K Number
K243945

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-24

(125 days)

Product Code
Regulation Number
876.1075
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.
The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.
The ALARA BMAC Needle is intended for use in harvesting core biopsy samples of cancellous bone and/or bone marrow.

Device Description

The ALARA BMA Neuro Access Kit utilizes a neuromonitoring platform in providing the surgeon with nerve / root location feedback during the approach to a pedicle to support a guidewire placement.

The ALARA BMAN and ALARA BMAC Kit includes various single-use needles designed for bone marrow aspiration and aspiration of autologous blood.

AI/ML Overview

This document is an FDA 510(k) clearance letter for a medical device called the "ALARA BMA Neuro Access Kit." This type of document primarily focuses on establishing "substantial equivalence" to predicate devices, rather than presenting detailed study results for a new AI/ML-driven diagnostic device.

Therefore, the provided text does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML diagnostic. The "Performance Testing" section lists various engineering and material tests (e.g., torque, pull, strength, sterilization, biocompatibility) relevant to the physical components of the device, not performance metrics for an AI algorithm.

Here's why the requested information cannot be extracted from this document:

  • No AI/ML Component: The device description and performance testing sections indicate a physical medical instrument (needles for biopsy, aspiration, and a neuromonitoring platform for "proximity feedback" during pedicle preparation). There is no mention of an AI or machine learning algorithm. The "proximity feedback" is likely an electrical or mechanical system, not an AI interpreting images or data.
  • Focus on Substantial Equivalence: A 510(k) clearance is primarily about demonstrating that a new device is "substantially equivalent" to a legally marketed predicate device. This often relies on comparing design, materials, intended use, and non-clinical performance (like the engineering tests listed). It does not typically involve the rigorous clinical efficacy studies or AI performance validation described in your prompt.
  • Lack of Diagnostic Claims: The device is for "aspiration of bone marrow or autologous blood," "pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback," and "harvesting core biopsy samples." These are procedural/interventional uses, not diagnostic uses that would involve interpreting data (image, signal, etc.) with an AI.

In summary, as the provided document pertains to a physical medical device and not an AI/ML-driven diagnostic, the questions about acceptance criteria for AI performance, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, and ground truth establishment for AI are not applicable to the content provided.

FDA 510(k) Clearance Letter - ALARA BMA Neuro Access Kit

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 24, 2025

SurGenTec LLC
Guilherme Pires
Vice President of Operations
911 Clint Moore Rd
Boca Raton, Florida 33487

Re: K243945
Trade/Device Name: ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW, PDQ
Dated: February 20, 2025
Received: February 20, 2025

Dear Guilherme Pires:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243945 - Guilherme Pires Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243945 - Guilherme Pires Page 3

Sincerely,

Jessica Carr -S

Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

K243945

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K243945

Please provide the device trade name(s).

ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)

Please provide your Indications for Use below.

The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.

The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.

The ALARA BMAC Needle is intended for use in harvesting core biopsy samples of cancellous bone and/or bone marrow.

Please select the types of uses (select one or both, as applicable). | ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) | Page 7 of 26

Page 5

K243945

510(k) Summary

Submitter: SurGenTec, LLC
911 Clint Moore Rd
Boca Raton, FL 33487
Phone: 561-990-7882
Email: gui@surgentec.com

Official Correspondent: Mr. Guilherme Pires
Vice President of Operations
SurGenTec, LLC
911 Clint Moore Rd
Boca Raton, FL 33487
Phone: 561-990-7882
Email: gui@surgentec.com

Date Prepared: April 24, 2025

Trade Name: ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)

Common Name: Aspiration Needle, BMA Needle

Classification: 21 CFR 876.1075, 21 CFR 874.1820

Product Code: KNW, PDQ

Predicate Device: ALARA Neuro Access Kit (K240664)
ALARA Neuro Access Kit (K190163)
Ranfac Bone Marrow Aspiration Needle (K131157)

Reference Device: Avitus Bone Harvester (K170539)

Device Description:
The ALARA BMA Neuro Access Kit utilizes a neuromonitoring platform in providing the surgeon with nerve / root location feedback during the approach to a pedicle to support a guidewire placement.

The ALARA BMAN and ALARA BMAC Kit includes various single-use needles designed for bone marrow aspiration and aspiration of autologous blood.

Indications for Use:
The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood.

Page 6

The ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.

The ALARA BMAC Needle is intended for use in harvesting core biopsy samples of cancellous bone and/or bone marrow.

Substantial Equivalence:

The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is substantially equivalent to the predicate device.

InformationSubject DevicePrimary PredicateSecondary PredicateReference Device
Trade NameALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) SurGenTec, LLCRanfac Bone Marrow Aspiration Needle Ranfac CorporationALARA Neuro Access Kit SurGenTec, LLCAvitus Bone Harvester
510(k) NumberSubject DeviceK131157K190163K170539
Regulation21 CFR 876.1075 21 CFR 874.182021 CFR 876.107521 CFR 874.182021 CFR 876.1075
ClassificationIIIIIIII
Product CodePDQ, KNWKNWPDQKNW
Indications for UseThe ALARA BMA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback. The ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is indicated for aspiration of bone marrow or autologous blood. The ALARA BMAC Needle is intended for use in harvesting core biopsy samples of cancellous bone and/or bone marrow.The Ranfac Bone Marrow Aspiration Needle is intended for use in aspirating bone marrowThe ALARA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.The Avitus Bone Harvester is intended to harvest cancellous bone and marrow.
Surgical ApproachMinimally InvasiveMinimally InvasiveMinimally InvasiveMinimally Invasive
MaterialStainless SteelStainless SteelStainless SteelStainless Steel
Diameter11 Gauge – 3mm8 Gauge, 11 Gauge11 Gauge – 3mmN/A

Page 7

Lengths12.8cm11cm – 15cm12.8cmN/A
Use of the NeedleNeedle depth stop at boneNeedle depth stop at boneNeedle depth stop at boneN/A
Packaged SterilizedYesYesYesYes
Single-Use DeviceYesYesYesYes

The ALARA BMAN and BMAC Kit are similar in materials and design to the predicate device. Both are manufactured from stainless steel and ABS. Both devices include a cannulated version and include similar lengths, and diameters. Both devices have plastic molded handles for the cannula and stylet that mate with each other to prevent separation.

Performance Testing:

The following tests have been performed on the ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit):

  • Cannula Handle Torque Test
  • Cannula Handle Pull Test
  • Stylet Handle Pull Test
  • Needle Strength from Impaction and Compression Tests
  • Luer Lock Verification per ISO 80369
  • Needle Particle Aspiration
  • Needle Cell Viability Test
  • Packaging Validation
  • Sterilization Validation
  • Biocompatibility Assessment

Conclusion:

Based on the similarities in materials, design, principles of operation, and results of the non-clinical testing, the ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit) is substantially equivalent to the predicate device for the requested indications for use.

Page 8

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.