Re: K253536
Trade/Device Name: Evala Nerve Stimulator (EPNR002)
Regulation Number: 21 CFR 874.1820
Surgical nerve stimulator/locator |
| Classification Name | Stimulator, Nerve |
| Regulation Number | 874.1820
Evala is a single-use device intended to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability.
The Evala Nerve Stimulator is a small handheld device used by a surgeon to deliver electrical stimulation intraoperatively to test nerve integrity and muscle excitability. This is a sterile disposable device designed to be simple to use with one-handed control.
The device employs a bipolar stimulating probe to deliver a charge-balanced, biphasic electrical stimulus to nerve tissue, allowing surgeons to monitor the resulting muscle responses. The device contains embedded firmware that controls the function of the stimulator. The firmware cannot be modified by the user. The stimulator is powered by an internal battery which is not user replaceable. The surgeon is notified of stimulation delivery and device status via the visual indicators at the front of the device and vibratory (haptic) feedback felt on the body of the device.
N/A
Ask a specific question about this device
(256 days)
K251672**
Trade/Device Name: NIM Essence™ EMG Endotracheal Tube
Regulation Number: 21 CFR 874.1820
Nerve |
| Regulation Name: | Surgical nerve stimulator/locator |
| Regulation Number: | 21 CFR 874.1820
Comparison |
|---|---|---|---|
| Product Code | ETN | ETN | Same |
| Regulation Number | 21 CFR 874.1820
| 21 CFR 874.1820 | Same |
| Regulation Description | Surgical nerve stimulator/locator | Surgical nerve
The EMG Tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use.
The Medtronic NIM Essence™ EMG Endotracheal Tube (ETT) is a flexible, reinforced PVC endotracheal tube with an inflatable cuff. FPC (Flexible Printed Circuit) is attached and wrapped around the main shaft of the endotracheal tube, and silicone cover tube overlays on the FPC. The FPC electrodes are exposed only for a short distance, approximately 30 mm, slightly superior to the cuff. The electrodes are designed to make contact with the laryngeal muscles around the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to an EMG monitoring device. Both the tube and the cuff are manufactured from material that allows the tube to readily conform to the shape of the patient's trachea with minimal trauma to tissues. The NIM Essence™ EMG Endotracheal Tube is packaged as a sterile single-use device.
The Medtronic NIM™ 3.0 (cleared via K083124 on February 27, 2009) and NIM Vital™ (cleared via K200759 on October 28, 2020) nerve monitors are the recommended EMG monitors for use with the EMG Endotracheal Tube. All references to connections and technical specifications for EMG monitor contained in the instructions are made with references to the NIM™ family of devices. Medtronic recommends that the user consult the NIM™ User's Guide for determining proper settings of the NIM™ console and instructions for intraoperative EMG monitoring and motor nerve location and stimulation.
The NIM Essence™ EMG Endotracheal Tube is Type BF applied part.
N/A
Ask a specific question about this device
(125 days)
System (K234092) |
| Regulation Number (21 CFR) | §882.1870, §882.1320, §882.1350, §890.1375, §874.1820
, §874.1820 | §882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820 | Identical to predicate
The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.
SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, Delta Dilators and Stimulating Probe and Delta Clip contained in various kits.
The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.
The provided FDA 510(k) clearance letter and summary for the SafeOp 3: Neural Informatix System (K252842) primarily describe the device's substantial equivalence to a predicate device (K234092) and outline a Predetermined Change Control Plan (PCCP) for future modifications. However, it does not include a detailed study proving the device meets specific acceptance criteria in the context of clinical performance or AI algorithm validation for the initial clearance.
The "Performance Data" section explicitly states: "Determination of substantial equivalence is not based on an assessment of clinical performance data." and refers to nonclinical performance testing (functional, system, software requirements) and usability testing.
Therefore, the following information is extracted and synthesized based on the available document, but a comprehensive study proving acceptance criteria for all aspects of neural informatic performance is not present. The document focuses on the future PCCP for algorithmic changes, which will have performance criteria.
Acceptance Criteria and Reported Device Performance
The core of the documentation for this 510(k) clearance is about demonstrating substantial equivalence to an existing predicate device (SafeOp 3™: Neural Informatix System, K234092). Therefore, the "acceptance criteria" for the cleared device itself are primarily its ability to match the predicate's performance and specifications, which is shown through a direct comparison.
The document does not report specific quantitative performance metrics for the SafeOp 3: Neural Informatix System for its initial clearance, as it relies on substantial equivalence.
However, the Predetermined Change Control Plan (PCCP) outlines future modifications to the SSEP Baseline Classification Algorithm and defines acceptance criteria for these future algorithmic changes. These are the only explicitly stated performance acceptance criteria in the document.
Table 1: Acceptance Criteria and Reported Device Performance (from PCCP for future algorithmic changes)
| Acceptance Criteria (for future SSEP algorithm modifications) | Reported Device Performance (for future SSEP algorithm modifications) | Notes |
|---|---|---|
| Modification #1 & #4 (SSEP Baseline Classification Algorithm Retraining/Maintenance): | No reported performance data for the current cleared device. | The document states these are criteria for future modifications of the SSEP algorithm, not for the initial clearance of the SafeOp 3 System. |
| Overall Positive Predictive Value (PPV) ≥ 90% | To be demonstrated with verification testing of modified algorithm. | These criteria are goals for future changes. |
| Overall Accuracy ≥ 85% | To be demonstrated with verification testing of modified algorithm. | |
| Overall F1-score ≥ 85% | To be demonstrated with verification testing of modified algorithm. | |
| Performance of each stimulation site-specific classifier to be demonstrated as non-inferior to the current SafeOp 3 System at the 95% confidence level. | To be demonstrated with verification testing of modified algorithm. | |
| Modification #2 (Removal of Recalculating Active Trace (RAT) State): | No reported performance data for the current cleared device. | These are criteria for a future modification involving signal processing. |
| Average noise content for the reconstructed Ensemble Average (EA) following restart to be demonstrated as non-inferior to that of the preceding EA(s) with 95% confidence. | To be demonstrated with internal testing and retrospective analysis of intraoperative data. | These criteria are goals for future changes. |
Study Details (Based on the document, for the initial clearance)
Because the 510(k) is based on substantial equivalence and explicitly states "Determination of substantial equivalence is not based on an assessment of clinical performance data," there is no detailed "study" proving the device meets acceptance criteria in the way a de novo or PMA submission might. The "study" mentioned below refers to the verification and validation activities conducted to demonstrate substantial equivalence and safety/effectiveness for the initial clearance, as described in the "Performance Data" section.
-
Sample size used for the test set and the data provenance:
- Not explicitly stated for clinical performance data for the initial clearance. The document mentions "nonclinical performance testing" for functional, system, and software requirements, and "usability testing." These typically involve controlled test cases and user interactions rather than a large clinical test set.
- For the future PCCP modifications of the SSEP algorithm, it mentions "an independent, firewalled dataset" and "additional labeled data" for re-training and testing, but no sample size or provenance is specified within this document.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated for the initial clearance. Since the clearance is based on substantial equivalence and does not assess clinical performance data, detailed expert ground truth establishment for a clinical test set is not described.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated for the initial clearance. Not applicable given the reliance on substantial equivalence and lack of clinical performance data assessment.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. The document does not describe an MRMC study. The device is a neural informatic system, and while it involves signal processing, it's not positioned as an AI-assisted diagnostic tool that human readers would interpret with assistance in the traditional sense of an MRMC study. The SSEP algorithm mentioned in the PCCP appears to be a standalone classification, not a human-in-the-loop assistance tool subject to such a study.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- For the initial clearance, the document refers to "nonclinical performance testing" demonstrating the system meets functional, system, and software requirements. This implies standalone testing of device functions.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), the acceptance criteria (PPV, Accuracy, F1-score) clearly indicate standalone algorithm performance is intended to be evaluated. However, no such standalone performance data is provided for the initially cleared device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not explicitly stated for the initial clearance. For the "nonclinical performance testing" and "usability testing," ground truth would typically be established based on engineering specifications and validated operational procedures.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," which implies a ground truth established by experts or other definitive methods, but the specific type is not detailed in this document.
-
The sample size for the training set:
- Not explicitly stated for the initial clearance. No AI algorithms are reported as being evaluated for the initial clearance.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it mentions the algorithm "will be re-trained using site-specific data" and "periodically re-trained as additional labeled data become available," but no sample size is specified within this document.
-
How the ground truth for the training set was established:
- Not explicitly stated for the initial clearance.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," but the method of labeling (establishing ground truth) is not detailed in this document.
Ask a specific question about this device
(213 days)
Re: K243636
Trade/Device Name: Neuralytix iD3 System (NTX-9001)
Regulation Number: 21 CFR 874.1820
Common/Usual Name:** Surgical Nerve Stimulator/Locator
Device Classification: Class II per 21 CFR §874.1820
This device is intended for use in surgical procedures to assist in locating and mapping motor nerves through the use of mechanomyographic (MMG) signals and electrical stimulus of nerves. The device provides information directly to the surgeon to help assess a patient's neurophysiologic status by measuring and comparing MMG signals throughout a surgical procedure.
The device is intended to identify relative changes in the conduction and neural transmission ability of the nerve throughout a surgical procedure by measuring and comparing the minimum amount of electrical stimulation current (mA) required to induce a measurable MMG response (MMG nerve response threshold).
Examples of surgical applications which may require mechanomyographic (MMG) monitoring:
- Minimally invasive and open spinal surgery involving spinal fusion cages, screws, rods, plates, discs and biologics.
- Minimally invasive, open and endoscopic, direct and indirect nerve decompressions, discectomies, laminectomies, laminotomies, facetectomies, foraminotomies.
- Treatment of nerve compression, stenosis, degenerative disc disease, disc herniation, spondylolisthesis.
The Neuralytix iD3 System is a multichannel device for locating, mapping and assessing the status of motor nerves during surgical procedures. Neuralytix alerts the user of recorded mechanical activity (termed mechanomyography, or MMG) from muscles innervated by affected nerves, which may originate from operator applied electrical stimulus. The device will assist with nerve identification and assessment by alerting the surgeon when monitored nerves are activated. The device will also assist with tracking the status of monitored nerves throughout the course of surgical intervention. Neuralytix is especially useful in helping assess a patient's neurophysiologic status by measuring and identifying nerve response thresholds, or the minimum amount of electrical current (mA) necessary to elicit an MMG response. Neuralytix enables intuitive controls for measuring, recording and comparing nerve response thresholds throughout a surgical procedure to provide insights to the user as to how the conduction and neural transmission ability of a nerve may change throughout surgery.
The provided FDA 510(k) clearance letter and summary for the Neuralytix iD3 System offer details on device acceptance criteria and performance testing. However, it's important to note that the document primarily focuses on demonstrating substantial equivalence to a predicate device through engineering and performance characteristics, rather than a clinical multi-reader multi-case (MRMC) study or standalone AI performance evaluation as might be found for imaging AI devices.
Based on the provided text, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed within the context of demonstrating substantial equivalence to a predicate device (Sentio MMG Gen 2) and conformity to recognized consensus standards. The performance testing outlined is primarily engineering and bench-top validation rather than clinical studies with human readers or AI-only performance.
Here's a breakdown of the information as requested, with details extracted from the provided text and limitations noted where information is not present:
Acceptance Criteria and Device Performance for Neuralytix iD3 System
The acceptance criteria for the Neuralytix iD3 System are primarily defined by its ability to demonstrate substantial equivalence to its predicate device (Sentio MMG Gen 2) in terms of technological characteristics and performance, as well as adherence to relevant consensus standards for medical devices. The "study" proving acceptance is a compendium of non-clinical, bench-top performance tests, biocompatibility evaluations, and material characterizations.
The core function being tested for performance is "MMG-detection performance."
1. Table of Acceptance Criteria and Reported Device Performance
| Test / Attribute | Acceptance Criteria (Implied by "passed" as meeting standards/equivalence) | Reported Device Performance (Results/Conclusions) |
|---|---|---|
| MMG-Detection Performance (Comparative) | >99% agreement with predicate device (Sentio MMG Gen 2) when subjected to MMG-positive and MMG-negative waveforms. | All samples passed the acceptance criteria. Overall correlation was evaluated and confirmed to be >99% agreement. Substantially Equivalent. |
| Plug and Wire Disconnect Force Testing | At least 2.2N of retention force (normal to plug). | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Hydrogel Patch Adhesion Testing | At least 4N of retention force to HDPE (skin-approximating) test surface when pulled normal from surface by Sensor Wire or Ground Wire. | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Sterility | Sterility Assurance Level (SAL) of 10⁻⁶ (gamma irradiation per ISO 11137-1, 11137-2, 11137-3). | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Sterile Package Integrity (Initial & Real Time Aging) | Conformity to ISO 11607-1, ISO 11607-2, ASTM F88/F88M-21 (Seal Strength), ASTM F2096-19 (Bubble-Leak). | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Accelerated Aging (Non-Sterile Patches) | At least 4N retention force after 1 year accelerated aging per ASTM F1980-21, and PSTC 101 Type F pull testing. | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Environmental and Transport Stability | Withstand stated transportation conditions (ASTM D4332, ASTM D4169 Cycle 13) with subsequent functional testing and package integrity checks. | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| EMC and Electrical Safety | Compliance to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-40, and IEC TR 60601-4-2. | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Biocompatibility (Stim Probe) | Non-Cytotoxic, Non-sensitizer, Non-Irritant, Non-Toxic, Non-Pyrogenic, Non-Hemolytic for tissue/bone/CSF contact (ISO 10993 series). | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Biocompatibility (Sensor/Ground Kit) | Non-Cytotoxic, Non-sensitizer, Non-Irritant for skin contact (ISO 10993 series). | All samples passed the acceptance criteria. Design inputs satisfied. Substantially Equivalent. |
| Sensor Characterization | Performance within specification with respect to low-pass filtering and % error when compared to NIST traceable reference accelerometer at multiple frequencies. | All samples passed the acceptance criteria. Design Inputs satisfied. Substantially Equivalent. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Performance Tests: The document frequently uses the phrase "All samples passed the acceptance criteria," but it does not specify numerical sample sizes for most of the individual performance (bench-top) tests. For the "Comparative Performance Evaluation" of MMG-detection, it states "a statistically significant sample" was captured, but the exact number is not provided.
- Data Provenance: The data provenance is not explicitly stated in terms of country of origin or whether a clinical dataset was retrospective or prospective. Given the nature of the tests (bench-top, signal simulation), it's not applicable in the same way as for clinical image-based AI studies. The "clinically relevant MMG signal simulation" implies simulated data rather than data from actual human patients.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. The "Comparative Performance Evaluation" uses "clinically relevant MMG signal simulation" and compares the subject device's output to that of the predicate device, not to human expert interpretation of MMG signals. Therefore, there is no mention of experts establishing a ground truth for a test set in the traditional sense of clinical labeling or diagnostic interpretation.
4. Adjudication Method for the Test Set
- Not Applicable. No human adjudication method (e.g., 2+1, 3+1) is mentioned, as the primary performance evaluation is a direct comparison of the subject device's signal detection performance against a predicate device using simulated waveforms, not human interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. An MRMC comparative effectiveness study was not conducted based on the provided summary. The comparison is between the subject device's MMG detection performance and the predicate device using simulated waveforms, not with human readers.
- Effect Size: Not applicable, as no MRMC study was performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, implicitly. The "Comparative Performance Evaluation" section describes testing the Neuralytix iD3 System's "MMG-detection performance" by subjecting it (and the predicate device) to "MMG-positive and MMG-negative waveforms." This evaluates the device's technical ability to detect MMG signals in a controlled environment, which can be considered a form of standalone technical performance (algorithm/system only). However, this is distinct from testing a standalone AI algorithm's diagnostic performance on a clinical dataset.
7. The Type of Ground Truth Used
- The ground truth for the "Comparative Performance Evaluation" was based on "clinically relevant MMG signal simulation" (i.e., known MMG-positive and MMG-negative waveforms) which were then fed into both the subject and predicate devices. This is a technical ground truth based on simulated signals, not a clinical ground truth from patient outcomes, pathology, or expert consensus on patient data.
8. The Sample Size for the Training Set
- The document does not mention a training set or any deep learning/machine learning model training process. The device's function as described (measuring mechanomyographic signals and electrical stimulation thresholds) sounds like a traditional signal processing system rather than a machine learning-based AI system that would require a distinct training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as no training set or machine learning model for which a ground truth would need to be established is mentioned.
Summary Limitations:
The provided FDA summary primarily addresses the substantial equivalence of the Neuralytix iD3 System to a predicate device via detailed technical and engineering comparisons and bench-top performance testing. It does not describe clinical trials, MRMC studies, or specific AI model validation processes that involve human interpretation or large, diverse clinical datasets. The "Comparative Performance Evaluation" is a technical test against simulated signals, not an evaluation of diagnostic accuracy in a clinical setting by human readers or an AI.
Ask a specific question about this device
(266 days)
I.D. 7.5, O.D. 10.3 (520846), ALM Tube I.D. 8.0, O.D. 11.0 (520847)
Regulation Number: 21 CFR 874.1820
ALM Tube
Regulation Medical Specialty: Ear, Nose & Throat
Classification Regulation: 21 CFR 874.1820
The ALM Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative neuromonitoring by recording the EMG activity of the laryngeal musculature when connected to an appropriate neuromonitoring device.
This device must be used in connection with the C2 Xplore® or any approved IEC 60601-1, compatible EMG monitoring system with 42802 DIN compatible connectors.
The ALM Tube is an endotracheal intubation tube which combines the function of airway ventilation and intraoperative neuromonitoring purposes regarding recording of electrophysiological signals. It combines the recording of EMG signals from the vocal cords to monitor the recurrent laryngeal nerve or the vagal nerve function during surgical procedures. The product is used intraoperatively and is connected to an appropriate neuromonitoring device for signal recording outside of the sterile field via a connection cable. The connection cable is not part of the delivered package.
This FDA 510(k) clearance letter details a medical device, the ALM Tube, but it does not describe an AI/ML powered device or a study comparing its performance against acceptance criteria in the manner requested.
The questions in your prompt are geared towards evaluating AI/ML device performance and related studies (e.g., sample size for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment). The ALM Tube is an endotracheal intubation tube with integrated electrodes for neuromonitoring, which is a hardware device.
Therefore, many of the requested details cannot be extracted from this document because they are not applicable to the type of device described.
Here's an attempt to answer the questions based on the provided text, highlighting where information is absent or irrelevant:
1. A table of acceptance criteria and the reported device performance
The document describes the device's adherence to various international standards and internal requirements rather than specific, quantifiable acceptance criteria with corresponding performance metrics like sensitivity/specificity for an AI model.
| Acceptance Criteria Category (implied from standards/testing) | Reported Device Performance |
|---|---|
| Biocompatibility (ISO 10993 series) | Complies with applicable standards. |
| Sterilization (ISO 11135) | Complies with applicable standards. |
| Packaging & Shelf-life (ISO 11607 series) | Complies with applicable standards. |
| Performance Testing – Bench (ISO 5356-1, ISO 5361) | Complies with applicable standards, except for clause 6.7.1 of IEC 5361:2023. (No further detail on this exception). |
| Electrical Insulation | Electrical insulation on all surfaces until the head of the electrode. |
| Electrode Impedance | Average value of impedance measured within neuromonitoring comparative testing: 0.27 kΩ. (Comparable to predicate's 0.30 kΩ). |
| Software Functionality | Not applicable (device does not contain software). |
| Electrical Safety | Not electrically powered. Must be used with IEC 60601-1 and IEC 60601-1-2 compliant neuromonitoring system. |
| Electromagnetic Compatibility | Not electrically powered. Must be used with IEC 60601-1 and IEC 60601-1-2 compliant neuromonitoring system. |
| Human Factors / Usability | User interface evaluation conducted as a knowledge test with representative US citizens. |
The study that proves the device meets the acceptance criteria is a compilation of various engineering, bench, biocompatibility, sterilization, and packaging tests, as well as a human factors evaluation. This is stated in the "Summary of Performance Testing" section and the "Conclusion," which claims: "Verification and validation activities and human factors engineering testing were undertaken in order to establish the performance and safety characteristics of the ALM Tube. The results of these activities demonstrate that the ALM Tube is as safe, effective and perform as well as the predicate device, despite the differences in technology between the subject and predicate device."
Missing/Not Applicable Information based on the provided document:
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- N/A. The document states: "No additional clinical testing was performed for the ALM tube. Therefore, this section does not apply." The performance evaluation relies on bench testing, biocompatibility, sterilization, and packaging validation. For the human factors testing, the 'sample size' is described as "representative US citizens" for a knowledge test, but no specific number is provided.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- N/A. No "ground truth" establishment by medical experts is described, as this is not an AI/ML diagnostic or prognostic device requiring expert consensus on clinical cases.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set
- N/A. No expert adjudication of clinical cases is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- N/A. The device is a physical endotracheal tube with neuromonitoring capabilities, not an AI assistance tool for human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- N/A. This is not an algorithm-only device. It requires connection to an "appropriate neuromonitoring device" and is used in a clinical setting by medical professionals.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- N/A. For this type of physical medical device, "ground truth" typically refers to established engineering specifications, material properties, and standardized test methodologies. The document indicates adherence to relevant ISO and IEC standards for these aspects.
-
The sample size for the training set
- N/A. This device does not use a training set as it is not an AI/ML product.
-
How the ground truth for the training set was established
- N/A. This device does not use a training set.
Ask a specific question about this device
(125 days)
BMAC Kit)
Common Name: Aspiration Needle, BMA Needle
Classification: 21 CFR 876.1075, 21 CFR 874.1820
Number** | Subject Device | K131157 | K190163 | K170539 |
| Regulation | 21 CFR 876.1075 21 CFR 874.1820
| 21 CFR 876.1075 | 21 CFR 874.1820 | 21 CFR 876.1075 |
| Classification | II | II | II | II |
|
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.
The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
The ALARA BMAC Needle is intended for use in harvesting core biopsy samples of cancellous bone and/or bone marrow.
The ALARA BMA Neuro Access Kit utilizes a neuromonitoring platform in providing the surgeon with nerve / root location feedback during the approach to a pedicle to support a guidewire placement.
The ALARA BMAN and ALARA BMAC Kit includes various single-use needles designed for bone marrow aspiration and aspiration of autologous blood.
This document is an FDA 510(k) clearance letter for a medical device called the "ALARA BMA Neuro Access Kit." This type of document primarily focuses on establishing "substantial equivalence" to predicate devices, rather than presenting detailed study results for a new AI/ML-driven diagnostic device.
Therefore, the provided text does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML diagnostic. The "Performance Testing" section lists various engineering and material tests (e.g., torque, pull, strength, sterilization, biocompatibility) relevant to the physical components of the device, not performance metrics for an AI algorithm.
Here's why the requested information cannot be extracted from this document:
- No AI/ML Component: The device description and performance testing sections indicate a physical medical instrument (needles for biopsy, aspiration, and a neuromonitoring platform for "proximity feedback" during pedicle preparation). There is no mention of an AI or machine learning algorithm. The "proximity feedback" is likely an electrical or mechanical system, not an AI interpreting images or data.
- Focus on Substantial Equivalence: A 510(k) clearance is primarily about demonstrating that a new device is "substantially equivalent" to a legally marketed predicate device. This often relies on comparing design, materials, intended use, and non-clinical performance (like the engineering tests listed). It does not typically involve the rigorous clinical efficacy studies or AI performance validation described in your prompt.
- Lack of Diagnostic Claims: The device is for "aspiration of bone marrow or autologous blood," "pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback," and "harvesting core biopsy samples." These are procedural/interventional uses, not diagnostic uses that would involve interpreting data (image, signal, etc.) with an AI.
In summary, as the provided document pertains to a physical medical device and not an AI/ML-driven diagnostic, the questions about acceptance criteria for AI performance, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, and ground truth establishment for AI are not applicable to the content provided.
Ask a specific question about this device
(29 days)
.: | 21 CFR Part 882.4250 and 874.1820
The iovera® System is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera System is not indicated for treatment of central nervous system tissue.
When stimulation compatible components are used, the iovera System can also facilitate target nerve location by conducting electrical nerve stimulation from a compatible 3rd party nerve stimulator.
The iovera System is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, NoO) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization. The iovera8 System may be used in conjunction with standard of-the-shelf nerve stimulator device in applications where precise nerve location is desired.
The device is comprised of four main components:
- A reusable Handpiece .
- A Charging Dock ●
- An assortment of single-patient-use Smart Tips ●
- A Cartridge containing nitrous oxide ●
The provided text describes the iovera® System, a cryogenic surgical device cleared by the FDA (K243677). The document focuses on demonstrating substantial equivalence to predicate devices (K220656 and K211334) for a new Smart Tip configuration (STT21180STIM) that also facilitates nerve location.
Here's a breakdown of the acceptance criteria and study information, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily relies on adherence to recognized standards and satisfactory completion of design verification testing. Specific quantitative acceptance criteria for each test (e.g., specific thresholds for biocompatibility, precise tolerances for mechanical tests) are not explicitly detailed in this summary, but the general outcome of "meets all ISO10993-1 and USP acceptance criteria" or "complied with standards" is stated.
| Acceptance Criterion Category | Reported Device Performance (Summary) |
|---|---|
| Biocompatibility | Conformed to ISO 10993-1:2022. No signs of acute systemic toxicity and pyrogenic response, sensitization, or skin irritation. All extractables below Analytical Evaluation Threshold (AET), no unexpected contaminants or material degradation. Negligible toxicological risk with a significant margin of safety. |
| Electrical Safety & EMC | Complied with IEC 60601-1:2005+AMD1:2012+AMD2:2020 Edition 3.2, IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Edition 3.2, and IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 Edition 2.2 for safety (e.g., electric shock), usability, and alarms. Met all EMC requirements per IEC TS 60601-4-2:2024 (specific clauses listed). |
| Software V&V | Documentation provided per FDA guidance. Software classified as "Class B" concern. Followed IEC 62304:2006/AMD1:2015 and relevant FDA guidance documents for software development, risk management, software life cycle processes, and cybersecurity. Specific performance metrics are not provided but the implication is that all V&V activities were completed satisfactorily. |
| Functional & Product Perf. | Testing performed on Handpiece, stim cable, cartridges, and Smart Tip STT21180STIM. Indicated that the device meets product specifications and is safe and effective. Specific performance metrics (e.g., temperature ranges, cooling times) are listed in the comparison table for different Smart Tips, and implied to be met for the new STT21180STIM tip. For example, cooling time for STT21180STIM is 70 seconds. Post-heat time is 15 seconds. Skin Warmer Set Point is 15°C. |
| Usability Engineering/UI | Conformed to IEC 62366-1 Ed. 1.1. Assessed usability by identifying hazards and hazardous situations associated with the user interface, ensuring safety features are intuitive and support proper usage. Simulated use testing (Human Factors Study) with the Smart Tip and Stim cable assembly attached was performed. |
| Mechanical & Resistance | Resistance Testing (difficulty in securing connection between needle and PCBA), Flex Introducer and Leak Test (to ensure needle does not break), Heater Block-to-Needle Bond Strength Test. All tests were conducted successfully, indicating mechanical strength and reliability. Specific quantitative acceptance criteria are not mentioned beyond "ensuring mechanical strength and reliability." |
| Packaging Evaluation | Conformed to ASTM D4332-22 and ASTM D4169. Package Integrity Peel Strength Testing, Inspection, and bubble test performed. Validated packaging integrity and compliance with ISO 11607 requirements for sterile barrier systems. ASTM F1886/F1886M-16 and ASTM F88/F88M-23 are referenced. |
| Sterility Assurance | Conformed to ASTM F1980-21 and ISO 11135:2014. Sterilization validation performed (EO). Smart Tip individually packaged and sterile with a SAL of 10-6. Ensured compliance. Also referenced EN ISO 10993-7:2008+A1:2022 (Ethylene oxide sterilization residuals) and BS EN ISO 11737-1:2018+A1:2021 (Microbiological methods-Determination of a population of microorganisms on products). |
2. Sample Size Used for the Test Set and Data Provenance:
The document summarizes design verification testing rather than a clinical study with a "test set" of patient data. Therefore, information about sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or provided in this regulatory summary. The testing focuses on engineering verification (e.g., number of units tested for mechanical properties, number of samples for biocompatibility, etc.), not patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
Again, this pertains to a clinical study with patient data and does not apply to the design verification testing described. The "ground truth" for the engineering tests would be the compliance with recognized standards and specified performance parameters.
4. Adjudication Method for the Test Set:
Not applicable, as this is design verification testing rather than a clinical study requiring consensus from experts on case interpretation.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence through non-clinical performance testing and comparison to predicate devices, not on evaluating human reader performance with or without AI assistance. The device in question is a physical cryogenic surgical device, not an AI diagnostic tool that assists human readers with interpretations.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:
Not applicable. The iovera® System is a physical medical device, not an algorithm. Performance is assessed through engineering tests of the device itself. While it facilitates nerve location, its primary function is direct tissue destruction via freezing, with a human operator always in the loop.
7. The Type of Ground Truth Used:
For the design verification testing, the "ground truth" used is:
- Compliance with recognized international and national standards (e.g., ISO 10993-1, IEC 60601-1, IEC 62366-1, ISO 11135, ASTM standards).
- Pre-defined product specifications and design input requirements for physical, electrical, mechanical, software, and biological properties.
- Laboratory test results (e.g., chemical analysis for extractables, temperature measurements for cooling performance, force measurements for bond strength).
8. The Sample Size for the Training Set:
Not applicable. This is not a machine learning or AI device that requires a training set of data.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as there is no training set for this type of medical device submission.
Ask a specific question about this device
(448 days)
06800 Turkey
Re: K233001
Trade/Device Name: Bioscope Neuromonitor Device Regulation Number: 21 CFR 874.1820
Common Name | Nerve Monitor |
| Regulation Number | 21 CFR 874.1820
The Bioscope Neuromonitor Device is intended to help surgeons locate, identify, and preserve cranial motor nerves during surgery. Intra-operative monitoring and stimulation of cranial peripheral motor nerves.
Indications for Bioscope Neuromonitor Device Procedures include: thyroidectomy and parathyroidectomy, mastoidectomy.
Bioscope Neuromonitor Device is a nerve locator and monitor. The device has 3 infrastructure;
- Electronic
- Mechanic
- Software
These infrastructures are integrated with each other and work with the principle of stimulating current at 0.01-10 mA levels, conceiving EMG signals from related muscles groups. Collected signals are interpreted by the device and auditory and visual notifications will be given.
The device basically consists of stimulation and EMG subsystems. Both stimulation and EMG lines can send and collect signals from 2 channels. Depending on the operation, an EMG Endotracheal Tube or needle electrode is inserted into the muscles innervated by the relevant nerves for EMG reading. The electrode parts of the EMG tube are placed in contact with the muscles attached to the vocal cords. Needle electrodes are also inserted into the relevant muscle. Appropriate current is sent with the stimulation probe. The electrical changes in the muscles in the region and are transmitted to the device as a signal. The device gives audible and visual warnings.
Channels works on the principle of potentially stimulating the critical region through monopolar or bipolar probe connection and completing the circuit, allowing the device to make a nerve-tissue separation audibly and visually. It allows the detection of nerves such as the recurrent laryngeal nerve, which are connected to the vocal cords. The doctor brings the probe into contact with the area of potential nerve risk. The current sent to the patient through the probe causes a change in the electrical activity of the nerves and the muscles to which the nerves are connected, and this change can be detected by the device. Based on the collected data, the device gives clear audible alerts for nerve-tissue separation. The EMG signal collected from the patient is displayed as a both voltage value and latency times on the output. These values can be recorded and the interface can be used to select specific details, such as the right and left sides of the vagus, and recurrent laryngeal nerves.
The Bioscope Neuromonitor Device is intended to help surgeons locate, identify, and preserve cranial motor nerves during surgery. Intra-operative monitoring and stimulation of cranial peripheral motor nerves.
Indications for Bioscope Neuromonitor Device Procedures include: thyroidectomy and parathyroidectomy, mastoidectomy.
Here's an analysis of the acceptance criteria and study information provided:
1. Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state quantitative performance acceptance criteria (e.g., minimum sensitivity, specificity, or accuracy targets) for the Bioscope Neuromonitor Device. Instead, the acceptance criteria are implied through adherence to recognized medical device standards and the demonstration of "similar technical features" and "safety and effectiveness" compared to predicate and reference devices.
The reported device performance is primarily described as compliance with various international standards for medical electrical equipment, software life cycle processes, and specific requirements for nerve and muscle stimulators. The conclusion from nonclinical tests is that the device is "as safe, as effective, and perform as well as the legally marketed predicate device."
Table of Acceptance Criteria (Implied) and Reported Device Performance:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Safety and Essential Performance | IEC 60601-1 Ed. 3.2 2020-08 CONSOLIDATED VERSION – Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (with US national differences applied). |
| Electromagnetic Compatibility (EMC) | ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] – Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance (collateral standard: electromagnetic disturbances – requirements and tests). |
| Usability | ANSI AAMI IEC 60601-1-6 Ed. 3.2 2020-07 CONSOLIDATED VERSION – Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. |
| Software Life Cycle Processes | ANSI AAMI IEC 62304:2006/A1:2016 – Medical device software - Software life cycle processes [Including Amendment 1 (2016)]. Validation of software in accordance with EN 62304. |
| Electrical Equipment Safety for Measurement Control | ANSI AAMI IEC 61010-1 Ed. 3.1 2017-01 CONSOLIDATED VERSION – Safety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements. |
| Nerve and Muscle Stimulator Specific Requirements | IEC 60601-2-10 Ed. 2.1 2016-04 – Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. |
| Alarm Systems in Medical Electrical Equipment | IEC 60601-1-8 Ed. 2.2 2020-07 CONSOLIDATED VERSION – Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. |
| General Performance (Functionality, Accuracy, Compatibility) | System Performance test conducted, testing: device opening, closing, parameter change, accuracy of current output, accuracy of signal detection, compatibility with consumables. Results demonstrated compliance. |
| Overall Equivalence to Predicate Device | Nonclinical tests concluded the device is "as safe, as effective, and perform as well as the legally marketed predicate device." The identified differences (hardware, bandwidth, software, screen size, stimulation range/type/frequency, print capacity) were justified as not posing significant risks to safety or performance. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The provided document describes non-clinical performance testing. This type of testing typically involves laboratory-based evaluations, bench testing, and simulations to demonstrate compliance with standards and functional requirements. It does not mention clinical studies involving patient data or human subjects. Therefore, there is no information provided regarding a "test set" in the context of patient data (e.g., sample size, country of origin, retrospective/prospective).
The "test set" in this context refers to the device itself and its components undergoing various engineering, electrical, and software validation tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
As the testing described is non-clinical performance testing (bench testing, standard compliance), the concept of "ground truth" established by experts for a test set of patient cases is not applicable here. The "ground truth" in this context would be the specifications and expected performance values defined by the relevant standards and the device design. The "experts" would be the engineers, quality assurance personnel, and regulatory specialists conducting and reviewing these tests. No specific number or qualifications of such experts are mentioned beyond the general reference to "Test results demonstrate that the Bioscope Neuromonitor Device complies with the applicable standards."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Again, since this is non-clinical performance testing and not a clinical study involving human assessment of cases, the concept of an "adjudication method" for a test set is not applicable. Test results are typically evaluated against established technical specifications and standard requirements, not through expert consensus on case interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was mentioned or performed. This submission describes a conventional medical device that provides intra-operative monitoring and stimulation. It is not an AI-assisted diagnostic or interpretative device that would typically involve human readers or AI assistance for image interpretation.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The Bioscope Neuromonitor Device is an active medical device that provides real-time information to a surgeon during surgery. It is not an "algorithm only" device or an AI system that operates without human-in-the-loop performance. Its function is to assist the surgeon by providing auditory and visual notifications based on EMG signals. Therefore, the concept of a "standalone (algorithm only)" performance study without human interaction is not applicable to this device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance testing, the "ground truth" used is the defined technical specifications, functional requirements, and the pass/fail criteria established by the international standards listed (e.g., IEC 60601-1, IEC 62304, etc.). For instance, for accuracy of current output, the ground truth would be the known, calibrated output of measurement equipment used to verify the device's output.
8. The sample size for the training set
No information about a "training set" is provided. As this is a conventional neuro-monitoring device based on electrical stimulation and EMG signal detection, it does not appear to involve machine learning or AI models that would require a training set in the typical sense. The software validation is mentioned in the context of IEC 62304, which focuses on software life cycle processes and validation, not on training an AI model.
9. How the ground truth for the training set was established
Since no training set is mentioned or applicable to this type of device, there is no information on how its ground truth would have been established.
Ask a specific question about this device
(154 days)
|
| Classification: | 21 CFR 876.1075, 21 CFR 874.1820
|
| Regulation | 21 CFR 876.107521 CFR 874.1820
876.1075 | 21 CFR 874.1820
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.
The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
The ALARA BMA Neuro Access Kit utilizes a neuromonitoring platform in providing the surgeon with nerve / root location feedback during the approach to a pedicle to support a guidewire placement.
The ALARA BMAN and ALARA BMAC Kit includes various single-use needles designed for bone marrow aspiration and aspiration of autologous blood.
The provided text describes the regulatory clearance of the ALARA BMA Neuro Access Kit, focusing on its substantial equivalence to predicate devices based on design, materials, and non-clinical testing. However, it does not include acceptance criteria, reported device performance, or details of a study that proves the device meets specific performance criteria.
Therefore, I cannot populate the table or answer questions 2 through 9 based on the provided document. The document primarily focuses on regulatory approval based on comparison to existing legally marketed devices, rather than a detailed performance study against predefined acceptance criteria.
Ask a specific question about this device
(260 days)
andRegulation: | Class IIOLO - 21 CFR §882.4560HRX - 21 CFR §888.1100PDQ - 21 CFR §874.1820
TELIGEN Navigation Ready Instruments:
The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.
The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:
· the use of stereotactic surgery may be appropriate, and
· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.
These procedures include but are not limited to spinal fusion.
TELIGEN Clear and TELIGEN Access Probe, when used with UNAS Navigation Rings, can be precalibrated with:
- · the VELYS Spine System using the VELYS Spine Instrument Arrays,
- · the Brainlab Navigation System using the UNAS Arrays.
TELIGEN Access Probe in conjunction with UNAS can also be manually calibrated with other navigation systems, using tracking arrays supplied by the navigation system manufacturer.
The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.
Discectomy Navigation Ready Instruments:
The Discectorny Navigation Ready Instruments when used with the compatible Universal Navigation Adaptor Set are intended to assist the surgeon in locating anatomical structures to facilitate disc space preparation, including discectomy or bony resection. These are indicated for use in surgical spinal procedures, in which:
- · the use of stereotactic surgery may be appropriate, and
· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy using a navigation system and associated navigation arrays.
These procedures include but are not limited to spinal fusion. The Discectomy Navigation Ready Instruments can be pre-calibrated with the VELYS Spine System (only for TELIGEN Graft Delivery Cannula and the curettes) as well as the Brainlab Navigation System.Universal Navigation Adaptor Set:
The Universal Navigation Adaptor Set (UNAS) is intended for use with the compatible DePuy Synthes Navigation Ready Instruments to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which:
· the use of stereotactic surgery may be appropriate, and
· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.
These procedures include but are not limited to spinal fusion. The DePuy Synthes Navigation Ready Instruments, when used with UNAS, can be:
· pre-calibrated with the VELYS Spine System using VELYS Spine System Instrument Arrays,
- · pre-calibrated and/or manually calibrated with the Brainlab Navigation System,
where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.
TELIGEN Navigation Ready Instruments
TELIGEN Navigation Ready Instruments are part of TELIGEN Kits.
The TELIGEN Kits are sterile, single use kits intended for use in surgical spinal procedures allowing for access, visualization, discectomy, graft delivery, navigation and peripheral motor nerve stimulation.
The TELIGEN Kits include a camera, ports and port holder, TELIGEN Clear, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments. Additionally, the TELIGEN Procedure Kit Pro includes an Access Probe.
TELIGEN Access Probe and TELIGEN Clear are part of the DePuy Synthes Navigation Ready Instruments Portfolio and are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS). For further details on UNAS, refer to the UNAS labeling.
Discectomy Navigation Ready Instruments
The Discectomy Navigation Ready Instruments are reusable instruments used for disc space preparation, including discectomy or bony resection. These instruments are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS) and associated navigation arrays. For further details on UNAS, refer to the UNAS labeling.
Universal Navigation Adaptor Set
The Universal Navigation Adaptor Set (UNAS) contains reusable spine surgical instruments used to aid in determining the correct location and trajectory of spinal instruments and implants. The UNAS has an interface between third-party navigation systems and the DePuy Synthes Navigation Ready Instruments. The UNAS can only be used with the VELYS Spine System as well as Brainlab and Medtronic StealthStation® navigation systems. The UNAS includes:
- Brainlab compatible UNAS Navigation Arrays,
- . VELYS Spine/Brainlab compatible Navigation Rings and
- Medtronic compatible Navigation Ring ST. ●
The Navigation Rings and Navigation Ring ST mate with compatible DePuy Synthes Navigation Ready Instruments. These instruments include implant site preparation and implant insertion instruments as well as access and discectomy instruments.
When the VELYS Spine/Brainlab compatible Navigation Ring is attached to the Navigation Ready Instrument:
- . VELYS Spine System Instrument Array can be attached and the instrument can be used with the VELYS Spine System as pre-calibrated instrument, or
- . UNAS Navigation Array can be attached and the instrument can be used with the Brainlab Navigation System as either a manually calibrated and/or pre-calibrated instrument.
When the Navigation Ring ST is attached to the Navigation Ready Instrument, a Medtronic SureTrak® II Universal Tracker Fighter array (SureTrak II array) can be attached, and the instrument can be manually calibrated only with the Medtronic StealthStation Navigation System.
This submission pertains to the TELIGEN System Navigation Ready Instruments, which include the TELIGEN Access Probe, TELIGEN Clear, TELIGEN Graft Delivery Cannula, and associated curettes, along with the Universal Navigation Adaptor Set (UNAS). The primary purpose of this 510(k) submission is to expand the compatibility of these instruments with the VELYS Spine System.
1. Table of Acceptance Criteria and Reported Device Performance:
The provided document does not explicitly list quantitative acceptance criteria in a tabular format. However, it states the following performance aspects were evaluated:
| Acceptance Criteria Category | Reported Device Performance (as described) |
|---|---|
| Accuracy Verification | Verified through engineering analysis. |
| Navigation System Instrument Accuracy Requirements | Fulfilled navigation systems instrument accuracy requirements as stated by the navigation manufacturer. |
| CAD Model Evaluation | Evaluation performed. (Implies satisfactory evaluation) |
| Simulated Use Evaluation | Evaluation performed. (Implies satisfactory evaluation) |
| Technological Characteristics (Design, Materials, Performance) | Consistent with predicate devices. |
| Intended Use | Consistent with predicate devices. |
| Safety and Effectiveness | Does not raise new questions of safety and effectiveness based on application of recognized consensus standards and design controls. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify the exact sample size or number of "test cases" for the evaluations (Accuracy Verification, CAD Model Evaluation, Simulated Use Evaluation). It refers to "engineering analysis" and "simulated use evaluation" which typically involve a set of defined tests rather than a statistical sample size of patient data.
The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. Given the nature of the device (surgical instruments with navigation capabilities), the evaluations would likely be laboratory-based and simulated environments.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
The document does not mention the use of experts to establish ground truth for the test set. The evaluations described (engineering analysis, meeting manufacturer's requirements, CAD model, simulated use) suggest a focus on design validation and performance specifications rather than clinical evaluation with expert-defined ground truth on patient data.
4. Adjudication method for the test set:
The document does not describe any adjudication method. This type of evaluation, for surgical instruments, typically relies on direct measurement and adherence to specifications rather than subjective expert assessment requiring adjudication.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:
No mention of an MRMC comparative effectiveness study, or any studies involving human readers, is found in the provided text. The device is a set of surgical instruments, not an AI diagnostic tool that would typically involve human reader studies.
6. Standalone (algorithm only without human-in-the-loop performance) Study:
Not applicable. The device is a set of physical surgical instruments with navigation capabilities, not a standalone algorithm. Its performance is intrinsically tied to its use by a surgeon within a navigation system.
7. Type of Ground Truth Used:
Based on the evaluation types:
- For Accuracy Verification and Navigation Systems Instrument Accuracy Requirements, the ground truth would be the established engineering specifications, dimensional tolerances, and the accuracy claims/requirements published by the navigation system manufacturers (e.g., VELYS Spine System, Brainlab, Medtronic StealthStation®).
- For CAD Model Evaluation, the ground truth would be the design specifications and theoretical performance derived from the CAD models.
- For Simulated Use Evaluation, the ground truth would be the defined objective criteria for successful surgical simulation (e.g., precise instrument placement, proper interaction with simulated anatomy).
8. Sample Size for the Training Set:
Not applicable. The device is a set of physical surgical instruments, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for this device.