Search Results
Found 208 results
510(k) Data Aggregation
(30 days)
5S4
Canada
Re: K254114
Trade/Device Name: Baylis Connector Cable
Regulation Number: 21 CFR 870.2900
Common Name: Cable, transducer and electrode, patient, including connector
Classification Name: 21 CFR 870.2900
of the cable and updated list of compatible generators |
| Classification | Identical | 21 CFR 870.2900
The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture generators to separately cleared compatible RF devices.
The Baylis Connector Cable is a sterile, single-use device designed to connect Baylis-approved Radiofrequency (RF) generators (including the PrecisePath RF Puncture Generator and RFP-100A Generator) to separately cleared compatible RF devices (e.g., RF guidewires).
The Baylis Connector Cable is 10 feet long with a multi-pin positive locking Generator Connector at one end and a spring-loaded wire connection connector at the other end which allows connection to the RF delivery device. Between the connector ends is a single insulated conductor cable.
Principle of Operation: Enables RF power transfer from generator to RF puncture device; no changes from predicate device.
N/A
Ask a specific question about this device
(87 days)
Re: K252827
Trade/Device Name: IBP cables and IBP adapter cables
Regulation Number: 21 CFR 870.2900
description: Patient transducer and electrode cable (including connector)
Regulation number: 21 CFR §870.2900
The IBP cables and IBP adapter cables are intended to forward electrical signals from connected IBP pressure transducer sets to the patient monitor.
This traditional 510(k) submission comprises six distinct IBP cables and IBP adapter cables [IBP (adapter) cables] to transmit electrical signals acquired from an IBP transducer to the Draeger Vista 300 Patient Monitor.
The subject devices are passive electrical conductors in the context of invasive blood pressure (IBP) monitoring and do not modify the signal but convey it in an unchanged manner to the Draeger Vista 300 Patient Monitor.
N/A
Ask a specific question about this device
(123 days)
cable, Codan/DPT, 2m
- IBP transducer adapter cable, Codan/Xtrans, 0.3m
Regulation Number: 21 CFR 870.2900
name:** Patient transducer and electrode cable (including connector)
Regulation number: 21 CFR §870.2900
Number | Applying | K162768 | N/A |
| Device Classification | 2 | 2 | Same |
| Regulation Number | 870.2900
| 870.2900 | Same |
| Product Code | DSA | DSA | Same |
| Indications for Use | The reusable Dräger
The reusable Dräger IBP adapter cables are intended to forward electrical signals from pressure transducers to Dräger Hemo devices (Pods).
This traditional 510(k) submission covers 15 distinct IBP transducer adapter cables [also referred to as IBP adapter cables] developed and manufactured by Shanghai Draeger Medical Instrument Co., Ltd., based in Shanghai, China. These devices are passive electrical conductors in the context of invasive blood pressure (IBP) monitoring used to transmit electrical signals acquired from an IBP transducer (not part of this submission) to an external IBP Module (Pod) within a patient monitoring system (not part of this submission). They do not modify the signal but convey it in an unchanged manner to the monitoring device.
The Pod is referring to the MPod- QuadHemo and Mcable- DualHemo.
N/A
Ask a specific question about this device
(222 days)
. |
| DSA | 21 CFR 870.2900 Patient transducer and electrode cable (including connector). |
| MSX, DRT
Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in a hospital environment and during patient transport inside hospitals.
Not intended for home use. Intended for use by health care professionals.
The Telemetry Monitor 5500 is a battery-operated patient worn monitor with a touchscreen display. The Telemetry Monitor 5500 is intended to monitor and record, and to generate alarms for, multiple physiological parameters like ECG, SpO2, and respiration rate. It is equipped with a 1.4 GHz radio to enable wireless bi-directional data flow with Philips' Patient Information Center (PIC) iX.
This 510(k) clearance letter pertains to the Philips Telemetry Monitor 5500 Release 4.0. The provided document focuses on demonstrating substantial equivalence to a predicate device and includes information regarding performance testing against established standards. However, it does not contain specific details about acceptance criteria for particular performance metrics, nor does it describe studies proving the device meets those specific acceptance criteria in the format often associated with AI/ML device clearances (e.g., sensitivity, specificity, AUC).
Instead, the document primarily cites compliance with general medical device standards and internal testing to support its claims of performance. Therefore, many of the requested sections below cannot be fully populated as the information is not present in the provided text.
Acceptance Criteria and Device Performance
The document does not explicitly state quantitative acceptance criteria for device performance (e.g., a specific minimum sensitivity or accuracy percentage). Instead, it states that the device was assessed for conformity with relevant standards and that "results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards."
| Performance Metric | Acceptance Criteria (from document) | Reported Device Performance (from document) |
|---|---|---|
| SpO₂ Accuracy | Range between ± 1-3 % for measurements between 70-100 % SpO₂ (implied by predicate, identical claim for subject device) | Accuracy claims (results) are sensor specific and range between ± 1-3 % for measurements between 70-100 % SpO₂. |
| Electrical Safety | Conformity with IEC 60601-1 Edition 3.2 2020-08 | Found to comply |
| EMC | Conformity with ANSI AAMI IEC 60601-1-2:2014 [Including AMD1:2021] | Found to comply |
| Usability | Conformity with IEC 60601-1-6 Edition 3.2 2020-07 and ANSI AAMI IEC 62366-1:2015+AMD1:2020 | Found to comply; Summative usability testing has shown that the performance of the subject device does not raise questions of safety and effectiveness relative to the predicate device. |
| Alarm Systems | Conformity with ANSI AAMI IEC 60601-1-8:2006 and A1:2012 [Including AMD 2:2021] | Found to comply |
| ECG Monitoring Equipment | Conformity with ANSI AAMI IEC 60601-2-27:2011(R) 2016/IEC 60601-2-27:2011-03 | Found to comply |
| Multifunction Patient Monitoring | Conformity with IEC 80601-2-49 | Found to comply |
| Pulse Oximeter Equipment | Conformity with ISO 80601-2-61 Second edition 2017-12 | Found to comply |
| Software Life Cycle Processes | Conformity with ANSI AAMI IEC 62304:2006/A1:2016 and FDA's Guidance "Content of Premarket Submissions for Device Software Functions" | Development and testing conducted in accordance with these standards/guidance. |
Study Details:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated for any specific performance metric beyond general statements of testing compliance. For the "Clinical Studies" section related to SpO2 accuracy, it only notes that studies were conducted to "support accuracy performance" and "meet the acceptance criteria laid out in the associated protocols." No sample size for patients or data points is provided.
- Data Provenance: Not specified. The document mentions "Philips conducted clinical studies" but does not detail the country of origin of the data or whether it was retrospective or prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. The document describes performance testing against established standards and internal "accuracy specifications". For the SpO2 accuracy claims, ground truth would typically be established by invasive measurements (e.g., CO-oximetry of arterial blood samples), not by expert review of device output. The document does not provide details on how ground truth was established for the clinical studies supporting SpO2 accuracy.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable/Not Specified. Since expert review for ground truth is not indicated, adjudication methods are not relevant in the context described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The provided text does not mention any MRMC comparative effectiveness study. This device is a telemetry monitor for physiological parameters, not an AI-assisted diagnostic tool for human readers. "Summative usability testing" was done, indicating human interaction with the device, but not in a comparative effectiveness study involving AI assistance for human "readers."
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone algorithm performance is implied for certain functions, but not detailed. The device uses "Philips proprietary monitoring algorithms for ECG arrhythmia monitoring" (EASI, Hexad, ST/AR Arrhythmia Monitoring, ST/AR ST Analysis Algorithm, ST/AR QT/QTc Interval Monitoring). The performance of these algorithms would constitute "standalone" performance, but the document only states "Functional and system level testing... was performed. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards." No specific performance metrics for these algorithms are provided.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not explicitly stated for all performance metrics.
- For SpO₂ accuracy, ground truth is typically established by comparing the device's SpO₂ readings to invasive arterial blood gas analysis (CO-oximetry). The document only states "Philips conducted clinical studies... to support accuracy performance."
- For other performance metrics related to compliance with standards (e.g., electrical safety, EMC, usability), the "ground truth" is adherence to the standard's requirements, demonstrated through specific tests.
- Not explicitly stated for all performance metrics.
-
The sample size for the training set
- Not Applicable/Not Specified. The document mentions "Philips proprietary monitoring algorithms," which would have been developed using training data. However, the size or nature of any training set is not disclosed in this 510(k) summary, as it's not a primary requirement for demonstrating substantial equivalence for this type of device.
-
How the ground truth for the training set was established
- Not Applicable/Not Specified. The document does not provide information on how the ground truth for any potential training sets used in the development of "Philips proprietary monitoring algorithms" was established.
Ask a specific question about this device
(218 days)
102401
China
Re: K242986
Trade/Device Name: SpO2 Extension Cable
Regulation Number: 21 CFR 870.2900
Including Connector)
Classification: II
Product Code: DSA, DQA
Regulation Number: 21 CFR 870.2900
name | SpO2 Extension Cable | SpO2 Extension Cable K222370 | / |
| Classification Regulation | 21 CFR 870.2900
| 21 CFR 870.2900 | SAME |
| Classification | II | II | SAME |
| Product Code | DSA, DQA | DSA | SAME
SpO2 Extension Cables are intended to be used to connect SpO2 sensors, placed at appropriate sites on the patient to a monitoring device for SpO2 monitoring by health care professional.
SpO2 Extension Cable consists of two parts: the SpO2 patient cable and the adapter. The SpO2 patient cable is used for connecting the monitors, and the adapter is used for connecting the SpO2 patient cable and the SpO2 sensor.
The SpO2 Extension Cable is the replacements for similar cables manufactured by Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors.
By using the same types of construction and technological characteristics to the compatible patient monitor and SpO2 sensor, the SpO2 Extension Cable can avoid measured data corrupted.
The proposed device was modification from the legally marketed (existing) device "SpO2 Extension Cable"/K222370, manufactured by "Beijing Rongrui-Century Science & Technology Co., Ltd.", same as the sponsor of this submission.
Compared to the initial product (K222370), only the structure of the SpO2 Extension Cable has changed. Originally a non-separable complete extension cable, now an extension cable can be separated into two parts, called the SpO2 patient cable and the adapter.
The type of instrument connector (which connects one end of the monitoring device) for the SpO2 patient cable is 20pin, there are four types pairing interfaces between SpO2 patient cable and adapter, the pairing interfaces name for the SpO2 patient cable and adapter connection is DB9F-LNCS, MLNCS-MLNCS, LNOP-LNOP, RD-RD and the other end of the adapter connecting to SpO2 sensor is DB9F. Adapters are divided into two categories: with cable and without cable. Adapters for each type of interface are available with and without cables.
The provided FDA 510(k) Clearance Letter for the "SpO2 Extension Cable" is for a Class II medical device, specifically an extension cable used to connect SpO2 sensors to monitoring devices. Based on the content of the letter, the submission is about demonstrating substantial equivalence to a predicate device, not necessarily proving that an AI/ML algorithm meets specific performance criteria.
Therefore, many of the requested points, such as AI/ML specific acceptance criteria, sample sizes for training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI/ML, are not applicable to this type of device clearance. This clearance focuses on electrical safety, EMC, biocompatibility, and functional compatibility of a hardware component.
Here's a breakdown of the requested information based on the provided document, highlighting what is applicable and what is not:
Analysis of Acceptance Criteria and Study for SpO2 Extension Cable (K242986)
This submission (K242986) pertains to an SpO2 Extension Cable, which is a hardware component. The clearance is based on demonstrating substantial equivalence to a previously cleared predicate device (K222370), rather than establishing performance of an AI/ML algorithm. Therefore, many of the typical acceptance criteria and study details associated with AI/ML devices are not relevant here.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this type of hardware device are primarily based on compliance with recognized consensus standards and functional compatibility. The "performance" here refers to meeting these engineering and safety specifications, rather than clinical accuracy metrics typically seen for diagnostic algorithms.
| Acceptance Criterion (Standard Compliance / Test) | Reported Device Performance (Compliance) |
|---|---|
| Biocompatibility: | |
| ISO 10993-5: In Vitro Cytotoxicity | Meets standard |
| ISO 10993-10: Irritation and Skin Sensitization | Meets standard |
| Electrical Safety & Performance: | |
| IEC 60601-1 (Edition 3.2): General requirements | Complies |
| ISO 80601-2-61: Pulse Oximeter Equipment specifics | Complies |
| ANSI AAMI EC53: ECG trunk cables/leadwires (likely for general cable integrity/materials, given context) | Complies |
| Electromagnetic Compatibility (EMC): | |
| IEC 60601-1-2 (Edition 4.1): EMC requirements | Complies |
| Functional Performance: | |
| Compatibility testing (no measured data corrupt during communication between SpO2 sensors and host monitors) | Performance accuracy evaluated and proven |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated in terms of a "test set" like in AI/ML validation. Compliance is shown through testing of engineering samples of the device itself and its components against the specified standards.
- Data Provenance: Not applicable in the context of clinical data for an AI/ML algorithm. The "data" here comes from laboratory testing of the physical cable. The device manufacturer is Beijing Rongrui-Century Science & Technology Co., Ltd. in China. The testing would have been conducted in a controlled lab environment.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not Applicable. Ground truth, in the sense of expert annotation for clinical data, is not established for this type of hardware device. The device's performance is verified against engineering standards and functional specifications.
4. Adjudication Method for the Test Set
- Not Applicable. As there's no "ground truth" derived from expert consensus on clinical cases, no adjudication method is relevant.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. An MRMC study is designed to assess the impact of an AI algorithm on human reader performance. This device is an extension cable, not an AI algorithm.
6. Standalone (Algorithm Only) Performance
- Not Applicable. This is a physical cable, not a software algorithm.
7. Type of Ground Truth Used
- Not Applicable. The "ground truth" for this device's acceptance is its adherence to recognized engineering, electrical, biocompatibility, and functional performance standards. This is established through non-clinical testing results, not clinical outcomes data or expert consensus on patient cases.
8. Sample Size for the Training Set
- Not Applicable. No training set exists as this is a hardware product, not a machine learning model.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. No training set or associated ground truth for training.
Summary of the K242986 Submission's Core Argument:
The core of the K242986 submission is to demonstrate that the modified structure of the SpO2 Extension Cable (now separable into two parts) does not negatively impact its safety and effectiveness when compared to the predicate device (K222370), which was a non-separable cable. This is achieved by showing continued compliance with relevant performance, safety, and compatibility standards. The primary difference noted is a structural change, and the manufacturer successfully demonstrated that this change "will not affect the safety and effectiveness of the proposed device" through non-clinical testing.
Ask a specific question about this device
(165 days)
: BZQ / 21 CFR868.2375• Secondary: DQA / 21CFR 870.2700• Secondary: DSA / 21 CFR870.2900
Airmod, when used in conjunction with Accursound Electronic Stethoscope AS-101, is a software as medical device intended to be used for the continuous, non-invasive monitoring of respiratory rate (RR) in adult patients who are subjected to procedural sedation and/or anesthesia.
Airmod is intended for use by healthcare professionals in hospitals and healthcare facilities who are legally credentialed to perform procedural sedation and/or anesthesia.
Airmod is intended for Android-based devices only.
Airmod 114 is an Android-based software application designed to aid healthcare professionals by monitoring a sedated and/or anesthetized patient's breathing in real time. The device has an AIbased algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals of breathing sounds collected by AccurSound Electronic Stethoscope AS-101. Airmodid for the continuous monitoring of respiratory rate (RR) in adults who are subjected to procedural sedation. Airmod™ is designed for use in hospitals and healthcare facilities performing procedural sedation/anesthesia. The device is not intended for patients who are not anesthetized/sedated.
Here's a breakdown of the acceptance criteria and study details for the Airmod device, based on the provided FDA 510(k) summary:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance (Airmod™) |
|---|---|
| Root Mean Square Error (RMSE) < 3 Breaths Per Minute (BPM) when compared to manual-scored capnography (mancRR) | RMSE = 2.689 BPM (95% CI: 2.529, 2.695) |
Study Details
1. Sample Size Used for the Test Set and Data Provenance:
* Sample Size: 270 participants
* Data Provenance: The study included both "outside of US (OUS) and US populations." This indicates a combination of retrospective and prospective data, though the document does not explicitly state the temporal nature of the data collection for each region.
2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
* The document mentions "manual scored capnography (mancRR)" as the reference for ground truth. However, it does not specify the number of experts used or their qualifications for establishing this ground truth.
3. Adjudication Method:
* The document does not specify the adjudication method used for establishing the ground truth from manual-scored capnography.
4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
* No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The clinical testing focused on the non-inferiority of Airmod™'s respiratory rate measurements compared to a capnography system. There is no mention of human readers improving with or without AI assistance.
5. Standalone Performance Study:
* Yes, a standalone performance study was done. The clinical testing evaluated the performance of Airmod™ (algorithm only, as a software as medical device) against the Capnostream™35 capnography system. The results of the RMSE directly reflect the algorithm's standalone accuracy.
6. Type of Ground Truth Used:
* The ground truth used was manual-scored capnography (mancRR). This provides a direct, physiological measurement of respiratory rate that is then manually interpreted or verified.
7. Sample Size for the Training Set:
* The document does not specify the sample size used for the training set of the AI-based algorithm. The 270 participants relate to the clinical testing (test set) for validation.
8. How the Ground Truth for the Training Set Was Established:
* The document does not provide details on how the ground truth for the training set was established. It only mentions that the device has an "AI-based algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals."
Ask a specific question about this device
(63 days)
5 Leadset, Grabber, Chest, AAMI, OR; M1680A, 6 Lead Set Grabber AAMI, ICU Regulation Number: 21 CFR 870.2900
Classification Name | Panel & Name: Cardiovascular DevicesSubpart & Division: 21 CFR §870.2900
Philips IntelliVue reusable ECG lead sets/trunk cables and Philips OR ECG trunk cables are indicated for continuous monitoring of cardiac signals for both diagnostic and monitoring purposes. These devices are limited by the indications for use of the connected monitoring and diagnostic equipment in healthcare facilities. These devices are intended to interact with patient intact skin only.
Philips Reusable ECG Lead Sets and Trunk Cables are non-invasive reusable medical devices designed to transfer ECG signals from the patient to the monitor used in the healthcare facilities, which includes intensive care unit (ICU) for both adults and neonates and the operating room (OR). The ECG cables and lead sets are used to forward an electrical cardiac signal via electrical wires from the electrode attached on patient skin to the ECG measurement hardware in the patient monitor and to support other outputs as needed. The portfolio consists of two types of trunk cables and two types of lead sets: ICU and OR, including various electrode connectors with AAMI and IEC coloring. The ECG Trunk cables and lead sets are suitable for multiple patient use. The Philips Reusable ECG Lead Sets and Trunk Cables have a service life of 18 months. The Philips Reusable ECG Lead Sets and Trunk Cables can be used with any Philips monitors including Philips monitor/defibrillator product lines and Philips monitoring or diagnostic equipment which claim compatibility to Philips Reusable ECG Lead Sets and Trunk Cables.
This document is a 510(k) premarket notification for Philips Reusable ECG Lead Sets and Trunk Cables. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a standalone study with clinical accuracy metrics. Therefore, many of the requested details about acceptance criteria, clinical study design, and ground truth are not applicable or explicitly stated in this type of submission.
However, I can extract the information that is present regarding performance testing and standards.
1. Table of Acceptance Criteria and Reported Device Performance
Instead of specific quantitative clinical acceptance criteria and device performance metrics, this submission relies on compliance with recognized harmonized standards. The "performance" is demonstrated through successful completion of tests outlined in these standards.
| Acceptance Criteria (Compliance with Standard) | Reported Device Performance |
|---|---|
| IEC 60601-1:2005+A1:2012+A2:2020 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. |
| ANSI AAMI EC53:2013/(R)2020 (ECG trunk cables and patient lead wires) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. |
| IEC 60601-2-25:2011/(R)2016 (Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. |
| IEC 60601-2-27:2011(R)2016 (Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic equipment) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. |
| IEC 60601-1-12 Edition 1.1 2020-07 CONSOLIDATED VERSION (Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. (Specifically, ICU Lead Sets and Trunk cables comply with this additional standard). |
| IEC 62366-1:2015+AMD1:2020: (Part 1: Application of usability engineering to medical devices) | The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. |
| Reprocessing Standards: ISO 17664-2:2021, AAMI TIR12 (2020)/(R)2023, and ANSI AAMI ST98:2022, and FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" | The subject device adheres to these standards, and its proposed reprocessing instructions are as effective as the predicate device, meeting regulatory requirements for reusable devices. |
| Biocompatibility: (ISO 10993-1, as inferred from comparison to predicate) | Both the subject and predicate devices meet standards for biocompatibility. |
| Electrical Conductivity, Mechanical Strength: (as inferred from comparison to predicate) | Both the subject and predicate devices meet standards for electrical conductivity and mechanical strength. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This document explicitly states: "The subject devices, like the primary predicate devices, did not require clinical trials." Therefore, there is no "test set" in the context of a clinical study for external validation of performance. Performance is primarily demonstrated through compliance with recognized standards via non-clinical bench testing. The data provenance would be from internal lab testing, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. No clinical test set requiring expert ground truth was used as per the statement above.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. No clinical test set requiring expert adjudication was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an ECG cable and lead set, a hardware accessory. It is not an AI algorithm for interpreting medical images or signals, nor does it involve "human readers" or "AI assistance" in that sense.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a hardware accessory, not an algorithm. Performance is assessed through its physical and electrical properties, not algorithmic output.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable in the sense of clinical ground truth. For non-clinical bench testing, the "ground truth" would be established by the specifications and measurement techniques defined within the harmonized standards (e.g., electrical resistance, mechanical durability measured against stated tolerances, biocompatibility testing results).
8. The sample size for the training set
Not applicable. This is a hardware accessory, not a machine learning model, so there is no training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set.
Ask a specific question about this device
(246 days)
Branford, Connecticut 06405
Re: K240087
Trade/Device Name: RT Carbon ECG Leads Regulation Number: 21 CFR 870.2900
transducer and electrode cable (including connector) |
| Classification Regulation: | 21 CFR 870.2900
tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900
The Ivy Biomedical RT Carbon ECG Leads are intended for use in X-ray imaging environments. They connect between ECG electrodes that are placed on the patient by a trained healthcare professional and an ECC trunk cable that connects to the ECG monitoring device. The RT Carbon ECG Leads are reusable and nonsterile. The RT Carbon ECG Leads are intended for adult, pediatric and neonatal patient populations.
The RT Carbon ECG Leads are intended to connect electrodes placed on the patient to a physiological detection module or monitor (CTM-300, CTM-400, CTM-500, 7000 Series and 3000 Series). The design includes 3 and 4 lead designs with overall lengths of 24", 30" and 36".
The provided text describes a 510(k) premarket notification for the Ivy Biomedical RT Carbon ECG Leads. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing rather than presenting novel clinical efficacy or diagnostic performance data. Therefore, many of the requested categories related to AI performance studies are not applicable.
Here's the information extracted and categorized as requested:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria alongside reported device performance in the format of specific thresholds and measured values for the RT Carbon ECG Leads. Instead, it states that the device was "thoroughly tested from design requirements through device verification and validation testing as required by 21 CFR 870.2900" and in accordance with EC53. The reported performance is summarized as:
"The performance data support the safety of the device and hardware verification and validation demonstrate that the RT Carbon ECG Leads should perform as intended in the specified use conditions and is substantially equivalent to the predicate."
This implies that the device met the acceptance criteria derived from the listed tests.
| Acceptance Criterion (Implicit from tests performed) | Reported Device Performance |
|---|---|
| Biocompatibility: | |
| Cytotoxicity | Met (implies no cytotoxicity) |
| Sensitization | Met (implies no sensitization) |
| Irritation | Met (implies no irritation) |
| Systemic Toxicity | Met (implies no systemic toxicity) |
| Pyrogen Testing | Met (implies no pyrogenicity) |
| Performance Testing (per EC53): | |
| Patient Leadwire to Trunk Cable Interconnection | Met (implies proper interconnection) |
| Flex life | Met (implies satisfactory durability) |
| Tensile strength | Met (implies satisfactory strength) |
| Connector mating/unmating cycles | Met (implies satisfactory durability) |
| Connector retention force | Met (implies satisfactory retention) |
| Contact resistance | Met (implies satisfactory electrical contact) |
| Leadwire resistance | Met (implies satisfactory electrical resistance) |
| Dielectric withstand voltage | Met (implies satisfactory insulation) |
2. Sample size used for the test set and the data provenance
The document does not specify the sample size for individual tests conducted on the RT Carbon ECG Leads. The testing was described as "conducted at Ivy" (implying in-house lab testing) and does not involve human data, thus the provenance (country of origin, retrospective/prospective) is not applicable in the context of clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI/diagnostic device that requires expert ground truth establishment for a test set. The tests are engineering and biological evaluations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not an AI/diagnostic device that requires adjudication of interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an ECG lead, not an AI system or diagnostic tool for interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an ECG lead, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The 'ground truth' for the performance testing is based on established engineering standards (EC53) and biocompatibility standards (ISO 10993-1). Compliance with these standards serves as the benchmark against which the device's performance is measured.
8. The sample size for the training set
Not applicable. This is a medical device (ECG lead), not an AI model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is a medical device (ECG lead), not an AI model.
Ask a specific question about this device
(90 days)
Redmond, Washington 98052
Re: K240377
Trade/Device Name: LIFEPAK® ECG Cables Regulation Number: 21 CFR 870.2900
---|--------------|
| Cable, Transducer and Electrode, Patient,(Including Connector), Class II (870.2900
The LIFEPAK ECG cables are reusable electrode cable systems intended to transmit signals from patient electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes. LIFEPAK ECG cables are intended for use in outdoor and indoor emerqency care settings.
LIFEPAK ECG Cables:
- LIFEPAK 3-wire ECG Cable -
- LIFEPAK 5-wire ECG Cable i
- -LIFEPAK 4-wire ECG cable
- LIFEPAK 3-wire extended precordial ECG Cable l
- -LIFEPAK 6-wire expandable precordial ECG cable
The LIFEPAK ECG cables are an accessory designed to be used with Physio-Control LIFEPAK 35 monitor/defibrillator. The LIFEPAK ECG cables are intended for use by Advanced Life Support (ALS) and Basic Life Support (BLS) providers to transmit signals from patient electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes.
The LIFEPAK ECG Cables are reusable ECG trunk cables and lead wires that connect to the LIFEPAK 35 monitor/defibrillator for ECG monitoring. There are several types or configurations of ECG cables: 3-wire, a 4-wire cable assembly to be used with the LIFEPAK 6-wire precordial cable and/or the LIFEPAK 3-wire precordial cable to perform a 12 or 15-lead ECG Cable assembly. When the LIFEPAK 4-wire ECG cable is configured with a 6-wire expandable precordial cable, a 12-lead ECG is constructed. Likewise, when the LIFEPAK 4-wire ECG cable is configured with a 6-wire expandable precordial cable and a 3-wire extended precordial ECG cable a 15-lead ECG is constructed.
The 4- wire ECG cable is available in two lengths, 5 feet or 8 feet. Additionally, the 3-, 4-, and 5-wire ECG Cables and the 6-wire precordial ECG cable are available with color coding according to the American Heart Association (AHA) or International Electrotechnical Commission (IEC) standards.
The LIFEPAK ECG cables are a reusable electrode cable system intended to transmit signals from patient electrodes to the LIFEPAK 35 monitor/defibrillator for both diagnostic and monitoring purposes. These accessories consist of conductor wires that are shielded and insulated by a polyurethane jacket material. The lead wires are color coded according to which electrode location they are attached to on the patient.
The LIFEPAK ECG cables are not intended for use with other manufacturers' defibrillators. The LIFEPAK ECG cables are intended only for use with LIFEPAK 35 monitor/defibrillator.
The provided text describes the 510(k) premarket notification for LIFEPAK® ECG Cables. This is a device for transmitting ECG signals, not an AI/ML-based medical device. Therefore, many of the requested criteria (such as MRMC studies, AI/ML-specific ground truth, training set information, etc.) are not applicable to this submission.
The document focuses on demonstrating substantial equivalence to a predicate device (Multi-Link™ X2 ECG Cable and Leadwire System) primarily through performance testing of the hardware components and comparison of technical characteristics.
Here's an attempt to answer the questions based only on the provided document, acknowledging that much of it is not relevant for an AI/ML device:
1. A table of acceptance criteria and the reported device performance
The document lists performance tests conducted against applicable standards:
| Test Performed | Applicable Standard | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Connector mating/unmating, Retention force | 60601-2-25:2011, 60601-2-27:2011, ANSI AAMI EC53:2013/(R)2020 | Meet standard requirements | Met |
| Inspection of Air Clearance | 60601-2-25:2011, 60601-2-27:2011 | Meet standard requirements | Met |
| Cable Noise Testing | ANSI AAMI EC53:2013/(R)2020 | Meet standard requirements | Met |
| Flex/Twist Life Testing | ANSI AAMI EC53:2013/(R)2020 | Meet standard requirements | Met |
| Tensile Strength Testing | ANSI AAMI EC53:2013/(R)2020 | Meet standard requirements | Met |
| Defibrillation Protection and Energy Reduction | IEC 60601-1:2005/AMD2:2020, IEC 60601-2-25:2011, IEC 60601-2-27:2011 | Meet standard requirements | Met |
| Dielectric Withstand Testing | IEC 60601-1:2005/AMD2:2020, ANSI AAMI EC53:2013/(R)2020 | Meet standard requirements | Met |
| Leakage Current Testing | IEC 60601-1:2005/AMD2:2020 | Meet standard requirements | Met |
| Storage Conditioning and Drop Test | IEC 60601-1:2005/AMD2:2020, IEC 60601-2-27:2011 | Meet standard requirements | Met |
| Material Resistance for Cleaning and Disinfection Stress | IEC 60601-1:2005/AMD2:2020 | Meet standard requirements | Met |
| Electromagnetic Compatibility | IEC 60601-1-2:2014 | Meet standard requirements | Met |
| Shock, Vibration and IP Classification | IEC 60601-1:2005/AMD2:2020, IEC 60601-2-27:2011, IEC 60529: 2013 | Meet standard requirements | Met |
| Biocompatibility Assessment | ISO 10993-1 | Meet standard requirements | Yes - Assessed |
| ASCA Test Reports | N/A (General term) | Meet standard requirements | Yes |
| Design Verification Testing | N/A (General term) | Meet standard requirements | Yes |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each physical test. It confirms that "Performance testing has been completed to demonstrate that the proposed LIFEPAK ECG Cables meet the safety and performance requirements established in the design specifications." These are typically laboratory-based engineering tests on device samples, not studies on patient data. The provenance for these tests is the manufacturer (Physio-Control Inc.). The tests are "non-clinical performance testing."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a hardware device undergoing engineering and safety performance testing, not an AI/ML algorithm requiring expert interpretation of medical images or data for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are typically for human-read or AI-generated interpretations of medical data, not for physical performance tests of a cable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML-based device. No human clinical studies were submitted as part of this 510(k) Premarket Notification.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML-based device.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is typically adherence to established engineering, electrical, and biocompatibility standards (e.g., IEC 60601, ANSI AAMI EC53, ISO 10993-1). Compliance with these standards confirms the device's technical specifications and safety profile, rather than a clinical ground truth like a pathology report.
8. The sample size for the training set
Not applicable. This is not an AI/ML-based device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML-based device.
Ask a specific question about this device
(102 days)
Lifesciences TruWave Invasive Pressure Transducers (or equivalent)) (11230-000024) Regulation Number: 21 CFR 870.2900
------|------------------------|--------------|
| Patient transducer and electrode cable(21 CFR 870.2900
|
| Regulation Number | 21 CFR 870.2900
| 21 CFR 870.2900
The LIFEPAK Invasive Pressure Adapter Cable is intended to connect invasive pressure transducers from the patient to the LIFEPAK monitor/defibrillator for invasive pressure monitoring purposes and/or aid in diagnostic evaluation by a health care professional. The LIFEPAK Invasive Pressure Adapter Cable is intended for use in outdoor and indoor emergency care settings.
The LIFEPAK Invasive Pressure Adapter Cable is a reusable, insulated, shielded, electrical cord (trunk cable) with a proximal connector (to connect with the LIFEPAK 35 device) and a main yoke that houses three distal leans with connectors (to connect with invasive pressure transducers). The LIFEPAK Invasive Pressure Adapter Cable is designed to transmit an electrical signal (data) between devices (i.e., to connect invasive pressure transducers from the patient to the LIFEPAK 35 monitor/defibrillator). The invasive pressure cable is designed to connect up to three pressure transducers to the front panel of the LIFEPAK 35 monitor/defibrillator for invasive pressure monitoring. Three channels are available for invasive pressure (IP) monitoring, with labels P1, P2, and P3. Each channel also has a user-selectable label for the line type through the LIFEPAK 35 monitor/defibrillator.
Invasive Pressure monitoring involves the conversion of fluid pressure into an electrical signal. The conversion is accomplished with a pressure transducer (i.e., IP Probe). The invasive pressure cable passes the electrical signal from the transducers to the LIFEPAK 35 monitor/defibrillator.
The LIFEPAK Invasive Pressure Adapter Cable is available in two different models that are compatible with either ICU Medical Transpac® IV Disposable Pressure Transducers or Edwards Lifesciences TruWave® Disposable Pressure Transducers. These pressure transducers provide industry standard sensitivity and defibrillation protection of at least 360 joules.
The LIFEPAK Invasive Pressure Adapter Cable is not intended for use with other manufacturers' defibrillators and/or monitors. The LIFEPAK Invasive Pressure Adapter Cable is intended only for use with LIFEPAK 35 monitor/defibrillator.
The provided document describes a 510(k) premarket notification for the "LIFEPAK® Invasive Pressure Adapter Cable." It does not contain information about acceptance criteria for a device's performance, nor does it detail a study proving the device meets specific performance criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device through general performance testing.
Therefore, most of the requested information cannot be extracted from this document. However, I can provide the available details:
1. A table of acceptance criteria and the reported device performance:
This information is not provided in the document. The document states "Performance testing has been completed to demonstrate that the proposed LIFEPAK Invasive Pressure Adapter Cable meets the safety and performance requirements established in the design specifications." However, the specific acceptance criteria and the numerical results of these tests are not detailed.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the document. The document only mentions "Biocompatibility Evaluation" and "Design Verification Testing" were completed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. The document does not describe a study involving expert-established ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. The document does not describe a study with adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The device is an adapter cable, not an AI-powered diagnostic tool, and no MRMC study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a physical cable, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
Not applicable. The document does not describe a study that uses ground truth in this context. The "performance data" refers to engineering and biocompatibility testing.
8. The sample size for the training set:
Not applicable. The device is a physical cable, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. The device is a physical cable, not a machine learning algorithm.
Ask a specific question about this device