloodline set which connects to a patient's vascular access system when used with a hemodialysis machine equipped with ultrafiltration control. During hemodialysis, blood is pumped from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzers contain semipermeable membranes that allow for diffusion and/or ultrafiltration to transport toxins and excess fluid from the blood compartment (fiber lumen) to the dialysate compartment. Dialyzers utilize a counter-current flow in which dialysate and blood flow in opposite directions in the dialyzer during hemodialysis. The counter-current flow maintains the concentration gradient across the membrane for waste and fluid removal.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243920

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-19

    (273 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K243920**
    Trade/Device Name: Purema® H Hemoconcentrator - Pediatric
    Regulation Number: 21 CFR 876.5860
    System
    Regulatory Class: Class 2
    Product Code: KDI
    Regulation Number: 21 CFR Part 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Purema® H Hemoconcentrator - Pediatric is designed to remove excess fluid from the blood in order to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass.

    It is intended to be used in a hospital setting, in connection with a suitable circuit for extracorporeal blood circulation and a pump that regulates its flow. There are no other accessories.

    The Purema® H Hemoconcentrator - Pediatric model DP03HC is intended to be used for pediatric patients.

    Device Description

    The Purema® H Hemoconcentrator - Pediatric is a single use hemoconcentrators containing filters composed of a hollow fiber made of polyethersulfone (PUREMA® H). The Purema® H Hemoconcentrator - Pediatric is a hemoconcentrators that can be used in treatments which require the removal of liquids that are in excess. Blood is pumped through a membrane that has high permeability, and the pressure gradient, through the transmembrane pressure (TMP) determines the passage of water and molecules with a mechanism that is similar to glomerular filtration (convective mechanism). The fraction of filtrated liquid depends on the osmotic pressure, hydrostatic transmembrane pressure, the surface, membrane permeability and patient hematocrit. The Purema® H Hemoconcentrator - Pediatric is designed to be used in a healthcare facility.

    Purema® H Hemoconcentrator - Pediatric is sterilized using Ethlyene Oxide (EtO). The EtO sterilized device consists of hollow fiber made of polyethersulfone (PUREMA® H), cartridge, rings, connectors, potting (fiber closure seals), and O-rings. The EtO sterilized Purema® H Hemoconcentrator - Pediatric is available in model DP03HC.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250508

    Validate with FDA (Live)

    Date Cleared
    2025-08-01

    (161 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K250508
    Trade/Device Name: AK 98 Dialysis Machine (955607)
    Regulation Number: 21 CFR§ 876.5860
    Dialysis Machine (955607)
    Classification Panel: Gastroenterology and Urology
    Regulation Number: 21 CFR 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Baxter AK 98 dialysis machine is intended to be used for intermittent hemodialysis and/or isolated ultrafiltration treatments of patients with chronic or acute renal failure or fluid overload upon prescription by a physician. The AK 98 dialysis machine is indicated to be used on patients with a body weight of 25 kg or more. The AK 98 dialysis machine is intended to be used by trained operators when prescribed by a physician, in a chronic care dialysis or hospital care environment. The Baxter AK 98 dialysis machine is not intended for Selfcare or home use.

    Device Description

    The AK 98 Dialysis Machine is a standalone hemodialysis machine intended for use as a single patient dialysis machine to perform HD treatments of patients with renal failure or fluid overload. The Vantive AK 98 Dialysis Machine is intended to be used in a chronic dialysis or hospitale care environment for intermittent hemodialysis and/or isolated ultrafiltration treatments of patients with chronic or acute renal failure or fluid overload upon prescription by a physician. The AK 98 Dialysis Machine is intended to be used on patients with a body weight of 25 kg or more.

    To perform hemodialysis therapy, the AK 98 system needs a high purity water source and the appropriate consumables for treatment, including but not limited to: cleaning products, ultrafilters, acid and base dialysate concentrates, bloodlines, and dialyzers.

    In hemodialysis therapies, the AK 98 is used to pump the blood through the extracorporeal circuit and to monitor the arterial and venous pressure, but the blood directly contacts only the disposable bloodline and dialyzer, not the monitor itself.

    The fluid unit of the AK 98 Dialysis Machine is used to produce the dialysis fluid (with the correct temperature, flow, and composition) from reverse osmosis water and concentrates (dry or liquid) and to transport the dialysis fluid through the dialyzer. The fluid unit offers Profiling of Ultrafiltration and/or conductivity, it also offers isolated Ultrafiltration. The fluid unit also includes "Clearance Measurement" (Diascan). The fluid unit maintains the dialysis fluid flow through the dialyzer and controls ultrafiltration. If a fault occurs, the machine enters a patient safe state, which, depending upon the fault, can include actions such as bypassing the dialyzer.

    The fluid path of the AK 98 is composed of a variety of different materials, including silicone tubing.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the AK 98 Dialysis Machine (K250508) do not contain the details typically associated with Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria for AI/ML-driven medical devices.

    The document describes a submission for a dialysis machine, and the key changes and testing detailed focus on physical components and performance attributes of a medical device (e.g., changes in silicone tubing curing process, ultrafiltration accuracy, blood flow rate accuracy). It specifically highlights non-clinical tests rather than clinical studies with human subjects or AI-specific validation.

    Therefore, many of the requested categories in the prompt (e.g., sample sizes for test and training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone AI performance) are not applicable to this type of device and submission as described in the provided text.

    The information that can be extracted relates to the device's functional performance and how it maintains equivalence to its predicate.

    Here's an attempt to address the prompt based only on the provided text, while noting the limitations:


    Acceptance Criteria and Device Performance for AK 98 Dialysis Machine

    Based on the provided 510(k) summary for the AK 98 Dialysis Machine (K250508), the device is demonstrated to be substantially equivalent to its predicate device (AK 98 Dialysis Machine, K232467). The assessment of differences focuses on ensuring the proposed device maintains similar performance and safety characteristics despite a material change in its silicone tubing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The concept of "acceptance criteria" in this context refers to maintaining the performance specifications of the predicate device. The "reported device performance" is implicitly that the proposed device performs identically to the predicate device for all listed features, having passed non-clinical functional testing. The key change evaluated was the transition from peroxide-cured and platinum-cured tubing to solely platinum-cured tubing.

    Feature / Performance MetricAcceptance Criteria (from Predicate Device K232467)Reported Device Performance (Proposed Device K250508)
    Intended Use/Indications for UseSame as predicate (intermittent hemodialysis/isolated ultrafiltration for patients 25 kg+ with renal failure/fluid overload, used by trained operators in chronic care/hospital, not for self-care/home use)Same
    Treatment ModalitiesHemodialysis (HD DN/SP, HD SN/SP)Same
    Dialysate Conductivity MonitoringYesSame
    Isolated UFYesSame
    Ultrafiltration ControlYesSame
    Ultrafiltration SupervisionYes, in accordance with IEC 60601-2-16, 4th editionSame
    Ultrafiltration AccuracyBetween ±50 and ±100 g/h (depending on UF rate); 50 mL or ± 50 mL/h x passed or treatment time (h) or ± 2.5% of accumulated UF volume, whichever is largest. Worst-case accuracy at 4 L/h UF rate is 100 g/h.Same
    Air DetectorYesSame
    Blood Leak DetectorYesSame
    Temperature MonitoringYesSame
    Fail-safe response during power failureYesSame
    Prescription ProfilingConductivity profiling (Na, HCO3)Same
    Disinfection ProgramsHeat, ChemicalSame
    Anticoagulant Administration Rate0 – 10.0 ml/hSame
    Anticoagulant Bolus0 – 10.0 mlSame
    Blood Flow Rate20 – 600 ml/minSame
    Blood Flow Rate AccuracyFor pre-pump pressure range -200 mmHg to 0 mmHg: ±10 ml/min or ±10% of set point, whichever is largestSame
    Dialysate Flow Rate300 – 800 ml/minSame
    Dialysate Flow Rate Accuracy±10% or 50 ml/min, whichever is largestSame
    Transmembrane Pressure-200 – +500 mmHg (calculated value) set of LIMITSSame
    Net Fluid Removal Rate0 – 4 L/hSame
    Dialysate Temperature33 – 40 °CSame
    Dialysate Conductivity Set Range9 – 16 mS/cmSame
    Arterial Pressure-400 - +300 mmHgSame
    Venous Pressure+10 - +500 mmHgSame
    Blood Pressure Measurements (BPM)YesSame
    IT ConnectivityYes, Integrates with CIS using HL7 protocolSame
    Silicone TubingPeroxide-cured and Platinum-cured tubingPlatinum-cured tubing

    Note on Silicone Tubing Difference: The crucial "acceptance criteria" here for the tubing change was that "Performance, biocompatibility, and extractables testing demonstrates that the difference in the curing process does not impact the performance or safety of the AK 98 Hemodialysis System." The reported performance is that this testing was successfully conducted, confirming no impact.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical functional performance testing, biocompatibility testing, and extractables and leachables assessment. It does not mention a "test set" in the context of patient data or clinical trials. Therefore, sample sizes for such data are not applicable here. The data provenance would be laboratory testing results from the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This submission pertains to a physical medical device and primarily relies on engineering and laboratory testing for equivalence, not AI/ML-driven analysis of medical images or patient data requiring expert adjudication. Thus, this information is not applicable.

    4. Adjudication Method for the Test Set

    As there is no mention of a clinical test set requiring human interpretation or AI output evaluation, an adjudication method is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done, as this device is a physical hemodialysis machine, not an AI/ML diagnostic or assistive tool for human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    No, a standalone algorithm performance study was not done, as this device does not contain a discrete AI/ML algorithm whose performance is being evaluated in isolation. Its "performance" refers to the mechanical and fluid dynamic functions of the dialysis machine itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation lies in established engineering standards, device specifications, and the performance characteristics of the legally marketed predicate device. The change in the silicone tubing's curing process required "Performance, biocompatibility, and extractables testing" to demonstrate continued safety and effectiveness. This testing serves as the "ground truth" method for verifying that the new component does not negatively impact the device's adherence to its established specifications.

    8. The Sample Size for the Training Set

    This concept is not applicable to the submission of this physical medical device. There is no AI/ML model being trained with a dataset.

    9. How the Ground Truth for the Training Set was Established

    This concept is not applicable as there is no training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    W/O CDX W/bibag BLUESTAR (191130)
    Regulation Number: 21 CFR 876.5860
    Regulation Name: High
    Regulation Name:** High Permeability Hemodialysis System
    Regulatory Class: Class II per 21 CFR §876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.

    Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20kg and ≤ 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤ 40 kg. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.

    Device Description

    The 2008T BlueStar Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
    The 2008T BlueStar Hemodialysis Machine provides hemodialysis treatment by controlling and monitoring both the dialysate circuit and the extracorporeal blood circuit. The machine pumps blood from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzer contains a semi-permeable membrane that uses diffusion to transfer toxins and ultrafiltration to transport excess water from the blood into the dialysate circuit. In this separate dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, and delivered to the dialyzer. Balancing chambers control the incoming flow and outgoing flow of the dialysate fluid during ultrafiltration. During hemodialysis, the extracorporeal blood circuit is monitored for venous and arterial blood pressures as well as for the presence of air and blood.
    The 2008T BlueStar Hemodialysis Machine has automation features (independent internal conductivity testing, auto priming, auto start, CDX auto on, assisted reinfusion) to minimize user workload and improve user experience. The 2008T BlueStar Hemodialysis Machine includes SLED (Sustained Low Efficiency Dialysis), Low Volume, and Isolated Ultrafiltration therapy options. The 2008T BlueStar Hemodialysis Machine also accommodates the use of the Patient Card system, a dialysis treatment information system.

    AI/ML Overview

    Based on the provided FDA 510(k) Clearance Letter, the device in question is the 2008T BlueStar Hemodialysis Machine. This document is a Class II Medical Device submission, which means it relies on substantial equivalence to a predicate device, rather than a clinical trial proving new efficacy. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are focused on demonstrating substantial equivalence to a legally marketed predicate device, rather than establishing new clinical effectiveness benchmarks through a de novo study.

    The primary study type proving the device meets the "acceptance criteria" for 510(k) clearance is a comparison to a predicate device and supporting non-clinical performance and safety testing.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" for a 510(k) submission are typically related to demonstrating that the new device meets the same performance specifications as the predicate and does not raise new questions of safety or effectiveness. The reported device performance is presented as its features and specifications.

    Feature / Acceptance Criteria (Derived from Predicate)Reported Device Performance (2008T BlueStar Hemodialysis Machine)
    Intended Use EquivalenceIdentical to predicate (K231125): "The 2008T Machine is intended for use in acute and chronic hemodialysis therapy."
    Indications for Use EquivalenceIdentical to predicate. Includes acute and chronic dialysis therapy in a healthcare facility, with options for Isolated Ultrafiltration, SLED, and low volume hemodialysis (patients ≥20kg and ≤40kg). Accommodates low/high flux dialyzers. Restrictions: not for SLED on patients ≤40kg, plasma replacement, patients <20kg, or renal therapies using substitution fluid.
    Design Specifications Equivalence"Design Specifications" are listed as equivalent to the predicate.
    Technological Characteristics Equivalence"Technological Characteristics" are listed as equivalent to the predicate.
    Principle of Operation Equivalence"Principle of Operation" is listed as equivalent to the predicate.
    Performance Requirements Equivalence"Performance Requirements" are listed as equivalent to the predicate.
    BiocompatibilityTested per ISO 10993-1:2018 and FDA guidance. Evaluated for Cytotoxicity, Sensitization, Irritation, Pyrogenicity, Hemocompatibility, Chemical Characterization, Toxicological Risk Assessment. (Specific pass/fail criteria are implicit in compliance with standards).
    Electrical Safety and EMCTested per IEC 60601-1-2 (2020). (Specific pass/fail criteria are implicit in compliance with standards).
    Software V&V (if changes)"No changes to the 2008T BlueStar Hemodialysis Machine's software." Therefore, software V&V was not required. (Implicit acceptance criteria: software stability/no new risks).
    Key Performance Specifications (Specifics compared)Blood Flow Rates: - 8 mm: 20–600 mL/min* (*Not available with Low Volume)- 6.35 mm: 20–465 mL/min- 4.8 mm: 10–274 mL/min- 2.6 mm: 6–86 mL/min- Accuracy: ±10% at -200 mmHg
    Maximum Dialysate Flow Rate: - Selectable: (0)/100†‡/150†‡/200†‡/300†/400/500/600/700/800 mL/min- Auto Flow (based on Qb): Ranges from 300 mL/min (for Qb 0-165 mL/min) up to 800 mL/min (for Qb 481+ mL/min)- Dialysate flow adjusts if blood pump adjusted ~15-20 mL/min.
    Net Fluid Removal: 0–4000 mL/hr- Accuracy: ± (1% UF rate + 18 mL/hr) at 100 mL/min Qd- Accuracy: ± (1% UF rate + 30 mL/hr) at 500 mL/min Qd- Accuracy: ± (1% UF rate + 48 mL/hr) at 800 mL/min Qd
    Dialysis Time: - Hemodialysis: 0–9:59 hours (adjustable)- SLED: Fixed at 12 hours- Accuracy: + 1 second per hour
    Dialysis Fluid Composition: - Volumetric, selectable- Acid adjustment: 130–155 mEq/L Na+- Bicarbonate adjustment: 20–40 mEq/L Bicarbonate- Monitoring conductivity average accuracy: ±1.5%
    Dialysis Fluid Temperature: - Range: 35°C–39°C (alarm window ±2°C)- Accuracy: ±0.3°C
    Heparin Delivery Rate: - 0–9.9 mL/hr- Accuracy: ±5%
    Human Factors Impact"Changes do not impact the usability of the device. Human Factors Validation Testing is not required."

    2. Sample Size Used for the Test Set and Data Provenance

    For a 510(k) of this nature (a hemodialysis machine), the "test set" does not refer to a clinical data set in the same way it would for an AI/ML algorithm. Instead, it refers to the units of the device that underwent various engineering and biocompatibility tests.

    • Sample size for test set: Not explicitly stated as a number of specific units in a formal statistical sense, but implies a sufficient number of devices were tested to demonstrate compliance with standards and specifications for various non-clinical tests (e.g., electrical safety, EMC, performance specifications). For biocompatibility, representative materials were tested.
    • Data provenance: Not directly applicable in the sense of patient data. The provenance of the testing is internal company testing (Fresenius Medical Care Renal Therapies Group, LLC). All testing is prospective in the sense that the tests were conducted specifically for this submission. The origin of the testing would be within the manufacturer's facilities or accredited testing labs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable to this type of 510(k) submission. This clearance is for a hardware device (hemodialysis machine) with embedded software, not an AI/ML diagnostic or prognostic device that relies on expert interpretation of medical images or patient data to establish ground truth for a test set. The "ground truth" for the device's performance is established by engineering specifications, validated sensors, and calibrated measurement systems.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "test set" of patient cases requiring adjudication by multiple experts for a hemodialysis machine clearance. Engineering tests are pass/fail based on predefined thresholds and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is not an AI/ML diagnostic or imaging device. There were no human readers or AI assistance studies conducted or required for this clearance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While the device contains software, it is an electromechanical device, not a standalone algorithm. Its performance is intrinsically linked to the physical machine. Therefore, a "standalone algorithm performance" study is not relevant or applicable. The software controls the machine's functions, and its performance is evaluated as part of the total system.

    7. The Type of Ground Truth Used

    The "ground truth" for the 2008T BlueStar Hemodialysis Machine's performance is established by:

    • Engineering Specifications: The design and expected performance metrics (e.g., flow rates, temperatures, accuracies) are the primary ground truth for the device's functional integrity.
    • International and Industry Standards: Compliance with standards like IEC 60601-1-2 for EMC and ISO 10993-1 for biocompatibility serves as the ground truth for safety and foundational performance aspects.
    • Predicate Device Performance: The underlying assumption for a 510(k) is that the predicate device's performance is the established "ground truth" for safe and effective hemodialysis delivery. The new device demonstrates equivalence to this established standard.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243505

    Validate with FDA (Live)

    Date Cleared
    2025-05-30

    (199 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Massachusetts 02451

    Re: K243505
    Trade/Device Name: 5008X Hemodialysis System
    Regulation Number: 21 CFR 876.5860
    Regulation Name:** High Permeability Hemodialysis System
    Regulatory Class: Class II per 21 CFR §876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 5008X Hemodialysis System is intended for use in acute and chronic hemodialysis therapy. Therapy options include hemodialysis (HD), hemodiafiltration (HDF), hemofiltration (HF), and isolated ultrafiltration (ISO).

    The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.

    The 5008X™ Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine. The bibag system is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag system is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.

    The 5008X™ Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.

    The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

    The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, nonpyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).

    The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

    The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.

    The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.

    The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.

    The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.

    The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.

    The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

    CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.

    Device Description

    The 5008X Hemodialysis System consists of the following:

    • The 5008X Hemodialysis Machine
    • Two (2) DIASAFE®plusUS Filters (hereinafter referred to as the "Diasafe filter(s)")
    • The 5008X Bloodlines are available in five (5) configurations:
      • 5008X HD/HDF – Standard Blood Tubing Set (hereinafter referred to as the "HD/HDF Standard Bloodline")
      • 5008X HD/HDF – With Twister Blood Tubing Set (hereinafter referred to as the "HD/HDF Twister Bloodline")
      • 5008X HD/HDF – With CLiC Blood Tubing Set (hereinafter referred to as the "HD/HDF CLiC Bloodline")
      • 5008X HD Pre-Flush – Standard Blood Tubing Set (hereinafter referred to as the "HD Standard Bloodline")
      • 5008X HD Pre-Flush – With Twister Blood Tubing Set (hereinafter HD Twister Bloodline")
    • CitraSure™ Disinfectant

    The 5008X Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.

    The Diasafe filter is a non-sterile dialysis fluid filter that produces ultrapure dialysate and sterile, non-pyrogenic substitution fluid as defined in ISO 23500-5. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment.

    The 5008X Bloodlines are single-use, ethylene oxide (EO) sterilized bloodlines.

    CitraSure™ disinfectant is a liquid chemical disinfectant for chemical heat disinfection of the 5008X Hemodialysis Machine.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the 5008X Hemodialysis System pertains to a medical device for renal treatment. The document does not describe an AI/ML device, a comparative effectiveness study (MRMC), or a standalone AI algorithm performance study. Therefore, sections of your request related to those topics (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, ground truth types, human reader improvement with AI) are not applicable and cannot be extracted from this document.

    However, I can provide the acceptance criteria and a summary of the performance testing conducted to prove the device meets these criteria based on the provided text for the various components of the 5008X Hemodialysis System.

    Acceptance Criteria and Device Performance Summary for 5008X Hemodialysis System

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document lists "Key Performance Specifications/Characteristics" which serve as the acceptance criteria for each component. The "Reported Device Performance" column summarizes the claims made in the document regarding the device meeting these specifications through various tests.

    Feature/Test ObjectiveAcceptance Criteria (Specification/Characteristic)Reported Device Performance (Summary of Test Results)
    5008X Machine
    Maximum Blood Flow Rate600 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Dialysate Flow Rate1000 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Substitution Flow RateAutoSub plus: 400 mL/min; Manual substitution: 600 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Net Fluid Removal0–4000 mL/hr; Accuracy: ± (1%UF + 0.15% of balanced fluid volume)System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis TimeIntermittent, typical time 4 hoursSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis Fluid CompositionVolumetric, selectable: Acid 125–151 mEq/L Na+; Bicarbonate 25–40 mEq/L; Monitoring conductivity average accuracy: ± 1.5%System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis Fluid TemperatureRange 34°C–39°C. Fixed alarm window 33°C (or 32°C during BTM recirculation measurement and preparation) or above 40°C.System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Heparin Administration Rate0.1 to 9.9 mL/hr; Accuracy: ± 0.5mL or ± 6%System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Electrical Safety and EMCIEC 60601-1-2 Edition 4.1 2020-09 complianceEMC testing conducted in accordance with IEC 60601-1-2.
    Software Verification & ValidationFunctional and Performance Verification, Regression Testing, Code ReviewsUnit, integration, and system level software verification testing performed to demonstrate efficacy and confirm operation.
    Human Factors ValidationSafe and effective use (per FDA guidance)Human Factors Validation Testing conducted.
    Diasafe Filter
    Bacterial and Endotoxin FiltrationProduces sterile, non-pyrogenic substitution fluid from dialysis fluid with max incoming water quality < 200 CFU/mL (bacteria) and < 2 EU/mL (endotoxin)Performance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Number Disinfection Cycles100 CitraSure cycles; 13 Pure Bright Bleach cyclesPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Use Life90 days maximum or if max disinfection cycles reachedPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    5008X Bloodlines
    Maximum Blood Flow Rate600 mL/minBlood Pump Endurance test performed. Results support substantial equivalence, safety, and efficacy.
    Minimum Labeled Arterial Pressure−300 mmHgStructural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions.
    Maximum Labeled Venous Pressure+500 mmHgStructural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions.
    Arterial Pump Segment8.0 mm/12.0 mm (ID/OD)Pump Segment Performance test evaluated characteristics over range of inlet pressures and flow rates.
    Twister ComponentFacilitates measurement of access flow by reversing blood flow while maintaining closed circuitTwister Rotation Test (180° ± 1°) and Twister Torque Test (≤ 7 inch-pound) performed.
    CLiC Chamber (Optional) – Hematocrit (HCT)Standard deviation X2 ≤ 3%, average bias ≤ 1% (of average HCT) on control blood chambers; measurement range 10–60 HCTFunctional CLiC Chamber Test (Hematocrit) performed.
    CLiC Chamber (Optional) – Oxygen Saturation (O2 Sat)Standard deviation X2 ≤ 3%, average bias ≤ 2% (of average O2 Sat) on control blood chambers; measurement range 30%–100% (when HCT ≥ 10)Functional CLiC Chamber Test (O2 Sat) performed.
    Structural Integrity (Bloodlines & Accessories)Withstand 1.5X max positive/negative pressures; Withstand pressures from faulty machine conditionsTesting performed.
    Pump Segment PerformanceEvaluate characteristics over range of inlet pressures (0 to -250 mmHg), flow rates, and treatment timeTesting performed.
    Blood Pump EnduranceNo break or detachment at max blood flow rates & pressures for 12 hoursTesting performed.
    Simulated UseNo tubing failures (kinking, collapsing, disconnection) under simulated use for ≥ 4 hours; components removable without external leaksTesting performed.
    Needle Access PortWithstand 1.5X recommended max positive/negative pressures after multiple accessesTesting performed.
    Luer Connectors (Performance)Meet ISO 80369-7 2021, Section 6Testing performed.
    Luer Connectors (Dimensions)Meet ISO 80369-7: 2021, Section 5Testing performed.
    Luer Connectors (Raw Material)Nominal modulus of elasticity > 700 MPaTesting performed.
    Transparency of Transducer ProtectorClear to allow visual inspection of blood contaminationTesting performed.
    Transducer Protector Leak TestMaintain secure and leak-free connection when subjected to 2X max labeled pressureTesting performed.
    Arterial Pressure Dome Leak TestMaintain secure and leak-free connection when subjected to 2X max labeled pressureTesting performed.
    CLiC Blood Chamber Lens Gap0.078 ± 0.005 in.Testing performed to verify optical property.
    Tubing Compliance TestCapable of being occlusively clamped by venous line clamp of 5008X machineTesting performed.
    Clamp Occlusion TestAir and liquid tightness occlusion at desired pressure and timeTesting performed.
    Check Valve Cracking PressureLower than 400 mbarTesting performed.
    Check Valve Pressure DropSpecific range when substitution fluid pumped at certain flow rateTesting performed.
    Tube TransparencyObserve interface of air and liquid during passage of air bubblesTesting performed.
    Tube Resistance to KinkingNo kinking at worst-case bend radius when packaged (evaluate % flow reduction)Testing performed.
    Tube ClampingResist clamp occlusions by subjecting to desired pressure and cyclesTesting performed.
    Tube PatencyMaintain patency after clamping and reopening (measure flow reduction)Testing performed.
    Air Bubble Trapping (Venous Chamber)Able to trap incoming air bubbles larger than 20 µLTesting performed.
    Recirculation Connector PackagingBurst strength, peel strength, dye penetration, microbial resistanceTesting performed.
    Viral Retentiveness (Transducer Protector)Prevent passage of bacteriophage (PhiX174) from patient to machine side up to 600 mmHg for 1 hrTesting performed.
    Sterility (Bloodlines)Sterility assurance level (SAL) 10-6Sterilized by 100% ethylene oxide (EO).
    EO Residuals (Bloodlines)< 4.6 mg/device for EO and EChTested per AAMI/ANSI/ISO 10993-7:2008/(R)2012.
    Bacterial Endotoxin (Pyrogenicity) (Bloodlines)Non-pyrogenic (< 20 EU/device)Tested per ANSI/AAMI/ST72:2019.
    Sterile Barrier Testing (Bloodlines)Maintain sterility of fluid pathStructural integrity test adapted from ISO 8637-2:2018 performed on samples after aging and distribution.
    CitraSure Disinfectant
    Minimum Disinfection Temperature70ºCPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Minimum Contact Time10 minutesPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Use Life90 DaysPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Shelf Life2 yearsPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.

    2. Sample Size for the Test Set and the Data Provenance

    As this is not an AI/ML device, the concept of "test set" in the context of diagnostic AI algorithms (e.g., patient images or clinical data) is not directly applicable. The performance data presented refers to engineering and quality assurance tests on the physical devices/systems.

    • Sample Size for Physical Device Testing: The document does not specify exact sample sizes for each individual engineering performance test (e.g., how many bloodlines were tested for structural integrity, how many machines for fluid flow accuracy). It generally states "Testing performed" or "Testing conducted."
    • Data Provenance: Not applicable in the context of patient data provenance (e.g., country of origin, retrospective/prospective). These are laboratory and simulated use tests on hardware components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    Not applicable. The "ground truth" for the performance of a hemodialysis system is based on direct physical measurements against defined engineering specifications and regulatory standards (e.g., flow rates, pressure tolerances, material properties, sterility levels), not expert interpretation of complex data like medical images.

    4. Adjudication Method for the Test Set

    Not applicable for a medical device's engineering performance testing. Adjudication methods like 2+1 or 3+1 typically refer to resolving discrepancies in expert interpretations (e.g., radiologists reviewing medical images for ground truth labeling in AI studies).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The document explicitly states: "No clinical studies were performed for the 5008X Hemodialysis System." MRMC studies inherently involve human readers (clinicians) assessing cases, often with or without AI assistance, to measure diagnostic performance and comparative effectiveness. This type of study was not conducted as this is not an AI-assisted diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. The device described is a physical hemodialysis system, not a standalone AI algorithm. While it contains software, the performance evaluation is for the integrated system, and there is no mention of a standalone AI algorithm being assessed.

    7. The Type of Ground Truth Used

    For the various performance tests, the "ground truth" is defined by:

    • Engineering Specifications and Standards: Numeric values for flow rates, pressures, temperatures, accuracies, dimensions, and material properties derived from established engineering principles and relevant international standards (e.g., ISO 8637-2:2018 for bloodlines, ISO 23500-5 for dialysate quality, IEC 60601-1-2 for EMC).
    • Regulatory Requirements: Compliance with FDA guidance documents and regulations (e.g., for biocompatibility, sterility, software verification and validation, human factors).
    • Absence of Failure: Demonstrating that components do not break, kink, leak, or fail under specified conditions (e.g., blood pump endurance, simulated use).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set. The software for the 5008X Hemodialysis Machine would have undergone development, verification, and validation, but this is distinct from AI model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI training set, there is no ground truth established in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242155

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-15

    (296 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)
    Regulation Number: 21 CFR 876.5860
    accessories

    Regulation Name: Dialyzer, High Permeability Dialysate System

    Regulation Number: 21 CFR 876.5860
    Hemodialysis System

    Regulation Name: High Permeability Hemodialysis System

    Regulation Number: 21 CFR 876.5860
    Delivery Device

    Regulation Name: High Permeability Hemodialysis System

    Regulation Number: 21 CFR 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended indications for use of the PICO 14 is for individuals with mild to moderate pain associated with carpal tunnel syndrome. It is designed to provide support and compression to the wrist, helping to alleviate discomfort and promote healing.

    Device Description

    A very compact and versatile smart plug with Wi-Fi connectivity, energy monitoring, and scheduling capabilities.

    AI/ML Overview

    The provided text describes the FDA clearance for the DBB-06 PRO Hemodialysis Delivery System. However, it does not contain detailed information regarding acceptance criteria for specific device performance metrics (e.g., sensitivity, specificity, accuracy) nor does it describe specific studies that prove the device meets such criteria in the context of an AI/algorithm-driven medical device.

    The document focuses on:

    • Safety and Performance of a Physical Hemodialysis Machine: It outlines various engineering specifications, material compatibility, sterilization validation, electrical safety, and software verification for a hemodialysis delivery system.
    • Substantial Equivalence: The primary goal of the 510(k) submission is to demonstrate that the DBB-06 PRO is substantially equivalent to predicate devices (DBB-06 Hemodialysis Delivery System and 2008T Bluestar Hemodialysis Machine).
    • General Performance Testing: The "Performance – Bench Testing" section lists various types of tests conducted (e.g., "System Level Performance Testing," "Disinfection Validation Testing," "Bacterial Retention Testing"), but it does not provide acceptance criteria for these tests nor the specific performance results in a way that would allow for the population of the requested table for an AI model.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to a physical medical device (hemodialysis machine) rather than an AI-driven diagnostic or prognostic algorithm. Specifically, sections 1-7 of your request are not applicable as presented because there is no AI/algorithm being evaluated for diagnostic or prognostic performance with associated metrics like sensitivity, specificity, or reader studies.

    The document does mention "Software controls the machine during hemodialysis treatment" and "Software Verification and Validation Testing," indicating software components are present, but these are for controlling the machine's operation, not for diagnostic interpretation or prediction in an AI context.

    Here's what can be extracted, and where limitations exist:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Limitation: The document provides "Key Performance Characteristics" (Table 1) for various mechanical and electrical aspects of the hemodialysis machine. These are specifications (e.g., "Flow rate accuracy: Set value ±10%") rather than acceptance criteria for an AI model's diagnostic/prognostic performance (like sensitivity, specificity, AUC). No reported performance results against specific acceptance criteria for AI are available.
    FeatureAcceptance Criteria (Specification/Characteristic)Reported Device Performance
    Blood PumpSetting range: 40 to 600 mL/min (ID 8.0mm) Flow rate accuracy: Set value ±10% (Inflow Pressure –150mmHg ≤ P ≤ +150mmHg); Set value -20 to 0% (Inflow Pressure -200mmHg ≤ P < -150mmHg) Flow rate reduction due to pump segment fatigue: -10% to 0% for typical treatment duration. Inflow pressure: -150 mmHg minimum, +150 mmHg maximum. Outlet pressure: +500 mmHg maximum. Alarm Limit (Flow rate): Upper Limit: Set value +10%, Lower Limit: Set value -10% Reminder alarm (BP OFF, BP cover open, blood flow set 0): 20 secondsImplied to meet these specifications as part of clearance. No specific measured values reported.
    Dialysis Flow RateSetting range: Single ETRF 300 to 800 mL/min; Double ETRF 300 to 700 mL/min Flow rate accuracy: Set value ±10% Protective system (Ultrafiltration) Method: Monitoring of Duplex pump rate with cell; Alarm limit: Upper +10%, Lower -10% Protective system (Ultrafiltration) Method: Monitoring of Duplex pump valve leak with cell; Alarm limit: 0.7V Dialysate Flow Rates can be automatically set based on provider's prescription (optional)Implied to meet these specifications as part of clearance. No specific measured values reported.
    Net Fluid Removal0 to 4000 (mL/hr), Profiled UFImplied to meet these specifications as part of clearance. No specific measured values reported.
    Dialysis TimeHemodialysis: 0 to 10 hours, time can be adjusted manuallyImplied to meet these specifications as part of clearance. No specific measured values reported.
    Dialysis Fluid Comp.Total conductivity range (mS/cm): 12.7 to 15.2 Monitoring conductivity average accuracy: ±-0.2mS/cm Maximum concentrate deviation alarm limit: ±5% Acid Setting range: 12.7 to 15.2 mS/cm, Measurement accuracy: ±0.2 mS/cm (±2 mmol/L) Bicarbonate Setting range: 2.3 to 7.0 mS/cm, Measurement accuracy: ±0.1 mS/cmImplied to meet these specifications as part of clearance. No specific measured values reported.
    Dialysis Fluid Temp.Setting range: 33.0 to 40.0°C (91.4 to 104 °F) Measurement range: 10.0 to 45.0 °C (50 to 113 °F) Measurement accuracy: Measurement value ±0.8°C (±1.4 °F) Fixed alarm limit: 41 °C (105.8 °F) Auto alarm limit: Set value +1 °C (+1.8 °F), Lower limit: Set value -1 °C (-1.8 °F)Implied to meet these specifications as part of clearance. No specific measured values reported.
    Heparin Delivery RateSetting range: 0.0 to 9.9 mL/h Output rate accuracy: Set value ±10% Back pressure: +500 mmHg Alarm limit: Upper +20%, Lower -20%Implied to meet these specifications as part of clearance. No specific measured values reported.
    Blood Volume Monitor (BVM)dBV monitoring accuracy: ±2.3dBV% (Double needle)Implied to meet this specification as part of clearance. No specific measured values reported.
    Blood Pressure Monitor (BPM)Pressurization-type: Pressure display accuracy: Less than ±3 mmHg Blood Pressure measurement accuracy: Conform to ISO 81060-2:2018/Amd.1:2020 Average within ±5mmHg, Delta standard deviation within 8 mmHg. Pulse rate accuracy: ±2% or ±2 beats (whichever is greater) Depressurization-type: Pressure display accuracy: Less than ±3 mmHg Blood Pressure measurement accuracy: Conform to ISO 81060-2:2018/Amd.1:2020 Average within ±5mmHg, Delta standard deviation within 8 mmHg. Pulse rate accuracy: ±2% or ±2 beatsImplied to meet these specifications as part of clearance. No specific measured values reported.
    Daily Dose Monitor (DDM)Kt/V monitoring accuracy: ±0.1 (Kt/V 0 to 1) ±10% (Kt/V 1 to 3) URR monitoring accuracy: ±5%Implied to meet these specifications as part of clearance. No specific measured values reported.
    Sterilization (EF-02D)Sterility Assurance Level (SAL) of 10⁻⁶"Validated to ensure the appropriate Steility Assurance Level (SAL) of 10⁻⁶". No specific log reduction or sterility test result values are provided.
    Bacterial Retention (EF-02D)Capable of producing ultrapure dialysate from spiked dialysis fluid, exceeding allowable limit of < 100 CFU/mL (ANSI/AAMI ISO23500-5:2019)Verified that the filter "can produce ultrapure dialysate". No specific CFU/mL reduction reported.
    Endotoxin Retention (EF-02D)Capable of producing ultrapure dialysate from spiked dialysis fluid, exceeding allowable limit of < 0.5 EU/mL (ANSI/AAMI ISO23500-5:2019)Verified that the filter "can produce ultrapure dialysate". No specific EU/mL reduction reported.

    Regarding AI/Algorithm-Specific Questions (2-7):

    The provided text does not contain any information related to an AI/algorithm for diagnostic or prognostic purposes, and therefore, these questions cannot be answered from the given document. The software mentioned "controls the machine during hemodialysis treatment" rather than performing analytical or interpretive tasks characteristic of AI in medical imaging or diagnostics.

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable. This is a physical device, not a diagnostic AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.

    Regarding Training Set Information (8-9):

    The document mentions "Software Verification and Validation Testing" but does not describe a "training set" in the context of machine learning or AI. The software is for controlling the hemodialysis machine's functions.

    1. The sample size for the training set: Not applicable. The software performs control functions for the machine, not an AI model that requires a training set of data.
    2. How the ground truth for the training set was established: Not applicable for the same reasons as above. The software is validated against functional requirements and design specifications, not against a "ground truth" for a learned AI task.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243264

    Validate with FDA (Live)

    Date Cleared
    2025-04-25

    (192 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ; DHF 0.6 Hemoconcentrator (DHF 06); SH 14 Hemoconcentrator (SH 14)
    Regulation Number: 21 CFR 876.5860
    14)
    Classification Name: High Permeability Hemodialysis System
    Regulation Number: 21 CFR 876.5860
    Dialyzer, high permeability with or without sealed dialysate system |
    | Regulation Number: | 21 CFR 876.5860
    Dialyzer, high permeability with or without sealed dialysate system |
    | Regulation Number: | 21 CFR 876.5860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hemoconcentrators are intended for use in cardiopulmonary bypass circuits for hemoconcentration and consequent restoring of patient physiological hematocrit. The choice of hemoconcentrator depends on the protocol being used and required filtration speed. The device is intended to be used for six hours or less.

    Device Description

    DHF and SH Hemoconcentrators are single-use, non-toxic and nonpyrogenic fluid path devices; they are supplied sterile and individually packaged. The devices are made of plastic materials and are recommended for use in cardiopulmonary bypass circuits for hemoconcentration and consequent restoring of patient's physiological hematocrit: The choice of hemoconcentrator depends on the protocol being used and required filtration rate. The device can be used up to 6 hours.

    The DHF and SH hemoconcentrators are the modified version of the disposables currently marketed in the Dideco DHF hemoconcentrators (K021732) and the SH 14 hemoconcentrators (K081313).

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the DHF and SH Hemoconcentrators does not describe a study that proves the device meets specific acceptance criteria in the manner of an AI/ML algorithm or diagnostic device.

    Instead, this document details a change to an existing, cleared medical device (hemoconcentrators used in cardiopulmonary bypass circuits). The core of the 510(k) submission is to demonstrate substantial equivalence to previously cleared predicate devices, not to establish performance against new, quantitative clinical acceptance criteria as one would find for a novel diagnostic or AI-powered system.

    The "study" described here is a non-clinical performance evaluation focused on demonstrating that a material change (from Santoprene to Silicone for O-rings) does not introduce new questions of safety or effectiveness.

    Therefore, many of the specific questions you've asked (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable to this type of device clearance.

    Here's an analysis based on the provided document, highlighting what is (and isn't) present:


    Analysis of Acceptance Criteria and Device Performance for DHF and SH Hemoconcentrators

    The information provided describes a 510(k) clearance for hemoconcentrators, which are physical medical devices, not an AI/ML or diagnostic software. The "acceptance criteria" and "study" are therefore framed around demonstrating substantial equivalence to existing predicate devices, particularly after a material change to a component, rather than performance metrics for a diagnostic algorithm.

    This document explicitly states: "No clinical testing was conducted in support of the DHF and SH hemoconcentrators, as the indications for use and technical characteristics are unchanged with respect to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use."

    The "study" instead focuses on non-clinical performance data to ensure that the device still complies with applicable standards and performs as expected after the specified design change.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of device and submission, acceptance criteria are generally met through compliance with recognized standards and demonstrating that the device's fundamental characteristics and performance are maintained despite the change. The document does not list specific quantitative performance criteria in a table format, but rather states compliance.

    Acceptance Criteria (Inferred from documentation)Reported Device Performance
    Material Biocompatibility and Safety:New silicone O-ring material demonstrated to be safe and biocompatible. (Implied by clearance)
    Mechanical Integrity/Functionality:Device continues to function as intended (e.g., maintain integrity, proper fluid path, non-pyrogenic). "Passed all testing in accordance with national and international standards."
    Sterility:Ethylene Oxide sterilized; non-pyrogenic fluid path maintained.
    Substantial Equivalence:The DHF and SH hemoconcentrators are deemed substantially equivalent to their predicate devices, raising no new questions of safety or effectiveness.
    Compliance with Voluntary Standards:Complies with all applicable voluntary standards related to Dialyzers.
    Intended Use Maintained:Intended for use in cardiopulmonary bypass circuits for hemoconcentration and restoring physiological hematocrit, for 6 hours or less. (Unchanged from predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a number of "cases" or "patients" because this was non-clinical performance testing (e.g., bench testing of prototypes or manufacturing samples), not a clinical study on patient data.
    • Data Provenance: The testing was conducted by Sorin Group Italia S.R.L. and is "non-clinical," implying laboratory or bench testing. The country of origin would be Italy (where Sorin Group Italia S.R.L. is located). It is not retrospective or prospective in the sense of a clinical trial; it is product performance verification.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not Applicable. This is a physical device, and the testing involved demonstrating compliance with engineering and safety standards, not establishing a "ground truth" for diagnostic purposes by human experts.


    4. Adjudication Method for the Test Set

    Not Applicable. There was no human adjudication process involved in assessing diagnostic performance. The evaluation was based on engineering and performance testing.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was NOT done. This type of study is relevant for diagnostic devices (especially imaging-based AI) to assess how human readers perform with and without AI assistance. This device is a hemoconcentrator, not an imaging or diagnostic AI system.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is not an algorithm. Performance was assessed for the physical device itself.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by engineering specifications, recognized national and international standards (e.g., for dialyzers), and performance parameters (e.g., filtration rates, material integrity, biocompatibility, sterility assurance). It is not based on expert consensus, pathology, or outcomes data from patients in the context of a new diagnostic claim. The "ground truth" is that the device, with the new material, still meets the same performance and safety requirements as the predicate device.


    8. The Sample Size for the Training Set

    Not Applicable. This device is not an AI/ML algorithm, so there is no concept of a "training set."


    9. How the Ground Truth for the Training Set was Established

    Not Applicable. As there is no AI/ML algorithm or training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 24