Search Results
Found 2 results
510(k) Data Aggregation
(98 days)
KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofial skeletal system.
KLS Chin Plate System (K943348): The KLS Chin Plate System is indicated for 1) Fractures, 3) Reconstruction procedures of the craniomaxillofacial skeletal system.
KLS-Martin Micro Osteosynthesis System (1.0MM) (K944561): The KLS-Martin Micro Osteosynthesis System (1.0MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
KLS-Martin Micro Osteosynthesis System (1.5MM) (K944565): The KLS-Martin Micro Osteosynthesis System (1.5MM) is used in oral-maxillo-cranio-facial surgery to stabilize fractures. The bone segments are attached to the plate with screws to prevent movement of the segments.
KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bone fragments. The addition of the self drilling feature is the only difference between the submitted device and the predicate device referenced.
KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is only intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to remove malignant or benign the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMF or treatment of temporomandibular joint disease (TMD).
KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System is intended to be surgically placed in the mouth for use an an anchor for orthodontic procedures.
KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) are implants intended to be surgically placed in the mouth for use as an anchor for orthodontic procedures in patients.
KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile is in sterile packaging, osteosynthesis products with the following indications for use:
K051236: The RESORB-X® SF Sonotrode is only intended for use for insertion of the RESORB-X® SF pins.
K032442: The KLS Martin Mandibular/Reconstruction System II is intended for use in the stabilization of mandibular fractures and mandibular reconstruction.
K971297: The KLS Martin Centre-Drive Drill-Free screws are in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates which are contoured to fit the bony surface and stabilize the bone fragments. The addition of the self drilling feature is the only difference between the predicate device reference
K944565: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
K944561: The KLS-Martin Micro Osteosynthesis System is used in oral-maxillo-cranio-facial surgery to stabilize fractured bone segments. The bone segments are attached to the plate with screws to prevent movement of the segments.
KLS Martin Drill-Free MMF Screw (K042573): The KLS Martin Drill-Free MMF Screws is intended for use in maxilonandibular fixation of fractures of the maxilla, mandible, or both.
Drill Free MMF Screw (K083432): The Drill Free MMF Screw is intended for use in maxillomandibular fixation of fractures of the maxilla, mandible, or both.
KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is intended for temporary stabilization of maxillary fractures. It is designed to maintain proper occlusion during intraoperative bone healing (app. 6-8 weeks). It is indicated for the temporary treatment of maxillomandibular fixation (MMF) in adults or adolescents who have permanent teeth present (ages 12 and older).
KLS Mini Osteosynthesis System (K943347): The KLS Mini Osteosynthesis System consists of titanium non-locking plates ranging in thickness from 0.6mm - 2.5mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
KLS Chin Plate System (K943348): The KLS Chin Plate System consists of titanium plates ranging in thickness of 0.6mm and titanium screws ranging in diameter from 1.5mm - 2.3mm.
KLS-Martin Micro Osteosynthesis System (1.0mm) (K944561): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates, bone plates, bone screws and accessories of various shapes and sizes for use in oral-maxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 1.2mm.
KLS-Martin Micro Osteosynthesis System (1.5mm) (K944565): The KLS-Martin Micro Osteosynthesis System is designed to aid in the alignment and stabilization of the skeletal system after a facial fracture or surgery. The bone plates and screws of various shapes and sizes for use in oralmaxillo-cranio-facial surgery. The bone plates are manufactured from CP Titanium and range in thickness from 0.3mm - 0.6mm. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.5mm - 1.8mm.
KLS Martin Centre-Drive Drill-Free Screw (K971297): The KLS Martin Centre-Drive Drill-Free Screws are designed to eliminate the need for pre-drilled pilot holes. They are self-tapping with one step insertion. They are intended for use in rigid internal fixation of the oral-maxillo-cranio-facial bones. The bone screws are used to anchor plates where are contoured to fit the bone fragments. The bone fragments. The bone screws are manufactured from Titanium Alloy and range in diameter from 1.0mm - 2.0mm.
KLS-Martin Temporary Condylar Implant (K990667): The KLS-Martin Temporary Condylar Implant is a solid condylar head which attaches with fastening screws to a KLS-Martin Fracture/ Reconstruction Plate. The implant is available for left and right placement. The KLS-Martin Temporary Condy intended for temporary reconstruction of the mandibular condyle in patients who have undergone resective procedures to benign tumors requiring the removal of the mandibular condyle. This device is not for permanent implantation, for patients with TMJ or traumatic injuries, or for treatment of temporomandibular joint disease (TMD).
KLS-Martin Mandibular/Reconstruction System II (K032442): The KLS-Martin Mandibular/Reconstruction System II includes several different designs of titanium plates and screws intended for use in the stabilization and fixation of mandibular fractures and reconstruction. The plates are manufactured from either CP Titanium or Titanium Alloy and range in thickness from 1.0mm - 3.0mm. The screws are manufactured from either CP Titanium Alloy and range in diameter from 2.0mm - 3.2mm.
KLS-Martin Ortho Anchorage System (K033483): The KLS-Martin Ortho Anchorage System consists of a titanium screw designed to aid in dental movement by providing a rigid skeletal fixation point. The screw is intended to be surgically placed in the mouth for orthodontic procedures. The screws are manufactured from either CP Titanium or Titanium Alloy.
KLS-Martin Ortho Anchorage System (Plates) (K040891): The KLS-Martin Ortho Anchorage System (Plates) consists of titanium non-locking plates to aid in dental movement by providing a rigid skeletal fixation point. The plates are manufactured from either CP Titanium Alloy and are fixated with titanium screws and are utilized as an anchor for orthodontic procedures in the palatal, maxilla or mandible region.
KLS-Martin Drill-Free MMF Screw (K042573): The KLS-Martin Drill-Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The KLS-Martin Drill-Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, or both. The screws are manufactured from either CP Titanium Alloy and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
KLS Martin Rigid Fixation - Sterile (K060177): The KLS Martin Rigid Fixation - Sterile includes titanium plates of various shapes and thickness, titanium screws of various length and diameter, stainless steel twist drills of various length and stainless steel sonotrode tips that are provided in sterile packaging. The KLS Martin Rigid Fixation - Sterile is intended to provide KLS Martin's previously cleared osteosynthesis products in sterile packaging.
Drill Free MMF Screw (K083432): The Drill Free MMF Screw provides temporary occlusal and fracture stabilization. These screws may be applied prior to or after exposure of the fracture. The Drill Free MMF Screw is in maxillomandibular fixation to provide stabilization of fractures of the maxilla, mandible, or both. The screws are manufactured from Stainless Steel and are provided in 2.0mm diameter with lengths ranging from 8mm - 12mm.
KLS Martin L1 MMF System (K173320): The KLS Martin L1 MMF System is a bone-borne maxillomandibular fixation (MMF) system consisting of metalic archbars with sliding locking plates that attach to the dental arches with screws. The system is intended to provide temporary stabilization of mandibular and maxillary fractures as well as maintain properative bone fixation and postoperative bone healing (app. 6-8 weeks). The patient is brought into occlusion by wiring around the archbar wire hooks. The L1 MMF system plates are manufactured from CP Titanium (ASTM F67), are available in either a 7-hole siding plate configuration with two different lengths, and are 0.5mm in plate thickness. The L1 MMF system sliding locking plates are fixated with either 2.0 x 6 mm selfdrilling locking screws manufactured from Ti-6Al-4V (ASTM F136). Implants are available both sterile. The system also includes the necessary instruments to facilitate placement of the implants.
The document describes the KLS Martin Oral-Max Implants - MR Conditional, a bundled submission of various osteosynthesis systems and screws intended for use in craniomaxillofacial surgery. The purpose of this submission is to support the conditional safety and labeling modification of these devices in the magnetic resonance (MR) environment.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria are implied by the non-clinical tests conducted to support MR Conditional safety, aligning with relevant ASTM standards and FDA guidance. The reported device performance is that the devices can be safely scanned under specified conditions.
Acceptance Criteria (from ASTM/FDA Guidance) | Reported Device Performance (Summary from Submission) |
---|---|
Magnetically induced displacement force within acceptable limits (ASTM F2052-21) | Not explicitly quantified but implied as acceptable for MR Conditional labeling. |
Magnetically induced torque within acceptable limits (ASTM F2213-17) | Not explicitly quantified but implied as acceptable for MR Conditional labeling. |
Image artifacts within acceptable limits (ASTM F2119-07, R2013) | Not explicitly quantified but implied as acceptable for MR Conditional labeling. |
RF-induced heating (ASTM F2182-19e2) resulting in a temperature rise below 6 ℃ | Achieved under specified scanning conditions: 1.5 T/64 MHz and 3 T/128 MHz at a whole-body averaged specific absorption rate (wbSAR) of 2 W/kg or head SAR of 3.2 W/kg for an hour-long scanning session. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The testing involved computational modeling and simulation (CM&S). The "test set" in this context refers to the simulated scenarios and device configurations.
- Sample Size for Test Set: "the entire portfolio of KLS Martin maxillofacial implants" was simulated. This implies that all devices grouped under "KLS Martin Oral-Max Implants - MR Conditional" were included in the simulations. The document also mentions "various in-vivo device positions and landmarks," "worst-case single and multiple devices," and simulations in "10 cm increments." This suggests a comprehensive set of simulated scenarios rather than a traditional physical sample size.
- Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective" in the same way clinical data is. The data is generated through computational modeling and simulation using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and the Duke virtual human anatomy. This is a form of prospective simulation data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable in the traditional sense for this type of non-clinical, simulation-based study. The "ground truth" for the RF-induced heating simulations is derived from the established physics and engineering principles embedded in the FDA-qualified MDDT and the Duke virtual human anatomy model. The expertise lies in the development and validation of these computational tools and the interpretation of the simulation results by experts in MR safety and medical device engineering at MED Institute and the submitting company. The document does not specify the number or qualifications of individual experts validating the computational model, but implies that the MDDT itself is "FDA-qualified," indicating a level of expert review and agreement on its methodology.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like "2+1" or "3+1" are typically used for consensus building among human expert readers for clinical studies, especially when establishing ground truth from image interpretation. This study is based on physical property testing and computational simulations, not human interpretation of clinical data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not a study involving human readers or AI-assisted diagnostic performance. It focuses on the physical safety of implants in an MR environment.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The RF-induced heating assessment involved "Computational modeling and simulation (CM&S) ... using MED Institute's FDA-qualified Medical Device Development Tool (MDDT) and in a clinically relevant position within the Duke virtual human anatomy." This is a standalone algorithm/model-based assessment without a human-in-the-loop for the performance evaluation itself. Human experts design the simulations, configure the models, and interpret the results, but the "performance" (temperature rise, SAR calculations) is computed by the algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical tests:
- Magnetically induced displacement force, torque, and image artifacts: The "ground truth" is based on the physical properties of the materials and device designs, measured or calculated according to established ASTM standards (F2052-21, F2213-17, F2119-07).
- RF-induced heating: The "ground truth" for the simulations is derived from the established electromagnetic physics and thermal dynamics principles implemented in the FDA-qualified Medical Device Development Tool (MDDT) and applied to the Duke virtual human anatomy model. The MDDT's qualification process by the FDA implicitly establishes the reliability of its results as a form of "ground truth" for simulation-based assessments.
8. The sample size for the training set
Not applicable. This is a non-clinical study for MR safety assessment, not a machine learning model requiring a training set in the typical sense. The "training" for the MDDT is its initial validation and qualification against known physical phenomena and experimental data, which is a separate process from this submission.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for a machine learning model in this context. The "ground truth" for qualifying the simulation tool (MDDT) would have been established through extensive validation against experimental measurements and recognized physical theories.
Ask a specific question about this device
(279 days)
The subject devices are intended to be place in the mouth for use as an anchor in orthodontic procedures.
The anchor plates for attachment to the orthodontic device offer four types (T-Type, Y-Type and W-Type). This plate has three portions; head, arm and body, and there are two types of head portion. Anchor plate is developed to provide intraoral absolute anchorage for the intrusion or molars. Anchor Plate does not disturb any kind of tooth movement because they are placed outside the dentition. The anchor plate consists of pure titanium alloy, which is suitable for osseointegration and also tissue integration.
The provided document is a 510(k) premarket notification for a medical device called "Anchor plate." This type of document is generally used by manufacturers to demonstrate that their device is substantially equivalent to a predicate device already legally marketed in the U.S. It primarily focuses on comparing the new device against existing predicates based on technological characteristics, intended use, and performance data to establish safety and effectiveness.
This document does not describe the acceptance criteria and study that proves the device meets the acceptance criteria in the format requested in the prompt.
Specifically, it lacks:
- A table of acceptance criteria with reported device performance against those criteria.
- Details on sample size used for a "test set" (as this is not a diagnostic device, there isn't a traditional test set of clinical cases).
- Information on data provenance for a clinical test set.
- Number and qualifications of experts for ground truth establishment.
- Adjudication methods.
- Multi-reader multi-case (MRMC) comparative effectiveness study.
- Standalone (algorithm only) performance study.
- Type of ground truth used (pathology, outcomes data, etc.) for clinical evaluation.
- Training set size or how ground truth for a training set was established (again, not a diagnostic AI device).
Instead, the document focuses on non-clinical performance data and a comparison to predicate devices, which is typical for a 510(k) submission for a mechanical implantable device.
Here's an extraction of the relevant performance data and comparisons that are available in the document:
1. A table of acceptance criteria and the reported device performance:
The document mentions comparative performance testing but does not present a formal table of acceptance criteria. Instead, it compares the proposed device's mechanical properties to those of the primary predicate device. The "acceptance criteria" are implicitly that the proposed device performs comparably or better than the predicate device.
Performance Metric | Acceptance Criteria (Implied "Similar to Predicate") | Reported Device Performance (Anchor Plate) | Reported Predicate Performance (K091051) |
---|---|---|---|
Bending strength (Avg.) | Comparable to K091051 (97.125 N·mm) | 102.428 N·mm | 97.125 N·mm |
Tensile force (Avg.) | Comparable to K091051 (176.782 N) | 301.814 N | 176.782 N |
Biocompatibility | Compliance with ISO 10993-1 | Compliant per ISO 10993-1 | Compliant per ISO 10993-1 |
Sterilization Efficacy | Needs autoclave prior to use (specific temp/time) | Autoclave at 121°C for 30 min exposure, 15 min dry | Gamma Radiation (ISO 11137-1) |
Note: The document states "the results showed that the properties considered were similar when used for intended use" despite the numerical differences in bending strength and tensile force, suggesting that the observed differences were deemed clinically acceptable.
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated for performance testing. Mechanical tests typically involve a certain number of units, but the specific quantity is not reported. This is a non-clinical device; there isn't a "test set" of clinical data.
- Data Provenance: The tests were conducted by the manufacturer, BioMaterials Korea, Inc., (South Korea). It is preclinical (bench and biocompatibility) data, not clinical patient data. The tests are non-clinical (bench testing and biocompatibility).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. This is a pre-clinical (bench) study for a mechanical device, not a diagnostic or clinical study requiring expert ground truth for imaging or clinical outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This is a pre-clinical (bench) study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-enabled device requiring MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For mechanical properties: The "ground truth" is measured physical properties against established standards (e.g., ISO 9585 for bending strength) or internal validated test methods.
- For biocompatibility: The "ground truth" is established by adherence to recognized standards like ISO 10993.
8. The sample size for the training set:
- Not applicable. This is a mechanical device, not a machine learning model requiring a training set.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
Page 1 of 1