K Number
K242798

Validate with FDA (Live)

Device Name
Airmod
Date Cleared
2025-02-28

(165 days)

Product Code
Regulation Number
868.2375
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Airmod, when used in conjunction with Accursound Electronic Stethoscope AS-101, is a software as medical device intended to be used for the continuous, non-invasive monitoring of respiratory rate (RR) in adult patients who are subjected to procedural sedation and/or anesthesia.

Airmod is intended for use by healthcare professionals in hospitals and healthcare facilities who are legally credentialed to perform procedural sedation and/or anesthesia.

Airmod is intended for Android-based devices only.

Device Description

Airmod 114 is an Android-based software application designed to aid healthcare professionals by monitoring a sedated and/or anesthetized patient's breathing in real time. The device has an AIbased algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals of breathing sounds collected by AccurSound Electronic Stethoscope AS-101. Airmodid for the continuous monitoring of respiratory rate (RR) in adults who are subjected to procedural sedation. Airmod™ is designed for use in hospitals and healthcare facilities performing procedural sedation/anesthesia. The device is not intended for patients who are not anesthetized/sedated.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Airmod device, based on the provided FDA 510(k) summary:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance (Airmod™)
Root Mean Square Error (RMSE) < 3 Breaths Per Minute (BPM) when compared to manual-scored capnography (mancRR)RMSE = 2.689 BPM (95% CI: 2.529, 2.695)

Study Details

1. Sample Size Used for the Test Set and Data Provenance:
* Sample Size: 270 participants
* Data Provenance: The study included both "outside of US (OUS) and US populations." This indicates a combination of retrospective and prospective data, though the document does not explicitly state the temporal nature of the data collection for each region.

2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
* The document mentions "manual scored capnography (mancRR)" as the reference for ground truth. However, it does not specify the number of experts used or their qualifications for establishing this ground truth.

3. Adjudication Method:
* The document does not specify the adjudication method used for establishing the ground truth from manual-scored capnography.

4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
* No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The clinical testing focused on the non-inferiority of Airmod™'s respiratory rate measurements compared to a capnography system. There is no mention of human readers improving with or without AI assistance.

5. Standalone Performance Study:
* Yes, a standalone performance study was done. The clinical testing evaluated the performance of Airmod™ (algorithm only, as a software as medical device) against the Capnostream™35 capnography system. The results of the RMSE directly reflect the algorithm's standalone accuracy.

6. Type of Ground Truth Used:
* The ground truth used was manual-scored capnography (mancRR). This provides a direct, physiological measurement of respiratory rate that is then manually interpreted or verified.

7. Sample Size for the Training Set:
* The document does not specify the sample size used for the training set of the AI-based algorithm. The 270 participants relate to the clinical testing (test set) for validation.

8. How the Ground Truth for the Training Set Was Established:
* The document does not provide details on how the ground truth for the training set was established. It only mentions that the device has an "AI-based algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals."

{0}------------------------------------------------

February 28, 2025

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Heroic Faith International Ltd. % Pierre Bounaud Principal Consultant Ram+ 2790 Mosside Blvd, Suite 800 Monroeville, Pennsylvania 15146

Re: K242798

Trade/Device Name: Airmod Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZO Dated: September 13, 2024 Received: September 16, 2024

Dear Pierre Bounaud:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242798

Device Name Airmod

Indications for Use (Describe)

Airmod, when used in conjunction with Accursound Electronic Stethoscope AS-101, is a software as medical device intended to be used for the continuous, non-invasive monitoring of respiratory rate (RR) in adult patients who are subjected to procedural sedation and/or anesthesia.

Airmod is intended for use by healthcare professionals in hospitals and healthcare facilities who are legally credentialed to perform procedural sedation and/or anesthesia.

Airmod is intended for Android-based devices only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

DATE PREPARED

February 28, 2025

MANUFACTURER AND 510(k) OWNER

Heroic Faith International Ltd. 18F., No. 460, Sec. 4, Xinyi Rd., Xinyi Dist. Taipei City, 110 Taiwan (R.O.C.) Telephone: +886-22659-0291 Official Contact: Yuan-Ren Cheng, CEO

REPRESENTATIVE/CONSULTANT

Pierre Bounaud, Ph.D. Allison Komiyama, Ph.D. RAC ROM+ Email: pbounaud(@rqmplus.com, akomiyama(@rqmplus.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Airmod™
Regulation Name:Breathing frequency monitor
Regulation Number:21 CFR 868.2375
Class:II
Product Code:BZQ
Premarket Review:OHT1/ Division of Anesthesia, Respiratory, and SleepDevices (DHT1C)
Review Panel:Anesthesiology

PREDICATE DEVICE IDENTIFICATION

Airmod 104 is substantially equivalent to the following predicates:

510(k) NumberPredicate Device Name / ManufacturerPrimaryPredicateReferenceDevice
K111933Covidien Nellcor Respiration Rate Software, Version 1.0 and Covidien Nellcor Adult Respiratory Sensor / Covidien
K120984Masimo Acoustic Monitoring Sensors / Masimo Corporation

The predicate devices have not been subject to a design related recall.

DEVICE DESCRIPTION

Airmod 114 is an Android-based software application designed to aid healthcare professionals by monitoring a sedated and/or anesthetized patient's breathing in real time. The device has an AIbased algorithm that can detect inhalation acoustics and provides respiratory rates based on the analysis of the acoustic signals of breathing sounds collected by AccurSound Electronic Stethoscope AS-101. Airmodid for the continuous monitoring of respiratory rate (RR) in adults who are subjected to procedural sedation. Airmod™ is designed for use in hospitals and healthcare facilities performing procedural sedation/anesthesia. The device is not intended for patients who are not anesthetized/sedated.

{5}------------------------------------------------

INDICATIONS FOR USE

Airmod™, when used in conjunction with AccurSound Electronic Stethoscope AS-101, is a software as medical device intended to be used for the continuous, non-invasive monitoring of respiratory rate (RR) in adult patients who are subjected to procedural sedation and/or anesthesia. Airmod TM is intended for use by healthcare professionals in hospitals and healthcare facilities who are legally credentialed to perform procedural sedation and/or anesthesia. Airmod1M is intended for Android-based devices only.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Heroic Faith International believes that Airmod™ is substantially equivalent to the predicate device based on the information summarized here:

The subject device has the same intended use and a similar design as the primary predicate device cleared in K111933. i.e., a software intended to be used with sensors as input to provide continuous, non-invasive monitoring of respiratory rates in a hospital setting. While the primary predicate was cleared as a system with specific sensors, the subject device is a software only device that is designed to accept sound files from a digital stethoscope as input.

The subject device has a different intended population than the primary predicate, i.e., it is intended for adults undergoing procedural sedation or anesthesia while the predicate is intended for adults in hospitals and hospital-type facilities. This intended population of the subject device is a subset of the intended population of the primary predicate and does not constitute a new intended use.

The subject device shares some technological characteristics with the primary predicate (K111933). Both are software with data from sensors as input and they both provide a respiratory rate measurement as output with a similar range and accuracy. The main differences in technological characteristics between the two devices are:

Principle of operation. Airmod™ uses acoustic signals produced by the turbulent airflow in the upper airway that occurs during inhalation and exhalation for input. Those signals are recorded by microphones on a digital stethoscope. Signal processing algorithms then convert the acoustic patterns into breath cycles to calculate a respiratory rate. The primary predicate device uses pulse oximetry technology, sensors and workflows to derive respiration rate based on the changes in the photoplethysmogram waveform that occur as a result of breathing. Heroic Faith has conducted performance clinical testing to validate that the type of signals used by Airmod™ is compatible with the intended use of the device. Due to the similarities in technology (i.e., acoustic sensor, sensor application site), Masimo Acoustic Monitoring Sensors cleared in K120984 is used as a reference device to further support the use of acoustic signals for respiratory rate monitoring in a clinical setting.

Sensor application site. The site of measurement on the patient is based on the mode of operation of the device. Since Airmod™ uses acoustic signals generated in the upper airway, the input sensor is placed on the patient's neck, while the primary predicate device has the sensor placed on the patient's finger to leverage pulse oximetry technology. Heroic Faith has conducted performance clinical testing to validate that the site of measurement used by Airmod 100 is compatible with the intended use of the device. Due to the similarities in technology (i.e., acoustic sensor, sensor application site), Masimo Acoustic Monitoring Sensors cleared in K120984 is used as a reference device to further support the use of acoustic signals for respiratory rate monitoring in a clinical

{6}------------------------------------------------

setting.

Software interface/platform. Airmod™ is a software as medical device intended as a mobile application to be installed on compatible operating systems such as Android 10 and 11. Heroic Faith has conducted software V&V testing to validate Airmod™ as a mobile application for Android devices. Due to the similarities in technology (i.e., acoustic sensor, sensor application site), Masimo Acoustic Monitoring Sensors cleared in K 120984 is used as a reference device to further support the use of acoustic signals for respiratory rate monitoring in a clinical setting.

A summary of the technological characteristics of the subject device and the predicate devices is presented in the table below.

Subject DevicePredicate DeviceReference Device
Airmod™Covidien NellcorRespiration Rate Software,Version 1.0Masimo AcousticMonitoring Sensors
K111933K120984
Indications for UseAirmod™, when used inconjunction withAccurSound ElectronicStethoscope AS-101, is asoftware as medical deviceintended to be used for thecontinuous, non-invasivemonitoring of respiratoryrate (RR) in adult patientswho are subjected toprocedural sedation and/oranesthesia.Airmod™ is intended for useby healthcare professionals inhospitals and healthcarefacilities who are legallycredentialed to performprocedural sedation and/oranesthesia.Airmod™ is intended forAndroid-based devicesonly.The Covidien NellcorRespiration Rate Software,when used in conjunctionwith a Nellcor pulseoximeter and a NellcorRespiration Rate Sensor, isintended to be used for thecontinuous, non-invasivemonitoring of respirationrate in adults in hospitalsand hospital-type facilities.The Rainbow AcousticMonitoring sensors andcables are indicated forcontinuous, noninvasivemonitoring of respiratoryrate (RRa).The RAS-125 is indicatedfor adult patients inhospitals, hospital-typefacilities, mobile and homeenvironments.The RAS-125c is indicatedfor adult and pediatricpatients in hospitals,hospital-type facilities,mobile and homeenvironments.
Product Codes /Regulation NumberBZQ / 21 CFR 868.2375• Primary: BZQ / 21 CFR868.2375• Secondary: DQA / 21CFR 870.2700• Secondary: DSA / 21 CFR870.2900• Primary: DQA / 21 CFR870.2700• Secondary: BZQ / 21CFR 868.2375
Patient PopulationAdults undergoing proceduralsedation and/or anesthesiaAdultsAdults and pediatric
Use EnvironmentsHospitals and healthcarefacilitiesHospital and hospital-likefacilitiesHospitals, hospital-typefacilities, mobile and homeenvironments
Device ComponentsSoftware• Hardware (sensors)• Software• Hardware (sensors)• Software

{7}------------------------------------------------

Principle ofOperationCompatible stethoscopewith microphone recordsacoustic signals produced bythe turbulent airflow in theupper airway that occursduring inhalation andexhalation, while signalprocessing algorithmsconvert the acoustic patternsinto breath cycles tocalculate respiration rateThe Respiration RateSoftware uses pulseoximetry technology,sensors and workflows toderive respiration rate basedon the changes in thephotoplethysmogram (pleth)waveform that occur as aresult of breathing.Sensor detects acousticsignals produced by theturbulent airflow in theupper airway that occursduring inhalation andexhalation, while signalprocessing algorithmsconvert the acoustic patternsinto breath cycles tocalculate respiration rate
Mode of OperationContinuous monitoringContinuous monitoringContinuous monitoring
Sensor ApplicationSiteNeckFingerNeck
Rate MeasurementRespiratory rate (RR) perminRespiratory rate (RR) perminRespiratory rate (RR) permin
RR Range andAccuracyClinical: 4-35 ±2.7 breathper minClinical: 4-40 ±1 breath perminBenchtop: 4-70±1 breathper minClinical:· Adult: up to 30 breathper minute· Pediatric: up to 50breath per minute
Notifications/AlarmsUser-adjustable notificationsfor RR lower/upper limitsand speed of notifications(every 1 or 2 seconds)User-adjustable alarms forRR lower/upper limitsUnknown
SoftwareInterface/DisplayMobile devicePatient monitorPatient monitor
Software PlatformMobile application forAndroid 10 and 11 operatingsystemsIntegrated softwareIntegrated software

SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate safety and performance based on current industry standards:

  • Software testing per IEC 62304 and FDA's guidance Content of Premarket Submissions for . Device Software Functions.
  • Cybersecurity testing per FDA's guidance Cybersecurity in Medical Devices: Oualitv . System Considerations and Content of Premarket Submissions.
  • Bench performance testing, including validation of the active noise cancellation functionality.

SUMMARY OF CLINICAL TESTING

Clinical testing was conducted to evaluate the non-inferiority of Airmod™ compared to the Capnostream™35 capnography system during procedural sedation. The study included both outside of US (OUS) and US populations, totaling 270 participants across multiple healthcare facilities. Primary endpoints assessed the difference in respiratory rate measurements between Airmod™ (airRR) and manual scored capnography (mancRR) with a performance goal of Root Mean Square Error (RMSE) < 3 breaths per minute (BPM). Secondary endpoints examined performance during periods of abnormal end-tidal CO2 graphs (ETF).

{8}------------------------------------------------

Test setMean Difference(95%CI)Lower LOA(95%CI)Upper LOA(95%CI)RMSE(95% CI)
N=270-0.457(-0.519,-0.395)-6.211(-6.437,-5.994)5.297(5.105,5.501)2.689(2.529,2.695)

Summary Table for All Point Estimates (airRR-mancRR)

(95% CI based on bootstrap data)

Overall. Airmod TM demonstrated a mean difference of -0.457 breaths per minute (BPM) compared to mancRR (95% CI: -0.519 to -0.395). Additionally, the RMSE value for Airmod™ was found to be 2.689 BPM (CI: 2.529, 2.695) when compared to manCRR, thus meeting the primary outcome and demonstrating statistical equivalence to the reference device in the study. Bland-Altman analysis, concordance correlation coefficients, and measures of bias and precision further supported Airmod™s performance.

CONCLUSION

Based on the testing performed, including software testing, cybersecurity testing, bench performance testing and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Airmod™ are assessed to be substantially equivalent to the predicate devices.

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).