    510k Summary Text (Full-text Search) :

    K253373
    Trade/Device Name: Southern Maxillofacial (MF) System
    Regulation Number: 21 CFR 872.4760
    K253373
    Trade/Device Name: Southern Maxillofacial (MF) System
    Regulation Number: 21 CFR 872.4760
    Device Classification Name:** Plate, Bone
    Classification & Regulation #: Class II per 21 CFR §872.4760
    LORENZ 1.0mm, 1.5mm, 2.0mm System (K953385) | Comparison |
    |---|---|---|---|---|
    | Regulation | 872.4760
    , 872.4880 | 872.4760, 872.4880 | 872.4760 | Same. |
    | Product Code | JEY, DZL | JEY, DZL | JEY |

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Southern MF System is a maxillofacial plate and screw system intended for osteotomy, reconstruction, and fracture fixation in midface, maxillofacial, and mandibular surgeries.

    Device Description

    The Southern MF System includes a series of plates, meshes, and screws intended to stabilize and immobilize bone segments in the maxillofacial, midface, and mandibular regions, thereby facilitating osseous fusion in a set orientation. The Southern MF System consists of bone plates made from Grade 2 commercially pure titanium (ASTM F67) in various shapes and sizes to accommodate various surgical requirements. It also includes bone screws made from Ti-6Al-4V ELI (ASTM F136) Titanium Alloy. The bone screws are made available in a variety of lengths and diameters, and in self-drilling and self-tapping variations. The Subject Device is not intended for use in the orbital roof and can only be used if no exposure of the intracranial compartment is presented.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Predicate Device) | (Predicate Device) | (Predicate Device) |
    | Classification | Class II | 21 CFR 872.4760
    , Class II | 21 CFR 888.3030, Class II | 21 CFR 872.4760, Class II | 21 CFR 882.5330, Class II | 21 CFR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KLS-Martin Micro Osteosynthesis System (1.5MM): The KLS-Martin Micro Osteosynthesis System (1.5MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone structures. The bone segments are attached to the plate with screws to prevent movement of the segments.

    KLS Martin Centre-Drive Drill-Free Screw: The KLS Martin Centre-Drive Drill-Free Screws are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bony surface and stabilize the bone fragments. The addition of the self drilling feature is the only difference between the submitted device and the predicate device referenced.

    KLS Martin Rigid Fixation - Sterile: The KLS Martin Rigid Fixation - Sterile is intended to provide, in sterile packaging, osteosynthesis products with the following indications for use:

    • The RESORB-X® SF Sonotrode is only intended for use for insertion of the RESORB-X® SF pins.
    • The KLS Martin Mandibular/Reconstruction System II is intended for use in the stabilization and fixation of mandibular fractures and mandibular reconstruction.
    • The KLS Martin Centre-Drive Drill-Free screws are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bony surface and stabilize the bone fragments. The addition of the self drilling feature is the only difference between the submitted device and the predicate device reference
    • The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
    • The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.

    Individual Patient Solutions - Ti (IPS-Ti)*: Individual Patient Solutions - Ti (IPS-Ti) is intended to replace bony voids in the mandibular, maxillofacial or craniofacial skeleton.

    Individual Patient Solutions - PEEK (IPS-PEEK)*: Individual Patient Solutions - PEEK (IPS-PEEK) is intended to replace bony voids in the cranial and/or craniofacial skeleton.

    Individual Patient Solutions - (IPS-PEEK):** The Individual Patient Solutions - (IPS-PEEK) is intended to replace bony voids in the cranial skeleton.

    Device Description

    KLS-Martin Micro Osteosynthesis System (1.5mm): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the bones of the skeletal system after a facial fracture or surgery. The bone plate system is composed of bone plates and screws of various shapes and sizes for use in oral-maxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 1.0mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.5mm - 1.8mm.

    KLS Martin Centre-Drive Drill-Free Screw: The KLS Martin Centre-Drive Drill-Free Screws are designed to eliminate the need for pre-drilled pilot holes. They are self-tapping with one step insertion. They are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates where are contoured to fit the bony surface and stabilize the bone fragments. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 2.0mm.

    KLS Martin Rigid Fixation - Sterile: The KLS Martin Rigid Fixation - Sterile includes titanium plates of various shapes and thickness, titanium screws of various length and diameter, stainless steel twist drills of various length and diameter and stainless steel sonotrode tips that are provided in sterile packaging. The KLS Martin Rigid Fixation - Sterile is intended to provide KLS Martin's previously cleared osteosynthesis products in sterile packaging.

    Individual Patient Solutions - Ti (IPS-Ti)*: The Individual Patient Solutions - Ti (IPS-Ti) is a preformed implant manufactured from Titanium material. The implant is pre-shaped to fit the anatomy of the patient using a CT-based model of the patient's skull. The PCI-PEEK is fixated to native bone using previously cleared KLS Martin titanium plates and screws.

    Individual Patient Solutions - PEEK (IPS-PEEK)*: The Individual Patient Solutions - PEEK (IPS-PEEK) is a preformed implant manufactured from PEEK material. The implant is pre-shaped to fit the anatomy of the patient using a CT-based model of the patient's skull. The PCI-PEEK is fixated to native bone using previously cleared KLS Martin titanium plates and screws.

    Individual Patient Solutions - (IPS-PEEK)**: The Individual Patient Solutions - (IPS-PEEK) is a preformed implant manufactured from PEEK material offered to the end-user as a sterile product. The implant is pre-shaped to fit the anatomy of the patient using a CT-based model of the patient's skull. The PCI-PEEK is fixated to native bone using previously cleared KLS Martin titanium plates and screws.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251441

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-07

    (182 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Fixation System; Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System
    Regulation Number: 21 CFR 872.4760
    number:** | 9710629 |
    | Device common name: | Bone plate |
    | Device Regulation and name: | 872.4760
    #** | K010352, K122890, K013039, K022981, K020266 |
    | Predicate Device Classification & Name | 872.4760
    | 872.4760 | Identical |
    | Product Code | JEY | JEY | Identical |
    | Intended Use | INION CPS
    Bone Plate872.4880 Screw, fixation, intraosseous | 872.4760 Bone Plate872.4880 Screw, fixation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INION CPS™ BABY 1.5 BIOABSORBABLE FIXATION SYSTEM implants are intended for use in trauma and reconstructive procedures in the mid-face and maxilla of infant (from 29 days to 2 years of age) and child (greater than 2 years to 12 years of age) patients.

    Specific indications:

    • congenital anomalies correction in the orbital rim, orbital floor, maxilla and midface
    • traumatic injuries.

    The INION CPS™ 1.5/2.0/2.5 BIOABSORBABLE FIXATION SYSTEM implants are intended for use in trauma and reconstructive procedures, and to maintain the relative position of bone grafts or bone graft substitutes, in orbital floor, medial and lateral orbital walls, orbital rim, mid-face, maxilla and mandible (in conjunction with appropriate maxillomandibular fixation) of child (greater than 2 years to 12 years of age), adolescent (greater than 12 years through 21 years of age) and adult patients. Specific indications: fixation of LeFort (I, II, III) osteotomies and mandibular orthognathic procedures

    Device Description

    Inion CPS™ and Inion CPS™ Baby implants are bioabsorbable plates, screws and mesh plates made of bioabsorbable co-polymers. Inion CPS™ 1.5/2.0/2.5 System implants are composed of L-lactic acid, D-lactic acid and Trimethylenecarbonate. Inion CPS™ 1.5 Baby System implants are composed of L-lactic acid, D-lactic acid and Poly-Glygolic acid. These polymers have a long history of safe medical use and they degrade in vivo by hydrolysis into alpha-hydroxy acids that are metabolised by the body. The implants retain sufficient strength to fulfil their intended function during the healing period of the fracture or osteotomy, and degrade gradually thereafter. Bioresorption of Inion CPS™ implants takes place within 2-4 years, and Inion CPS™ Baby implants within 2-3 years.

    The systems consist of fixation plates, meshes and screws offered in different sizes and designed to be used with the Inion CPS™ bone drill bits, bone taps, self-drilling bone taps, countersinks, screw drivers, plate bending pliers and heating device. Nonsterile instruments are intended to be cleaned and sterilized before initial use and after each use.

    The implants are provided sterile by gamma irradiation. They are intended for single use and shall not be re-sterilized or re-used. Implants are non-pyrogenic and fully synthetic. The Subject Device is not intended for use in the orbital roof and can only be used if no exposure of the intracranial compartment is presented.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251472

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-11

    (59 days)

    Product Code
    Regulation Number
    882.5360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Predicate Device Premarket Notification #:* K010352
    Predicate Device Classification & Name: 872.4760
    ----------|-------------------|
    | Company | Inion Oy | Inion Oy | - |
    | Regulation Number | 872.4760
    | 872.4760 | Identical |
    | Product Code | JEY, HWC | JEY | Identical |
    | Intended Use | INION
    | 872.4760 | Identical |
    | Product Code | JEY, HWC | JEY | Identical |
    | Intended Use | INION
    | 872.4760 | Identical |
    | Product Code | JEY, HWC | JEY | Identical |
    | Intended Use | INION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inion CPS™ 1.5 Baby Bioabsorbable Fixation System implants are intended for use in trauma and reconstructive procedures in the craniofacial skeleton (e.g., infant craniofacial surgery, craniosynostosis, congenital malformations, craniotomy flap fixation), mid-face and maxilla.

    The Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System implants are intended for use in trauma and reconstructive procedures and to maintain the relative position of bone grafts or bone graft substitutes in the cranium, including craniotomy flap fixation.

    Device Description

    Inion CPS™ and Inion CPS™ Baby implants are bioabsorbable plates, screws and mesh plates made of bioabsorbable co-polymers. Inion CPS™ 1.5/2.0/2.5 System implants are composed of L-lactic acid, D-lactic acid and Trimethylenecarbonate. Inion CPS™ 1.5 Baby System implants are composed of L-lactic acid, D-lactic acid and Poly-Glygolic acid. These polymers have a long history of safe medical use and they degrade in vivo by hydrolysis into alpha-hydroxy acids that are metabolised by the body. The implants retain sufficient strength to fulfil their intended function during the healing period of the fracture or osteotomy, and degrade gradually thereafter. Bioresorption of Inion CPS™ implants takes place within 2-4 years, and Inion CPS™ Baby implants within 2-3 years.

    The systems consist of fixation plates, meshes and screws offered in different sizes and designed to be used with the Inion CPS™ bone drill bits, bone taps, self-drilling bone taps, countersinks, screw drivers, plate bending pliers and heating device. The instruments are nonsterile and reusable. They are intended to be cleaned and sterilized before initial use and after each use.

    The implants are provided sterile by gamma irradiation. They are intended for single use and shall not be re-sterilized or re-used. Implants are non-pyrogenic and fully synthetic.

    AI/ML Overview

    This FDA 510(k) clearance letter is for bioabsorbable fixation systems, not an AI/ML powered medical device. Therefore, many of the requested fields are not applicable. I will provide information from the document as it relates to the general acceptance criteria and the study that proves the device meets them, while noting when a specific AI/ML related field is not relevant to this type of device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (General)Reported Device Performance (as per document)
    Material CompositionInion CPS™ 1.5/2.0/2.5 System implants: L-lactic acid, D-lactic acid, Trimethylenecarbonate.
    Inion CPS™ 1.5 Baby System implants: L-lactic acid, D-lactic acid, Poly-Glygolic acid.
    Bioresorption TimeInion CPS™ implants: 2-4 years.
    Inion CPS™ Baby implants: 2-3 years.
    Strength RetentionInion CPS™ 1.5/2.0/2.5: Retain minimum of 70% of initial strength 9 weeks after implantation.
    Inion CPS™ Baby 1.5: Retain minimum of 70% of initial strength 6 weeks after implantation.
    BiocompatibilityEvaluated in accordance with ISO 10993-1:2018 and FDA Guidance. No changes to materials or manufacturing methods.
    Non-pyrogenicityConstantly low endotoxin level, below acceptance limit (USP 2011 Chapter <161>). Passed Rabbit pyrogen tests (ISO 10993-11:2017, Annex G / USP General chapter <151 >).
    Clinical PerformanceSafe and effective use demonstrated through clinical literature and post-market follow-up data. Performance rate and overall complication rate are within the range of state-of-the-art devices.
    SterilityProvided sterile by gamma irradiation. Intended for single use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Description: Not directly applicable as this is not an AI/ML device with a specific "test set" in the context of algorithm validation. The "testing" refers to various evaluations:
      • Biocompatibility/Non-pyrogenicity Testing: Representative samples of the subject devices were selected for rabbit pyrogen tests. The exact number of samples or rabbits is not specified but it generally follows standard laboratory protocols. The provenance is likely from laboratory testing.
      • Clinical Evaluation: The clinical evaluation relied on published clinical studies.
        • Inion CPS™ Baby 1.5 Bioabsorbable Fixation System: 11 publications including 240 patients.
        • Inion CPS™ 1.5/2.0/2.5 System: 39 publications including 1755 patients.
      • Data Provenance for Clinical Studies: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective, beyond being "published clinical studies."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a physical implant, not an AI/ML diagnostic tool that requires experts to establish ground truth for image interpretation or similar tasks. The "ground truth" for its performance is assessed through physical testing (material properties, pyrogenicity) and clinical outcomes reported in published literature.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in AI/ML performance studies, where multiple human experts independently review cases and then resolve disagreements. This concept does not apply to the evaluation of a physical bioabsorbable fixation system.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are specific to evaluating human reader performance with and without AI assistance for tasks like medical image interpretation. This device is a physical surgical implant and does not involve "human readers" in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not have an algorithm for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" (or basis for acceptance) for this device is established through:

    • Physical Properties: Material composition, degradation profile, and strength retention are measured against pre-defined specifications.
    • Biocompatibility and Non-pyrogenicity: Demonstrated through laboratory testing (e.g., rabbit pyrogen tests) against established biological safety standards (ISO 10993-1, USP chapters).
    • Clinical Outcomes Data: Safe and effective use, performance rates, and complication rates are evidenced by published clinical literature and post-market follow-up data.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This device is not an AI/ML model that requires a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243637

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-21

    (88 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    10/2018 | |
    | Product code | JEY (21 CFR 872.4760
    | 08/08/2017 |
    | Product code | JEY (21 CFR 872.4760
    |
    | Product code | JEY (21 CFR 872.4760
    2022 |
    | Product code | JEY (21 CFR 872.4760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Materialise Personalized Guides for Craniomaxillofacial Surgery are intended to guide the marking of bone and or guide surgical instruments in facial surgery.

    CMF Titanium Guides are used during bone repositioning/reconstruction surgical operations for orthognathic and reconstruction (including bone harvesting) indications.

    CMF Titanium Guides are intended for children, adolescents and adults.

    CMF Titanium Guides are intended for single use only.

    CMF Titanium Guides are to be used by a physician trained in the performance of maxillofacial surgery.

    Materialise Personalized Models for Craniomaxillofacial Surgery are intended for visualization of the patient's anatomy, preparation of surgical interventions and fitting or adjustment of implants or other medical devices such as osteosynthesis plates or distractors, in mandibular and maxillofacial surgical procedures.

    CMF Plastic Models are intended for infants, children, adolescents and adults.

    CMF Plastic Models are intended for single use only.

    CMF Plastic Models are to be used by a physician trained in the performance of maxillofacial surgery.

    Device Description

    Materialise Personalized Guides and Models for Craniomaxillofacial Surgery combines the use of 3D preoperative planning software with patient-matched guides and models to improve and simplify the performance of surgical interventions by transferring the pre-operative plan to surgery. Materialise Personalized Guides and Models for Craniomaxillofacial Surgery are used in the facial skeleton or in maxillofacial surgeries.

    The surgical planning is based on medical images of the patient that are segmented in order to create a 3D representation of the patient's anatomy. The surgical treatment of the patient is simulated based on instructions provided by the surgeon and the patient-matched devices are tailored to the treatment and the patient's needs. The patient-matched devices are manufactured from commercially pure Titanium, polyamide, or clear acrylic by means of additive manufacturing technologies. The patient-matched devices are provided non-sterile.

    Materialise Personalized Guides and Models for Craniomaxillofacial Surgery include CMF Titanium Guides and CMF Plastic Models.

    AI/ML Overview

    The provided text is a 510(k) summary for the device "Materialise Personalized Guides and Models for Craniomaxillofacial Surgery." It details the device's indications for use, description, comparison to predicate and reference devices, and non-clinical performance data. However, it does not include information about AI/algorithm performance, acceptance criteria for such an algorithm, or a clinical study for proving the device meets those criteria. The document lists "non-clinical testing" and states that "no guide specific mechanical testing is performed but this is covered by mechanical analysis of CMF Titanium Plates."

    Therefore, I cannot extract the information required for an AI device acceptance criteria and study from this document. The document primarily focuses on the substantial equivalence of the physical, patient-matched guides and models to existing devices, covering aspects like material compatibility, mechanical properties, biocompatibility, and sterilization.

    To answer your request, the ideal information would be present in a document describing an AI/ML medical device, which would typically contain details regarding the algorithm's performance metrics (acceptance criteria), the dataset used for testing, ground truth establishment, and potential MRMC studies. This document does not describe such a device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241253

    Validate with FDA (Live)

    Date Cleared
    2024-12-12

    (223 days)

    Product Code
    Regulation Number
    878.3680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | TITANIUM MINIPLATESYSTEM | K951690 | 872.4760JEY
    | CMS CRANIO MAXILLOFACIAL TITANIUM IMPLANTSYSTEM | K951691 | 872.4760JEY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medicon Epiplating System is indicated for the attachment of an external restoration prosthesis for restoration of a physical defect when other means of attachment are inadequate. The Epiplating system provides the bone anchorage for the prosthetic attachment. These devices are indicated for use in the maxillo-craniofacial region (including ear, nose and eye).

    Device Description

    The Medicon Epiplating System is an implant system consisting of various plates used as the foundation for anchoring prostheses or epitheses in the craniofacial region. The system offers various implant sizes to fit the anatomical needs of a wide variety of patients. Components are manufactured from pure titanium (ASTM F67) or Ti6Al4V alloy (ASTM F136).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Medicon Epiplating System" and primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study of its performance against specific acceptance criteria for an AI/ML-based device.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment cannot be found in the provided document.

    The document is a regulatory submission for a physical medical device (an implant system), not an AI/ML software device. The "Performance Data" section discusses mechanical testing, sterilization validation, cleaning validation, and biocompatibility, which are standard for such devices, but not the type of performance metrics (e.g., sensitivity, specificity, AUC) typically associated with AI/ML solutions and their acceptance criteria.

    Summary of what can be extracted based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    • Not applicable/Not found. The document does not specify quantitative "acceptance criteria" in terms of clinical performance metrics that would be relevant for an AI/ML device (e.g., sensitivity, specificity). Instead, it refers to mechanical testing standards (ASTM F382) and biocompatibility standards (ISO 10993-1). The results are qualitative ("demonstrate substantial equivalence") rather than quantitative against specific clinical performance thresholds.

    2. Sample sized used for the test set and the data provenance:

    • Not applicable/Not found. This section is irrelevant for the type of device described. No "test set" of clinical data (images, patient records) is mentioned. "Performance data" refers to bench testing on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not found. This information pertains to studies with human expert interpretation, typically in the context of AI/ML evaluation. This document describes a physical implant system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not found. This is for clinical data interpretation, not for evaluating a physical implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable/Not found. This is specific to AI/ML devices that assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable/Not found. This is specific to AI/ML algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable/Not found. For this physical device, "ground truth" would be related to material properties meeting standards, mechanical performance under specified loads, and biocompatibility, which are evaluated against established engineering and biological standards, not clinical ground truth derived from expert consensus or pathology.

    8. The sample size for the training set:

    • Not applicable/Not found. There is no AI/ML training set.

    9. How the ground truth for the training set was established:

    • Not applicable/Not found. There is no AI/ML training set.

    Information that is mentioned in the document (but not directly addressing the AI/ML specific questions):

    • Device Type: Medicon Epiplating System, an implant system.
    • Purpose: Attachment of an external restoration prosthesis for physical defects in the maxillo-craniofacial region (ear, nose, eye).
    • Regulatory Mechanism: 510(k) premarket notification.
    • Substantial Equivalence: Demonstrated through comparison of technological characteristics (design, intended use, material composition, function) to predicate devices.
    • Performance Data (for this device type):
      • Mechanical Testing: Performed in accordance with ASTM F382 (proof of load, bending strength, bending stiffness, structural bending stiffness).
      • Sterilization Validation: Pre-vacuum (steam) sterilization in accordance with ANSI/AAMI/ISO 17665-1:2006/(R)2013.
      • Cleaning Validation: Manual and machine cleaning validation in accordance with AAMI TIR30:2011.
      • Biocompatibility: Based on previously cleared devices with identical material and manufacturing methods, tested according to ISO 10993-1.

    The document entirely pertains to a physical medical device and does not involve AI/ML technology, which is why the specific questions about AI/ML acceptance criteria and study design cannot be answered from the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241314

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-16

    (98 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | |
    | Regulation Number | 872.4760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofial skeletal system.
    KLS Chin Plate System (K943348): The KLS Chin Plate System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofacial skeletal system.
    KLS-Martin Micro Osteosynthesis System (1.0MM) (K944561): The KLS-Martin Micro Osteosynthesis System (1.0MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS-Martin Micro Osteosynthesis System (1.5MM) (K944565): The KLS-Martin Micro Osteosynthesis System (1.5MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bone fragments. The addition of the self drilling feature is the only difference between the submitted device and the predicate device referenced.
    KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is only intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to remove malignant or benign the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMF or treatment of temporomandibular joint disease (TMD).
    KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
    KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System is intended to be surgically placed in the mouth for use an an anchor for orthodontic procedures.
    KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) are implants intended to be surgically placed in the mouth for use as an anchor for orthodontic procedures in patients.
    KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile is in sterile packaging, osteosynthesis products with the following indications for use:
    K051236: The RESORB-X® SF Sonotrode is only intended for use for insertion of the RESORB-X® SF pins.
    K032442: The KLS Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
    K971297: The KLS Martin Centre-Drive Drill-Free screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bony surface and stabilize the bone fragments. The addition of the self drilling feature is the only difference between the predicate device reference
    K944565: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
    K944561: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
    KLS Martin Drill-Free MMF Screw (K042573): The KLS Martin Drill-Free MMF Screws is intended for use in maxilonandibular fixation of fractures of the maxilla, mandible, or both.
    Drill Free MMF Screw (K083432): The Drill Free MMF Screw is intended for use in maxillomandibular fixation of fractures of the maxilla, mandible, or both.
    KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is intended for temporary stabilization of maxillary fractures. It is designed to maintain proper occlusion during intraoperative bone healing (app. 6-8 weeks). It is indicated for the temporary treatment of maxillomandibular fixation (MMF) in adults or adolescents who have permanent teeth present (ages 12 and older).

    Device Description

    KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System consists of titanium non-locking plates ranging in thickness from 0.6mm - 2.5mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
    KLS Chin Plate System (K943348): The KLS Chin Plate System consists of titanium plates ranging in thickness of 0.6mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
    KLS-Martin Micro Osteosynthesis System (1.0mm) (K944561): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates, bone plates, bone screws and accessories of various shapes and sizes for use in oral-maxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 1.2mm.
    KLS-Martin Micro Osteosynthesis System (1.5mm) (K944565): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates and screws of various shapes and sizes for use in oralmaxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.5mm - 1.8mm.
    KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are designed to eliminate the need for pre-drilled pilot holes. They are self-tapping with one step insertion. They are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates where are contoured to fit the bone fragments. The bone fragments. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 2.0mm.
    KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is a solid condylar head which attaches with fastening screws to a KLS-Martin Fracture/ Reconstruction Plate. The implant is available for left and right placement. The KLS-Martin Temporary Condy intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to benign tumors requiring the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMJ or traumatic injuries, or for treatment of temporomandibular joint disease (TMD).
    KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II includes several different designs of titanium plates and screws intended for use in the stabilization and fixation of mandibular fractures and reconstruction. The plates are manufactured from either CP Titanium or Titanium Alloy and range in thickness from 1.0mm - 3.0mm. The screws are manufactured from either CP Titanium Alloy and range in diameter from 2.0mm - 3.2mm.
    KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System consists of a titanium screw designed to aid in dental movement by providing a rigid skeletal fixation point. The screw is intended to be surgically placed in the mouth for orthodontic procedures. The screws are manufactured from either CP Titanium or Titanium Alloy.
    KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) consists of titanium non-locking plates to aid in dental movement by providing a rigid skeletal fixation point. The plates are manufactured from either CP Titanium Alloy and are fixated with titanium screws and are utilized as an anchor for orthodontic procedures in the palatal, maxilla or mandible region.
    KLS-Martin Drill-Free MMF Screw (K042573): The KLS-Martin Drill-Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The KLS-Martin Drill-Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, or both. The screws are manufactured from either CP Titanium Alloy and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
    KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile includes titanium plates of various shapes and thickness, titanium screws of various length and diameter, stainless steel twist drills of various length and stainless steel sonotrode tips that are provided in sterile packaging. The KLS Martin Rigid Fixation - Sterile is intended to provide KLS Martin's previously cleared osteosynthesis products in sterile packaging.
    Drill Free MMF Screw (K083432): The Drill Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The Drill Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, mandible, or both. The screws are manufactured from Stainless Steel and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
    KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is a bone-borne maxillomandibular fixation (MMF) system consisting of metalic archbars with sliding locking plates that attach to the dental arches with screws. The system is intended to provide temporary stabilization of mandibular and maxillary fractures as well as maintain properative bone fixation and postoperative bone healing (app. 6-8 weeks). The patient is brought into occlusion by wiring around the archbar wire hooks. The L1 MMF system plates are manufactured from CP Titanium (ASTM F67), are available in either a 7-hole siding plate configuration with two different lengths, and are 0.5mm in plate thickness. The L1 MMF system sliding locking plates are fixated with either 2.0 x 6 mm selfdrilling locking screws manufactured from Ti-6Al-4V (ASTM F136). Implants are available both sterile. The system also includes the necessary instruments to facilitate placement of the implants.

    AI/ML Overview

    The document describes the KLS Martin Oral-Max Implants - MR Conditional, a bundled submission of various osteosynthesis systems and screws intended for use in craniomaxillofacial surgery. The purpose of this submission is to support the conditional safety and labeling modification of these devices in the magnetic resonance (MR) environment.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implied by the non-clinical tests conducted to support MR Conditional safety, aligning with relevant ASTM standards and FDA guidance. The reported device performance is that the devices can be safely scanned under specified conditions.

    Acceptance Criteria (from ASTM/FDA Guidance)Reported Device Performance (Summary from Submission)
    Magnetically induced displacement force within acceptable limits (ASTM F2052-21)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    Magnetically induced torque within acceptable limits (ASTM F2213-17)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    Image artifacts within acceptable limits (ASTM F2119-07, R2013)Not explicitly quantified but implied as acceptable for MR Conditional labeling.
    RF-induced heating (ASTM F2182-19e2) resulting in a temperature rise below 6 ℃Achieved under specified scanning conditions: 1.5 T/64 MHz and 3 T/128 MHz at a whole-body averaged specific absorption rate (wbSAR) of 2 W/kg or head SAR of 3.2 W/kg for an hour-long scanning session.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The testing involved computational modeling and simulation (CM&S). The "test set" in this context refers to the simulated scenarios and device configurations.

    • Sample Size for Test Set: "the entire portfolio of KLS Martin maxillofacial implants" was simulated. This implies that all devices grouped under "KLS Martin Oral-Max Implants - MR Conditional" were included in the simulations. The document also mentions "various in-vivo device positions and landmarks," "worst-case single and multiple devices," and simulations in "10 cm increments." This suggests a comprehensive set of simulated scenarios rather than a traditional physical sample size.
    • Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective" in the same way clinical data is. The data is generated through computational modeling and simulation using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and the Duke virtual human anatomy. This is a form of prospective simulation data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable in the traditional sense for this type of non-clinical, simulation-based study. The "ground truth" for the RF-induced heating simulations is derived from the established physics and engineering principles embedded in the FDA-qualified MDDT and the Duke virtual human anatomy model. The expertise lies in the development and validation of these computational tools and the interpretation of the simulation results by experts in MR safety and medical device engineering at MED Institute and the submitting company. The document does not specify the number or qualifications of individual experts validating the computational model, but implies that the MDDT itself is "FDA-qualified," indicating a level of expert review and agreement on its methodology.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like "2+1" or "3+1" are typically used for consensus building among human expert readers for clinical studies, especially when establishing ground truth from image interpretation. This study is based on physical property testing and computational simulations, not human interpretation of clinical data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving human readers or AI-assisted diagnostic performance. It focuses on the physical safety of implants in an MR environment.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The RF-induced heating assessment involved "Computational modeling and simulation (CM&S) ... using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and in a clinically relevant position within the Duke virtual human anatomy." This is a standalone algorithm/model-based assessment without a human-in-the-loop for the performance evaluation itself. Human experts design the simulations, configure the models, and interpret the results, but the "performance" (temperature rise, SAR calculations) is computed by the algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Magnetically induced displacement force, torque, and image artifacts: The "ground truth" is based on the physical properties of the materials and device designs, measured or calculated according to established ASTM standards (F2052-21, F2213-17, F2119-07).
    • RF-induced heating: The "ground truth" for the simulations is derived from the established electromagnetic physics and thermal dynamics principles implemented in the FDA-qualified Medical Device Development Tool (MDDT) and applied to the Duke virtual human anatomy model. The MDDT's qualification process by the FDA implicitly establishes the reliability of its results as a form of "ground truth" for simulation-based assessments.

    8. The sample size for the training set

    Not applicable. This is a non-clinical study for MR safety assessment, not a machine learning model requiring a training set in the typical sense. The "training" for the MDDT is its initial validation and qualification against known physical phenomena and experimental data, which is a separate process from this submission.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a machine learning model in this context. The "ground truth" for qualifying the simulation tool (MDDT) would have been established through extensive validation against experimental measurements and recognized physical theories.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240651

    Validate with FDA (Live)

    Device Name
    MRI Universal
    Date Cleared
    2024-07-12

    (127 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , D-79111
    GERMANY

    Re: K240651
    Trade/Device Name: MRI Universal
    Regulation Number: 21 CFR 872.4760
    GERMANY

    July 12, 2024

    Re: K240651
    Trade/Device Name: MRI Universal
    Regulation Number: 21 CFR 872.4760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Universal CMF System Intended use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. MP Lefort I Plates Intended Use The Stryker Leibinger Universal CMF and Universal 2.0 Mini Plating Systems are Craniomaxillofacial (CMF) plate and screw systems intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Indications for Use The MP LeFort I Plates as part of the predicate subgroup Universal 2.0 Mini Plating System are implants indicated for osteotomy, stabilization, and rigid fixation of LeFort I fractures of the maxillofacial skeleton. Stryker Upper-Face AXS Screws And Mid-Face AXS Screws Indications for Use The Stryker Universal CMF System is a Cranio-maxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Stryker MP, Mandible, HMMF And MMF AXS Screws -Universal CMF System: Intended Use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for Use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. -Stryker Universal SMARTLock Hybrid MMF System: Intended Use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for Use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. -Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Stryker Universal Orbital Floor System Intended Use The Stryker Universal Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit. Indications for Use The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older. Stryker Universal Smartlock Hybrid MMF System Intended use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion.

    Device Description

    This Bundled. Traditional 510(k) is for a single change the MR status that affects multiple devices and can be assessed during one review. There are no changes to the predicate devices themselves, just the addition to allow MR Conditional use of the implants in scope after MRI compatibility testing was completed. Associated labeling for each Predicate Device is also updated to reflect the MR conditional use of the implants.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233874

    Validate with FDA (Live)

    Device Name
    tmCMF Solution
    Manufacturer
    Date Cleared
    2024-07-11

    (217 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Knoxville, Tennessee 37922

    Re: K233874

    Trade/Device Name: tmCMF Solution Regulation Number: 21 CFR 872.4760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The tmCMF Solution is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT-based system. The input data file is processed by the tmCMF Solution, and the result is an output data file that may then be provided as digital models or used as input to a manufacturing portion of the system that produces physical outputs, including anatomical models, surgical guides, dental splints, and implants for use in maxillofacial, midface, and mandibular surgery. Implants are intended for bone fixation and reconstruction. restoration of bone defects, and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton, midface, mandible, and chin in adolescents (greater than 12 to 21 years of age) and adults. The tmCMF Solution is also intended as a preoperative software tool for simulating/ evaluating virtual surgical treatment options.

    Device Description

    TechMah CMF (tmCMF) Solution is a family of personalized product solutions for Reconstructive, Orthognathic, Trauma, and Augmentation procedures in the mandible and midface (including orbital floor, medial and lateral orbital walls). The solution is comprised of Surgeon Review Tool (SRT) software and maxillofacial and mandibular surgical instruments (surgical guides, anatomical models, and dental splints) and implants. The surgical instruments and implants are patient-specific devices and are designed utilizing CT and dental scan patient image data.

    Surgical guides are patient-specific devices or templates based on pre-operative software planning designed to fit a specific patient. These guides are used to assist a surgeon in transferring the pre-operative plan to the surgery by guiding the marking/drilling of bone for plate fixation screws and the position of the osteotomy marking/cutting slots. Guides can be used in conjunction with anatomical models to verify anatomical positioning and fit. Surgical guides can be used with either tmCMF Solution patient-specific implants or compatible off-the-shelf DePuy Synthes implants.

    Anatomical models are patient-specific models that are based on pre-operative anatomy and surqical planning specific to a patient. These models are used to assist a surgeon in transferring the pre-operative plan to the surgery by representing pre-operative, intra-operative, and post-operative anatomical models as quidance. Anatomical models can be used to check quide fit and facilitate the pre-bending of a non-patient-specific off-the-shelf plate.

    Dental splints are patient-specific devices or templates based on pre-operative software planning designed to fit a specific patient. These templates are used to assist a surgeon in transferring the pre-operative plan to the surgery by guiding the dental alignment of bone and teeth.

    Implants are patient-specific devices that are based on pre-operative software planning designed to fit a specific patient. These implants are integral components for CMF (craniomaxillofacial) procedures, facilitating bone repositioning, fixation, reconstruction, and the restoration of bone defects. They are designed according to the pre-operative surgical plan to ensure continuity in regions where bone is absent and to provide structural integrity to the maxillofacial skeletal components, midface, mandible, and chin. This could include stabilizing fractured or intraoperatively cut bone, fixating harvested graft segments, or augmenting bone defects. Implants can be used in conjunction with surgical guides and anatomical models to assist with anatomical positioning and fit.

    The Surgeon Review Tool (SRT) software is used by surgeons for the review of surgical plans, surgical instruments, and implant designs. The SRT software is accessed through a web interface.

    tmCMF Solution supports the following maxillofacial, midface (including orbital floor, lateral and medial orbital wall), and mandibular surgical procedures:

    • . Reconstructive
    • Orthognathic
    • Trauma
    • Augmentation ●
    AI/ML Overview

    The provided text describes the tmCMF Solution device and its substantial equivalence to predicate devices, but it does not contain the specific acceptance criteria or the study details proving the device meets those criteria, particularly for performance metrics like accuracy, sensitivity, or specificity for the software components.

    The "Benchtop Performance" section mentions that "Testing demonstrated that the tmCMF Solution and substantial equivalence with the predicate device" and lists several "Performance verification" items, but these are high-level statements rather than detailed acceptance criteria and reported performance values. For example, "Testing demonstrated that the surgical guide and implants meets the predetermined acceptance criteria" indicates that acceptance criteria exist but doesn't provide them or the specific performance results.

    Therefore, I cannot fulfill all parts of your request with the given input. I will extract the information that is present and indicate where information is missing.


    Here's a summary of what can be extracted based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not explicitly provided with specific numerical acceptance criteria or reported performance values in the document. The document generally states that "Testing demonstrated that... meets the predetermined acceptance criteria" for various aspects.

    Performance Verification ItemAcceptance Criteria (Not explicitly detailed)Reported Device Performance (Not explicitly detailed)
    Performance verificationPredetermined acceptance criteriaDevice meets criteria
    Locking screw compatibility verificationPredetermined acceptance criteriaDevice meets criteria
    Ti device manufacturing accuracy verificationPredetermined accuracy acceptance criteriaDevice meets criteria
    Torque through resistance verificationPredetermined acceptance criteriaDevice meets criteria
    Hardware verification inspection and analysisPredetermined acceptance criteriaDevice is compliant with criteria
    Surgical case report verificationPredetermined acceptance criteriaReports are compliant with criteria
    System validationPredetermined acceptance criteriaSystem has met user needs and is compliant

    Note: The document only states that the device "meets" or is "compliant" with predetermined acceptance criteria, without providing the specific quantitative or qualitative criteria themselves, nor the specific performance values (e.g., accuracy percentages, error margins).

    2. Sample Size for the Test Set and Data Provenance

    This information is not provided in the document. The document refers to "testing" but does not specify the sample size for any test set or the provenance (country of origin, retrospective/prospective) of data used for performance validation.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not provided in the document.

    4. Adjudication Method

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not provided in the document. The document focuses on the device's technical specifications and substantial equivalence to predicate devices, not on comparative effectiveness with human readers.

    6. Standalone (Algorithm Only) Performance Study

    The document does not explicitly present a standalone performance study with specific metrics (e.g., accuracy, sensitivity, specificity) for the imaging or segmentation algorithms. It mentions tmCMF Solution is "a software system and image segmentation system," and that "Software Verification and Validation Testing was conducted," but focuses on compliance with software standards (IEC 62304) and general system functionality rather than detailed standalone performance metrics against ground truth for clinical tasks.

    7. Type of Ground Truth Used

    The specific type of ground truth (e.g., expert consensus, pathology, outcomes data) used for the software's performance assessment is not explicitly stated in the document. It generally refers to "predetermined acceptance criteria" for various verification items.

    8. Sample Size for the Training Set

    This information is not provided in the document.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232889

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-30

    (225 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: METICULY Patient-specific titanium maxillofacial mesh implant Regulation Number: 21 CFR 872.4760
    |
    | Regulation: | 872.4760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    METICULY Patient-specific titanium maxillofacial mesh implant is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton and midface.

    Device Description

    The METICUL Y Patient-specific titanium maxillofacial mesh implant is a device designed individually for each patient and intended for use in selective trauma of the maxillofacial skeleton, maxillofacial surgery, and reconstructive procedures. It is specifically designed with a focus on applications of non-bending related scenarios. The implant is made of titanium alloys produced via additive manufacturing (Laser Powder Bed Fusion) and is intended to be used with titanium screws. All additive manufacturing and other post-processing steps are only to take place under Meticuly manufacturing control. The device is not intended to substitute for bone reconstruction in clinical situations where bone is needed for support and stability of the maxillofacial skeleton under functional loading conditions. It is intended for adults only (at least 22 years of age). The surgeon approves the design of the mesh implant prior to fabrication of the implant device by the sponsor. The proposed FaciMesh models are intended to treat the maxillofacial region, while the OrbiMesh models treat the Orbital region. The device is not intended for reconstruction of the orbital roof defects or for any frontal bone defects, such as the supra-orbital ridge. Additionally, the FDA-cleared commercially available titanium screws that can be used with the subject devices include the TITANIUM MINIPLATE SYSTEM (K951690) and MICRO TITANIUM PLATE SYSTEM (K951688).

    AI/ML Overview

    The METICULY Patient-specific titanium maxillofacial mesh implant underwent various performance tests to demonstrate its substantial equivalence to predicate devices.

    1. Acceptance Criteria and Reported Device Performance:

    The document broadly mentions compliance with various ASTM and ISO standards for material properties, manufacturing quality, biological evaluation, sterilization, and packaging. While specific quantitative acceptance criteria are not explicitly listed in a single table alongside reported values within the provided text, the conclusion states that "Verification and validation testing confirms that the product specifications have been met, demonstrating that the device will perform as intended. There were no unexpected results which indicate the suitable material used and manufacturing process compared to the standards for medical devices." This implies that the device met the requirements outlined in the referenced standards.

    Here's a breakdown of the performance aspects and the reported methods/outcomes:

    Acceptance Criteria CategoryReported Device Performance/Testing MethodImplied Acceptance Criteria (based on standards compliance)
    Material & ManufacturingEvaluated maintained composition, particle size, flowability, density, and visual characteristics of recycled/reused powder. Density of printed samples consistently within acceptance criteria across build plate locations. Dynamic fatigue strength testing to account for worst-case print orientation. Patient-matched device form output consistently matches digital design and CT scan data inputs.Virgin powder characteristics maintained after recycling. Consistent and acceptable density of printed implants. Sufficient dynamic fatigue strength for intended use (meets ASTM F3001-14). High fidelity between digital design, CT data, and fabricated implant (within specified tolerances).
    BiocompatibilityTested per ISO 10993-1, -3, -5, -6, -10, -11, -23, and USP-NF Pyrogen Test.Non-toxic, non-sensitizing, non-irritating, non-genotoxic, no unacceptable local effects after implantation, and non-pyrogenic.
    SterilizationValidated using Overkill method for steam sterilization (135°C for 10 minutes) to achieve SAL of 10^-6. Tested per ANSI AAMI ST72, USP-NF <85>, <161>, <71>, ISO 17665-1, ISO 11737-1, -2.Achieves a Sterility Assurance Level (SAL) of 10^-6. Complies with bacterial endotoxin limits and sterility verification.
    Packaging & TransportationTested per ASTM F88/F88M-15, D7386-16, F1886/F1886M-16, F1929-15.Maintains sterility and device integrity during storage and transport. Adequate seal strength, package performance, and seal integrity.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample sizes for the various performance tests conducted (e.g., number of powder samples, printed devices for fatigue testing, implants for biocompatibility).

    The data provenance is also not explicitly stated beyond the tests being conducted by the manufacturer to support the 510(k) submission. It doesn't mention country of origin for data or if it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not provided in the document. The studies described are primarily engineering and laboratory-based performance tests for the device itself (materials, manufacturing, sterility), not clinical studies requiring expert ground truth for interpretation of medical images or patient outcomes.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the described tests are laboratory-based device performance evaluations, not clinical studies requiring expert adjudication of results.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not done or reported in this document. The submission focuses on the chemical, physical, and mechanical properties of the device and its manufacturing process, not on its diagnostic or treatment efficacy as interpreted by human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This information is not applicable. The device is a physical implant, not an algorithm or software that performs a diagnostic or treatment function on its own. While the design process involves digital input (CAD data from CT scans), the "performance" described is of the physical mesh implant, not a standalone algorithm.

    7. Type of Ground Truth Used:

    For the performance tests, the "ground truth" is largely established by recognized industry standards (ASTM, ISO, USP-NF). For instance:

    • Material properties: Measured values compared against specified ranges in ASTM F3001-14.
    • Biocompatibility: Results of biological tests (cytotoxicity, irritation, sensitization, etc.) are compared against established biological endpoints defined by ISO 10993 series and USP-NF.
    • Sterilization: Achievement of a defined Sterility Assurance Level (SAL) and compliance with microbial limits as per ISO/ANSI/USP standards.
    • Manufacturing accuracy: Comparison of the physical implant to the digital design and CT scan data inputs, with acceptable tolerances implied by "consistently matches."

    No pathology, expert consensus on patient outcomes, or outcomes data is mentioned as ground truth for these device performance tests.

    8. Sample Size for the Training Set:

    This information is not applicable as the document describes a physical medical device, not an AI/ML algorithm that requires a training set. The device design process involves patient-specific CT data, but this is an input for individual device fabrication, not a training set in the context of AI/ML.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reasons as point 8.